Am Instru Q-FP 201 User manual

Instru Q-FP 201 Handhold Ultrasonic Flowmeter
User Manual
(Ver 8.5)
Ver 8.5
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

Instru Q-FP 201 Handheld Ultrasonic Flowmeter
-2-
Contents
1
Introduction
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
1
§1.1 Preface
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
1
§1.2 Main Features
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
1
§1.3 Principle of Measurement
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
2
§1.4 Packing List (Standard Configuration)
- - - - - - - - - - - - - - - - - - - - - - - - - - - - -
2
§1.5 Front view
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
3
§1.6 Typical Applications
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
4
§1.7 Data Integrity and Built-in Time-Keeper
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
4
§1.8 Product Identification
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
•
4
§1.9 Specifications
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
5
2. Starting Measurement
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
6
§2.1 Built-in Battery
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
6
§2.2 Power On
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
6
§2.3 Keypad
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
7
§2.4 Menu Windows
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
7
§2.5 Steps to Configure the Parameters
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
8
§2.6 Transducers Mounting Allocation
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
9
§2.7 Transducers Installation
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
10
§2.7.1 Transducers Spacing
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
11
§2.7.2 V-method Installation
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
11
§2.7.3 Z-method Installation
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
12
§2.7.4 W-method Installation
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
12
§2.7.5 N-method Installation
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
12
§2.8 Installation Checkup
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
13
§2.8.1 Signal Strength
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
13
§2.8.2 Signal Quality
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
13
§2.8.3 Total Transit Time and Delta Time
- - - - - - - - - - - - - - - - - - - - - - - - - - -
13
§2.8.4 Time Ratio between the Measured Total Transit Time and the Calculated Time
- - 14
3. Menu Window Details
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
15
§3.1 Menu Windows Arrangement
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
15
§3.2 Menu Window Details
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
15
4. How To
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
21
§4.1 How to judge if the instrument works properly
- - - - - - - - - - - - - - - - - - - - - - - -
21
§4.2 How to judge the liquid flowing direction
- - - - - - - - - - - - - - - - - - - - - - - - - -
•
21
§4.3 How to change between units systems
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
--
21
§4.4 How to select a required flow rate unit
-
•
21
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

-3-
Instru Q-FP 201 Handheld Ultrasonic Flowmeter
§4.5 How to use the totalizer multiplier
- - - - - - - - - - - - - - - - - - - - - - - - - - -
21
§4.6 How to open or shut the totalizers
- - - - - - - - - - - - - - - - - - - - - - - - - - -
22
§4.7 How to reset the totalizers
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
22
§4.8 How to restore the flow meter with default setups
- - - - - - - - - - - - - - -
22
§4.9 How to use the damper
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
22
§4.10 How to use the zero-cutoff function
- - - - - - - - - - - - - - - - - - - - - - - - - -
22
§4.11 How to setup a zero point
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
22
§4.12 How to get a scale factor for calibration
- - - - - - - - - - - - - - - - - - - - - - -
22
§4.13 How to use the operation locker
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
23
§4.14 How to use the built-in data logger
- - - - - - - - - - - - - - - - - - - - - - - - - -
23
§4.15 How to use the Frequency Output
- - - - - - - - - - - - - - - - - - - - - - - - - -
23
§4.16 How to use the Totalizer Pulse Output
- - - - - - - - - - - - - - - - - - - - - - -
24
§4.17 How to produce an alarm signal
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
24
§4.18 How to use the built-in Buzzer
- - - - - - - - - - - - - - - - - - - - - - - - - - - - -
24
§4.19 How to use the OCT output
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
25
§4.20 How to modify the built-in calendar
- - - - - - - - - - - - - - - - - - - - - - - - -
25
§4.21 How to adjust the LCD contrast
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
25
§4.22 How to use the RS232 serial interface
- - - - - - - - - - - - - - - - - - - - - - -
25
§4.23 How to view the Date Totalizers
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
25
§4.24 How to use the Working Timer
- - - - - - - - - - - - - - - - - - - - - - - - - - - - -
25
§4.25 How to use the manual totalizer
- - - - - - - - - - - - - - - - - - - - - - - - - - - -
25
§4.26 How to know how long the battery will last
- - - - - - - - - - - - - - - - - - - - -
26
§4.27 How to charge the built-in battery
- - - - - - - - - - - - - - - - - - - - - - - - - - -
26
§4.28 How to check the ESN and other minor details-
26
5.Troubleshooting
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
27
§5.1 Power-on Error Displays and Counter-Measures
- - - - - - - - - - - - - - - -
27
§5.2 Error Code and Counter-Measures
- - - - - - - - - - - - - - - - - - - - - - - - - - -
27
§5.3 Other Problems and Solutions
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
28
6. Communication Protocol
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
30
§6. 1 General - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
30
§6.2 Interface Pin-out Definition
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
30
§6.3 the Protocol
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- 30
§6.4 Protocol Prefix Usage
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
31
§6.5 Codes for the Keypad-
- 32
7. Service
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
33
§7.1 Service
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
33
§7.2 Software Upgrade Service
33
Appendix
34
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

