Allied AutoVent 2000 User manual

OPERATING
MANUAL
Version
1.5
AutoVent
ea-
AE
CCR
CR
ZA
人
本
本
4
Time
cycied,
constant
flow,
gos
powcred
:
ventilator
for
Emergency
Medical
Service.
>
É

AutoVent55S2/3S52
A
TABLE
OF
CONTENTS
Section
Page
Number
INTRODUCTION
1
GENERAL
DESCRIPTION
1
USE OF
THE
OPERATING
MANUAL
2
Design
Explanation
of
Warnings/Cautions/Notes
PERFORMANCE
CHARACTERISTICS
AND
FEATURES
3
PATIENT
VALVE
ASSEMBLY
3
Visual
Indicator
Pressure
Limit
Alarm
Module
Source
Gas
Inlets
Exhalation
Valve
CONTROL
MODULE
5
Source
Gas
Inlet
Patient
Valve
Outlets
Breaths
Per
Minute
(BPM)
Tidal
Volume
(V+)
Inspiratory
Time
(Ti)
Patient
Valve
Supply
Tubing
and
Oxygen
Line
Non-Rebreathing
Valve
and
Corrugated
Hose
UNPACKING
AND
INSPECTION
OF
THE
8
AUTOVENT
2000/3000
OPERATING
INSTRUCTIONS
8
Instructions
for
Use
of
the
AutoVent
2000/3000
with
Oxygen
Cylinders
Regulator
Attachment
Control
Module
Attachment
Instructions
for
Use
of
the
AutoVent
2000/3000
with
a
Wall
Outlet
Employing
a
Quick
Connection
Adapter
Instructions
for
use
of
the
AutoVent
2000/3000
with
an
Air/Oxygen
Blender
Ventilator
Check
out
PATIENT
USE
12
MAINTENANCE
OF
THE
LSP
AUTOVENT
2000/3000
14
Cleaning
and
Disinfecting
Equipment
Cleaning
and
Disinfecting
the
Control
Module
Cleaning
and
Disinfecting
the
Patient
Valve
Assembly
Cleaning
the
Non-Rebreathing
Valve
STORAGE
OF
THE
LSP
AUTOVENT
2000/3000
16
TROUBLESHOOTING
GUIDE
FOR
THE
16
LSP
AUTOVENT
2000/3000
WARRANTY
INFORMATION
18
APPENDIX
A:
Support
Equipment
20
B:
Patient
Valve
Specifications
C:
Control
Module
Specifications
D:
Oxygen
Pressure
Re
ulator
Specifications
E:
AutoVent
2000/3000
Altitude
Conversion
Chart
F:
AutoVent
2000/3000
Oxygen
Cylinder
Depletion Chart

AutoVe:
INTRODUCTION
THE
AUTOVENT
2000/3000
GENERAL
DESCRIPTION
Long
an
innovator
in
emergency
and
trauma
medicine,
Life
Support
Products
has
been
a
primary
manufacturer
of
oxygen
delivery
devices
for
field
use
in
the
United
States
and
worldwide
since
1979.
LSP
is
also
a
manufacturer
of
a
wide
range
of
trauma,
burn
and
rescue
devices.
Specifications
for
the
LSP
AutoVent
2000/3000,
including
the
Patient
Valve
Assembly,
Control
Module,
and
Regulators
intended
for
use
with
this
device,
are
included
in
the
Appendix
of
this
manual.
The
Life
Support
Products
AutoVent
2000/3000
represent
a
major
breakthrough
in
pneumatic
technology.
They
are
intended
for
the
ventilatory
assistance
of
patients
following
cardiac
arrest,
near
drowning,
trauma,
paramedical
transport,
and
other
circumstances
requiring
ventilatory
assistance.
The
LSP
Automatic
Ventilator
(AutoVent
2000/3000)
time-cycled,
constantflow,
gas-powered
ventilators
offer
controlled
ventilation
at
rates
from
8
to
20
breaths
per
minute
(BPM)
in
the
AutoVent
2000
version,
and
8
to
27
breaths
per
minute
in
the
AutoVent
3000
version.
The
attached
Patient
Valve
Assembly
allows
a
patient
to
draw
supplemental
gas
flow
(up
to
48
LPM)
with
spontaneous
effort.
Designed
for
transport
and
emergency
medical
use,
the
AutoVent
2000
delivers
from
400
ml
to
1200
ml
volume.
The
AutoVent
3000
delivers
from
200
to
1200
ml
in
volume.
Both
AutoVents
deliver
peak
pressures
up
to
60+5
em
H20
at
flow
flow
rates
from
16
to
48
liters
per
minute
(LPM).
They
are
small,
compact
units
ideally
suited
for
emergency
and
transport
situations
at
temperature
extremes
from
O°F
to
125°F.
Operating
power
is
obtained
from
standard
50
psi
source
gas.
They
are
simple
to
assemble
and
operate,
and
their
functions
are
easily
understood.
The
Ventilators
meet
or
exceed
the
American
Heart
Association
(AHA)
guidelines
for
resuscitation.
An
Audible
alarm
sounds
whenever
ventilatory
pressures
approach
the
preset
pressure
limit,
alerting
the
operator.
This
alarm
will
continue
to
sound
until
the
airway
pressure
drops
or
the
system
cycles
to
the
expiratory
phase.
WARNING:
Use
only
as
directed.
Improper
usage
or
unauthorized
modification
of
this
product
may
result
in
user
or
patient
injury.

AutoVent2SSS/200€
USE
OF
THE
OPERATING
MANUAL
Explanation
of
Warnings/Cautions/
Notes
FIGURE
1.
AUTOVENT2000/3000
This
operating
manual
for
the
Life
Support
Products
AutoVent
2000/3000
Automatic
Ventilators
has
been
designed
for
ease
of
use
in
the
paramedical
and
field
transport
setting.
Illustrations
are
provided
throughout
this
manual
to
provide
the
user with
both
a
graphic
and
narrative
description
of
the
ventilators’
operating
features.
This
operating
manual
contains
three
(3)
types
of
statements
with
which
the
user
should
be
aware,
and
are
defined
as
followings:
WARNING:
Potential
injury
to
the
patient
or
operator.
These
are
always
in
boxes
throughout!
the
text
of
the
manual.
CAUTION:
Potential
damage
to
the
ventilator,
breathing
circuit,
and/or
other
equipment
may
result.
These
are
always
in
brackets
throughout
the
text
of
the
manual.
**NOTE:
An
item
of
special
interest
concerning
the
use
and
operation
of
the
device(s)
or
feature
(s)
being
discussed
is
high-lighted
to
note
ease
of
use
or
understanding.
These
are
always
preceded
and
followed
by
asterisks
throughout
the
text
of
the
manual.
**
WARNINGS
AND
CAUTIONS
SHOULD
BE
READ
PRIOR
TO
OPERATING
THE
LSP
AUTOVENT
2000/3000.

