Aeroqual IQM 60 User manual

IQM Version 5.0 (Pump) revision October 2010
Aeroqual Limited
109 Valley Road, Mount Eden, Auckland 1024, New Zealand
phone +64 9 623 3013
fax +64 9 623 3012
web www.aeroqual.com
IQM 60
Indoor Air Quality Monitor
User Guide
Version 5.0

IQM 60 V5 Pump Revision October 2010
2
Table of Contents
Table of Contents ........................................................................................................................... 2
Foreword ........................................................................................................................................ 3
Statements of Compliance ............................................................................................................. 3
Warranty......................................................................................................................................... 4
For Your Safety .............................................................................................................................. 5
Technical Support .......................................................................................................................... 5
1.0 Monitor and System Components............................................................................................ 6
1.1 IQM 60 ................................................................................................................................. 6
1.2 System Components include: .............................................................................................. 6
2.0 Description .................................................................................................................... 6
2.1 Inlets.............................................................................................................................. 7
3.0 Operating Instructions .............................................................................................................. 8
3.1 Getting Started..................................................................................................................... 8
3.2 Connecting a computer........................................................................................................ 8
3.3 Reading the Datacard .......................................................................................................... 8
3.4 Testing ................................................................................................................................. 8
3.5 Maintenance ........................................................................................................................ 8
3.6 Care and Cleaning ............................................................................................................... 8
4.0 Troubleshooting........................................................................................................................ 9
4.1 Troubleshooting Guide.................................................................................................. 9
4.2 Diagnostics ................................................................................................................. 10
5.0 Calibration .............................................................................................................................. 11
5.1 Gas Module Calibration .............................................................................................. 11
5.1.1 Nafion Humidifier .................................................................................................... 11
5.1.2 Zero verification using Gas Treatment Module ...................................................... 12
5.1.3 Zero calibration ....................................................................................................... 12
5.1.3.1 Using Gas Treatment Module ............................................................................ 12
5.1.3.2 Laboratory Zero calibration ................................................................................. 12
5.1.4 Span calibration...................................................................................................... 12
5.1.4.1 Span gas ............................................................................................................. 12
5.1.4.2 Span calibration procedure ................................................................................... 13
5.2 Particle Counter Calibration ........................................................................................ 14
Flow rate .............................................................................................................................. 14
Zero Check .......................................................................................................................... 14
Span Calibration .................................................................................................................. 14
Appendix A AQM V5.x Software Instructions.................................................................... 15
Description ............................................................................................................................... 15
Setting up ................................................................................................................................. 15
Passwords ............................................................................................................................... 15
Menu Descriptions ................................................................................................................... 15
File ........................................................................................................................................... 15
Setup........................................................................................................................................ 17
Data.......................................................................................................................................... 19
Tools ........................................................................................................................................ 22
Diagnostics .............................................................................................................................. 26
Window .................................................................................................................................... 28
Help.......................................................................................................................................... 29
Additional Buttons .................................................................................................................... 29
Updating Sensor List................................................................................................................ 29
Appendix B Communication Protocol................................................................................ 30
Enclosure ..................................................................................................................................... 42
External View ........................................................................................................................... 42
Description........................................................................................................................... 42
Dimensions .......................................................................................................................... 42
Front View ................................................................................................................................ 43
Rear View................................................................................................................................. 43

