Violet Defense S.A.G.E UV Beam Guide

USER MANUAL & WARRANTY
S.A.G.E UV Beam
Violet Defense, LLC | www.violetdefense.com | 407.433.1104
S.A.G.E. UVS.A.G.E. UV
by Violet Defenseby Violet Defense

• Read these instructions prior to operating the unit.
• Keep and follow all instructions.
• Heed all warnings. Failure to use the equipment in the manner specied may impair the unit from providing the
desired protection.
• Plug unit into a grounded outlet. Ensure proper conditions for operation. The unit is designed to operate under
normal conditions indoors (temperature 5-40 degrees Celsius, RH less than 80% (non-condensing), at an altitude
<2000m and main supply uctuations +/- 10% of nominal voltage, overvoltage Category II.
• Allow unit to complete pre-programmed cycles prior to unplugging or shutting o power. This allows it to enter safe
mode.
• Do not allow the unit to get wet or be used in or near water.
• Do not block any ventilation openings. Use in accordance with these instructions.
• Do not attempt to open or tamper with the unit.
• Do not place or store the unit near any ammable materials or liquids.
• Do not defeat the safety purpose of the polarized or grounding-type plug. A polarized plug has two blades and a
third grounding prong. The wide blade or the third prong are provided for your safety. If the provided plug does not
t into your outlet, consult an electrician for replacement of the obsolete outlet.
• Protect the power cord from being walked on or pinched particularly at plugs, convenience receptacles, and the
point where they exit from the unit.
• Unplug this unit during lightning storms or when unused for long periods of time.
• Refer all servicing to qualied service personnel. Servicing is required when the unit has been damaged in any way,
such as power-supply cord or plug is damaged, liquid has been spilled or objects have fallen into the unit, the unit
has been exposed to rain or moisture, does not operate normally, or has been dropped.
• Unit is designed to operate only when the room is unoccupied. There are redundant safety systems to ensure the
unit does not activate while the room is occupied as the UV light is not recommended for excessive exposure. If the
sensor in the system fails, the system will shut down.
• Do not look directly at the unit due to the brightness of the light. UV light does not signicantly penetrate standard
glass, therefore, incidental exposure on the other side of a window or wall does not present any signicant risk.
THIS PRODUCT IS NOT A TOY. KEEP AWAY FROM CHILDREN.
IMPORTANT SAFETY INSTRUCTIONS
S.A.G.E. UV
USER MANUAL
CAUTION: TO REDUCE THE RISK OF ELECTRIC SHOCK,
DO NOT REMOVE COVER. NO USER-SERVICEABLE
PARTS INSIDE. REFER SERVICING TO QUALIFIED SERVICE
PERSONNEL.
The lightning ash with arrowhead symbol,
within an equilateral triangle is intended
to alert the user to the presence of
uninsulated “dangerous voltage” within the
product’s enclosure that may be of sucient
magnitude to constitute a risk of electric shock to persons.
WARNING: TO REDUCE THE RISK OF FIRE OR
ELECTRIC SHOCK, DO NOT EXPOSE THIS UNIT TO
RAIN OR MOISTURE.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own expense.
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
2
SAFETY WARNINGS