AutoVentZSS5/SSSS
PERFORMANCE
CHARACTERISTICS
AND
FEATURES
PATIENT
VALVE
ASSEMBLY
Visual
Indicator
The
AutoVent
2000/3000
are
time
cycled,
constant
flow,
gas
powered
ventilators.
This
feature
allows
the
ventilators’
automatic
rate
to
be
set
by
the
operator
from
8
to
20
breaths
per
minute
(BPM),
in
the
AutoVent
2000
version,
and
8
to
27
BPM
in
the
AutoVent
3000
version.
Should
the
patient
require
additional
breaths,
these
can
be
obtained
on
demand
by
making
an
inspiratory
effort
on
the
patient
valve.
compatible
with
masks
and
other
airway
devices.
The
patient
valve
allows
spontaneous
breathing
The
patient
valve
delivers
both
controlled
and
spontaneous
breaths
to
the
patient.
It
consists
of
a
Figure
2
indicates
the
individual
components
which
make
up
the
Patient
Valve
Assembly.
demand
valve,
a
visual
indicator,
pressure
limit
alarm,
and
exhalation
upon
demand
if
the
patient
makes
an
inspiratory
effort
**NOTE:
Since
the
patient
valve
and
control
module
of
-2
cm.
HO.
(The
BPM
control
may
be
turned
to
the
"O"
position
if
desired
for
this
purpose.)
are
a
matched
set
with
identical
serial
numbers,
do
not
separate.
If
used
with
other
units,
setting
may
not
be
accurate.
**
valve.
It
has
a
standard
15mm.
inside
diameter/22mm.
outside
diameter
adapter,
which
is
____—
VISUAL
INDICATOR
NI
WI
<
X
SOURCE
GAS
INLET
hy
PRESSURE
LIMIT
ALARM
TOP
VIEW
OUTLET
ADAPTER
(EXHALATION
VALVE
INSIDE)
SIDE
VIEW
FIGURE
2.
PATIENT
VALVE
ASSEMBLY
The
visual
indicator
is
located
on
the
top
of
the
Patient
Valve
Assembly.
This
indicator
displays
bright
green
as
y
gas
flows
during
inspiration.
During
expiration,
the
il
indicator
dome
is
clear.
/
п
[
FIGURE
3
VISUAL
INDICATOR

AutoVent5222/5522
Pressure
Limit
Alarm
Module
An
audible
pressure
limit
alarm
is
located
in
the
Patient
Valve
Assembly.
This
alarm
sounds
whenever
the
patient
airway
pressure
approaches
the
designed
pressure
limit.
The
Pressure
Limit
Alarm
will
continue
to
sound
during
the
inspiratory
phase
until
either
the
airway
pressure
decreases
or
the
ventilator
cycles
off
to
begin
the
expiratory
phase.
In
addition
to
functioning
as
a
high
pressure
alarm
the
alarm
module
also
provides
additional
air
entrainment
during
the
Intermittent
Mandatory
Ventilation
(IMV)
mode
should
the
patient's
inspiratory
flow
rate
exceed
the
flow
delivery
from
the
control
module.
The
entrainment
of
ambient
air
occurs
through
the
blue
rubber
diaphragm
located
on
the
side
of
the
alarm
module.
WARNING:
If
the
maximum
pressure
limit
is
reached,
the
pre-set
tidal
volume
may
not
be
delivered
to
the
patient.
Inspiratory
time
will
remain
constant,
however,
and
an
inspiratory
hold
will
be
maintained
with
no
additional
volume
being
delivered
until
the
ventilator
cycles
to
the
expiratory
phase.
This
warning
also
appears
under
Tidal
Volume
in
the
Performance
Characteristics.
WARNING:
Should
the
blue
rubber
diaphragm
blow
outward
from
the
alarm
module's
air
entrainment
ports,
remove
the
AutoVent
immediately
from
service,
and
contact
your
LSP
distributor.
Source
Gas
Inlets
inlets
connect
the
Patient
Valve
with
the
Control
Module.
The
top
inlet:
(1)
Supplies
the
actuator
assembly
and
is
a
nippled
connector.
The
bottom
inlet:
(2)
Supplies
source
gas
to
the
patient
at
a
constant
flow
and
is
a
diameter
index
safety
system
(DISS)
oxygen
connector.
Located
on
the
side
of
the
Patient
Valve
Assembly,
the
PRESSURE
LIMIT
ALARM
FIGURE
4.
FIGURE
5.

AutoVent20
99/3000
Exhalation
Valve
CONTROL
MODULE
The
exhalation
valve
is
an
internal
diaphragm
located
on
PATIENT
VALVE
the
inside
of
the
Patient
Valve
Assembly.
The
exhalation
/
valve
allows
the
patient
to
exhale through
the
Patient
Valve
Assembly
once
the
inspiratory
cycle
is
completed,
whether
ventilator-controlled,
or
on
demand.
This
valve
can
be
accessed
by
removing
the
outlet
adapter
on
the
Patient
Valve
Assembly.
The
Control
Modules
for
the
AutoVent
2000/3000
are
designed
to
be
compact,
durable
and
easy
to
use.
The
units
are
constructed
to
perform
in
the
difficult
environments
of
paramedic
or
transport
operations.
Their
features
include
an
impactresistant
case
with
shock
absorbing
bumpers
and
easy
to
read
controls
for
independently
setting
Breaths
Per
Minute
(BPM),
Tidal
Volume
and
Inspiratory
Time
(AutoVent
3000
only).
The reverse
side
of
the
case
also
has
simplified
operating
instructions
for
ease
of
operation.
Figure
7
indicates
the
individual
components
which
make
up
the
Control
Module.
A
PRESSURE
LIMIT
ALARM
EXHALATION
VALVE
OUTLET
ADAPTER
—
——
FIGURE
6.
TIDAL
VOLUME
SOURCE
GAS
INLET
>
PATIENT
VALVE
OUTLETS
M
AUTOVENT
2000
BREATHS
PER
MINUTE
CONTROL
AUTOVENT
3000
INSPIRATORY
TIME
CONTROL
FIGURE
7.
CONTROL
MODULE