IQM 60 V5 Pump Revision October 2010
3
Foreword
Copyright Aeroqual Limited. All rights reserved. Reproduction, transfer, distribution or storage of part or all of the
contents of this document in any form without the prior written permission of Aeroqual Limited is prohibited.
“Aeroqual” and “Aeroqual Limited – Making the Invisible Visible” are registered trademarks of Aeroqual Limited. Other
product and company names mentioned herein may also be trademarks or trade names.
Aeroqual operates a policy of continuous development. Aeroqual reserves the right to make changes and improvements
to any of the products described in this document without prior notice.
Under no circumstances shall Aeroqual be responsible for any loss of data or income or any special, incidental,
consequential or indirect damages howsoever caused.
The contents of this document are provided "as is". Except as required by applicable law, no warranties of any kind,
either express or implied, including, but not limited to, the implied warranties of merchantability and fitness for a particular
purpose, are made in relation to the accuracy, reliability or contents of this document.
Aeroqual reserves the right to revise this document or withdraw it at any time without prior notice. The availability of
particular products may vary by region. Please check with the Aeroqual dealer nearest to you.
©Aeroqual Limited 2008. All rights reserved.
Statements of Compliance
1. The Aeroqual IQM Monitor complies with EN 61000-6-1:2001
2. The Aeroqual IQM Monitor complies with EN 61000-6-3:2001
3. The Aeroqual IQM Monitor complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) these devices may not cause harmful interference, and (2) these devices must accept any
interference received, including interference that may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television reception, which
can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
—Consult the dealer or an experienced radio/TV technician for help.

IQM 60 V5 Pump Revision October 2010
4
Warranty
Thank you for purchasing this Aeroqual product. To get maximum use of the features of your new product we
recommend that you follow a few simple steps:
•Read the guidelines for safe and efficient use.
•Read all the terms and conditions of your Aeroqual Warranty.
oSave your original receipt. You will need it for warranty repair claims. Should your Aeroqual product
need warranty service, you should return it to the dealer from whom it was purchased or contact
Aeroqual.
Our Warranty
Aeroqual warrants this product to be free from defects in materials and workmanship for a period of 1 (one) year from the
date of shipment from the factory. Consumable items such as bearings, filters and gas scrubber media are excluded
from this warranty.
This warranty is expressly limited to the original owner who purchases the equipment directly from Aeroqual or from an
authorized Aeroqual dealer.
What we will do
If, during the warranty period, this product fails to operate under normal use and service, due to improper materials or
workmanship, Aeroqual subsidiaries, authorized distributors or authorized service partners will, at their option, either
repair or replace the product in accordance with the terms and conditions stipulated herein.
Conditions
1. The warranty is valid only if the original receipt issued to the original purchaser by the dealer, specifying the date of purchase, is
presented with the product to be repaired or replaced. Aeroqual reserves the right to refuse warranty service if this information has
been removed or changed after the original purchase of the product from the dealer.
2. If Aeroqual repairs or replaces the product, the repaired or replaced product shall be warranted for the remaining time of the
original warranty period or for ninety (90) days from the date of repair, whichever is longer. Repair or replacement may be via
functionally equivalent reconditioned units. Replaced faulty parts or components will become the property of Aeroqual.
3. This warranty does not cover any failure of the product due to normal wear and tear, damage, misuse, including but not limited to
use in any other than the normal and customary manner, in accordance with Aeroqual’s user guide for use, faulty installation,
calibration and maintenance of the product, accident, modification or adjustment, events beyond human control, improper
ventilation and damage resulting from liquid or corrosion.
4. This warranty does not cover product failures due to repairs, modifications or improper service performed by a non-Aeroqual
authorized service workshop or opening of the product by non-Aeroqual authorized persons.
5. The warranty does not cover product failures which have been caused by use of non-Aeroqual original accessories.
6. This warranty becomes void if a non-Aeroqual approved power supply is used.
7. Tampering with any part of the product will void the warranty.
8. Damage to the sensors can occur through exposure to certain sensor poisons such as silicones, tetraethyl lead, paints and
adhesives. Use of Aeroqual sensors in these environments containing these materials may (at the discretion of Aeroqual) void the
warranty on the sensor head. Exposure to gas concentrations outside of the design range of a specific Aeroqual sensor head can
adversely affect the calibration of that sensor head and will also void this warranty as it applies to the replacement of sensor
heads.
9. Aeroqual makes no other express warranties, whether written or oral, other than contained within this printed limited warranty. To
the fullest extent allowable by law all warranties implied by law, including without limitation the implied warranties of merchantability
and fitness for a particular purpose, are expressly excluded, and in no event shall Aeroqual be liable for incidental or consequential
damages of any nature whatsoever, however they arise, from the purchase or use of the product, and including but not limited to
lost profits or business loss.
10. Some countries restrict or do not allow the exclusion or limitation of incidental or consequential damage, or limitation of the
duration of implied warranties, so the preceding limitations or exclusions may not apply to you. This warranty gives you specific
legal rights, and you may also have other rights, which may vary from country to country.

