Vivendi ELGA PURELAB Ultra User manual


PURELAB Ultra
PURELAB Ultra
OPERATOR MANUAL
OPERATOR MANUAL
VERSION 1 UK –05/02
REF: MANU37248
VERSION 1 UK –05/02
REF: MANU37248

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page i
Copyright Note
ELGA®and PURELAB®are registered trademarks of Vivendi Water
Systems Ltd.
The information contained in this document is the property of ELGA
LabWater, a division of Vivendi Water Systems Ltd. and is supplied
without liability for errors or omissions.
No part of this document may be reproduced or used except as
authorized by contract or other written permission from Vivendi Water
Systems Ltd. The copyright and all restrictions on reproduction and
use apply to all media in which this information may be placed.
Vivendi Water Systems Ltd. pursue a policy of continual product
improvement and reserve the right to alter without notice the
specification, design, price or conditions of supply of any product or
service.
© Vivendi Water Systems Ltd. 2002
All rights reserved.
Publication ref: MANU37248
Version 1 - 05/02
ELGA is a trading name of Vivendi Water Systems and USFilter
(USA).

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page iii
TABLE OF CONTENTS
1. INTRODUCTION ............................................................1
1.1 Product Range ......................................................1
1.2 Use of this Manual ................................................1
1.3 Customer Support .................................................1
2. HEALTH AND SAFETY NOTES....................................2
2.1 Electricity...............................................................2
2.2 Ultraviolet Light .....................................................2
2.3 Sanitization Chemicals..........................................2
2.4 Control of Substances Hazardous to Health
(COSHH) ...............................................................2
3. PRODUCT AND PROCESS DESCRIPTION .................3
3.1 Product Description...............................................3
3.2 Process Description ..............................................4
3.3 Technical Specification .........................................9
4. CONTROLS .................................................................12
5. INSTALLATION INSTRUCTIONS ...............................13
5.1 Unpacking the PURELAB Ultra..........................13
5.2 Positioning the PURELAB Ultra .........................13
5.3 Connecting the PURELAB Ultra.........................15
5.4 Initial Controller Set-Up .......................................17
5.5 Setting Up Password / Resetting Replacement
Timers .................................................................22
5.6 Initial Start Up......................................................24
5.7 POU Filter Installation (LC134) - Optional
Accessory............................................................26
6. OPERATION ................................................................27
6.1 Intermittent or Continuous Recirculation .............27
6.2 Manual Dispense ................................................28
6.3 Volumetric Dispense ...........................................29
6.4 Automatic Calibration ..........................................30
6.5 Water Supply Flush.............................................30
7. MAINTENANCE ...........................................................31
7.1 Replacing the Purification Packs.........................31
7.2 Replacing the Ultra-Microfilter (LC109)...............33
7.3 Replacing the Ultrafilter (LC151).........................35
7.4 Replacing the Ultraviolet Lamp
(LC105 and LC118).............................................37
7.5 Replacing the Point-Of-Use Filter (LC134)..........38
7.6 Cleaning the Inlet Strainer (External) ..................39
TABLE OF CONTENTS - CONT'D

PURELAB Ultra Operator Manual ELGA
Page iv PURELAB Ultra Version 1 - 05/02
8. SANITIZATION PROCEDURES.................................. 40
8.1 CT1 Sanitization Tablet - Safety Information ...... 42
9. TROUBLE SHOOTING................................................ 43
10. CONSUMABLES AND ACCESSORIES ..................... 44
11. KEY TO CONTROL PANEL........................................ 45
11.1 Icons ................................................................... 45
11.2 Alarm Conditions................................................. 46
11.3 Replacement Timers........................................... 46
11.4 Quality Alarms..................................................... 47
12. WARRANTY/CONDITIONS OF SALE ........................ 48
13. USEFUL ADDRESSES ............................................... 50