AutoVent2522/552S
Source
Gas
Inlet
Patient
Valve
Outlets
Breaths
Per
Minute
(BPM)
Tidal
Volume
(Vi)
located
on
the
left
side
of
the
Control
Module
and
marked
with
an
arrow,
the
source
gas
inlet
is
a
standard
diameter
index
safety
system
(DISS)
male
oxygen
connector.
When
a
high
pressure
line
is
attached
to
a
50
psi
source
gas
from
a
cylinder
or
bulk
oxygen
source,
this
gas
is
delivered
to
the
Control
Module
and
cycled
for
delivery
to
the
Patient
Valve
Assembly.
An
Air/Oxygen
Blender
can
also
be
inserted
in
line
with
this
system,
between
the
source
and
the
ventilator,
to
deliver
a
specified
oxygen
concentration.
**NOTE:
Use
of
an
Oxygen
Analyzer
is
recommended
prior
to
patient
use
in
order
to
accurately
measure
the
desired
oxygen
concentration
to
be
delivered.
**
located
on
the
right
side
of
the
Control
Module
body
and
marked
with
arrows,
the
patient
valve
outlets
connect
the
Control
Module
with
the
Patient
Valve
Assembly.
The
top
outlet
(1)
Supplies
source
gas
to
the
Patient
Valve
Assembly
at
a
constant
flow
and
is
a
diameter
index
safety
system
(DISS)
oxygen
connector.
The
bottom
outlet:
(2)
Supplies
the
Patient
Valve
Actuator
and
is
a
unique
7/16
inch
threaded
female
connecior.
This
control
sets
the
ventilator
rate
from
8
to
20
BPM
in
the
AutoVent
2000
version,
and
from
8
to
27
BPM
in
the
AutoVent
3000
version.
Adjusting
the
knob
clockwise
decreases
the
breathing
rate
and
adjusting
counter
clockwise
increases
the
rate.
Source
gas
is
available
on
demand,
even
in
the
"O"
position
up
to
48
LPM
depending
on
Tidal
Volume
setting,
from
the
Patient
Valve
Assembly
to
allow
the
patient
to
breathe
spontaneously.
SOURCE
GAS
INLET
FIGURE
8.
FIGURE
9.
FIGURE
10.
AUTOVENT
2000
This
control
adjusts
the
volume
available
to
the
patient
during
a
breath
and
is
adjustable
from
400
ml
to
1200
ml
in
the
AutoVent
2000
version,
and
from
200
to
1200
ml
in
the
AutoVent
3000
version.
Turning
the
knob clockwise
increases
tidal
volume.
Turning
the
knob
counter
clockwise
decreases
tidal
volume.
Following
a
volume
adjustment
change,
the
tidal
volume
stabilizes
after
one
breath
and
remains
constant.
WARNING:
If
the
maximum
pressure
limit
is
reached,
the
pre-set
tidal
volume
may
not be
delivered
to
the
patient.
Inspiratory
time
will
remain
constant,
however,
and
an
inspiratory
hold
will
be
maintained
with
no
additional
volume
being
delivered
until
the
ventilator
cycles
to
the
expiratory
phase.
This
warning
also
appears
under
Pressure
Limit
Alarm
Module
in
the
Performance
Characteristics.
AUTOVENT
3000
FIGURE
11.
AUTOVENT
2000
AUTOVENT
3000

AutoVent2552/5555
Tidal
Volume
(Vi)
(Continued)
Inspiratory
Time
(Ti)
Patient
Valve
Supply
Tubing
and
Oxygen
Line
**NOTE:
It
is
recommended
that
you
periodically
check
the
performance
characteristics
of
the
AutoVent
2000/3000
during
maintenance
by
placing
a
pressure
manometer
in
line
with
the
patient
circuit
near
the
outlet
to
verify
inspiratory
pressures
and
the
accuracy
of
the
pressure alarm
limit.**
This
control
knob
in
the
center position
of
the
AutoVent
3000
allows
adjustment
of
the
patient's
inspiratory
time.
The
two
settings
allow
selection
of
Adult
and
Child
inspiratory
time
respectively.
The
inspiratory
time
for
the
circled
orange
child
setting
is
approximately
.75
seconds
and
when
selected
corresponds
fo
the
circled
orange
settings
on
the
BPM
and
Tidal
Volume
control
knobs.
The
inspiratory
time
for
the
white
Adult
setting
is
approximately
1.5
seconds
and
when
selected
corresponds
to
the
white
settings
on
the
BPM
and
Tidal
Volume
control
knobs.
The
AutoVent
2000
has
a
pre-set
inspiratory
time
of
approximately
1.5
seconds.
either
of
the
end
stops.
CAUTION:
When
you
select
either
Adult
or
Child
setting,
rotate
the
center
control
knob
to
the
appropriate
setting
and
position
it
against
C
^
FIGURE
11A.
li
Tidal
Volume
settings.
WARNING:
Should
the
inspiratory
time
control
knob
on
the
AutoVent
3000
be
adjusted
after
initial
setup,
it
will
alter
the
patient's
BPM
and
The
patient
valve
supply
tubing
is
a
specially
constructed
twin
polyurethane
hose
enclosed
in
a
PVC
jacket.
It
is
three
feet
in
length
and
has
one
portion
with
DISS
fittings
at
both
ends
while
the
other
section
is
designed
for
nipple
connection
at
one
end
and
for
connection
to
the
unique
7/16
inch
female
connector
(on the
Control
Module)
at
the
other
end.
The
oxygen
line
is
standard
oxygen
tubing
with
DISS
fittings
at
both
ends.
Non-Rebreathing
Valve
and
Corrugated
Hose
Part
number
L496
Non-Rebreathing
Valve
is
designed
to
allow
a
simple
method
for
providing
P.E.E.P.
(positive-end-expiratory-pressure]
to
a
patient.
Attach
one end
of
the
corrugated
flex
tube
to
the
patient
valve
assembly.
Attach
the
opposite
end
to
the
inlet
port
on
the
P/N
L496
(see
diagram)
valve.
Both
these
connections
are
friction
fit,
so
be
sure
to
slide
the
flex
tube
over
each
connection
as
far
as
possible.
At
this
point,
you
are
set
to
install
your
P.E.E.P.
product
in
the
appropriate
port.
WARNING:
The
P/N
L496
Non-rebreathing
valve
is
not
for
use
in
toxic
atmospheres
PATIENT
VALVE
SUPPLY
TUBING
FIGURE
12.
P.E.E.P.
PORT
As
NON-REBREATHING
VALVE
CORRUGATED
HOSE