IQM 60 V5 Pump Revision October 2010
5
For Your Safety
Read these simple guidelines. Ignoring these guidelines may be hazardous.
USE SENSIBLY
Use only as per this user guide.
USE AEROQUAL APPROVED SERVICE
Only approved service personnel must work on this product.
ACCESSORIES
Use only approved accessories. Do not connect incompatible products.
CONNECTING TO OTHER DEVICES
When connecting to any other device, read the appropriate user guide for detailed safety
instructions. Do not connect incompatible products.
HAZARDOUS ENVIRONMENTS
Do not use the monitor in or near volatile fuel or chemicals.
HEALTH AND SAFETY IN THE WORKPLACE
Aeroqual Monitors and Sensor Heads are used to monitor ambient gas concentrations. Aeroqual
does not guarantee user safety. In hazardous environments, an appropriate Health and Safety
plan should be in place.
WARNING Do not switch the monitor on before reading the User Guide.

IQM 60 V5 Pump Revision October 2010
6
1.0 Monitor and System Components
1.1 IQM 60
The Aeroqual IQM 60 Environmental Monitor is a custom-built indoor air quality instrument which can be
configured with up to six gas sensor modules as well as options such as a particle counter. The
configuration of the instrument supplied is determined by the sensor modules and options that were
ordered.
The IQM60 uses an internal pump to sample the air via the gas sensor inlet and the particle counter inlet if
fitted.
This manual describes the general use and operation of the IQM Environmental Monitor. The Product
Information Sheet supplied at the front of the folder will specify the “as-built” monitor and the packing list of
shipped items. Please check that all the components have been supplied as per your order and contact
your dealer or Aeroqual on email at: sales@aeroqual.com if any of the components are missing.
1.2 System Components include:
•IQM 60 Monitor
•12VDC Regulated Power Supply
•Temperature and humidity sensor extension cable
•Computer CD containing software for data logging and calibrating IQM
•Data Card (1GB)
•Data card reader
•Serial Cable
•Nafion humidifier
2.0 Description
NOTE 1. Insert the data card by pressing in gently until the card latches in place. To remove the data
card gently press inwards and then let go to release the latch. Always turn off the IQM 60
before removing or inserting the data card.
Gas sample inlet
SD data card
Power on/off
RS 232 serial port
Temperature and RH
sensor
Display
Particle counter inlet

IQM 60 V5 Pump Revision October 2010
7
2.1 Inlets
The IQM is supplied with an isokinetic probe for the particle counter inlet and a teflon filter for the gas
sensor inlet. These should be fitted before starting the IQM. There is also a zero filter supplied for zero
checking the particle counter.
isokinetic probe
for particle counter
zero particle filter
for zero checking
the particle
counter
inlet filter for gas
sensors

