Disinfection of the Probes with Alcohol
One recommended disinfection technique is to clean the probe assemblies with
70% isopropyl alcohol.
A 5- to 10- minute exposure is recommended. It is imperative that the alcohol
be given time to evaporate before applying a probe to a patient’s eye. Do not
completely immerse the probe or cable; only the tip of the probe should be placed in
the solution.
After cleaning, rinse the end of the probe thoroughly with distilled water to remove
all traces of alcohol.
Probe surfaces should be dried with a lint-free cloth.
High-Level Disinfection of the Probe
If high-level disinfection is required by your facility, the probe may be cleaned using
an FDA-cleared high-level disinfectant, such as Cidex OPA Activated Dialdehyde
Solution.
If your facility is located in the EU, Mikrozid wipes are a compatible method of high-
level disinfection for Accutome probes.
Note: Be sure to follow the disinfectant manufacturer’s written protocol when
using any antibacterial solution, including high-level disinfectants.
CAUTION: No abrasive or harsh cleaning solutions should be used while cleaning the
AccuPach VI or any of its components.
Electrical Hazard and Safety
The instrument, footswitch, and probe of the AccuPach VI are electrical/electronic
devices. Reasonable care should be taken when making an electrical connection
and handling electrically powered devices. Avoid the use of damaged electrical
equipment or frayed electrical cords. If repair or maintenance is to be performed
on the AccuPach VI, the equipment must be turned off and the power cord
disconnected.
The device covers must not be removed except by qualified personnel. There are no
user controls inside the unit. To avoid injury, do not operate the AccuPach VI
without protective covers.
The system is intended to operate from a power source of 100-240 VAC, 47-63 Hz.
Protective earth grounding through the grounding connection in the power cord is
essential for safe system operation.
Use only a hospital-grade power cord and the connectors supplied with the
AccuPach VI. Also, be sure the power cord and connector are in good condition.
Safety