Agilent Technologies MXE EMI Receiver N9038A User manual

Agilent X-Series
MXE EMI Receiver
N9038A MXE
Specifications Guide
(Comprehensive Reference Data)

Notices
© Agilent Technologies, Inc. 2013 No
part of this manual may be reproduced in
any form or by any means (including
electronic storage and retrieval or transla-
tion into a foreign language) without prior
agreement and written consent from Agi-
lent Technologies, Inc. as governed by
United States and international copyright
laws.
Manual Part Number
N9038-90010
Supersedes: October 2013
Print Date: November 2013
Printed in USA
Agilent Technologies, Inc.
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
The material contained in this doc-
ument is provided “as is,” and is
subject to being changed, without
notice, in future editions. Further,
to the maximum extent permitted
byapplicablelaw, Agilentdisclaims
all warranties, either express or
implied, with regard to this manual
and any information contained
herein, including but not limited to
the implied warranties of mer-
chantability and fitness for a par-
ticular purpose. Agilent shall not
be liable for errors or for incidental
or consequential damages in con-
nection with the furnishing, use, or
performance of this document or of
any information contained herein.
Should Agilent and the user have a
separate written agreement with
warranty terms covering the mate-
rial in this document that conflict
with these terms, the warranty
terms in the separate agreement
shall control.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance
of a U.S. Government prime contract or
subcontract, Software is delivered and
licensed as “Commercial computer soft-
ware” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S.
Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Gov-
ernment users will receive no greater than
Limited Rights as defined in FAR 52.227-
14 (June 1987) or DFAR 252.227-7015
(b)(2) (November 1995), as applicable in
any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly per-
formed or adhered to, could result
in damage to the product or loss of
important data. Do not proceed
beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indi-
cated conditions are fully
understood and met.
2

3
Warranty
This Agilent technologies instrument product is warranted against defects in material and workmanship for
a period of three years from the date of shipment. During the warranty period, Agilent Technologies will,
at its option, either repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilent
Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall
pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties,
and taxes for products returned to Agilent Technologies from another country.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about this receiver, including firmware
upgrades, application information, and product information, see the following URLs:
http://www.agilent.com/find/mxe
To receive the latest updates by email, subscribe to Agilent Email Updates:
http://www.agilent.com/find/emailupdates
Information on preventing receiver damage can be found at:
http://www.agilent.com/find/tips

4

Contents
5
1. Agilent MXE EMI Receiver
Definitions and Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Conditions Required to Meet Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Frequency and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Frequency Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Band Overlaps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Precision Frequency Reference. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Frequency Readout Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Frequency Span. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Triggers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Minimum settable level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Gated Sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Number of Frequency Sweep/Step Points (buckets). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Resolution Bandwidth (RBW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
RF Preselector Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Microwave
Preselector Bandwidth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Analysis Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Video Bandwidth (VBW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Amplitude Accuracy and Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Measurement Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Maximum Safe Input Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Display Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Marker Readout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Frequency Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
IF Frequency Response. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
IF Phase Linearity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Absolute Amplitude Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Input Attenuation Switching Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
RF Input VSWR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Resolution Bandwidth Switching Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Reference Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Total Measurement Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Display Scale Fidelity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Display Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Available Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Preamp Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Amplitude Probability Distribution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Dynamic Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Gain Compression. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
IF Prefilter Bandwidth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Displayed Average Noise Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Indicated Noise (Receiver) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
DANL and Indicated Noise Improvement with Noise Floor Extension. . . . . . . . . . . . . . . . . . . . . . . . . . 52