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 1
1. INTRODUCTION
1.1 Product Range
This operator manual has been prepared for the PURELAB Ultra
product models:
PURELAB Ultra Scientific
PURELAB Ultra Bioscience
PURELAB Ultra Genetic
PURELAB Ultra Ionic
PURELAB Ultra Analytic
1.2 Use of this Manual
This manual contains full details on installation, commissioning and
operation of the PURELAB Ultra unit. If the instructions in this
handbook are not followed then the performance of this product
and/or the safety of the user may be compromised.
1.3 Customer Support
Service support and consumable items are available from your local
supplier or distributor. Refer to customer service contact details
shown at the end of this publication.
PURELAB Ultra

PURELAB Ultra Operator Manual ELGA
Page 2 PURELAB Ultra Version 1 - 05/02
2. HEALTH AND SAFETY NOTES
PURELAB Ultra products have been designed to be safe, however, it
is important that personnel working on these units understand any
potential dangers. All safety information detailed in this handbook is
highlighted as WARNING and CAUTION instructions. These are used
as follows:
WARNING! WARNINGS ARE GIVEN WHERE FAILING
TO OBSERVE THE INSTRUCTION COULD
RESULT IN INJURY OR DEATH TO
PERSONS.
CAUTION! Cautions are given where failure to
observe the instruction could result in
damage to the equipment, associated
equipment and processes.
2.1 Electricity
It is essential that the electrical supply to the PURELAB Ultra is
isolated before any items are changed or maintenance work
performed.
The ON/OFF switch is located at the left-hand side of the unit. The
mains power lead is located just behind the ON /OFF switch.
WARNING! THIS APPLIANCE MUST BE EARTHED.
The main water supply should be isolated and residual pressure
released prior to removal of any Purification Packs or carrying out
work on the unit.
Switching off the electrical supply will isolate the source of mains
water pressure.
2.2 Ultraviolet Light
WARNING! UNDER NO CIRCUMSTANCES SHOULD
THE LAMP BE CONNECTED AND
ACTIVATED WHEN OUTSIDE THE
HOUSING.
The Genetic, Analytic and Ionic units are fitted with an ultraviolet lamp.
The UV lamp is enclosed in a stainless steel chamber ensuring that
the operator will not be exposed to UV light.
2.3 Sanitization Chemicals
During the sanitization cycle a CT1 sanitization tablet is used and
relevant safety information is included in this handbook. A safety data
sheet conforming to COSHH regulations is also provided with the
tablets and should be read before a tablet is used.
2.4 Control of Substances Hazardous to Health (COSHH)
Material safety data sheets covering the various replaceable
Purification Packs are available upon request. Contact your local
supplier or distributor.
Fuse
ON/OFF
switch
Mains
p
ower lead
Mains power socket
Mains Power Supply
UV Lamp
Ultra with UV Lamp

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 3
3. PRODUCT AND PROCESS
DESCRIPTION
3.1 Product Description
The PURELAB Ultra water purification unit has been specifically
designed to provide a supply of ultrapure water with very low levels
of impurities for laboratory, medical and industrial applications.
The PURELAB Ultra can be bench, underbench or wall mounted with
an optional wall mounting kit. A range of accessories is available to
complement the unit (see Section 10. - Consumables and
Accessories, for detail).
Control Panel Dispense Nozzle Dispense Controller
Door
PURELAB Ultra
The PURELAB Ultra range of water purification units has been
designed to provide the ultimate in high purity water for laboratory
applications. The products need to be fed with pre-treated water,
typically from a reverse osmosis supply. They can be operated directly
from a ringmain feed as point of use polishers, with or without local
buffer storage, using a Docking Vessel or suitable reservoir.
Alternatively, they can be supplied from a local pre-treatment/storage
system, for example a PURELAB Prima and Docking Vessel
combination.

