5. Unplug the ethernet and power supply cables
from the particulate sensor module PCB.
When replacing the particulate sensor module, complete
these steps in reverse.
To confirm that the instrument is not leaking after the
particulate sensor module replacement, check that the flow
rate is still within the required range before covering the
inlet with a finger. If there is a build-up of pressure
from blocking the end, the leak check has passed.
PARTICULATE SENSOR MODULE
REMOVAL & REPLACEMENT