Vapotherm Q50 User manual

Q50 Compressor
Instructions for Use


Quick Start Guide 2
Intended Use 3
Indications, Warnings and Cautions 4
Primary Indications 4
Contraindications 4
Warnings and Cautions 4
Components and Documents 5
Related Documents 5
Specifications 6
Main Specifications 6
Environmental Requirements 6
Life Expectancy of Compressor 7
Dimensions and Weight 7
High Altitude Considerations 7
Working Principle and Construction 9
Working Principle 9
System Overview 9
Set-Up for Use 13
Mounting Positions 13
Pole Mount 14
Install High Altitude / High Flow Kit 16
Connecting the Air Hose 17
Operating Instructions 18
Pre-operational Check 18
Starting the Compressor 19
Running 19
Shutting down the system 20
Electromagnetic Compatibility 21
Cleaning and Disinfection 22
Maintenance 25
Troubleshooting 26
Disposal 28
Warranty 29
Table of Contents

Quick Start Guide
Review all instructions and warnings prior to use.
1. Ensure the Precision Flow unit is unplugged and the power outlet
meets the appropriate power requirements.
2. Mount the compressor as needed. See “Mounting Positions,”
page 13.
3. Install and connect the High Altitude / High Flow Kit. Remove the
water trap from the air port on the back of the Precision Flow
and Install the auto-draining water trap assembly on the Precision
Flow.
4. AttachtheendoftheairhoseontotheDISSairttingontheback
of the Q50 Compressor and hand tighten.
5. AttachtheotherendoftheairhosetotheDISSairttingonthe
airlterassemblyonthebackofthePrecisionFlowunitandhand
tighten.
6. Connect an oxygen supply to the DISS oxygen tting on the
oxygen lter assembly on the back of the Precision Flow, if
oxygen concentrations above 21% are desired, and pressurize
the oxygen source.
7. Attach the Q50 power cord to the compressor power cord inlet
and plug into the power outlet.
8. Toggle the power inlet switch into the “I” position. The Q50
Compressor will now run.
9. Plug the Precision Flow into the power outlet and press the
run/standby button to begin start up.
WARNING: Verify that the water bottle reservoir has been
emptied prior to starting the compressor. Failure to do so
may result in the spillage of liquid and could pose a slip
hazard.
WARNING: Do not block or restrict the air exhaust
underneath the compressor. This could result in failure to
deliver therapy as intended. When the unit is placed on the
oor, be sure that any ooring, such as carpet, does not
restrict the air exhaust.
Page 2 3100772 Rev E

Intended Use
The Vapotherm Q50 air compressor with High Altitude / High Flow Kit
supplies air up to 40 L/min in order to support use of the Vapotherm
Precision Flow system in the absence of wall or tank supplied air.
The compressor was designed to be mounted on a Precision Flow
RollStand,orplaceddirectlyonaatleveloorthatdoesnotobstruct
theairexhaustunderneaththecompressor.Thelowprolefootprint,
light weight, and pole mounting capability allow for multiple use
congurationsandmodesoftransfer.
WARNING: The Q50 Compressor is intended for use only
with the Precision Flow system.
Page 3

Indications, Warnings and Cautions
Primary Indications
The Vapotherm Q50 Air Compressor with High Altitude / High Flow Kit
is designed for use only with the Vapotherm Precision Flow system.
The compressor is intended to supplement the air needs of the
Precision Flow in the event that wall or tank air is not available. It is
for institutional use only. It is not intended for home use.
GENERAL WARNING: Federal law (US) restricts the sales of this
device to, or on the order of, any physician. This device should
be used ONLY by a trained operator.
Contraindications
The Compressor is not intended for use in MRI environments.
For more information, see the Precision Flow Instructions for Use.
Warnings and Cautions
WARNING: indicates that a situation may occur which is
potentially harmful to the patient or user.
CAUTION: indicates a condition that may lead to equipment
damage, malfunction or inaccurate operation.
REFER TO INSTRUCTIONS FOR USE: indicates that additional
information is supplied in the Instructions for Use.
HOT SURFACE: indicates that the surface may become hot
under a single fault condition.
NOTE: indicates a point of emphasis to make operation more
ecientorconvenient.
Page 4 3100772 Rev E