6
Contents
Spurious Responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Second Harmonic Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Third Order Intermodulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Phase Noise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Nominal Phase Noise at Different Center Frequencies [Plot] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Power Suite Measurements (Preselector off only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Channel Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Occupied Bandwidth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Adjacent Channel Power (ACP). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Fast ACPR Test [Plot]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Multi-Carrier Adjacent Channel Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Power Statistics CCDF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Burst Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
TOI (Third Order Intermodulation) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Harmonic Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Spurious Emissions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Spectrum Emission Mask . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Inputs/Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
2. I/Q Analyzer
Specifications Affected by I/Q Analyzer: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Clipping-to-Noise Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Data Acquisition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Time Record Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
ADC Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
3. Option TDS - Time Domain Scan
Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
4. Option B25 - 25 MHz Analysis Bandwidth
Specifications Affected by Analysis Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Other Analysis Bandwidth Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
IF Spurious Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
IF Frequency Response. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
IF Phase Linearity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Data Acquisition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Time Record Length (IQ pairs). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
ADC Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
5. Option CR3 - Connector Rear, 2nd IF Output
Specifications Affected by Connector Rear, 2nd IF Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

Contents
7
Other Connector Rear, 2nd IF Output Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Aux IF Out Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Second IF Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
6. Option ESC - External Source Control
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Frequency Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Span Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Dynamic Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Power Sweep Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Measurement Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Supported External Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
7. Option YAS - Y-Axis Screen Video Output
Specifications Affected by Y-Axis Screen Video Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Other Y-Axis Screen Video Output Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
General Port Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Continuity and Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
8. Analog Demodulation Measurement Application
Pre-Demodulation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Demodulation Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Capture Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Post-Demodulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Maximum Audio Frequency Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Filters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Frequency Modulation - Level and Carrier Metrics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
FM Deviation Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
FM Rate Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Carrier Frequency Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Carrier Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Frequency Modulation - Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Residual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Absolute Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
AM Rejection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Residual FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Measurement Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Amplitude Modulation - Level and Carrier Metrics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
AM Depth Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
AM Rate Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Carrier Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Amplitude Modulation - Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Residual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Absolute Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
FM Rejection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Residual AM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117

8
Contents
Measurement Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Phase Modulation - Level and Carrier Metrics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
PM Deviation Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
PM Rate Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Carrier Frequency Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Carrier Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Phase Modulation - Distortion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Residual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Absolute Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
AM Rejection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Measurement Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
9. Noise Figure Measurement Application
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Noise Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Gain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Noise Figure Uncertainty Calculator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
10. Phase Noise Measurement Application
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Maximum Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Measurement Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Offset Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Amplitude Repeatability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Nominal Phase Noise at Different Center Frequencies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128

9
1 Agilent MXE EMI Receiver
This chapter contains the specifications for the EMI receiver. The specifications and characteristics for
the measurement applications and options are covered in the chapters that follow.

10 Chapter1
Agilent MXE EMI Receiver
Definitions and Requirements
Definitions and Requirements
This book contains EMC receiver specifications and supplemental information. The distinction among
specifications, typical performance, and nominal values are described as follows.
Definitions
• Specifications describe the performance of parameters covered by the product warranty (temperature
= 5° to 50°C, unless otherwise noted).
• 95th percentile values indicate the breadth of the population (≈2σ) of performance tolerances
expected to be met in 95% of the cases with a 95%confidence, for any ambient temperature in the
range of 20 to 30°C. In addition to the statistical observations of a sample of instruments, these values
include the effects of the uncertainties of external calibration references. These values are not
warranted. These values are updated occasionally if a significant change in the statistically observed
behavior of production instruments is observed.
• Typical describes additional product performance information that is not covered by the product
warranty. It is performance beyond specification that 80%of the units exhibit with a 95%confidence
level over the temperature range 20 to 30°C. Typical performance does not include measurement
uncertainty.
• Nominal values indicate expected performance, or describe product performance that is useful in the
application of the product, but is not covered by the product warranty.
Conditions Required to Meet Specifications
The following conditions must be met for the receiver to meet its specifications.
• The receiver is within its calibration cycle. See the General section of this chapter.
• Under auto couple control, except that Auto Sweep Time Rules = Accy.
• For signal frequencies < 10 MHz, DC coupling applied.
• Any receiver that has been stored at a temperature range inside the allowed storage range but outside
the allowed operating range must be stored at an ambient temperature within the allowed operating
range for at least two hours before being turned on.
• The receiver has been turned on at least 30 minutes with Auto Align set to Normal, or if Auto Align
is set to Off or Partial, alignments must have been run recently enough to prevent an Alert message. If
the Alert condition is changed from “Time and Temperature” to one of the disabled duration choices,
the receiver may fail to meet specifications without informing the user.
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment
from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to
the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s
calibration facility, and to the calibration facilities of other International Standards Organization
members.