IMPORTANT SAFETY INSTRUCTIONS
Disclaimer: The S.A.G.E. UV unit is not intended for use as a medical device and people should not be directly exposed to the
light generated by the unit. The actual disinfection rates on a specic space will vary.
GENERAL OPERATION OF THE UNIT
1. Position the unit in the targeted cleaning area. See page 4 for set-up instructions.
• It is recommended that you place the unit approximately 2 meters (~6 1/2 feet) or less
away from the priority areas. Note that the unit has been tested to kill E. coli, S. enterica, S.
aureus at distances of up to 4 meters (~13 feet) away, but the length of time necessary to
run the unit increases with the distance from targeted area.
• Leverage reective surfaces in your space, including standard glass, as the UV light will
reect o of these surfaces. However, surfaces such as paint will absorb the photons,
minimizing eects of the unit, so avoid placing the unit too close to walls in your area.
2. Adjust the height of the unit to optimize eects of the unit, minimizing any shadowed areas.
See page 5 for how to adjust the height of the unit.
3. Plug the unit into a standard 110V wall receptacle using a surge protector. At its peak, the
Beam unit will draw less than 7 amps (less than a standard vacuum cleaner) for 4 second
intervals. Ensure the breaker has available amp draw during operation.
4. While the unit is operational, all personnel should leave the room during the disinfection cycle.
Exit the room within 60 seconds for optimal functioning of the unit. The unit is programmed
to operate for 30-minutes per cycle. See page 7 for more information on recommended
cleaning protocols. The unit’s redundant safety systems will stop the unit from operating if
motion is detected in the space. The unit will resume operation approximately 60 seconds
after room is unoccupied and no additional motion is detected.
5. The unit will run a safety test to ensure unit is functioning properly and that there is no motion
detected in the space. If successful, unit will run its pre-programmed operational cycle as
described in step 4.
6. Allow the unit to complete its entire cycle before unplugging the unit. This allows the unit
to enter safe mode. If necessary to unplug the unit prior to completion of cleaning cycle, be
sure to keep unit away from water for optimal safety. It’s recommended that the units have a
15-minute cool down period after each 30-minute disinfection cycle.
PRODUCT FEATURES
Beam features the S.A.G.E.
UV unit that uses pulsed
Xenon technology to deploy
powerful bursts of UV-C, UV-
B, UV-A, and violet-blue light
with ~170⁰of coverage
Built-in motion sensor will actively
monitor the space and only allow
the unit to activate when the space
is unoccupied
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
The height of each unit
may be adjusted to
optimize the disinfecting of
your space
3
GENERAL OPERATION
Caster wheels allow for easy
mobility of unit and will lock
in place. Simply step on the
brake to lock each wheel in
place once in position.

CONFIGURING THE UNIT
The S.A.G.E. Wizard can be used to congure units to the preferred programming mode and
adjust related settings. This conguration should be completed before initial operation or as
changes need to be made. To begin, visit www.violetdefense.com/wizard to download the
S.A.G.E. Wizard and computer conguration instructions. After downloading, run the installer
and a shortcut will be placed on computer’s start menu and desktop.
1. Congure your computer per instructions.
2. Plug the S.A.G.E. unit into your computer using an Ethernet cable.
3. Open the S.A.G.E. Wizard and follow the on-screen instructions to congure the unit,
including operational mode, related settings such as length of motion time-out, runtime
for manual or scheduled mode, and scheduling options, such as start-time for scheduled
mode.
4. Once completed, unplug the cable and the unit is ready for use.
Each standard unit has 3 primary modes available to users. Fixtures must be programmed
before initial use.
• Single Cycle Mode: (MOST COMMON FOR MOBILE DEPLOYMENTS) In this method,
users will manually activate the unit to start a disinfection cycle. The unit will then
run its pre-programmed cycle (i.e. 15 or 30 minutes) one time before entering safe
mode until power is cycled on and o again. Users may select the timeout time,
which determines how long the unit will wait until resuming operation after motion is
detected.
• Autonomous Mode: In this mode, once plugged in, the unit will automatically
activate once the space is unoccupied and continue disinfection on a regular
schedule, adjusting frequency based on level of activity in the space based on the
motion sensor. Users may select the timeout time, which determines how long the
unit will wait until resuming operation after motion is detected.
• Scheduled Mode: In this mode, the unit will activate at selected time(s) of day
during each 24-hour period. Users may schedule up to 4 start times each day to
operate for a specied run time (i.e. 15 or 30 minutes). Users may select the motion
timeout, which determines how long the unit will wait until resuming operation after
motion is detected. The unit may also be congured to activate a disinfection cycle
immediately when it is powered and then continue with the scheduled operational
time(s).
OPERATIONAL MODES
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
4
OPERATIONAL MODES

CONFIGURING THE UNIT
OPERATIONAL MODES
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
5
ADJUSTING HEIGHT OF BEAM UNIT
2. Use an allen wrench to
loosen 4 screws (2 screws
on each side) on the unit.
1. Support unit with one hand.
3. Raise or lower unit to
desired height.
4. Using allen wrench, tighten
4 screws (2 on each side) on
the unit.
ADJUSTING HEIGHT