IQM 60 V5 Pump Revision October 2010
8
3.0 Operating Instructions
3.1 Getting Started
The IQM 60 is simple to operate:
•Ensure that the 12VDC power supply to the IQM 60 is connected (rear panel).
•Ensure the datacard is correctly inserted in the slot.
•Turn on the power switch (front panel).
•Remove the inlet cap and fit the supplied filter to the inlet
Note: The IQM 60 should be allowed to run for 60 minutes so that it is fully warmed up before taking
readings.
3.2 Connecting a computer
The connection of a computer to the IQM 60, the installation of the software and its use are described in
Appendix A.
3.3 Reading the Datacard
Data is stored on the data card as a daily file. The format of the data is a tab delimited file which can be
opened with standard graphical software such as MS Excel. Always turn off the IQM 60 before removing or
inserting the data card. To remove the data card gently press inwards and then let go to release the latch.
3.4 Testing
Zero and span checks should be performed with humidified air (detailed description provided in section 5).
Aeroqual supplies a Nafion gas humidifier. If the IQM is to be deployed in environments with low or high
absolute humidity it is recommended that the sensors be zero and span calibrated under those conditions,
since the sensors are slightly sensitive to changes in absolute humidity.
3.5 Maintenance
The IQM is designed to operate without maintenance for extended periods. The following table outlines a
periodic maintenance schedule for the IQM. This schedule is based on experience under normal operating
conditions and may need to be modified to suit specific operating conditions.
Interval
Item
Procedure
Weekly
Gas Sensor zero check
Particle Counter zero check
Perform using inbuilt Zero scrubber
Perform
1Month Inlet particulate filter
Particle Counter flow test
Check/replace
Perform
6Months Gas Sensor Span calibration
Perform
1Year Particle Counter Factory Service
Contact Aeroqual for instructions
3.6 Care and Cleaning
Keep the IQM dry. Avoid water and/or condensation as humidity and liquids containing minerals may
corrode electronic circuits.
Do not use or store in dusty, dirty areas.
Do not store or turn off the monitor in temperatures below 5°C.
This unit is designed for use at temperatures between 5°C and +45°C. If you wish to use the monitor
outside this range you will need an additional environmental enclosure. Sudden changes in
temperature will cause condensation that may damage the electronic components.
Do not drop, knock or shake as this could lead to internal damage.

IQM 60 V5 Pump Revision October 2010
9
4.0 Troubleshooting
4.1 Troubleshooting Guide
Symptom Possible Cause Fault isolation/Solution
Gas sensor readings incorrect Insufficient warm up
Incorrect zero calibration
Incorrect span calibration
Sensor module leaking
Sensor pump failed
Allow the sensors to fully warm up after power
down. This may require 60 minutes or more.
Repeat zero calibration with clean air. Check
media in Gas Treatment module.
Perform span check.
Check for leaks
Measure flow. If pump has failed replace.
NO2sensor reading very high
Incorrect zero calibration
O3Scrubber failed
Repeat zero calibration with clean air. Check
media in Gas Treatment module.
Challenge the NO2sensor with 0.1 ppm of O3.If
the response is large (>0.5 ppm) then the
scrubber has failed. Replace.
Gas Sensor readings noisy or
unstable
Leaks
Particulate filter
Leaks dilute the sample stream and can cause
low span readings and incorrect zero readings.
Perform a leak test.
Replace if dirty.
“NR” or “9999” in Data table
(No response)
Occasional NR or “9999” in Data
Table
No Response from Sensor module.
Not connected correctly.
Computer too slow
RF noise in environment.
Check the electrical connector on the sensor
module is firmly connected
Test with a faster computer.
Try to reduce RF noise
COMMUNICATION
No communication over RS232
RS232 cable disconnected
Control Module Failed
Reconnect cable
Replace Control Module
PC Data logging stopped IQM power blackout interrupted
data logging.
USB to serial hub not working
Close and restart IQM software.
Check USB connectors. Check the Moxa serial
hub is present on the device hardware menu in
the PC. If not reload driver software and re-
install.
No data on SD card
Card not correctly installed in slot.
Data logging interval set too large.
Data card module fault
Turn off IQM and install card in slot correctly.
Set logging interval in configuration to a lower
interval.
Send control module to factory for replacement
module.
RF communication fails RF modem incorrectly configured Follow modem manual instructions to
reconfigure RF modem

