Gas Safety
Adapter
CP 250
Operating
Instructions
Warnings:
When using gas, be extremely careful!
Operate the adapter CP 250 on a level surface and in
the showen position (horizontal) , only!
This safety adapter is only to be operated with a
Camping Gaz CP 250 valve gas cartridge. Attempting
to use another cartridge with the safety adapter can
be dangerous. We do not take any responsibility for
incidents resulting from the use of other
manufacturer's cartridges.
Always assemble the CP 250 safety adapter well
away from sources of ignition (e.g. open flames,
cigarettes, pilot lights, electric fires, etc.) and at a safe
distance away from other people.
Make sure that there is sufficient ventilation for correct
combustion and to avoid the formation of dangerous
unburned gas mixtures.
Before every use ensure that the two seals (5&6) of
the adapter are in good condition and in the proper
position. Never use an adapter which is damaged or
impaired in function.
Never refill cartridges.
If there is a leak in the adapter (smell of gas)
immediately bring it to a well ventilated location at a
safe distance from any source of flame in order to
search for and eliminate possible leaks. Never search
for leaks with an open flame. Use soapy water instead.
Failure to follow instructions strictly may results in
gas leaks, which if ignited could cause serious burns
to the user or people nearby.
Never throw away a cartridge which still contains gas.
(shake the cartridge and if you hear the sound of liquid
there is still gas remaining). Read also the instructions
of the gas cartridge.
Storage:
Store cartridges in a dry, cool location; and safe
distance away from sources of heat.
(Ambient temperature below 40°C).
Protect the environment:
Dispose of empty cartridges in a safe place.
Pay attention to the instruction for "technical rules
liquid gas" (TRF)
sales department:
SpandauerWeg1
D - 37085 Göttingen
Phone:+49(0)551/793789
Fax : +49 (0)551/793707
Internet: http://www.wld-tec.com
Production and Service:
Halle-Kasseler Straße 49
D - 37318 Arenshausen
Phone:+49(0)36081/68940
Fax : +49 (0)36081/68942
Email: [email protected]
WLD - TEC GmbH
Additional devices from this product family:
Fuego SCS basic Art.-No.: 8.001.000
laboratory burner with button function
and foot pedal
Fuego SCS Art.-No.: 8.000.000
laboratory burner with button function,
IR-Sensor and LC-Display
Fuego SCS pro Art.-No.: 8.004.000
laboratory burner with button function,
IR-Sensor, LC-Display, 2 rechargeable
batteries
Powerjet 2 Art.-No.: 6.006.000
Hand-held laboratory gas burner
with piezoelectric ignition
Consumption material:
Gas cartridges CP 250 Art-No. 6.000.192
250g butan gas with valve
(Pack = 6 pieces)
O-Ring-sealing, outer (5): Art.-No.: 6.010.820
O-Ring-sealing inner (6): Art.-No.: 6.011.820
04/10