PURELAB Ultra Operator Manual ELGA
Page 4 PURELAB Ultra Version 1 - 05/02
3.2 Process Description
Scientific
Pre-treated water enters via an inlet solenoid, passes
through a flow sensor and is pumped through the primary
purification pack, an intermediate water quality sensor, a
polishing purification pack and final temperature and water
quality sensors before being dispensed or recirculated
through a non-return valve back to the pump inlet.
The first purification pack removes most of the impurities
from the water and the intermediate water quality sensor then
measures the resistivity of the water from the first pack to
check when it needs to be replaced.
Any remaining ionic and organic impurities are removed by
the second polishing purification pack. Product water
resistivity and temperature are measured before dispense.
Solenoid
Valve
NRV
Feedwater Inlet
(Pre-purified Water) Flow
Sensor Pump
Solenoid Valve
Drain
Purified Water
Outlet
Primary
Purification
Pack
Polishing
Purification
Pack
0.2µm Point-of-use Filter
(Optional)
Dispense Controller
Outlet Water
Quality Sensor
Temperature
Sensor
Intermediate
Water
Quality Sensor
Solenoid Valve

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 5
Bioscience
Pre-treated water enters via an inlet solenoid, passes
through a flow sensor and is pumped through the primary
purification pack, an intermediate water quality sensor, a
polishing purification pack, an ultrafilter and final temperature
and water quality sensors before being dispensed or
recirculated through a non-return valve back to the pump
inlet.
The first purification pack removes most of the impurities
from the water and the intermediate water quality sensor then
measures the resistivity of the water from the first pack to
check when it needs to be replaced.
Any remaining ionic and organic impurities are removed by
the second polishing purification pack; the ultrafilter removes
pyrogens, bacteria and other microbial impurities as well as
particles. Product water resistivity and temperature are
measured before dispense.
0.2µm Point-of-use Filter
(Optional)
Purified Water
Outlet
Solenoid Valve
Flow
Sensor
Feedwater Inlet
(Pre-purified Water)
Intermediate
Water
Quality Sensor
Temperature
Sensor
Outlet Water
Quality Sensor
Solenoid Valve
Drain Drain
Primary
Purification
Pack
Polishing
Purification
Pack
Pump
NRV
Dispense Controller
Flush
Solenoid
Valve
Ultrafiltration
Membrane
Solenoid
Valve

PURELAB Ultra Operator Manual ELGA
Page 6 PURELAB Ultra Version 1 - 05/02
Genetic
Pre-treated water enters via an inlet solenoid, passes
through a flow sensor and is pumped through the primary
purification pack and an intermediate water quality sensor.
The water then flows through a UV chamber, further
resistivity and temperature sensors, a polishing purification
pack, an ultrafilter and final temperature and water quality
sensors before being dispensed or recirculated through a
non-return valve back to the pump inlet.
The first purification pack removes most of the impurities
from the water and the intermediate water quality sensor then
measures the resistivity of the water from the first pack to
check when it needs to be replaced.
Purified water flows directly through the UV chamber where it
is exposed to intense UV radiation at wavelengths of 254nm
and 185nm to provide continuous bacterial control and the
photo-oxidation of residual organic impurities. The second
temperature compensated water quality sensor provides data
for TOC monitoring.
Any remaining ionic and organic impurities are removed by
the second polishing purification pack; the ultrafilter removes
pyrogens, bacteria and other microbial impurities as well as
particles. Final water resistivity and temperature are
measured before dispense.
Solenoid Valve
Outlet Water
Quality Sensor
Intermediate
Water
Quality
Sensor
Primary
Purification
Pack
Polishing
Purification
Pack
Dispense
Controller
Drain
Solenoid
Valve
Pump
NRV
Temperature
Sensor
UV Lamp
Drain
Feedwater Inlet
(Pre-purified Water) Flow
Sensor
Solenoid Valve 0.2µm Point-of-use Filter
(Optional)
Purified Water
Outlet
Ultrafiltration
Membrane
Post UV
Quality
Sensor
Temperature
Sensor
Flush
Solenoid
Valve