IQM 60 V5 Pump Revision October 2010
10
4.2 Diagnostics
The IQM has a number of sources for diagnostic information if a problem arises. These are described
below:
Event log A large number of instrument events are logged on the event log file which is
located on the SD data card. This can be downloaded remotely (if a modem is
fitted) and in many cases can be used to determine whether a site visit is required.
Event Code Examples
Meanin
g
O3 :P.F. 2008/11/12 23:45
Pump failure
O3 module pump failure on 12 Nov
2008 at 23:45
NO2 :S.F. 2008/11/12 23:45
Sensor failure
NO2 module sensor failure on 12 Nov
2008 at 23:45
Power on: 2008/11/12 10:12
AQM Power turned on at 10:12 am on
12 November 2008
Zero Cal at 2008/11/12 11:24
Zero Calibration started at 11:24 on
12 November 2008
Config at 2008/11/12 13:12
0x01,05,30,40,50,B0,F8,00,00,00,00,00,00,00,00,00,
00
AQM configuration saved at 1:12 pm
on 12 November 2008
Sensor codes listed
O3 : NR 2008/11/12 12:42
O3 module no data report at 12:42 pm
on 12 November 2008
NO2 Scrubber FAILED: 2008/11/12 15:36
NO2 scrubber failed at 3:36 pm on 12
November 2008
Messages
This is located on the AQM PC software tool bar. If the AQM is connected to the computer then event
messages will be written to the Messages window in real-time.
Diagnostics
This is located in the AQM PC software. If the AQM is connected to the computer then individual sensor
modules can be interrogated to determine if there is a problem not picked up by normal fault detection as
well as fine tune the sensor performance. Only qualified personnel should use this menu as incorrect use
may result in sensor malfunction. Please consult Aeroqual Technical Support to understand how to use this
feature.

IQM 60 V5 Pump Revision October 2010
11
5.0 Calibration
5.1 Gas Module Calibration
All the Gas modules in the IQM can be calibrated in the unit by applying certified gas to the IQM inlet. All
Zero and span calibrations or checks should be performed with humidified air and use single gas mixtures
in a balance of air. If the IQM is to be deployed in environments with very low or high absolute humidity it is
recommended that the sensors be zero and span calibrated under those conditions, since the sensors are
slightly sensitive to changes in absolute humidity.
Zero and span adjustments on the IQM60 V5 are performed by adjusting the OFFSET and GAIN values
under the Calibration menu. The equation that relates the OFFSET and GAIN to the Gas reading is given
below:
Gas Reading = GAIN x (Module Reading - OFFSET)
All the Gas modules in the IQM can be calibrated in the unit by applying certified gas to the IQM inlet.
All zero and span calibrations or checks should be performed with gas mixtures in a balance of
air. Note: the calibration standard may be in a balance of N2 but it should be diluted with air
prior to presenting to the IQM.
Fluorocarbon (PTFE, FEP or PFA) tubing and fittings should be used for all sample lines.
Warm up the IQM for at least 12 hours before attempting a gas sensor calibration
5.1.1 Nafion Humidifier
ANafion tubing membrane humidifier is supplied with the IQM which can be used to humidify the air
delivered to the IQM during calibrations. Two lengths of 1/4” Teflon tubes are supplied to enable
connection to sample lines and the IQM inlet.
Span gas and zero air is typically dry (~0%RH) and the sensors used in the IQM require 50 +/-20% RH for
best performance. The Nafion membrane is selectively permeable to water vapor (although some other
compounds such as ammonia, alcohols, acetone, and acetic acid may also be transported across the
membrane see www.permapure.com for further information) and the direction of transport is driven by the
concentration gradient across the tubing wall. The humidifier is designed to extract water vapor from the
surrounding ambient air and transfer this into the dry span or zero gas source. The resultant humidity in the
sample gas is roughly half of that of the ambient air. If the surrounding air humidity is less than 50% it is
recommended that the membrane is positioned above a water bath to increase the humidity in the sample
air. Also, Make sure that the air surrounding the membrane is not contaminated with hydrocarbons that
could be transported across the nafion to the sensors. If in doubt contact Aeroqual.
Nafion humidifier