Components and Documents
Item Description Qty
• Q50 Compressor 1
• Replaceable Air Intake Filter* 2
• Pole Mounts* 2
• Power Cord 1
• High Altitude / High Flow Kit
• Auto-draining water trap assembly
• Water Bottle
• Water Bottle Holder
• Drain Tube
1
*Polemountsandoneairintakelterareinstalledduringmanufacturing.
Related Documents
• Warranty
• CerticateofConformity
• Packaging List
NOTE: Warranty and Certicate of Conformance are the
warrants of after-sales maintenance.
Page 5

Specications
Main Specications
• Power Supply: 115V~ ±10%, 60Hz
• Continuous Flow: 0-40 L/min.
• Input Power: <600VA
• Noise: 54 dB(A) on average
• Electrical Safety: Meets requirements as dened by IEC
60601-1 “medical electrical equipment, Part 1 general safety
requirement”.
Environmental Requirements
Relative
Temperature Humidity
Operation: +18°-+30°C 10-90%
Storage: -10°-+50°C 20-90%
CAUTION: If the storage conditions are outside of the stated
operating range, the Q50 Compressor must be
acclimated to the operating environment for a
minimum of 8 hours prior to use.
Page 6 3100772 Rev E

Life Expectancy of Compressor
The average life expectancy will be 20,000 hours, based upon
operating conditions and with maintenance being performed as
outlined on page 26.
Dimensions and Weight
Dimensions: 20.5in(L)X19in(H)X9in(W)
Weight: 44 lbs.
High Altitude Considerations
Thespecicationsprovidedaboveareforsealevelconditions.The
densityofairislowerathigheraltitudesandhasanimpactontheow
capacity of the compressor. The following table provides guidance for
operating at higher altitudes.
CAUTION: Operating above the ow rates shown for a given
altitude and FiO2 level may result in condensation
or water collecting in the water trap on the back of
the Precision Flow device. Monitor the water trap
regularly and empty the trap if necessary to prevent
possible water entrainment into the Precision Flow.
Page 7

Page 8
FiO2
(%)
Maximum Precision Flow Rate (L/min)
Sea
Level
1,000
ft
2,000
ft
3,000
ft
4,000
ft
5,000
ft
6,000
ft
7,000
ft
8,000
ft
21% 40 40 40 39 37 34 32 30 28
22% 40 40 40 39 37 35 33 30 28
23% 40 40 40 40 38 35 33 31 29
24% 40 40 40 40 38 36 34 31 29
25% 40 40 40 40 38 36 34 31 30
26% 40 40 40 40 39 36 34 32 30
27% 40 40 40 40 39 37 35 32 30
28% 40 40 40 40 40 37 35 33 31
29% 40 40 40 40 40 38 36 33 31
30% 40 40 40 40 40 38 36 33 31
31% 40 40 40 40 40 39 37 34 32
32% 40 40 40 40 40 39 37 34 32
33% 40 40 40 40 40 40 38 35 33
34% 40 40 40 40 40 40 38 35 33
35% 40 40 40 40 40 40 38 36 34
36% 40 40 40 40 40 40 39 36 34
37% 40 40 40 40 40 40 40 37 34
38% 40 40 40 40 40 40 40 37 35
39% 40 40 40 40 40 40 40 38 35
40% 40 40 40 40 40 40 40 38 36
41% 40 40 40 40 40 40 40 39 36
42% 40 40 40 40 40 40 40 39 37
43% 40 40 40 40 40 40 40 40 38
44% 40 40 40 40 40 40 40 40 38
45% 40 40 40 40 40 40 40 40 39
46% 40 40 40 40 40 40 40 40 39
47% 40 40 40 40 40 40 40 40 40
48% 40 40 40 40 40 40 40 40 40
49% 40 40 40 40 40 40 40 40 40
50% or
Greater
40 40 40 40 40 40 40 40 40
3100772 Rev E