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 7
Ionic
Pre-treated water enters via an inlet solenoid, passes
through a flow sensor and is pumped through the primary
purification pack and an intermediate water quality sensor.
The water then flows through a UV chamber, a further
resistivity and temperature sensor, a polishing purification
pack, an ultra microfilter and final temperature and water
quality sensors before being dispensed or recirculated
through a non-return valve back to the pump inlet.
The first purification pack removes most of the impurities
from the water and the intermediate water quality sensor then
measures the resistivity of the water from the first pack to
check when it needs to be replaced.
Purified water flows directly through the UV chamber where it
is exposed to intense UV radiation at a wavelength of 254nm
to provide continuous bacterial control.
Any remaining ionic and organic impurities are removed by
the second polishing purification pack; the ultra microfilter
removes bacterial impurities and particles. Product water
resistivity and temperature are measured before dispense.
Feedwater Inlet
(Pre-purified Water)
Solenoid
Valve
Pump
Flow
Sensor
NRV
Solenoid Valve
Drain
Primary
Purification
Pack
Polishing
Purification
Pack
UV Lamp
UMF
Intermediate
Water Quality
Sensor Outlet Water
Quality
Sensor
Temperature
Sensor
Solenoid Valve
Dispense Controller
0.2µm Point-of-use Filter
(Optional)
Purified Water
Outlet

PURELAB Ultra Operator Manual ELGA
Page 8 PURELAB Ultra Version 1 - 05/02
Analytic
Pre-treated water enters via an inlet solenoid, passes
through a flow sensor and is pumped through the primary
purification pack and an intermediate water quality sensor.
The water then flows through a UV chamber, further
resistivity and temperature sensors, a polishing purification
pack, an ultra microfilter and final temperature and water
quality sensors before being dispensed or recirculated
through a non-return valve back to the pump inlet.
The first purification pack removes most of the impurities
from the water and the intermediate water quality sensor then
measures the resistivity of the water from the first pack to
check when it needs to be replaced.
Purified water flows directly through the UV chamber where it
is exposed to intense UV radiation at wavelengths of 254nm
and 185nm to provide continuous bacterial control and the
photo-oxidation of residual organic impurities. The second
temperature compensated water quality sensor provides data
for TOC monitoring.
Any remaining ionic and organic impurities are removed by
the second polishing purification pack; the ultra microfilter
removes bacterial impurities and particles. Product water
resistivity and temperature are measured before dispense.
Feedwater Inlet
(Pre-purified Water)
Solenoid
Valve
Flow
Sensor
Solenoid Valve
Pump
Primary
Purification
Pack
UV Lamp Polishing
Purification
Pack
UMF
0.2µm Point-of-use Filter
(Optional)
Purified Water
Outlet
Solenoid Valve
Intermediate
Water
Quality
Sensor
Temperature
Sensor
Outlet Water
Quality
Sensor
Drain
NRV
Dispense Controller
Post UV
Quality
Sensor
Temperature
Sensor

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 9
3.3 Technical Specification
The Technical Specifications for the PURELAB Ultra are as follows:
Feedwater*
Parameter Limits
Source - Originally from potable supply, then
pre-treated
Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled.
Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion.
Fouling Index (max) 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds.
Service Deionisation (SDI) - MΩ-cm 1 MΩ-cm minimum resistivity at exhaustion.
Reverse Osmosis (RO) - µS/cm Ideally < 30 µS/cm
Free Chlorine 0.05ppm max.
TOC 50ppb max.
Carbon Dioxide 30ppm max.
Silica 2ppm max.
Particulates Filtration down to 0.2 micron advisable to protect internal and/or point of use filters
Temperature 1-40°C - Recommended 10-15°C
Flowrate (maximum requirement) 130 l/hr
Drain requirements (gravity fall with air gap).
Maximum during service
Up to 2 l/min
Feedwater Pressure 0.7bar (10psi), maximum, 0.07bar (1psi) minimum
* Contact technical support for advice on feedwaters outside the range listed.
Note: Different system configurations are available for different feedwater sources. See system set
up.
Dimensions
Height 490mm (19.3")
Width 410mm (16.2")
Depth 365mm (14.4")
Weight
Scientific 14.0kg (30.8 lb.)
Bioscience 14.5kg (32.0 lb.)
Ionic 14.5kg (32.0lb.)
Analytic 15.0kg (33.1lb.)
Genetic 15.0kg (33.1lb.)
Connections
Inlet-quick connect 8mm (5/16") OD
Drain 8mm (5/16") OD
Flush - Bioscience/ Genetic 8mm (5/16") OD
Positioning Wall, bench or under bench mounted.
Environment Clean dry indoor. Temp 5-40°C.
Humidity max 80% non-condensing.