Chapter 1 11
Agilent MXE EMI Receiver
Frequency and Time
Frequency and Time
Description Specifications Supplemental
Information
Frequency Range
Maximum Frequency
RF Input 1
Option 508 8.4 GHz
Option 526 26.5 GHz
Option 544 44 GHz
RF Input 2 1.0 GHz
Minimum Frequency
Preselector Off AC Coupleda
a. AC Coupled only applicable to Freq Options 508 and 526.
DC Coupled
Preamp Off 10 MHz 20 Hz
Preamp On 10 MHz 100 kHz
Preselector On AC CoupledaDC Coupled
Preamp Off 10 MHz 20 Hz
Preamp On 10 MHz 1 kHz
Band Harmonic Mixing
Mode LO Multiple (Nb)
b. N is the LO multiplication factor. For negative mixing modes (as indicated by the “−” in the “Har-
monic Mixing Mode” column), the desired 1st LO harmonic is higher than the tuned frequency by the
1st IF (5.1225 GHz for band 0, 322.5 MHz for all other bands).
Band Overlapsc
0 (20 Hz to 3.6 GHz) 1−1Options 508, 526
1 (3.5 GHz to 8.4 GHz) 1−1Options 508, 526
2 (8.3 GHz to 13.6 GHz) 1−2Options 526
3 (13.5 to 17.1 GHz) 2−2Option 526
4 (17.0 to 26.5 GHz) 2−4Option 526
5 (26.4 to 34.5 GHz) 2−4Option 544
6 (34.4 to 44 GHz) 4−8Option 544

12 Chapter1
Agilent MXE EMI Receiver
Frequency and Time
c. In the band overlap regions, for example, 3.5 to 3.6 GHz, the receiver may use either band for measure-
ments, in this example Band 0 or Band 1. The receiver gives preference to the band with the better
overall specifications (which is the lower numbered band for all frequencies below 26 GHz), but will
choose the other band if doing so is necessary to achieve a sweep having minimum band crossings. For
example, with CF = 3.58 GHz, with a span of 40 MHz or less, the receiver uses Band 0, because the
stop frequency is 3.6 GHz or less, allowing a span without band crossings in the preferred band. If the
span is between 40 and 160 MHz, the receiver uses Band 1, because the start frequency is above
3.5 GHz, allowing the sweep to be done without a band crossing in Band 1, though the stop frequency
is above 3.6 GHz, preventing a Band 0 sweep without band crossing. With a span greater than
160 MHz, a band crossing will be required: the receiver scans up to 3.6 GHz in Band 0; then executes
a band crossing and continues the sweep in Band 1.
Specifications are given separately for each band in the band overlap regions. One of these specifica-
tions is for the preferred band, and one for the alternate band. Continuing with the example from the
previous paragraph (3.58 GHz), the preferred band is band 0 (indicated as frequencies under 3.6 GHz)
and the alternate band is band 1 (3.5 to 8.4 GHz). The specifications for the preferred band are war-
ranted. The specifications for the alternate band are not warranted in the band overlap region, but per-
formance is nominally the same as those warranted specifications in the rest of the band. Again, in this
example, consider a signal at 3.58 GHz. If the sweep has been configured so that the signal at
3.58 GHz is measured in Band 1, the analysis behavior is nominally as stated in the Band 1 specifica-
tion line (3.5 to 8.4 GHz) but is not warranted. If warranted performance is necessary for this signal,
the sweep should be reconfigured so that analysis occurs in Band 0. Another way to express this situa-
tion in this example Band 0/Band 1 crossing is this: The specifications given in the “Specifications”
column which are described as “3.5 to 8.4 GHz” represent nominal performance from 3.5 to 3.6 GHz,
and warranted performance from 3.6 to 8.4 GHz.