SENSOR INDICATORS
TROUBLESHOOTING
1. If unit is ashing red, then an issue was detected by the unit. Unplug or power down the unit.
2. Plug the unit back into the wall or turn on power to unit. Exit the space within 60 seconds.
3. When you return to the space, identify the color of light on the sensor indicator. If unit is
green or solid red, the unit is working properly. If unit is still ashing red, contact customer
service at Violet Defense at 407-433-1104.
4. If unit is non-responsive (i.e. no indicator light showing), please contact Violet Defense for
technical support.
Each S.A.G.E. UV unit has indicator lights that
help you determine the status of the unit.
CLEANING THE UNIT
• Do not use abrasive cleaners to clean the unit.
• To clean the unit, wipe down the metal plates with stainless steel polish using a microber
cloth.
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
Color of Light Meaning
Flashing Green Unit is ready to begin cleaning when room
becomes unoccupied.
Solid Green Unit has successfully completed its cleaning
cycle. If unit detects motion, the unit will change
to solid red, even if cycle completed.
Solid Red Unit was interrupted during the cleaning cycle.
Cleaning will resume when room becomes
unoccupied again during the programmed times
for operation.
Flashing Red One of the safety sensors has detected an issue.
Please see troubleshooting section for further
information.
6
TROUBLESHOOTING

RECOMMENDED CLEANING PROTOCOLS
For most situations, the general cleaning procedures can be used to operate the S.A.G.E. UV
unit. However, there are special instructions recommended for blood spills, or when certain
organisms, such as C. di are known or suspected to be present.
General Cleaning
1. Complete standard cleaning procedures to remove any visible dirt, grease, or other
debris from space.
2. Activate the unit per operational instructions on page 3.
3. Upon returning to space, look at indicator light. Solid green light indicates cleaning cycle
completed. If motion was detected prior to observing the indicator light, the indicator
light will turn red. If users want to determine if unit completed the last cycle, prior to
powering o, stand out of range of the motion sensor for 60 seconds. The light will
resume solid green if cycle was completed. Otherwise, it will resume operation.
Cleaning When Blood or Bodily Fluids Are Present
UV light has the ability to penetrate liquid blood, which has not formed a dried crust, or
bodily uids up to a 1/16” thickness as long as there is no tissue or solid material present.
Therefore, UV light may be used to help protect sta before and after traditional cleaning
protocols for your facility.
1. Activate S.A.G.E. UV unit prior to commencing any cleaning procedures. If the blood has
already dried, apply hydrogen peroxide solution prior to running the unit.
2. Utilize normal procedures, including appropriate safety protocols to clean the space,
including removal of blood and bodily uids.
3. Activate a second cycle of the S.A.G.E. UV unit to target any remaining bacteria and
viruses.
Cleaning When C. di Is Of Concern
Clostridium dicile (C. di) is a bacteria that is extremely dicult to kill (hence its name) as
it is able to survive in unfavorable conditions. Furthermore, it is an anaerobic, endosporic
organism, which means it can survive without oxygen and form dormant spores to survive
harsh conditions. With the risk of severe illness associated with C. di, it is recommended
that you use a combination of approaches to reduce the risk of contracting C. di in your
facility.
1. Utilize your facility’s adopted environmental cleaning and disinfection strategy for dealing
with C. di.
2. Activate the S.A.G.E. UV unit for two consecutive 30-minute cycles at a range of less than
one meter for best results.
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
7
CLEANING PROTOCOLS
SENSOR INDICATORS

S.A.G.E. UV SAFETY OVERVIEW
Given the correct combination of user protocols and the built-in safety features, there are
minimal risks of any harmful eects from using the S.A.G.E. UV units.
RECOMMENDATIONS
• If in the vicinity of the S.A.G.E. UV unit during operation, do not look directly at the
light. This is similar to how one would avoid harmful eects from the sun by not
looking directly at it.
• Germicidal UV light does not signicantly penetrate standard glass.
• Per the built-in safety features described below, the unit is designed to not operate
while a room is occupied. However, operators are instructed in the user manual to
have all persons vacate a space before operation. Further optional protocols, such
as signage indicating the UV unit is in operation may be desired.
SAFETY FEATURES
The S.A.G.E. UV units have four primary aspects to their built-in safety features to ensure
the safety of anyone operating the unit or in its vicinity.
Limited Exposure Risk
Due to the nature of the programming on the S.A.G.E. UV units, over a 30-minute cycle
time for the unit, the actual exposure to UV (if there were no other safety systems in place)
would be no more than two seconds of exposure.
Motion Detection
Each S.A.G.E. UV unit has built-in passive infrared (PIR) sensor used to detect motion in
a space. The sensor is programmed to detect individuals walking into the deployment
space. The unit will not resume operation of the cleaning cycle until it has successfully
detected the space is free of motion.
Safe Stop
The third layer of protection built into the S.A.G.E. units is to ensure that if the motion
sensor or other components experience technical issues, the unit will safe stop by cycling
the unit o. The unit will ash red indicating that one should contact Violet Defense for
technical support. The unit will not resume operation until the issue detected has been
resolved. If unit is non-responsive (i.e. non-operational or no indicator light), contact Violet
Defense.
Frequency of Flashes
The S.A.G.E. UV units will ash every 6 seconds during operation. Violet Defense
collaborated with the Epilepsy Foundation when designing the product to ensure that the
frequency of ashes will not cause any issues for someone with photosensitive epilepsy.
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
8
UV SAFETY OVERVIEW