AutoVentESSS/33SS
8
PATIENT
VALVE
SUPPLY
TUBING
UNPACKING
AND
After
opening
your
new
LSP
AutoVent
2000/3000,
INSPECTION
OF
THE
examine
the
shipping
carton
and
contents.
Lay
out
contents
AUTOVENT
so
that
each
component
is
identifiable,
as
displayed
below
2000/3000
(Figure
13).
If
the
carton
is
crushed,
previously
opened,
or
shows
other
signs
of
damage,
notify
the
carrier
immediately
to
file
a
claim.
Do
not use
the
unit
on
a
patient
until
it
has
been
tested
and
performs
as
specified.
The
complete
LSP
AutoVent
2000,
P/N
L460,
and
AutoVent
3000
Automatic
Ventilator,
P/N
L461,
consists
of
the
following
component
parts:
Description
AV
2000
PartNo.
AV
3000
Part
No.
OXYGEN
LINE
One
Control
Module
[462-150
1463-150
One
Patient
Valve
Assy.
=
一
—
One
Patient
Valve
Ca
U
Supply.
Tubing
1535114
1535114
nni
i
One
Oxygen
Line
1535026
1535026
=
Non-Rebreathing
Valve
1496
1496
ЕАО
НЕЕ
&
Corrugated
Hose
1535124-010 1535124-010
Operator
Manual
1909005-225
1909005-225
FIGURE
13
**NOTE:
Appendix
A
provides
a
list
of
suggested
equipment
for
use
in
conjunction
with
or
in
support
of
the
LSP
AutoVent
2000/3000.
**
PATIENT
VALVE
ASSEMBLY
OPERATING
**NOTE:
Read
all
instructions
carefully
prior
to
setup
and
INSTRUCTIONS
operation
of
this
unit.
Particular
attention
should
be
paid
to
all
warnings,
cautions
and
notes
in
order
to
assure
proper
performance
during
use.**
WARNING:
Should
a
WARNING:
This
CAUTION:
In
order
to
mechanical
problem
device
operates
with
provide
optimal
develop
or
the
patient
medical gases
under
performance,
check
appears
to
be
pressure,
including
all
source
gas
supplies
experiencing
difficulty
oxygen.
Do
not
use
to
assure
only
clean,
while
connected
to
this
device
while
dry
gas
is
used,
free
this
unit,
disconnect
smoking
or
near
open
of
contaminants
the
unit
immediately
flames.
Do
not
use
and/or
liquids.
and
ventilate
by
other
oil
on
this
device
or
ーー
means.
If
unable
to
operate
near
determine
the
cause
flammable
materials.
of
the
problem,
the
unit
should
be
WARNING:
This
returned
to
an
device
should
only
be
authorized
AutoVent
operated
by
qualified
repair
center.
personnel
under
approved
medical
direction.

Instructions
for
Use
of
the
AutoVent
2000/3000
with
Oxygen
Cylinders
Regulator
Attachment
Remove
plastic
wrap
from
oxygen
cylinder
valve
outlet.
Point
the
cylinder
valve
in
a
safe
direction
before
opening
the
valve.
Remove
all
dirt
and
debris
from
cylinder
valve
by
"cracking"
the
cylinder
prior
fo
attaching
the
pressure
regulator.
("Cracking"
consists
of
slowly
opening
the
cylinder
valve
and
allowing
a
brief
flow
of
gas
to
occur
prior
to
attaching
the
regulator).
When
mounting
a
pin
index
regulator
(LSP
P/N
1270-220,
L270-230,
1735-260,
or
other
approved
regulator)
on
a
cylinder,
make
sure
the
gasket
is
properly
positioned
on
the
inlet
stem
to
prevent
oxygen
or
source
gas
leakage.
Tighten
the
regulator
yoke
by
hand
using
the
"T"
handle
assembly.
(The
use
of
tools
may
result
in
damage
to
the
regulator).
Control
Module
Attachment
Connect
the
other
end
of
the
oxygen
supply
line
to
the
source
gas
inlet
on
the
Control
Module.
Connect
the
Patient
Valve
supply
tubing
to
the
Patient
Valve
outlets
of
the
Control
Module.
Hand
tighten
the
DISS
connector
and
the
unique
7/16
inch
actuator
connector
to
the
Control
Module.
Connect
the
twin
hose
to
the
source
gas
inlets
on
the
Patient
Valve
Assembly.
The
fittings
allow
for
proper
connection
only.
Hand
tighten
the
DISS
connector
to
the
Patient
Valve
Assembly.
Examine
the
cylinder
pressure
gauge.
This
can
be
used
to
indicate
cylinder
contents
since
the
pressure
is
proportional
to
the
amount
of
remaining
oxygen.
A
portable
cylinder
is
essentially
empty
when
the
pressure
has
fallen
to
200
psig.
LSP
P/N
L280-220,
L160-260,
or
other
regulators
complying
with
Compressed
Gas
Association
(CGA)
guidelines,
mount
on
cylinders
with
CGA
540
connections.
Connect
the
oxygen
high
pressure
line
to
the
50
psig
gas
outlet
on
the
regulator.
FIGURE
14.