PURELAB Ultra Operator Manual ELGA
Page 10 PURELAB Ultra Version 1 - 05/02
Electrical Requirements
Mains input 100 -240V ac, 50-60Hz all models
System voltage 24V dc
Power consumption during recirculation 60VA
Power consumption during dispense 75VA
Fuses 2 x T6.3 Amp
Reservoir level connection Jack Plug 3.5mm
Noise level during recirculation <40dBA
User Interface
Display Continuous graphical quality display
Graphical flow schematic on screen with mimic display
Backlit display with Intuitive Icons
Adjustable settings Volumetric dispense Selectable dispense volumes (0.1 – 60.0 liters)
Recirculation mode Intermittent or continuous
Feed water type Selectable (RO/DI/SDI)
Display viewing angle Adjustable electronically
Water quality units Selectable (MΩ-cm or µS/cm)
Water quality alarm Selectable alarm setpoints
Date / time Adjustable
Auto restart after power failure Selectable (On/Off)
Audible alarm Selectable (On/Off)
Temperature control Selectable alarm setpoints
Night service Selectable (On/Off)
Indicators Product water quality Resistivity or conductivity
Intermediate water quality Resistivity or conductivity
Temperature Degrees Centigrade or Fahrenheit
Total Organic Carbon (TOC)* ppb C
Purification pack Replacement date
UV lamp* Replacement date
Filter* Replacement date
Alarms-Audiovisual Purified water purity Above set point alarm
Temperature**** Above set point alarm
UV lamp operation* Imminent lamp failure or failure to start
Calibration Calibration alarm – calibration required
Purification pack Change reminder
UV lamp* Change reminder
Filter* Change reminder
Reservoir**** Low level / level control disconnect alarm
Outputs RS232 printer connection
RS232 remote display connection
Volt free contact-internal
Remote dispense

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 11
Safety Features
Power fail safe
Water temperature alarm
Water quality alarm
Purification pack interlock
UV current monitoring
Timeout of manual dispense
Low operating voltage 24V
Dry contact alarm connection
Visual alarms
Audible alarms
Operational Features
Low noise levels – minimum intrusion
Continuous calibration to traceable standards
Drop dispense
Multistage monitoring
Restart on power interrupt
Volumetric dispense
Optional printer kit for record of operating parameters
Optional remote display
Purified Water Specification
Scientific Ionic Analytic Bioscience Genetic
Flowrate 2.0 l/min max.
Inorganic Resistivity 18.2 MΩ-cm (0.055µs/cm)
TOC **<15ppb **<5ppb **<2ppb **<10ppb **<3ppb
Bacteria ***<1cfu/ml <1cfu/ml
Pyrogens - <0.03Eu/ml <0.001Eu/ml
pH Effectively neutral
Particles ***0.2µm0.05µmUltrafiltration
RNase / DNase - Free
Capacity
(LC147)
56,000 liters >18MΩ-cm per single purification pack
µS/cm at pH7.0
Conforms to ASTM, CAP, NCCLS – TYPE 1 and BS3978 Grade 1 Specification.
* On some models only.
** Dependant on feedwater - recommended feed <50ppb TOC
*** With POU filter fitted.
**** Switches process off.
Purification Packs
Standard pack LC147 recommended for most applications. Refer to Section 10 for alternative cartridges.
As part of our policy of continual improvement we reserve the right to
alter the specifications given in this document.