Working Principle and Construction
Working Principle
The air compressor starts after being plugged in and the power
inlet switch toggled. The power inlet switch is located above the
power inlet on the unit. Ambient room air is pulled in through the
airintakelterontheunit,pasttheheatexchangecoilandintothe
compressor. (Note: Excess moisture is directed to the water bottle
reservoir.) Here it is compressed, moves through the heat exchange
coil in order to reduce the air temperature, through an auto-draining
airlter,pressurereliefvalve,andpressureregulator.Theairthen
exits the unit and is delivered to the Precision Flow via the air hose.
System Overview
The Front Panel (Fig. 1) contains:
1. Water bottle reservoir
Figure1
Water bottle
reservoir
Page 9

The Rear Panel (Fig. 2) contains:
1. Air Intake Filter
2. Power Inlet Switch and Fuse Drawer
3. Hour Meter
4. Serial Number Label
5. DISS Air Fitting
Figure2
Air Intake Filter
Power Inlet Switch and Fuse Drawer
Hour Meter
DISS Air Fitting
Serial Number Label
Page 10 3100772 Rev E

Power Inlet Switch
The Vapotherm Q50 compressor makes use of a medical grade
power inlet switch. The power inlet switch contains two (2) fuses.
• 5ET 8-R (5mm X 20mm, 8A, SLOW BLOW, GLASS FUSE,
250V)
WARNING: To avoid risk of electric shock, this equipment
must only be connected to supply mains with
protective earth.
Air Intake
The compressor utilizes air from its surrounding environment. It
passes through multiple stages of ltration prior to delivery to the
PrecisionFlow.Therstltrationstageisaccessiblefromtheoutside
and should be cleaned once per week according to maintenance
procedures.Furtherltrationisaccessibleonlybytrainedpersonnel
during preventative maintenance.
WARNING: Never restrict or block the air intake. This could
result in failure to deliver therapy as intended.
Page 11

Air Outlet
The air exiting the Q50 compressor is to be used only in conjunction
with the Vapotherm Precision Flow system. The compressor output is
designed to match the performance needs of the Precision Flow up
to 40 L/min and as such has no user adjustable settings or gauges.
Air Exhaust
The air exhaust is located on the bottom of the unit and vents the
warm air used to cool the compressor.
WARNING: Do not block or restrict the air exhaust underneath
the compressor. This could result in failure to
deliver therapy as intended. When the unit is
placed on the oor, be sure that any ooring, such
as carpet, does not restrict the air exhaust.
WARNING: Do not remove the compressor feet. These are
designed to reduce noise and provide the required
space to allow the compressor to circulate sucient
cooling air throughout the system. Removal of the
feet could result in overheating of the system and
excessive noise.
WARNING: The bottom of the unit may become uncomfortably
warm if air ow is obstructed or under a single
fault condition. Do not lift the unit from the bottom
immediately after use.
Page 12 3100772 Rev E

Set-Up for Use
Mounting Positions
Thecompressorisintendedforuseintwocongurations,stand-alone
ontheoorormountedtoaPrecision Flow Roll Stand. If the Q50
Compressor is intended to be used stand-alone, no further work is
required: the compressor comes with four compressor feet to limit
noise and vibrations coming from the unit and holds the compressor
the correct distance from the oor to accommodate cooling air to
exhaust from the unit.
WARNING: Do not remove the compressor feet. These are
designed to reduce noise and provide the required
space to allow the compressor to circulate sucient
cooling air throughout the system. Removal of the
feet could result in overheating of the system and
excessive noise.
WARNING: Do not modify or alter the Q50 Compressor. Failure
to follow this warning may result in improper
operation, reduced life of the compressor, or failure
to deliver therapy.
Page 13

Pole Mount
The Q50 Compressor comes with the ability to be mounted onto the
Precision Flow Roll Stand via the pole mounts located on the side of
the compressor. When mounting to a Precision Flow Roll Stand, the
bottom of the compressor must be placed on the Precision Flow Roll
Stand base and no greater than 8” from the ground. The compressor
also must be placed over two locking casters of the Precision Flow
Roll Stand. Improper mounting the Q50 Compressor on the Precision
Flow Roll Stand will result in a higher risk of tipping the IV pole unit.
The Q50 compressor is only to be mounted on a Precision Flow Roll
Stand.
Page 14 3100772 Rev E