Instru Q-FP 201
Handheld Ultrasonic Flowmeter
-1-
1. Introduction
§ 1.1 Preface
Welcome
to the ultrasonic flowmeter that has been manufactured
with patent technologies and is equipped with more functions and advanced
performance than our previousversions.
The ultrasonic flowmeter has been upgraded based on the previous ultrasonic
flowmeter which is still the main productline of the company. The new Version
retains most of the excellent featuresand functions of the previous versions:
the pulse measurement technology, the ultrasonic igniting and the small signal
receiving circuits etc. The main improvements are made on the battery supply
circuit and on the transmitting circuits. All other circuits are simply integratedinto
this new version without major modifications, due to the fact that we have already
applied the most advanced measurement technologies and attained a more
reliable model of ultrasonic flow meter.
Q-FP 201 flowmeter incorporates the latest ICs manufactured from the famous
semiconductor manufacturers like Philips, Maxim, TI, Winbond, and Xilinx.
Thehardware features the ease of operation, high accuracy and outstanding
reliability,while the software provides a very friendly user interface and much
more functions.lt employs a patent balanced lower voltage multi-pulse igniting
circuit whichincreases the anti-interference ability magnificently so that the
flow meter will work properly even in demanding industrial environments such
as those with power frequency transverter working nearby.
Other outstanding features:
-
---the signal receiving circuits feature self-adapting performance so as to
ensure that the user can easily operate the instrument without any
adjustment.
-
---the built-in rechargeable Ni-H battery can work continuously for more
than 12 hours without recharging.
The advanced circuit design, the integration of the latest semiconductors, the
user-friendly software interface both in English and Chinese languages and
small-sized PCB board, all these features combine to make the Q-FP 201
ultrasonic flowmeter the
best
and the biggest seller on the Chinese market.
Moreover, it is gaining more and more recognition on the international flow
meter market.
§ 1.2 Main Features
*
Large-screen
LCD
*
Built-in data-logger
*
High accuracy measuring
*
Non-contacting measuring
*
Built-in rechargeable battery
* Wide measuring range
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

Tdown
up
flow
-2-- Q-FP 201 Handheld Ultrasonic Flowmeter
* Small and light * RS-232 serial interface
§1.3 Principle of Measurement
The Q-FP 201 ultrasonic flowmeter is designed to measure the fluid velocity of liquid within a
closed conduit. The transducers are a non-contacting, clamp-on type, which will provide benefits
of non-fouling operation and easy installation.
The Q-FP 201 transit time flowmeter utilizes two transducers that function as both ultrasonic
transmitters and receivers. The transducers are clamped on the outside of a closed pipe at a specific
distance from each other. The transducers can be mounted in V-method where the sound
transverses the pipe twice, or W-method
where the sound transverses the pipe four times, or in
Z-method where the transducers are mounted on opposite sides of the pipe and the sound crosses
the pipe once. This selection of the mounting method depends on pipe and liquid characteristics.
The flow meter operates by alternately transmitting and receiving a frequency modulated burst of
sound energy between the two transducers and measuring the transit time that it takes for sound to
travel between the two transducers. The difference in the transit time measured is directly and
exactly related to the velocity of the liquid in the pipe, show as follows:
Downstream
transducer
V =
MD
x
AT
sin
20 T
v
• T
down
spacing
Where
0 is the include angle to the flow direction
M is the travel times of the ultrasonic beam
D is the pipe diameter
Tup is the time for the beam from upstream transducer to the downstream one
Tdown is the time for the beam from downstream transducer to the upstream one
AT=Tup
—Tdown
§1.4 Packing List (Standard Configuration)
I
Upstream
transducer
Check up the packing list carefully before installing
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