S.A.G.E. UV SAFETY OVERVIEW
Occupational Safety and Health Administration (OSHA)
The Occupational Safety and Health Administration (OSHA) does not have any mandated
exposure limits to ultraviolet light. OSHA only provides technical guidance regarding
protecting employees from ultraviolet laser exposure. While general information about
ultraviolet contained in that guidance is described below, it is important to note that Violet
Defense Technology does not currently deploy ultraviolet lasers. For more information on
OSHA’s guidelines, visit https://www.osha.gov/pls/oshaweb/owadisp.show_document?p_
table=INTERPRETATIONS&p_id=24755.
Ultraviolet radiation is divided into three regions: UV-A: 315-400 nanometers (nm), UV-B:
280-315 nm, and UV-C: 100-280 nm. UV can be associated with adverse health eects due
to prolonged exposure and the wavelength of light.
According to OSHA’s guidelines, “exposure in the shorter UV-C and longer UV-A ranges
seems less harmful to human skin. The shorter wavelengths are absorbed in the outer dead
layers of the epidermis and the longer wavelengths have an initial pigment-darkening eect
followed by erythema if there is exposure to excessive levels.”
“The hazards associated with skin exposure are of less importance than eye hazards.”
Exposure to light may cause photokeratitis or cataracts.
National Institute for Occupational Safety and Health (NIOSH)
The National Institute for Occupational Safety and Health (NIOSH) recommends limits to
exposure determined by the wavelength of UV light and intensity. NIOSH recommends that
the time of exposure to an intensity of 100 microwatts per square centimeter at wavelength
254 nm not exceed 1 minute. Per the programming, UV exposure from S.A.G.E. unit for a
30-minute cycle is less than two seconds. For more information, view the recommended
standards from NIOSH (https://www.cdc.gov/niosh/docs/73-11009/pdf/73-11009A.pdf).
Environmental Protection Agency (EPA)
The Environmental Protection Agency (EPA) is the governmental agency responsible for
regulating ultraviolet light products. It regulates chemical disinfectants along with devices,
such as equipment that generates UV light, used to control pests like bacteria and making
antimicrobial claims. For more information about EPA guidelines, visit https://www.epa.gov/
safepestcontrol/pesticide-devices-guide-consumers.
Food and Drug Administration (FDA)
The Food and Drug Administration (FDA) only regulates devices that are classied as medical
instruments, machines, and devices used to treat diagnosed medical conditions. Therefore,
Violet Defense does not fall under FDA guidelines.
GOVERNMENTAL SAFETY CONSIDERATIONS FOR ULTRAVIOLET LIGHT
© Copyright 2020 Violet Defense, LLC All Rights Reserved SUM-GEN2BEAM-UM2020
9
UV SAFETY OVERVIEW