ο.
~
AutoVentESS
10
Instructions
for
Use
of
the
AutoVent
2000/3000
with
Oxygen
Cylinders
(Continued)
Instructions
for
Use
of
the
AutoVent
2000/3000
with
a
Wall
Outlet
Employing
a
Quick
Connection
Adapter
**NOTE:
Check
all
line
and
tubing
connections
for
leaks.
If
a
leak
occurs,
check
that
the
previous
fitting
instructions
have
been
followed
correctly.
**
CAUTION:
Read
all
instructions
thoroughly
before
opening
the
cylinder
valve.
Connect
all
oxygen/source
gas
pressure
lines
to
the
LSP
AutoVent
2000/3000
and
Patient
Valve
Assembly
prior
to
use.
Assure
all
high
pressure
outlets
are
plugged
and
cylinders
turned
off
or
closed
CAUTION:
Always
make
sure
an
adequate
supply
of
oxygen
or
source
gas
is
available
for
patient
use
and
transport.
It
is
advisable
to
have
a
back-up
regulator
available
to
facilitate
change-over
in
the
event
a
cylinder
transfer
needs
to
be
made.
when
not
in
use.
Use
standard
approved
quick
release
connectors
intended
for
use
with
cylinder
banks
or
transport
gas
supplies,
attached
to
50
psig
high
pressure
lines.
Connect
the
other
end
of
the
oxygen
supply
high
pressure
line
to
the
source
gas
inlet
port
of
the
Control
Module.
Connect
the
Patient
Valve
supply
tubing
to
the
Patient
Valve
outlets
of
the
Control
Module.
Hand-tighten
the
DISS
connector
and
the
unique
7/16
inch
actuator
connector
to
the
Control
Module.
Connect
the
Patient
Valve
supply
tubing
and
the
actuator
supply
tubing
to
the
inlet
ports
on
the
Patient
Valve
Assembly.
The
fittings
allow
for
proper
connection
only.
Hand
tighten
the
DISS
connector
to
the
Patient
Valve
Assembly.
Insert
the
quick-release
adapter
into
the
corresponding
wall
outlet
and
assure
it
is
properly
in
place
with
an
audible
snap
or
click.
Pull
firmly
on
the
adapter
to
check
its
proper
insertion.
**NOTE:
Immediately
check
all
pressure
lines
and
tubing
for
leaks.
If
a
leak
occurs,
check
that
the
previous
fitting
instructions
have
been
followed
correctly.
Check
that
all
line
and
tubing
connections
have been
adequately
hand
tightened.
* *
CAUTION:
Always
verify
that
the
cylinder
valve
is
in
the
closed
or
off
position
(fully
clockwise)
prior
to
disconnecting
the
tubing
assembly
or
removing
the
regulator
from
the
oxygen
cylinder.
FIGURE
15,

AutoVent2522/2555
11
Instructions
for
Use
of
the
AutoVent
2000/3000
with
an
Air/Oxygen
Blender
Instructions
for
this
application
are
essentially
identical
to
those
for
use
with
either
cylinder
or
wall
outlet
systems.
The
high-flow
blender
system
should
be
connected
to
the
source
gases
as
per
manufacturers
instructions.
The
blender
then
becomes
the
attachment
site
for
the
source
gas
supply
line
to
the
Control
Module.
CAUTION:
Always
follow
the
blender
manufacturer's
instructions,
contained
in
the
blender product
manual,
for
exact
connection
of
the
blender
to
cylinders
or
wall
sources.
Always
use
a
high-flow
blender
(15
LPM
to
150 LPM)
for
ventilatory
application.
**NOTE:
Make
sure
a
compressed
air
source
and
oxygen
source
are
available
prior
to
using
a
blender.
**
Connect
the
other
end
of
the
oxygen
supply
high
pressure
line
(source
gas
supply
line)
to
the
source
gas
inlet
port
of
the
Control
Module.
Connect
the
Patient
Valve
supply
tubing
to
the
Patient
Valve
outlets
of
the
Control
Module.
Hand-tighten
the
DISS
connector
and
the
unique
7/16
inch
actuator
connector
to
the
Control
Module.
Connect
the
Patient
Valve
supply
tubing
and
the
actuator
supply
tubing
to
the
inlet
ports
on
the
Patient
Valve
Assembly.
The
fittings
allow
for
proper
connection
only.
Hand
tighten
the
DISS
connector
to
the
Patient
Valve
Assembly.
**NOTE:
Immediately
check
all
pressure
lines
and
tubing
for
leaks.
If
a
leak
occurs,
check
that
the
previous
fitting
instructions
have
been
followed
correctly.
Check
that
all
line
and
tubing
connections
have
been
adequately
hand
tightened,
**
**NOTE:
Assure
the
delivery
of
precise
oxygen
concentrations
when
using
a
blender
by
inserting
an
oxygen
monitor
probe
in
the
gas
delivery
system
at
the
patient
valve
outlet
prior
to
patient
use.
**
WARNING:
Always
check
or
change
the
source
gases
if
a
low
pressure
blender
alarm
sounds,
distinguished
by
a
continuous
high-pitched
hum.