Instructions for Installation on Precision Flow Roll Stand:
1. Facing the back of the Vapotherm Precision Flow unit and roll
stand and ensuring two locking casters are on the right side,
place the Vapotherm compressor on the right of the roll stand
base. (Note: The side with the water bottle reservoir should face
front.)
2. Place one half of a pole clamp between the Vapotherm compressor
mounting hardware and the Precision Flow roll stand.
3. Position the other half of the pole mount as shown and install the
supplied pole clamp screws. Do not tighten yet.
Page 15

4. Repeat for the lower pole mount.
5. Firmly tighten both pole mounts.
Install High Altitude / High Flow Kit
1. Remove the manual draining water
trap from the air inlet to the PF. Using
a pair of needle nose pliers, push
theretainingringoftheairttingon
the back of the Precision Flow unit in
towards the Precision Flow. Pushing
inthisringwillreleasethettingon
the water trap. Pull the water trap
away from the Precision Flow.
2. Installthedraintubeintothettingon
the bottom of the water trap. Gently
pullonthetubetoconrmthatitis
capturedbythetting.
3. Install the Auto draining water trap
into the Air inlet of the Precision Flow
bypressingthettingonthewater
trapfullyintotheairttingonthe
back of the Precision Flow. To ensure
thatthettinghavecompletely
engaged, gently pull on the water
trap.
Page 16 3100772 Rev E

4. Press the water bottle holder on to the
Vapotherm Roll Stand approximately
four inches below the auto draining
water trap.
5. Insert the tube attached to the bottom
of the water trap into the hole in the
cap of the water bottle and press
the water bottle into the water bottle
holder.
Connecting the Air Hose
1. Push the end of the air hose onto
theDISSairttingonthebackofthe
Q50 Compressor until it reaches the
threads. Rotate the collar on the air
hose, clockwise, so that the threads
engagewiththeDISSairtting.
Tighten by hand.
2. Push the other end of the air hose on
theDISSairttingonthebackofthe
Precision Flow unit. Rotate the collar
on the air hose so that the threads
engagewiththeDISSairtting.
Tighten by hand.
CAUTION: The air outlet hose should be connected prior to
turning the compressor on.
Page 17

Operating Instructions
Pre-operational Check
Prior to starting the Q50 unit, the user should verify the following:
1. Check the power supply
Ensure that the power cord is installed into an outlet with the
appropriate power requirements.
2. CheckthatthePrecisionFlowunitispoweredo
Disconnect the Precision Flow power.
3. Check the outlet hose to the Precision Flow unit
Verify that the outlet hose that supplies the Precision Flow unit
with air is fully connected to the compressor and Precision
Flow unit before turning the unit on.
4. Connect oxygen supply to Precision Flow (if applicable)
Connect an oxygen supply as directed in the Precision Flow
Instructions for Use if oxygen concentrations above 21% are
desired.
WARNING: Verify that the water bottle reservoir and water
trap have been emptied prior to starting the compressor.
Failure to do so may result in the spillage of liquid and
could pose a slip hazard.
WARNING: Verify that the air intake on the back of the
unit and air exhaust on the bottom of the units are
unobstructed.
NOTE: When connecting to a wall air source, use the manual
draining water trap and gas hose provided with the
Precision Flow.
Page 18
Other manuals for Q50
1
Table of contents
Popular Compressor manuals by other brands

Gardner Denver
Gardner Denver ELECTRA-SAVER OPERATING & SERVICE MANUAL

Challenge Xtreme
Challenge Xtreme CX 1500/1 operating instructions

Scheppach
Scheppach HC 53dc Operating and safety instructions

Metabo
Metabo ClassicAir 255 Original operating instructions

Coltri Compressors
Coltri Compressors LP-300 NITROX Use and maintenance manual

BENDIX
BENDIX BA-921 COMPRESSOR-STD-CLOSED ROOM Service data sheet

Quincy
Quincy QR-25 Series instruction manual

SABROE
SABROE SAB 193 S Engineering manual

EKOM
EKOM DK50 2V S Instructions for use and service manual

Prebena
Prebena Mobilo 300 Original instructions

Sparmax
Sparmax TC-620X operating instructions

Gardner Denver
Gardner Denver INTEGRA EFD99B Operating and service manual