PURELAB Ultra Operator Manual ELGA
Page 12 PURELAB Ultra Version 1 - 05/02
4. CONTROLS
The PURELAB Ultra operates with a tactile membrane control panel,
which has a graphics display window and four multi-purpose control
buttons.
Details of how to use the controls will be given in the appropriate
sections.
The PURELAB Ultra control panel has a range of control icons as
follows:
Button Icon Function
PROCESS Turns the process
on/off
ESCAPE from any
feature
Menu
Scroll
(0.10)* Preset dispense
volume
LEFT
Reset
Accept
Volumetric dispense
CENTRE
(0.50)* Preset dispense
volume
Mute Alarm
Printer
RIGHT
(1.00)* Preset dispense
volume
Intermittent
recirculation
* Example only - Ref. to Section 6.3 Step 3.
Process button
Left hand Centre Right hand
Control button Control button Control button
Control Panel

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 13
5. INSTALLATION INSTRUCTIONS
5.1 Unpacking the PURELAB Ultra
The following items are supplied with your PURELAB Ultra.
1. PURELAB Ultra unit.
2. Sanitization by-pass blocks LA638
(2 off, fitted in unit).
3. 6 metres of 8mm (5/16") O/D tube.
4. 1 off Pack of CT1 tablets.
5. 1 off Strainer Assembly.
6. Operator Manual in English.
7. Multilingual Operator Manual on CD.
8. Mains Lead.
9. Reservoir Level Connecting Lead.
10. Icon Directory, fitted on inside of door.
5.2 Positioning the PURELAB Ultra
Before installation and operation of the PURELAB Ultra unit, please
read and observe the following points.
Environment
The unit should be installed on a flat, level surface, in a
clean, dry environment. The unit can also be wall mounted
against a vertical wall capable of supporting the weight (for
this we recommend the use of the wall mounting kit Part No.
LA610 on substantial brick/concrete walls or LA622 for stud
partition walls).
WARNING! IF THE UNIT IS TO BE WALL MOUNTED,
ENSURE IT IS MOUNTED USING THE
CORRECT WALL MOUNTING KIT AND
THAT THE WALL IS CAPABLE OF
SUPPORTING THE OPERATING WEIGHT
OF THE SYSTEM. ALWAYS CAREFULLY
FOLLOW THE INSTRUCTIONS INCLUDED
IN THE KIT.
Note: Refer to Specifications for unit weights
(Section 3.4 - Technical Specification).
The unit is designed to operate safely under the following conditions:
•Indoor Use
•Altitude up to 2000m
•Temperature Range 5 - 40°C
•Maximum Relative Humidity 80% @ 31°C decreasing
linearly to 50% @ 40°C, non-condensating
The unit is in Installation Category II, Pollution Degree 2, as per
IEC1010-1.
Rear Mounting Points
Unit Rear Mounting Points