10
WARRANTY
STATEMENT OF LIMITED WARRANTY
FOR VIOLET DEFENSE S.A.G.E. PRODUCTS
04/2020
Thank you for purchasing your Violet Defense S.A.G.E Product (the “Product”).
THE TERMS OF THE ONE (1) YEAR LIMITED WARRANTY (THE “WARRANTY”) APPLICABLE TO THE PRODUCT ARE
SET FORTH BELOW.
HOW CONSUMER LAW RELATES TO THIS WARRANTY
THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS AND YOU MAY ALSO HAVE OTHER RIGHTS, WHICH VARY
FROM STATE TO STATE (OR BY COUNTRY, PROVINCE OR OTHER JURISDICTION). OTHER THAN AS PERMITTED
BY LAW, VIOLET DEFENSE, LLC (“VIOLET DEFENSE”) DOES NOT EXCLUDE, LIMIT OR SUSPEND OTHER RIGHTS
YOU MAY HAVE, INCLUDING THOSE THAT ARISE FROM NONCONFORMITY OF THE PRODUCT. FOR A FULL
UNDERSTANDING OF YOUR RIGHTS, YOU SHOULD CONSULT THE LAWS OF YOUR COUNTRY, PROVINCE, STATE
OR OTHER JURISDICTION.
WARRANTY LIMITATIONS AND DISCLAIMERS (SUBJECT TO CONSUMER LAW)
TO THE EXTENT PERMITTED BY LAW, THIS WARRANTY AND THE REMEDIES SET FORTH HEREIN ARE
EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, REMEDIES AND CONDITIONS, WHETHER ORAL,
WRITTEN, STATUTORY, EXPRESS OR IMPLIED. VIOLET DEFENSE DISCLAIMS ALL STATUTORY AND IMPLIED
WARRANTIES, INCLUDING WITHOUT LIMITATION, WARRANTIES OF MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE AND WARRANTIES AGAINST HIDDEN OR LATENT DEFECTS, TO THE EXTENT
PERMITTED BY LAW. IN SO FAR AS SUCH WARRANTIES CANNOT BE DISCLAIMED, VIOLET DEFENSE LIMITS THE
DURATION AND REMEDIES OF SUCH WARRANTIES TO THE WARRANTY PERIOD (AS DEFINED BELOW) AND, AT
VIOLET DEFENSE’S OPTION, THE REMEDIES SET FORTH HEREIN. SOME STATES (COUNTRIES, PROVINCES AND
OTHER JURISDICTIONS) DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY (OR CONDITION)
MAY LAST, SO THE LIMITATION DESCRIBED ABOVE MAY NOT APPLY TO YOU.
WHO MAY USE THIS WARRANTY?
Violet Defense extends this Warranty only to the end-user purchaser who originally purchased the Product from
Violet Defense or a Violet Defense authorized distributor or reseller. This Warranty is non-transferable and is not
intended for the benet of any person or entity other than the original end-user purchaser.
WHAT DOES THIS WARRANTY COVER?
VIOLET DEFENSE WARRANTS THE VIOLET DEFENSE PRODUCT TO BE FREE FROM DEFECTS IN MATERIALS
AND WORKMANSHIP WHEN USED NORMALLY IN ACCORDANCE WITH VIOLET DEFENSE’S PUBLISHED
GUIDELINES FOR A PERIOD OF ONE (1) YEAR FROM THE DATE OF ORIGINAL RETAIL PURCHASE BY THE END-
USER PURCHASER, AS EVIDENCED BY THE PROOF OF PURCHASE (THE “WARRANTY PERIOD”).
WHAT DOES THIS WARRANTY NOT COVER?
This Warranty does not apply to any appliance, assembly, or product other than the Product, even if the Product is
incorporated in, or packaged or sold with, any such appliance, assembly or other product.
This Warranty does not apply to any condition or damage due to external causes such as: (1) acts of nature; (2)
physical damage; (3) exposure to water either through submersion or water entering the unit; (4) exposure to
adverse or hazardous chemicals, including the use of reactive cleaning agents or harsh chemicals to clean the
Product; (5) environmental conditions; (6) vandalism; (7) re; (8) power failure, improper power supply, power
surges or dips; (9) induced vibration; (10) negligence or more culpable conduct by the purchaser, its employees,