Chapter 1 13
Agilent MXE EMI Receiver
Frequency and Time
Description Specifications Supplemental
Information
Precision Frequency Reference
Accuracy ±[(time since last adjustment
×aging rate) + temperature
stability + calibration
accuracya]b
Temperature Stability
20 to 30°C±1.5 ×10−8
Full temperature range ±5 ×10−8
Aging Rate ±5 ×10−10/day (nominal)
Total Aging
1 Year ±1 ×10−7
2 Years ±1.5 ×10−7
Settability ±2 ×10−9
Warm-up and RetracecNominal
300 s after turn on ±1 ×10−7of final frequency
900 s after turn on ±1 ×10−8of final frequency
Achievable Initial Calibration Accuracyd±4 ×10−8
Standby power to reference oscillator Not supplied
Residual FM
(Center Frequency = 1 GHz
10 Hz RBW, 10 Hz VBW)
≤0.25 Hz ×Nep-p in 20 ms
(nominal)
a. Calibration accuracy depends on how accurately the frequency standard was adjusted to 10 MHz. If the
adjustment procedure is followed, the calibration accuracy is given by the specification “Achievable
Initial Calibration Accuracy.”
b. The specification applies after the receiver has been powered on for four hours.
c. Standby mode does not apply power to the oscillator. Therefore warm-up applies every time the power
is turned on. The warm-up reference is one hour after turning the power on. Retracing also occurs
every time warm-up occurs. The effect of retracing is included within the “Achievable Initial Calibra-
tion Accuracy” term of the Accuracy equation.
d. The achievable calibration accuracy at the beginning of the calibration cycle includes these effects:
1) Temperature difference between the calibration environment and the use environment
2) Orientation relative to the gravitation field changing between the calibration environment and the
use environment
3) Retrace effects in both the calibration environment and the use environment due to turning the
instrument power off.
4) Settability
e. N is the LO multiplication factor.

14 Chapter1
Agilent MXE EMI Receiver
Frequency and Time
Description Specifications Supplemental
Information
Frequency Readout Accuracy ±(marker freq ×freq ref accy. + 0.25%
×span + 5%×RBWa+ 2 Hz + 0.5 ×
horizontal resolutionb)
Single detector onlyc
Example for EMCd±0.0032%(nominal)
a. The warranted performance is only the sum of all errors under autocoupled conditions. Under
non-autocoupled conditions, the frequency readout accuracy will nominally meet the specification
equation, except for conditions in which the RBW term dominates, as explained in examples below.
The nominal RBW contribution to frequency readout accuracy is 2%of RBW for RBWs from 1 Hz to
390 kHz, 4%of RBW from 430 kHz through 3 MHz (the widest autocoupled RBW), and 30%of RBW
for the (manually selected) 4, 5, 6 and 8 MHz RBWs.
First example: a 120 MHz span, with autocoupled RBW. The autocoupled ratio of span to RBW is
106:1, so the RBW selected is 1.1 MHz. The 5%×RBW term contributes only 55 kHz to the total fre-
quency readout accuracy, compared to 300 kHz for the 0.0.25%×span term, for a total of 355 kHz. In
this example, if an instrument had an unusually high RBW centering error of 7%of RBW (77 kHz) and
a span error of 0.20%of span (240 kHz), the total actual error (317 kHz) would still meet the computed
specification (355 kHz).
Second example: a 20 MHz span, with a 4 MHz RBW. The specification equation does not apply
because the Span: RBW ratio is not autocoupled. If the equation did apply, it would allow 50 kHz of
error (0.25%) due to the span and 200 kHz error (5%) due to the RBW. For this non-autocoupled RBW,
the RBW error is nominally 30%, or 1200 kHz.
b. Horizontal resolution is due to the marker reading out one of the sweep points. The points are spaced
by span/(Npts –1), where Npts is the number of sweep points. For example, with the factory preset
value of 1001 sweep points, the horizontal resolution is span/1000. However, there is an exception:
When both the detector mode is “normal” and the span > 0.25 ×(Npts –1) ×RBW, peaks can occur
only in even-numbered points, so the effective horizontal resolution becomes doubled, or span/500 for
the factory preset case. When the RBW is autocoupled and there are 1001 sweep points, that exception
occurs only for spans > 750 MHz.
c. Specifications apply to traces in most cases, but there are exceptions. Specifications always apply to the
peak detector. Specifications apply when only one detector is in use and all active traces are set to Clear
Write. Specifications also apply when only one detector is in use in all active traces and the "Restart"
key has been pressed since any change from the use of multiple detectors to a single detector. In other
cases, such as when multiple simultaneous detectors are in use, additional errors of 0.5, 1.0 or 1.5
sweep points will occur in some detectors, depending on the combination of detectors in use.
d. In most cases, the frequency readout accuracy of the receiver can be exceptionally good. As an exam-
ple, Agilent has characterized the accuracy of a span commonly used for Electro-Magnetic Compatibil-
ity (EMC) testing using a source frequency locked to the receiver. Ideally, this sweep would include
EMC bands C and D and thus sweep from 30 to 1000 MHz. Ideally, the analysis bandwidth would be
120 kHz at −6 dB, and the spacing of the points would be half of this (60 kHz). With a start frequency
of 30 MHz and a stop frequency of 1000.2 MHz and a total of 16168 points, the spacing of points is
ideal. The detector used was the Peak detector. The accuracy of frequency readout of all the points
tested in this span was with ±0.0032%of the span. A perfect receiver with this many points would have
an accuracy of ±0.0031%of span. Thus, even with this large number of display points, the errors in
excess of the bucket quantization limitation were negligible.