AutoVent2S5:
Ventilator
Check-Out
PATIENT
USE
1.
Check
the
ventilator
system
for
proper
function
by
performing
the
following
tests:
ce
®
Set
the
BPM
control
knob
to
the
setting
marked
"12".
(Adult
for
AutoVent
3000)
©
Set
the
Tidal
Volume
(Vt)
control
knob
to
800
ml.
*
Set
the
Inspiratory
Time
control
knob
to
the
adult
setting
1]
ト
|
on
the
AutoVent
3000.
(Inspiratory
time
is
pre-set
on
the
AutoVent
2000).
Rotate
the
control
knob
clockwise
until
it
is
against
the
end
stop
on
the
adult
setting.
FIGURE
16.
e
Count
the
number
of
complete
ventilator
cycles
for
a
60
SECONDS
full
minute.
At
the
checkout
setting,
there
should
be
12
INSPIRATORY
TIME
BPM
delivered,
with
a
1.5
second
inspiration
and
3.5
м
1.5
SEC.
пай
12
ВРМ
second
expiration
per
breath.
EXPIRATORY
TIME
2.
Occlude
the
outlet
of
the
patient
Valve
Assembly.
An
3.5
SEC.
audible
pressure
limit
alarm
should
sound
after
the
FIGURE
17.
ventilator
cycles,
indicating
the
designed
pressure
limit
has
been
reached.
**NOTE:
The
pressure
limit
alarm
should
sound
throughout
the
latter
portion
of
the
breath
after
reaching
the
pressure
limit,
and
stop
when
the
ventilator
cycles
to
expiration.
**
*“NOTE:
AutoVents
&
patient
Valves
are
serialized
and
calibrated
to
work
together
and
should
remain
together
for
the
life
of
the
products
FIGURE
18.
3.
Test
the
unit
for
proper
function
prior
to
each
patient
use.
Refer
to
Maintenance
section
for
this
procedure
(page
14).
4.
Should
the
unit
fail
to
operate
properly
at
any
time,
refer
to
the
Troubleshooting
Guide
(page
16.)
Disconnect
the
patient
from
the
ventilator
any
time
the
unit
does
not
appear
to
be
operating
properly.
Wunoble
to
determine
the
cause
of
problem,
contact
life
Support
Products
for
service.
5.
Clean
the
unit
after
each
use
(refer
to
the
Maintenance
section
for
detailed
instructions
on
page
14).
6.
Always
store
the
unit
in
a
clean,
dry
place.
**NOTE:
if
the
LSP
Automatic
Ventilator
is
to
be
powered
by
a
cylinder,
be
sure
fo
turn
on
the
cylinder
valve
slowly.
**
WARNING:
This
device
should
only
be
operated
by
qualified
personnel
under
approved
medical
direction.
Check
for
obstructions
in
the
patient's
throat
or
mouth
(vomitus,
foreign
bodies,
broken
dentures,
etc.),
and
remove
if
present,
in
accordance
with
prevailing
standards.
Set
the
volume
to
equal
8
to
10
ml. for
every
kg.
of
body
weight;
e.g.,
70
kg.
patient
equal
700
ml.
volume.
WARNING:
The
AutoVent
3000
is
not
recommended
for
use
with
patients
less
than
20
kg.
The
AutoVent
2000
is
not
recommended
for
use
with
patient
less
than
40
kg.

AutoVentZSSS,
13
PATIENT
USE
(Continued)
Set
the
BPM
control
knob
to
the
desired
setting.
Refer
to
the
quick
set-up
instructions
on
the
back
of
the
Control
Module
for
guidelines.
Set
the
inspiratory
time
control
knob
to
the
desired
adult
or
child
position.
Rotate
the
control
knob
to
either
position
until
it
is
against
the
end
stop.
(AutoVent
3000)
Set
the
Tidal
Volume
control
knob
to
the
desired
volume.
Occlude
the
outlet
port
of
the
Patient
Valve
Assembly.
Allow
the
ventilator
to
cycle
to
ensure
proper
operation
of
the
valve
and
pressure
limit
alarm.
Use
with
a
standard
resuscitation
mask:
after
initial
Control
Module
settings
have been
made
and
a
patient
airway
is
established,
install
the
mask
on
the
outlet
adapter
of
the
Patient
Valve
Assembly
and
place
on
the
patient.
**NOTE:
Follow established
procedural
guidelines
for
opening
and maintaining
a
patient
airway.
**
Use
on
patients
with
an
endotracheal
tube
or
tracheostomy
tube
in
place:
afer
initial
control
Module
settings
have
been
made,
connect
the
Patient
Valve
Assembly
directly
to
the
endotracheal
or
tracheostomy
tube
adapter
(15mm.
inside
diameter/22mm.
outside
diameter
dimensions
allow
this
connection).
**NOTE:
A
humidification
device
is
recommended
if
the
patient
has
an
endoctracheal
tube
or
tracheostomy
tube
in
place.
**
Check
the
following
parameters
immediately
after
connecting
the
patient
to
the
ventilator.
(1)
BPM
-
Using
the
second
hand
on
your
watch,
count
the
number
of
breaths
delivered
to
the
patient
for
one
full
minute.
If
you
need
to
increase
or
decrease
the
rate,
adjust
the
BPM
control
knob
accordingly,
checking
the
rate
again
once
the
adjustment
has
been
made.
(2)
Tidal
Volume
-
Observe
patient
for
adequate
chest
rise
and
fall
(chest
excursion}.
Chest excursion
should
be
normal
and
equal
on
both
sides.
If
the
chest
does
not
rise,
check
the
airway
and
evaluate
for
other
injuries
to
the
thoracic
area.
Recheck
the
tidal
volume
setting.
(3)
Inspiratory
Time
-
With
the
AutoVent
3000
ensure
that
the
NH
Time
control
knob
is
set
appropriately
and
all
the
way
against
the
appropriate
end
stop.
(AutoVent
3000
only.)
**NOTE;
If
the
pressure
İncrease
the
volume
limit
alarm
sounds
during
the
inspiratory
phase
and
adequate
chest
movement
does
not
occur,
an
increase
in
airway
resistance,
a
blocked
airway
and/or
a
stiff
lung
is
indicated.
delivered
to
the
patient,
until
adequate
chest
movement
occurs,
by
rotating
the
Tidal
Volume
control
knob
in
a
clockwise
direction.
Disconnect
the
patient
from
the
ventilator
FIGURE
20.
5
Le
FIGURE
21.
AUTOVENT
2000
FIGURE
22.
AUTOVENT
3000
FIGURE
23.
and
attempt
to
ventilate
via
other
means
if
adjustments
do
not
result
in
satisfactory
ventilation
of
the
patient.
For
additional
information,
refer
to
the
Troubleshooting
Guide.
**
(page
16)