PURELAB Ultra Operator Manual ELGA
Page 14 PURELAB Ultra Version 1 - 05/02
Electrical
The units can be connected universally to any electrical
supply in the range of 100 - 240V and 50 - 60Hz. The mains
lead is supplied with a molded plug on one end and a molded
connector to the unit on the other. The unit should be
connected to an earth. The unit includes a battery which will
require changing on a periodic basis, typically every 3-5
years. (Contact local Service provider)
WARNING! DANGER OF EXPLOSION IF BATTERY IS
INCORRECTLY REPLACED. REPLACE
ONLY WITH THE SAME OR EQUIVALENT
TYPE RECOMMENDED BY THE
MANUFACTURER (3 VOLTS, LITHIUM
CR2032). DISPOSE OF USED BATTERIES
ACCORDING TO THE MANUFACTURER’S
INSTRUCTIONS.
Drain/Flush
A semi-rigid flexible connection to a sink or suitable drain
capable of at least 2.0l/min is required. The drain point
should have a gravity fall below the level of the unit and any
connections direct to drain should have an air-break device
fitted. On the Genetic & Bioscience units a separate flush line
to drain is required. The drain and flush drain lines should run
individually to a convenient drain point.
Feed Water
The feedwater should be potable water, pre-purified using
reverse osmosis, deionisation, or distillation. If using a supply
other than reverse osmosis treated, it is strongly
recommended that a 0.2 micron membrane filter is installed
prior to remove colloidal impurities. The feedwater should
enter the PURELAB Ultra via an 8mm (5/16") O/D semi rigid
tube, and should be in the temperature range 1 to 40°C (34
to 104°F). To ensure optimum performance it is
recommended that the feedwater temperature be between 10
to 15°C.
CAUTION! Operating with feed water temperatures
outside the range from 1 to 40°C (33.8 to
104°F) will cause damage to the PURELAB
Ultra unit.
For pressurized feeds, the minimum direct inlet pressure is
0.07bar (1.0psi) and maximum inlet pressure is 0.7bar
(10psi). Higher feedwater pressures must be reduced using a
pressure regulator valve (Part No. LA575).
Reservoir feeds to the PURELAB Ultra unit should be
positioned at the same height, or above the unit, to provide a
positive flooded inlet pressure of approximately 0.07bar
(1psi).
Fuse
ON/OFF
switch
Mains
p
ower lead
Mains power socket
Electrical Connections

ELGA PURELAB Ultra Operator Manual
PURELAB Ultra Version 1 - 05/02 Page 15
5.3 Connecting the PURELAB Ultra
Once the PURELAB Ultra unit has been positioned either on a wall or
on a bench, it should be connected as follows:
•Mains water inlet tube
•Drain tube - to drain
•Flush tube (Bioscience and Genetic) - to drain
Step 1 - Remove Transit Plugs
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on an 8mm OD semi
rigid drain tube.
4. PUSH tube into connector.
Step 2 - Connect Water Inlet
1. Connect 8mm tube from the PURELAB Ultra
inlet to the inlet strainer.
2. Connect the inlet strainer to the pre-purified
water supply.
Step 3 - Connect Drains
1. Connect 8mm tube from the drain and UF drain
(if fitted) and direct to a suitable drainage point.
CAUTION! The drain and flush lines should allow a
gravity fall to drain with no restrictions.
CAUTION! If the water supply is at a pressure greater
than 0.7bar (10psi) fit a pressure regulator
(LA575).
UF Drain
Drain
Inlet
Inlet and Drain Connection
Direction of
water flow
Collar
Mesh filter
Collar
Strainer
Inlet
Feed Water Inlet
Connector Collet
Blanking
transit plug
Pull plug
Depress
collet
Clean square
cut end
Tube 8mm OD
Simply push in tube to attach
Tube in secured position
Fitting tubes
Table of contents
Popular Water System manuals by other brands

Zip
Zip HydroTap CS Commercial Installation and operating instructions

Reclaim
Reclaim REHP-CO2-160GL Owners & installation manual

King Technology
King Technology AQUA SMARTE Plus MINERAL ACTIVATOR Operational instructions

JABSCO
JABSCO 31765-0092 manual

Brita
Brita PROGUARD Coffee manual

Sterilite
Sterilite Cobalt SC200-DWS11 owner's manual

3M
3M Aqua-Pure CS-S quick guide

Kenmore
Kenmore 625.384650 manual

Field Controls
Field Controls CLEARWAVETM WATER CONDITIONER CW-HD-4 installation manual

Zip
Zip HydroTap G4 Installation and operating instructions

OSMOSEA
OSMOSEA NEW12 ES60 Instructions for installation, use and maintenance manual

CEFLA
CEFLA PURE 500 operating manual