AEB G5A-80A User manual

G5-ABE Battery
For use with Cardiac Science™ Powerheart® G5
Models: G5A-80A, G5A-80C, G5S-80A, G5S-80C
Operation Manual
Revision 3.4

Product model
The G5-ABE is indicated for use with the Cardiac Science™ Powerheart® G5, models G5A-80A, G5A-80C, G5S-80A, and G5S-
80C.
Please see the Cardiac Science™ Powerheart® G5, models G5A-80A, G5A-80C, G5S-80A, and G5S-80C User’s Guide for
complete information on AED use.
Indications for use
The automated external defibrillator (AED) battery supplies power to an AED as required during self maintenance, automated
diagnoses, and defibrillation. The G5-ABE is indicated for use with the Cardiac Science™ Powerheart® G5, models G5A-80A,
G5A-80C, G5S-80A, and G5S-80C.
Contraindications
Automated external defibrillators should not be used when a patient is conscious or breathing normally.
Battery specifications
Battery Type: 12VDC, 7.5Ah, lithium sulfur dioxide, primary cells.
Lithium Content: approximately 9.2 grams
Capacity: Typically 420 shocks or 16 hours of operating time at 25°C (77°F). Actual battery operating life depends on device settings,
usage, and environmental factors.
Charge Time: Initiation of rhythm analysis to ready to shock: 15 seconds (typical); 45 seconds (maximum) With a fully charged
battery. Initiation of rhythm analysis to ready to shock, used battery: 15 seconds typical); 45 seconds (maximum) with a battery that
has been used for 15 shocks.
Shelf Life (prior to insertion): A minimum of 5 years from date of refurbishing when stored from 20° to 30°C (68° to 86°F), 10% to
75% RH (non-condensing).
Standby Life (after insertion/between use): 4 years when operated from 20° to 30°C (68° to 86°F), 10% to 75% RH (non-
condensing).
Operating Conditions: 0°C to 50°C (32° to 122°F), 0% to 95% RH (non-condensing). Lithium Content: approximately 9.2 g.
Safety Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-4
Enclosure Protection: IP55 when housed in AED.
Transportation: UN38.3 tested: T1 –Altitude, T2 –Thermal, T3 –Vibration, T4 –Shock, T5 –External Short

Important Warnings and Reminders!!
Caution: Federal law restricts this device to sale by or on the order of a physician or practitioner licensed by law of
the state in which he/she practices to use the device.
Battery is not rechargeable. Do not attempt to recharge.
Do not expose battery to high heat or open flames. Do not incinerate battery.
Waste Electronic Electrical Equipment (WEEE). Separate collection for waste electrical and electronic equipment
This alert identifies hazards that may cause personal injury, product damage, or property damage.
Pressurized contents: Do not short circuit, puncture, deform, or expose to temperatures above 65°C (149°F).
Not tested for airborne applications
Inspect battery contacts on insertion into AED and annually. Clean contacts with gold contact cleaner if any sign
of oxidation or corrosion is present.
Always keep a spare battery on hand in addition to the battery currently in operation.
Use one of these solutions to clean the case of the AED: soapy water, denatured ethanol, or 91% isopropyl
alcohol.
Use one of these solutions to clean the plastic housing of the battery: soapy water, denatured ethanol, or 91%
isopropyl alcohol.
If the AED indicates that the battery is low and needs to be replaced please recycle or dispose of the lithium
battery in accordance with all federal, country, state, and local laws. Help our environment, recycle, and send
your depleted batteries to AED Battery Exchange:
AED Battery Exchange
1000 Brown Street, Ste 206
Wauconda, IL 600487

About the AED
AEDParts
Pad expiration window
Rescue Ready® status indicator
Lid
Battery compartment (on bottom)
Pad package holders
Latch (press to open)
AED is not Rescue Ready.
AEDisRescueReady.RescueReadytechnology
assessestheAED,battery,andpadstoensure
they are ready to perform a rescue.
The Display Panel
Speaker
Padsocket/USBport
Display(detailsbelow)
Informationdisplay Defibrillation padsindicator
◆
Number ofshocksdelivered
SHOCKS1
00:00 Lights when the pads are:
◆
Rescue timer
◆
RescuepromptsandCPR
countdown timer
NOW SAFE
TO TOUCH PATIENT
◆
improperly connected to theAED
◆
cold, dried, ordamaged
◆
detachedfromthepatientduringarescue
Function buttons
Presstoenterdiagnosticmodeorto
changethepromptlanguage.
Smartgauge™ battery status indicator
ThegreenLEDsdisplaythecapacityof
thebattery.Withuse,theLEDsturnoff
as battery capacity decreases. When
thegreenLEDsturnoffandtheredLED
lights, replace thebattery.
Shock button (semi-automatic model
only)
◆
LightsredwhentheAEDisreadyto
deliveradefibrillationshock.
◆
Presstodelivertherapytothepatient.
Service indicator
LightswhentheAEDdetectsaneedfor
maintenance orservice.

1
2
3
4
5
Battery Installation:
Note:TheABEbattery(G5-ABE)automaticallystoreshistoryofitsoperatinglife.
ThebatteryhistorycanbereviewedwithAEDManagerSoftware.
How to replace the battery:
Presstab Lift out Insert new Pressdowntohearclick Ensure RescueReady
is green
Note: Ensure that the battery is at room
temperaturebeforeinsertingitintotheAED.