IQM 60 V5 Pump Revision October 2010
12
5.1.2 Zero verification using Gas Treatment Module
The Gas Treatment Module contains a zero air scrubber which can be automatically or manually turned on
to periodically measure the zero of the instrument. The automatic zero verification is set in the
Configuration file under Zero Readings. This turns on the gas treatment module for 30 minutes during
which time the sensors continue to measure. This procedure is repeated at a user defined interval between
1and 255 hours. It is recommended that this procedure is used when making field measurements.
5.1.3 Zero calibration
5.1.3.1 Using Gas Treatment Module
You may use the zero air scrubber in the Gas Treatment Module to undertake a manual zero calibration.
Before doing this you should check that the media are in good condition or if possible replace with fresh
media.
1. Turn on the zero scrubber manually by clicking on PC software, Tools- >Turn On Zero Scrubber.
2. Allow the AQM to sample until stable readings are obtained (about 30 - 60 minutes).
3. Adjust the OFFSET values for each gas sensor via the Calibration --> Calibrate OFFSET-->enter
password “password”--> select gas--->enter new OFFSET
new OFFSET = old OFFSET + AQM60 gas reading
5.1.3.2 Laboratory Zero calibration
The IQM can be re-zeroed via the PC software using a zero air source with relative humidity of 50% +/-
20%. Aeroqual can supply a humidifier that uses a Nafion membrane to humidify the air.
1. Connect a source of zero air to the IQM through the sample port using ¼ inch Teflon tubing. Use a
Tfitting to ensure the IQM is sampling the gas flow at atmospheric pressure. The zero air should
be at least 1.5 LPM (check there is excess flow at the exhaust of the T fitting).
2. Allow the IQM to sample until stable readings are obtained (about 30-60 minutes).
3. Adjust the OFFSET values for each gas sensor via the Calibration --> Calibrate OFFSET-->enter
password “password”--> select gas--->enter new OFFSET
new OFFSET = old OFFSET + AQM60 gas reading
5.1.4 Span calibration
The sensors can be manually span calibrated by modifying the individual gain factors for each sensor.
Introduce a known concentration of span gas into the instrument and adjust the gain factor to set the
correct concentration reading. Note: a sensor zero point calibration should always be performed before
undertaking a span calibration.
5.1.4.1 Span gas
Span gas mixtures should be single gas mixtures that are certified and be in a balance of air.
DO NOT USE MIXTURES THAT CONTAIN MORE THAN ONE SPAN GAS COMPONENT
DO NOT USE A SPAN GAS IN A BALANCE OF NITROGEN UNLESS DILUTION IS GREATER
THAN 50 TIMES WITH ZERO AIR.
The mixture should also be humidified to about 50 +/-20% RH by either using a Nafion membrane
humidifier or by bubbling the zero grade air through water prior to mixing with the gas standard. Aeroqual
can supply a Nafion humidifier which is suitable for IQM units containing O3, NO2, CO, SO2 and VOC
sensor modules. Excessive gas concentrations should be avoided as these may damage the sensors or
cause contamination of the internal tubing. Recommended span points for gas sensors are given in the
table below.

IQM 60 V5 Pump Revision October 2010
13
Gas
Span Point / ppm
O3 0.100
NO2 0.100
CO 25
VOC (isobutylene) 10
NMHC (isobutylene) 10
CO2 1000
PID (isobutylene) 10
Certified calibration standards that can be used with the IQM include certified gas bottles, diluent
calibrators, or ozone calibrators.
5.1.4.2 Span calibration procedure
1. Perform zero calibration.
2. Connect a source of span calibration gas to the IQM through the sample port. Use a T fitting to
ensure the IQM is sampling the gas flow at atmospheric pressure. The span gas flow should be at
least 1.5 LPM (check there is excess flow at the exhaust of the T fitting).
3. Allow the IQM to sample the gas until a stable reading is obtained (about 30 minutes).
4. Adjust the IQM60 sensor gain factor to set the correct concentration reading
Calibration -->Calibrate Gain-->enter password “password”--> select gas--->enter new gain factor
new Gain factor = old Gain factor x Span gas concentration / AQM60 gas reading
Please note: The change is logged in the EVENTLOG file on the SD card.
Example of Span Calibration Equipment
Calibrator
Calibration
Gas zero
air Tfitting
exhaust
RS232 serial
cable
nafion
humidifier