11
WARRANTY
contractors, representatives or other users of the Product; (11) improper or unauthorized use, installation, handling,
storage, alteration, maintenance or service, including failure to abide by any product classications or certications,
unauthorized repair, or failure to comply with any applicable standards, codes, recommendations, product
specication sheets, instructions, and safety warnings; (12) use of the Product(s) with other products, equipment,
processes or materials supplied by any end user or third party; (12) transportation; (13) normal wear and tear; (14)
accidents; or (15) events or occurrences beyond Violet Defense’s control.
WHAT IS THE PERIOD OF COVERAGE?
This Warranty starts on the date of the original purchase of the Product by an end-user purchaser, and lasts for one
(1) year. The Warranty Period is not extended if we replace the Product. Violet Defense may change the availability or
terms of this Warranty at its discretion, but any changes will not be retroactive.
WHAT ARE YOUR REMEDIES UNDER THIS LIMITED WARRANTY?
During the Warranty Period, if the Product fails to conform to the foregoing Warranty, Violet Defense, at its sole
discretion, will replace the Product (or the defective portion thereof). The foregoing Warranty excludes labor and
equipment required to remove the non-conforming Product (or part(s) thereof) and/or re-install the replacement
Product(s) (or part(s) thereof).
HOW DO YOU OBTAIN WARRANTY SERVICE?
Warranty claims must be submitted in writing to Violet Defense within thirty (30) days of the discovery of the
nonconformity, or the nonconformity will be deemed to have been waived. To obtain Warranty service, you must call
407.433.1104 during the Warranty Period to obtain a Return of Materials Authorization (“RMA”) number. No Warranty
service will be provided without an RMA number.
The nonconforming Product(s) may be required to be returned to Violet Defense for inspection and verication of
the nonconformity. Violet Defense will not be responsible for any costs and expenses incurred in connection with
shipment of the nonconforming Product(s) (or part(s) thereof) to Violet Defense, however, Violet Defense shall bear
the expenses of shipping the replacement Product(s) (or part(s) thereof) to the end-user purchaser in the event that
Violet Defense replaces the Product(s) (or part(s) thereof).
SOLE REMEDIES; LIMITATION OF LIABILITY
THE REMEDIES DESCRIBED ABOVE ARE YOUR SOLE AND EXCLUSIVE REMEDIES AND OUR ENTIRE
RESPONSIBILITY FOR ANY BREACH OF THIS WARRANTY. OUR LIABILITY SHALL UNDER NO CIRCUMSTANCES
EXCEED THE ACTUAL AMOUNT PAID BY YOU FOR THE NONCONFORMING PRODUCT. EXCEPT AS PROVIDED
IN THIS WARRANTY AND TO THE MAXIMUM EXTENT PERMITTED BY LAW, WE ARE NOT LIABLE FOR DIRECT,
INDIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES RESULTING FROM ANY BREACH OF
WARRANTY OR CONDITION, OR UNDER ANY OTHER LEGAL THEORY, INCLUDING BUT NOT LIMITED TO LOSS
OF USE; LOSS OF REVENUE; LOSS OF ACTUAL OR ANTICIPATED PROFITS (INCLUDING LOSS OF PROFITS ON
CONTRACTS); LOSS OF THE USE OF MONEY; LOSS OF ANTICIPATED SAVINGS; LOSS OF BUSINESS; LOSS OF
OPPORTUNITY; LOSS OF GOODWILL; LOSS OF REPUTATION; OR ANY INDIRECT OR CONSEQUENTIAL LOSS OR
DAMAGE HOWSOEVER CAUSED INCLUDING THE REPLACEMENT OF EQUIPMENT AND PROPERTY.
SOME STATES (COUNTRIES, PROVINCES AND OTHER JURISDICTIONS) DO NOT ALLOW THE EXCLUSION OR
LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT
APPLY TO YOU.
GENERAL
No Violet Defense distributor, reseller, agent, or employee is authorized to make any modication, extension, or
addition to this Warranty. If any term is held to be illegal or unenforceable, the legality or enforceability of the
remaining terms shall not be aected or impaired. This Warranty is governed by and construed under the laws of the
country in which the Product purchase took place.

®
189 S Orange Avenue | Suite 1400 | Orlando, FL 32801
Table of contents
Other Violet Defense Lighting Equipment manuals
Popular Lighting Equipment manuals by other brands

Titanium
Titanium TT-FL04-24W quick start guide

Meteor
Meteor ELIPSCAN User Manual / Operation Instruction

Prolux
Prolux MATRIX BAR 2 IP Wi user manual

Clevertronics
Clevertronics ARGO-1200 EM Series Installation and maintenance instruction

HEICO lighting
HEICO lighting Polyoptik Technical bulletin

Perfect Pool
Perfect Pool PP2 installation manual