Chapter 1 15
Agilent MXE EMI Receiver
Frequency and Time
Description Specifications Supplemental
Information
Frequency CounteraSee noteb
Count Accuracy ±(marker freq ×freq ref accy. + 0.100 Hz)
Delta Count Accuracy ±(delta freq. ×freq ref accy. + 0.141 Hz)
Resolution 0.001 Hz
a. Instrument conditions: RBW = 1 kHz, gate time = auto (100 ms), S/N ≥50 dB, frequency = 1 GHz
b. If the signal being measured is locked to the same frequency reference as the receiver, the specified
count accuracy is ±0.100 Hz under the test conditions of footnote a. This error is a noisiness of the
result. It will increase with noisy sources, wider RBWs, lower S/N ratios, and source frequencies >
1 GHz.
Description Specifications Supplemental
Information
Frequency Span
Range
Swept and FFT
Option 508 0 Hz, 10 Hz to 8.4 GHz
Option 526 0 Hz, 10 Hz to 26.5 GHz
Option 544 0 Hz, 10 Hz to 44 GHz
Resolution 2 Hz
Span Accuracy
Stepped ±(0.25%×span + horizontal resolutiona)
Swept ±(0.25%×span + horizontal resolutiona)
FFT ±(0.1%×span + horizontal resolutiona)
a. Horizontal resolution is due to the marker reading out one of the sweep points. The points are spaced
by span/(Npts −1), where Npts is the number of sweep points. For example, with the factory preset
value of 1001 sweep points, the horizontal resolution is span/1000. However, there is an exception:
When both the detector mode is “normal” and the span > 0.25 ×(Npts −1) ×RBW, peaks can occur
only in even-numbered points, so the effective horizontal resolution becomes doubled, or span/500 for
the factory preset case. When the RBW is auto coupled and there are 1001 sweep points, that exception
occurs only for spans > 750 MHz.

16 Chapter1
Agilent MXE EMI Receiver
Frequency and Time
Description Specifications Supplemental Information
Sweep Time
Range
Span = 0 Hz
Span ≥10 Hz
1 μs to 6000 s
1 ms to 4000 s
Accuracy
Span ≥10 Hz, swept
Span ≥10 Hz, FFT
Span = 0 Hz
±0.01%(nominal)
±40%(nominal)
±0.01%(nominal)
Sweep Trigger Free Run, Line, Video,
External 1, External 2, RF
Burst, Periodic Timer
Delayed Triggera
Range
Span ≥10 Hz, swept 0 to 500 ms
Span = 0 Hz or FFT −150 ms to +500 ms
Resolution 0.1 μs
a. Delayed trigger is available with line, video, RF burst and external triggers.