IQM 60 V5 Pump Revision October 2010
14
5.2 Particle Counter Calibration
Flow rate
The sampling flow rate should be checked monthly and adjusted if required. To test the flow rate:
1. Remove the isokinetic probe and connect a flow meter to the barb fitting of the particle inlet using a short
piece of flexible 1/8 “ ID tubing, Tygon tubing is a good choice.
2. Turn on the IQM and note the flow meter reading. The flow rate should be 1.0 +/- 0.05 LPM
3. The flow rate can be adjusted by a screw flow adjuster located underneath the bottom panel of the IQM
(see below). To remove the bottom panel, remove the two screws located at the bottom rear corners of the
casing and the bottom panel lifts off. Use a screw diver to adjust the screw in the T piece fitting. Turn the
screw clockwise to decrease the flow and counter-clockwise to increase the flow.
Zero Check
It is recommended to periodically perform a zero count test as described below. False counts caused by air
leaks or spurious noise will cause errors that are especially apparent when sampling relatively clean
aerosols.
1. Attach a filter with a rating of 0.45µm or better to the particle inlet nozzle. The filter must be large enough
that it does not create a restriction and load down the small vacuum pump inside the IQM. One zero
particulate filter is supplied with the IQM.
2. Since the air passing through the IQM is now virtually particle free the output particle count should be
zero. If it is not, check for leaks around the tubing to the particle counter.
Span Calibration
There are no user serviceable components inside the particle counter instrument. The cover on the particle
counter should not be removed or opened for servicing, calibration or any other purpose except by a factory
authorized person. To do so will void warranty and may result in exposure to invisible laser radiation that
can cause blindness. Contact Aeroqual for service information. Calibrating particle sensors like the one in
the IQM requires specialized equipment and a skilled technician. The particle counter should be factory
calibrated on a yearly basis.
Particle Counter flow
adjustment screw

IQM 60 V5 Pump Revision October 2010
15
Appendix A AQM V5.x Software Instructions
Description
The AQM V5.x Software is designed to be a simple interface for communicating with the IQM monitor via a
PC. It can be used to configure the IQM, initiate the zero calibration routine, modify gain factors, poll data
and display the data in either table or graphical format. The software runs in Java VM1.5 (supplied with
software) and the database is an open source HSQLDB Java database. Data can be easily exported to a
spreadsheet for manipulation. The software also incorporates a GSM modem connection which can be
used to connect to and operate an IQM unit remotely.
The software is compatible with IQM firmware versions 4.0 and higher only.
Computer Requirements
•CD-ROM Drive
•RS232 port
•Windows OS version 2000 or later
•100 Mb of spare hard drive space
•recommended 256 Mb RAM
•recommended 1 GHz processor speed
Setting up
1. Install IQM CD into drive and extract the “Aeroqual AQM V5.x for Vista” or launch the
AQM_4.x.exe software if you have earlier versions of Windows OS.
2. Launch software
Click Setup---> COM Port.
Select correct COM port settings Port, 38400, 8, 1, N, N
Click OK
Passwords
The factory set passwords for the IQM are:
System password: password
Diagnostic password: george
Menu Descriptions
File
Search monitor
Searches for available IQM monitors.
Start Data Logging Starts data logging from the IQM monitor.
Stop Data Logging Stops data logging from the IQM monitor.
Export Logged Data Exports logged data from the IQM monitor.
Import Files Imports file (aql file).

IQM 60 V5 Pump Revision October 2010
16
Download Files Downloads files from the IQM; including daily files and event log (see
below)
Update sensor list updates the sensor list on the database.
Zip database zips the database.
Unzip database Unzips the database (prompted by a warning message). (See below)
Exit Ctrl X Exits the AQM software.

IQM 60 V5 Pump Revision October 2010
17
Setup
Configuration Ctrl P Configures the IQM (System password required). (See below)

IQM 60 V5 Pump Revision October 2010
18
Operations Changes the operation of the IQM. (See below)
Test Connection Tests the connection to the IQM monitor. (See below)

IQM 60 V5 Pump Revision October 2010
19
COM Port Ctrl M Configures the COM port for communicating with the IQM Monitor. (See below)
Data

IQM 60 V5 Pump Revision October 2010
20
Graph --> Real Time Graphs real time data from the IQM Monitor. (See below)
Graph --> Logged Data Graphs Logged data from the IQM Monitor. (See below)
Table of contents
Other Aeroqual Monitor manuals