AutoVent2555/555S
PATIENT
USE
(Continued)
MAINTENANCE
OF
THE
LSP
AUTOVENT
2000/3000
Cleaning
and
Disinfecting
Equipment
If
the
patient
is
being
ventilated
by
mask, check
the
patient
frequently
for
signs
of
vomiting.
Should
vomiting
occur,
remove
the
mask
to
prevent
aspiration
which
may
cause
airway
obstruction.
Immediately
clear
the
mask
and
Patient
Valve
Assembly
of
any
foreign
material,
reestablish
the
patient's
airway,
and
resume
ventilation.
If
unable
to
resume
ventilation
with
the
Patient
Valve
Assembly,
use
a
resuscitator
bag
or
perform
mouth-to-mask
resuscitation.
**NOTE:
If
a
compressed
gas
cylinder
is
used,
check
the
cylinder
contents
frequently;
should
the
cylinder
require
**NOTE:
Gloves
and
protective
coverings
are
recommended
when
performing
maintenance
and
cleaning
of
patient
care
equipment,
**
**NOTE:
Clean
and
disinfect
the
ventilator
after
each
use.
Re-certify
calibration
of
AutoVent
once
a
year.**
*Note:
AutoVents
should
be
checked
for
calibration
annually.
1.
Cleaning
and
Disinfecting
the
Control
Module
**NOTE:
Water
will
not
affect
the
operation
of
the
Control
Module.
**
replacement,
perform
maneuver
with
minimal
interruption
to
ventilation
of
the
patient.
**
Should
patient
begin
breathing
spontaneously
{an
effort
of
-2
cm.H,0
will
activate
the
demand
valve]
it
may
be
desirable
to
14
WARNING:
Monitor
the
patient
closely
while
using
the
demand
mode.
Should
the
patient's
respirations
slow,
become
shallow
or
labored,
return
to
initial
automatic
decrease
or
turn
the
ventilator
settings
ventilator
rate
(BPM)
to
the
immediately.
"O"
position.
This
will
allow
the
patient
to
breathe
TABLE
1,
spontaneously.
ni
TV
SETTİNG
|
FLOW
(LPM)
The
ventilator
will
deliver
100%
source
gas
to
the
ADULT
|
CHILD
patient
on
demand,
up
to
400
|
200
16
48
LPM
depending
on
the
600
|
300
24
tidal
volume
setting
(See
800
|
400
32
Table
1.).
Any
volume
1000
|
500
40
required
by
the
patient
in
οσο
|
16b
48
excess
of
the
indicated
|
|
source
gas
flow
rate
(see
Source
gas
flow
rates
upon
patient
demand.
Table
|.)
will
be
supplied
by
ambient
air.
WARNING:
Cleaning
procedures
must
be
performed
in
an
environment
free
of
oil
and
petroleum-
based
products.
Leave
hoses
connected
so
you
do
not
get
water
inside.
Follow
established
protocol
regarding frequency
of
cleaning.
Do
not
submerge
the
Control
Module
when
cleaning.
Take
a
clean
cloth
soaked
in
a
detergent
solution
and
wipe
off
any
residue
from
surface.
Wipe
thoroughly.
Take
a
clean
cloth
soaked
in
an
80%
isopropyl
alcohol
solution
or
a
cold
chemical
disinfecting
solution,
and
wipe
entire
surface
of
Control
Module.
Rinse
Thoroughly
being
careful
not
to
get
any
liquid
inside
the
control
module.
Take
a
clean
cloth
and
dry
surface
of
Control
Module.
Test
Control
Module
before
use.

AutoVent5SS
2/3555
15
Cleaning
and
2.
Cleaning
and
WARNING:
Clean
and
disinfect
the
Patient
Valve
Disinfecting
Eguipment
Disinfecting
the
Patient
Assembly
after
every
use.
(Continued)
Valve
Assembly
Remove
the
outlet
adapter
and
exhalation
valve
assembly
from
the
Patient
Valve
Assembly,
Leave
tubing
assembly
connected.
Clean
all
foreign
matter
from
the
components
with
a
mild
soap
solution,
being
careful
not
to
get
any
liquid
inside
the
Patient
Valve
Assembly.
Rinse
the
parts
thoroughly
in
clean
water.
Immerse
the
outlet
adapter
and
the
exhalation
valve
assembly
from
the
Patient
Valve
in
a
disinfectant
or
bacteriocidal
solution
for
a
minimum
of
10
minutes.
Remove
the
outlet
adapter
and
the
exhalation
valve
assembly
from
the
solution
and
rinse
thoroughly
with
water.
Rinse
repeatedly
to
assure
that
all
the
solution
is
removed.
Place
the
Patient
Valve,
outlet
side
down,
into
a
shallow
container
with
not
more
than
1/2
inch
of
disinfectant
or
bacteriocidal
solution.
The
Patient
Valve
should
remain
in
this
solution
for
a
minimum
of
10
minutes.
Leave
tubing
assembly
connected.
Remove
the
Patient
Valve
from
the
solution
and
rinse
thoroughly
with
water.
Dry
assembly
using
approved
standard
methods
such
as
hot
air
drying.
After
drying,
carefully
examine
the
parts
of
the
Patient
Valve
Assembly.
Discard
any
cracked
or
damaged
parts
and
replace
as
necessary.
Prior
to
reassembling
the
entire
unit,
inspect
all
lines
and
tubing
filters
for
contaminants,
replacing
as
necessary.
Reconnect
the
tubing
assembly
to
the
fittings.
Check
the
exhalation
valve
assembly
to
assure
the
flapper
valve
is
not
twisted
and
the
locating
bosses
are
properly
positioned.
(Figure
25)
FIGURE
24.

AutoVent2500/3055
16
Cleaning
and
Disinfecting
Eguipment
(Continued)
FIGURE
26.
STORAGE
OF
THE
LSP
AUTOVENT
2000/3000
TROUBLESHOOTING
GUIDE
FOR
THE
LSP
AUTOVENT
2000/3000
LOCATING
BRACKETS
FIGURE
25.
FIGURE
27.
3.
Cleaning
the
Non-
Rebreathing
Valve
CAUTION:
If
the
flapper
valve
is
twisted
or
the
locating
bosses
are
not
properly
positioned,
the
Patient
Valve
Assembly
will
not
function
properly.
Always
make
sure
the
valve
is
flat
and
properly
seated.
Immediately
after
cleaning,
reassemble
the
Patient
Valve
Assembly
and
connect
to
the
Control
Module.
Turn
on
the
oxygen
supply
and
allow
the
ventilator
to
cycle
several
times
to
blow
out
any
liquid
which
may
have
gotten
inside
during
the
cleaning
process.
Test
the
unit
for
proper
function
prior
to
each
patient
use.
Attach
an
inflatable
test
lung,
P/N
L109,
to
the
Patient
Valve
Assembly
outlet
and
complete
the
test
at
different
rates
and
tidal
volumes.
Clean
after
each
use.
All
components
of
the
P/N
L496
NonRebreathing
Valve
are
autoclavable.
To
disassemble,
unscrew
the
valve
inlet
from
the
outlet
and
remove
the
duckbill
diaphragm.
The
individual
components
can
now
be
cleaned.
If
autoclaving
systems
are
not
available,
you
may
also
sterilize
via
cidex
or
other
bacteriocidal
solution.
Be
sure
to
rinse
and
dry
all
parts
thoroughly
before
re-
assembling.
After
cleaning,
inspect
all
parts
for
damage
or
breakage.
Replace
any
damaged
or
broken
parts.
Store
the
unit
in
a
clean,
dry
area
within
a
temperature
range
of
-40'F
to
160°F.
After
long
periods
of
storage,
the
unit
should
be
fully
tested
before
use
in
accordance
with
the
checkout
procedures
in
this
manual.
INDICATION
Decreased
tidal
volume
or
decreased
chest
expansion
PROBABLE
CAUSE
SOLUTION
leak
around
mask
or
Patient
Valve
tubing
Inappropriate
volume
setting
Inappropriate
inspiratory
Time
setting
(AV
3000)
Decreased
lung
compliance
and/or
increased
airway
resistance
Airway
secretions
Check
all
connections
for
leaks
Check
Control
Module
setting
and
adjust
as
required
Check
Control
Module
setting
and
adjust
as
required
Evaluate
patient
and
correct
as
required
by
adjusting
Control
Module
settings
Clear
airway
of
secretions