Chapter 1 17
Agilent MXE EMI Receiver
Frequency and Time
Description Specifications Supplemental Information
Triggers Additional information on some of the triggers
and gate sources
Video Independent of Display Scaling and Reference
Level
Minimum settable level −170 dBm Useful range limited by noise
Maximum usable level Highest allowed mixer levela+ 2 dB (nominal)
Detector and Sweep Type
relationships
Sweep Type = Swept
Detector = Normal, Peak,
Sample or Negative Peak Triggers on the signal before detection, which is
similar to the displayed signal
Detector = Average Triggers on the signal before detection, but with
a single-pole filter added to give similar
smoothing to that of the average detector
Sweep Type = FFT Triggers on the signal envelope in a bandwidth
wider than the FFT width
RF Burst
Level Range −50bto −10 dBm plus attenuation (nominal)
Level Accuracy ±2 dB + Absolute Amplitude Accuracy
(nominal)
Bandwidth (−10 dB)
Most cases 16 MHz (nominal)
Sweep Type = FFT;
FFT Width = 25 MHz;
Span ≥8 MHz
30 MHz (nominal)
Frequency Limitations If the start or center frequency is too close to
zero, LO feedthrough can degrade or prevent
triggering. How close is too close depends on
the bandwidth listed above.
External Triggers See “Trigger Inputs (Trigger 1 In, Trigger 2 In)”
on page 77
a. The highest allowed mixer level depends on the IF Gain. It is nominally –10 dBm for Preamp Off and
IF Gain = Low.
b. Noise will limit trigger level range at high frequencies, such as above 15 GHz.

18 Chapter1
Agilent MXE EMI Receiver
Frequency and Time
Description Specifications Supplemental Information
Gated Sweep
Gate Methods Gated LO
Gated Video
Gated FFT
Span Range Any span
Gate Delay Range 0 to 100.0 s
Gate Delay Settability 4 digits, ≥100 ns
Gate Delay Jitter 33.3 ns p-p (nominal)
Gate Length Range
(Except Method = FFT) 100 ns to 5.0 s Gate length for the FFT method is fixed at
1.83/RBW, with nominally 2% tolerance.
Gated Frequency and
Amplitude Errors Nominally no additional error for gated
measurements when the Gate Delay is
greater than the MIN FAST setting
Gate Sources External 1
External 2
Line
RF Burst
Periodic
Pos or neg edge triggered
Description Specifications Supplemental Information
Number of Frequency Sweep/Step
Points (buckets)
Factory preset 1001
Range 1 to 500,001 Zero and non-zero spans