Read these instructions carefully to familiarize yourself with the
product before fitting the CP 250 gas cartridge into the adapter.
Please retain these operating instructions for future reference.
1. Use: The CP 250 safety gas adapter has been developed for
the operation of WLD-TEC products independent of a central gas
supply. The adapter is fitted with a pressure regulator which
reduces the cartridge pressure to 50 mbar. Temperature range
when using cartridges is +10°C - +40°C. Avoid temperatures
above +50°C. Operate the adapter CP 250 on a level surface
and in the showen position (horizontal) , only! Pay attention
to the instruction for "technical rules liquid gas" (TRF).
1.1 Delivery:
Safety Adapter CP 250 Art.-No. 6.000.820
with 0.5 m of DVGW hose
Operating Instructions
1.2 Technical Specifications:
Adapter for Cartridges: CP 250 from
Camping Gaz
250g Butane gas
Pressure reduction adapter: 50mbar
Gas burner continuous operation: ca. 210 min.
per cartridge
Dimensions: (L x W x H) 260 x 110 x 80 mm
with cartridge
Weight: (without cartridge) 660 g
DVGW Gas hose: 0.5m / G 1/4" left-hand
thread
2. Connection:Before connecting or changing the gas cartridge
adapter turn off the laboratory gas burner.
A nozzle for natural gas (08)or (N) has been installed into the gas
burner at the factory. The nozzle is to be replaced by one suitable
for the CP 250 cartridge, butane gas (06) or (P). (The suitable
nozzle (06 / P) is in the delivery of the gas burner included).
Please see the instruction manual of the laboratory gas burner for
further information.
Connect one end of the DVGW gas hose (1)to the gas connection
(2) of the adapter and the other end to the gas connector of the
laboratory gas burner. Ensure that all gas connections are gas-
tight and that they are screwed tight (left-handed threads). Ensure
gas proofness with a suitable test fluid / equipment. See warnings.
2.1 Fitting a new cartridge: Pay attention to the warning on the
back. Do not use any cartridges which are dented. Ensure that the
gas burner is switched off. Ensure that the two seals (5 & 6) of the
adapter are in good condition and in the proper position. In case
of damages order replacement sealings:
-O-Ring-sealing, outer (5):Art.-No.: 6.010.820
- O-Ring-sealing inner (6): Art.-No.: 6.011.820
Before inserting the gas cartridge (3) turn the cartridge
connector (4) into correct position by turning it a few degrees
counter-clockwise to the stop (see Fig. 2). Then, hold the
adapter tightly and insert the cartridge CP 250 (3) plain into the
cartridge connector (4) of the adapter. Take care that the slot
(8) of the cartridge fits to the guide lug (7) of the adapter.
To interlock the cartridge, turn the cartridge connector (4) with gas
cartridge (3) a few degrees clockwise to the end position (see
Fig. 4). The end position of the locked cartridge is shown in Fig. 1.
If you suspect a leak, follow the safety instructions on the back.
2.2 Operating the unit: Operate the gas safety adapter CP 250
only in a horizontal position. When operating the burner for the first
time turn the gas regulation knob two revolutions to the left and turn
the air regulator to full air supply (+). There is additional operating
information in "Operating Instructions" of the laboratory gas burner
beginning in the paragraph "Operating Procedure". When operating
this unit please read instructions thoroughly.
2.3 Taking the unit out of operation: After completing your work
or if you have to leave the burner unattended for a longer period of
time, switch off the gas burner and disconnect the cartridge by
removing the gas cartridge (3) from the adapter.
Therefore turn the cartridge connector (4) with gas cartridge (3)a
few degrees counter-clockwise to the stop and take the cartridge
off (see Fig . 5).
2.4 Changing an empty cartridge: The cartridge (3) can also be
replaced before it is empty, since this type of cartridge has a built-
in valve.
The replacement is always to be performed at a well-ventilated
location and away from any source of ignition.Ensure that the gas
burner is switched off.
To remove the gas cartridge (3), turn the cartridge connector (4)
with the gas cartridge (3) a few degrees counter-clockwise to the
stop and take the cartridge off (see Fig . 5).
Never throw away a cartridge which still contains gas.(shake the
cartridge and if you hear the sound of liquid there is still gas
remaining). Read also the instructions of the gas cartridge.
Fitting a new cartridge see paragraph 2.1
3. Warranty: 2 - year warranty on parts and labor excluding
damage caused by improper operation, unauthorized changes in
design and unqualified service.
31
6 & 7
1
3
7 & 8
2
4
1 - DVGW Gas hose
2 - Gas connection
3 - Gas cartridge CP250
4 - Cartridge connector
5 - O-Ring-sealing, outer
6 - O-Ring-sealing, inner
7 - Guide lug
8 - Slot / Gas cartridge
Fig. 1
Fig. 2
Fig. 3
Fig. 4
5
6
4
7
8
3
87
4
Cartridge disconnected
Fig. 5
This manual suits for next models
1
Table of contents
Popular Adapter manuals by other brands

Viking Electronics
Viking Electronics PA-IP quick start guide

Honeywell
Honeywell OneWireless Adapter installation guide

APM
APM IEEE 802.11g Wireless LAN USB Adapter... Specifications

Lindemann
Lindemann Limetree Bridge manual

TRENDnet
TRENDnet TBW-104UB Specifications

Cisco
Cisco Linksys WUSB54G Quick installation