4utoVentS
ら
らら
/3
つ
らら
1Z
TROUBLESHOOTING
Increased
tidal
volume
or
GUIDE
increased
chest
expansion
(Continued)
Pressure
limit
alarm
at
beginning
of
inspiratory
phase
Pressure
limit
alarm
during
inspiratory
phase
Failure
of
the
ventilator
to
cycle
Failure
of
the
pressure
limit
alarm
Volume
setting
to
high
Increased
lung
compliance
Inappropriate
inspiratory
Time
setting
(AV
3000)
Airway
blockage,
kinked
tubing,
and/or
increased
airway
resistance
Increased
airway
resistance
Decreased
lung
compliance
Coughing
Increased
airway
secretions
Gas
source
failure
Cylinder
valve
closed
BPM
control
knob
in
"O"
position
Loose
connections
Disconnected
actuator
tubing
Kinked
oxygen
supply
line
and/or
actuator
tubing
Regulator
failure
Malfunctioning
Control
Module
Alarm
outlet
is
plugged
with
debris
or
has
malfunctioned
Check
Control
Module
settings
and
adjust
volume
as
required
Evaluate
patient
and
correct
as
required
by
adjusting
Control
Module
Settings
Check
Control
Module
setting
and
adjust
as
required
Clear
airway
of
secretions
or
foreign
matter;
check
endotracheal
tube;
check
ventilator
tubing
Evaluate
patient
and
adjust
ventilators
as
required
Evaluate
patient
and
correct
as
required
by
adjusting
Control
Module
settings
Attempt
to
alleviate
coughing
Clear
airway
secretions
Change
oxygen
cylinder
if
being
used,
or
evaluate
and
check
gas
source
outlet.
Open
cylinder
valve
fully
Adjust
BPM
knob
to
desired
rate
Tighten
connections
Reconnect
tubing
Straighten
tubing
Change
regulator
Remove
from
patient
and
ventilate
by
alternate
means.
Remove
and
clean,
or
replace

AutoVent2525/355S
WARRANTY
INFORMATION
SUMMARY
OF
WARNINGS
AND
CAUTIONS
Warranty
Repair
Service
Please
complete
and
return
the
Warranty
Registration
card
inclosed
with
your
AutoVent
2000/3000
as
soon
as
possible.
Please
read
the
following
limited
warranty
carefully:
In
the
event
your
LSP
AutoVent
2000/3000
Automatic
Ventilator
needs
servicing,
the
following
steps
will
help
to
ensure
that
the
repair
service
is
processed
promplly.
Contact
your
authorized
Life
Support
Products
distributor,
or
Life
Support
Products
before
returning
product
for
repair/service.
Mailing
Address:
Shipping
Address:
Life
Support
Products
Life
Support
Products
1720
Sublette
Avenue
St.
Louis,
MO
63110
Telephone:
(800)
444-3954
Repackage
the
Control
Module
and
Patient
Valve
Assembly,
providing
adequate
packaging
material
to
protect
the
module
during
shipment.
This
warranty
is
not
valid
if
the
Control
Module
or
Patient
Valve
Assembly
show
signs
of
misuse,
being
opened,
altered
or
modified
in
any
way
other
than
its
intended
use.
**NOTE:
Some
warnings
and
cautions
appear
more
than
once
throughout
the
manual.
They
appear
in
this
summary
to
help
direct
the
user
to
the
proper
page
and
section
of
this
manual.
**
Page
Number
4
Warnings
If
the
maximum
pressure
limit
is
reached,
the
pre-set
tidal
volume
may
not
be
delivered
to
the
patient.
Inspiratory
time
will
remain
constant,
however,
and
an
inspiratory
hold
will
be
maintained
with
no
additional
volume
being
delivered
until
the
Ventilator
cycles
to
the
expiratory
phase.
This
warning
also
appears
under
Tidal
Volume
in
the
Performance
Characteristics.
Should
the
blue
rubber
diaphragm
blow
outward
from
the
alarm
module's
air
entrainment
ports,
remove
the
AutoVent
immediately
from
service.
And
contact
your
LSP
Distributor.
If
the
maximum
pressure
limit
is
reached,
the
pre-set
tidal
volume
may
not
be
delivered
to
the
patient.
Inspiratory
time
will
remain
constant,
however,
and
an
inspiratory
hold
will
be
maintained
with
no
additional
volume
being
delivered
until
the
ventilator
cycles
to
the
expiratory
phase.
This
warning
also
appears
under
the
Pressure
Limit
Alarm
Module
in
the
Performance
Characteristics.
This manual suits for next models
1
Table of contents
Other Allied Fan manuals
Popular Fan manuals by other brands

Inline
Inline SmartHome 40160 manual

Suntec Wellness
Suntec Wellness KLIMATRONIC CoolBreeze 1500 TV manual

Vortice
Vortice VORT HR 300 NETI Instruction booklet

Koenic
Koenic KTF 121 user manual

Ridder Drive Systems
Ridder Drive Systems PolyDrive RPD 506 Series product manual

wurth
wurth WSH AIR III Translation of the original operating instructions