Chapter 1 19
Agilent MXE EMI Receiver
Frequency and Time
Description Specifications Supplemental
Information
Resolution Bandwidth (RBW)
Range (−3.01 dB bandwidth) 1 Hz to 8 MHz
Bandwidths above 3 MHz are 4, 5,
6, and 8 MHz.
Bandwidths 1 Hz to 3 MHz are
spaced at 10%spacing using the
E24 series 24 per decade: 1.0, 1.1,
1.2, 1.3, 1.5, 1.6, 1.8, 2.0, 2.2, 2.4,
2.7, 3.0, 3.3, 3.6, 3.9, 4.3, 4.7, 5.1,
5.6, 6.2, 6.8, 7.5, 8.2, 9.1 in each
decade.
CISPR Standard Bandwidths 200 Hz, 9 kHz, 120 kHz, 1 MHz –6 dB, subject to CISPR
mask
MIL Standard Bandwidths 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100
kHz, 1 MHz –6 dB
Power bandwidth accuracya
RBW Range CF Range
1 Hz to 750 kHz All ±1.0%(0.044 dB)
820 kHz to 1.2 MHz <3.6 GHz ±2.0%(0.088 dB)
1.3 to 2.0 MHz <3.6 GHz ±0.07 dB (nominal)
2.2 to 3 MHz <3.6 GHz ±0.15 dB (nominal)
4 to 8 MHz <3.6 GHz ±0.25 dB (nominal)
Accuracy (−3.01 dB bandwidth)b
1 Hz to 1.3 MHz RBW ±2%(nominal)
1.5 MHz to 3 MHz RBW
CF ≤3.6 GHz
CF > 3.6 GHz
±7%(nominal)
±8%(nominal)
4 MHz to 8 MHz RBW
CF ≤3.6 GHz
CF > 3.6 GHz
±15%(nominal)
±20%(nominal)
Selectivity (−60 dB/−3 dB) 4.1:1 (nominal)
a. The noise marker, band power marker, channel power and ACP all compute their results using the
power bandwidth of the RBW used for the measurement. Power bandwidth accuracy is the power
uncertainty in the results of these measurements due only to bandwidth-related errors. (The receiver
knows this power bandwidth for each RBW with greater accuracy than the RBW width itself, and can
therefore achieve lower errors.) The warranted specifications shown apply to the Gaussian RBW filters
used in swept and zero span analysis. There are four different kinds of filters used in the receiver:
Swept Gaussian, Swept Flattop, FFT Gaussian and FFT Flattop. While the warranted performance only
applies to the swept Gaussian filters, because only they are kept under statistical process control, the
other filters nominally have the same performance.

20 Chapter1
Agilent MXE EMI Receiver
Frequency and Time
b. Resolution Bandwidth Accuracy can be observed at slower sweep times than auto-coupled conditions.
Normal sweep rates cause the shape of the RBW filter displayed on the receiver screen to widen by
nominally 6%. This widening declines to 0.6%nominal when the Swp Time Rules key is set to Accu-
racy instead of Normal. The true bandwidth, which determines the response to impulsive signals and
noise-like signals, is not affected by the sweep rate.
Description Specifications Supplemental Information
RF Preselector Filters
Filter Band Filter Type 6 dB Bandwidth (Nominal)
20 Hz to 150 kHz Fixed lowpass 310 kHz
150 kHz to 1 MHz Fixed bandpass 1.7 MHz
1 to 2 MHz Fixed bandpass 2.4 MHz
2 to 5 MHz Fixed bandpass 7.5 MHz
5 to 8 MHz Fixed bandpass 10 MHz
8 to 11 MHz Fixed bandpass 9.5 MHz
11 to 14 MHz Fixed bandpass 9.5 MHz
14 to 17 MHz Fixed bandpass 10 MHz
17 to 20 MHz Fixed bandpass 9.5 MHz
20 to 24 MHz Fixed bandpass 9.5 MHz
24 to 30 MHz Fixed bandpass 9.0 MHz
30 to 70 MHz Tracking bandpass 10 MHz
70 to 150 MHz Tracking bandpass 24 MHz
150 to 300 MHz Tracking bandpass 28 MHz
300 to 600 MHz Tracking bandpass 50 MHz
600 MHz to 1 GHz Tracking bandpass 60 MHz
1 to 2 GHz Tracking bandpass 180 MHz
2 to 3.6 GHz Fixed highpass 1.89 GHz (–3 dB corner frequency)
Other manuals for MXE EMI Receiver N9038A
3
Table of contents
Other Agilent Technologies Receiver manuals

Agilent Technologies
Agilent Technologies MXE EMI Receiver N9038A User manual

Agilent Technologies
Agilent Technologies 83434A User manual

Agilent Technologies
Agilent Technologies MXE EMI Receiver N9038A User manual

Agilent Technologies
Agilent Technologies MXE EMI Receiver N9038A User manual

Agilent Technologies
Agilent Technologies N5264A User manual

Agilent Technologies
Agilent Technologies 83446A Manual