INSTRU Q-FP 201 Handheld Ultrasonic Flowmeter
-3-
Name
QTY (pcs)
Main Unit
1
Medium transducer
2
Ultrasonic signal cable
2
Data line
AC adapter
1
Coupling agent
1
Belt stretcher
2
Tape
1
User manual
1
Carrying case
1
§1.5 Front view
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

-6-
INSTRU Q-FP 201 Handheld Ultrasonic Flowmeter
2. Starting Measurement
§2.1 Built-in Battery
The instrument can operate either from the built-in Ni-H rechargeable battery, which will last over
12 hours of continuous operation when fully recharged, or from an external AC/power supply from
the battery charger.
The battery charging circuits employ a scheme of constant-current and constant-voltage. It has a
characteristic of fast charging at the beginning and very slow charging when the battery approaches
to full recharge. Generally, when the green LED starts coming on, the battery would be nearly 95%
recharged and when the red LED is off, the battery would be 98% recharged.
Since the charging current becomes tapered when the battery recharge is nearly completed, i.e. the
charging current becomes smaller and smaller, therefore, there should be no over-recharging
problem. That means the charging progress can last very long. The charger can be connected to the
handset all the time when an around-the-clock measurement is required.
When fully recharged, the terminal voltage reaches around 4.25V. The terminal voltage is displayed
on window M07. When the battery is nearly consumed, the battery voltage drops to below 3V. The
user can obtain an approximate battery working time from the battery voltage.
A software battery working time estimator is integrated in this instrument based on the terminal
voltage. Please note that the estimator may have relatively bigger errors in the estimated working
time, especially when the voltage is in the range of around 3.70 to -3.90 volt.
§2.2 Power On
key to switch on the instrument and press the I OFF to turn off the power.
Once the flow meter is switched on, it will run a self diagnostic program, checking first the
hardware and then the software integrity. If there is any abnormality, corresponding error messages
will display.
Generally, there should be no display of error messages, and the flow meter will go to the most
commonly used Menu Window Number 01 (short for M01) to display the Velocity, Flow Rate,
Positive Totalizer, Signal Strength and Signal Quality, based on the pipe parameters configured last
ti
me by the user or by the initial program.
The flow measurement program always operates in the background of the user interface. This
means the flow measurement will keep on running regardless of any user menu window browsing
or viewing. Only when the user enters new pipe parameters will the flow meter change
measurement to the new parameter changes.
Press the
ON
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

00041)
INSTRU Q-FP 201 Handheld Ultrasonic Flowmeter
-7-
When new pipe parameters have been entered or when the power has been just switched on, the
flow meter will enter an adjusting mode to make the signals magnified with proper amplification.
By this step, the flow meter is going to find the best threshold of receiving signal. The user will see
the progress by the number 1, 2, or 3, which are indicated on the right lower corner of the LCD
display.
When the transducers have been adjusted on the pipe by the user, the flow meter will re-adjust the
signal automatically.
Any user-entered configuration value will be retained into the NVRAM of the flow meter, until it is
modified by the user.
§2.3 Keypad
The keypad for the operation of the flow meter has 16+2 keys, as shown by the right picture.
Keys
1
01—
9
and . 1 are keys to enter numbers
Key
♦/+
is the going UP key, when the user wants to go to the
upper menu window. It also works as the "+" key when entering
numbers
Key
♦1-
is the going DOWN key, when the user wants to go
down-sided menu window. It also works as the "—" key when
entering numbers.
Key
.4
is backspace key, when the user wants go left or wants
backspace the left character that is located to the left of the cursor.
Key
Key MENU is the key for the direct menu window jump over.
Whenever the user wants to proceed to a certain menu window, the
user can press this key followed by 2-digit numbers.
key is shortened as the `M' key afterward when referring to the menu windows.
key is for the power on.
key is for the power off.
§2.4 Menu Windows
The user interface of this flow meter comprises about 100 different menu windows that are
numbered by MOO, MO1, MO2 ... M99.
There are 2 methods to enter certain menu window:
(1) Direct going/entering. The user can press the
MENU
key followed by two-digit number keys.
For example, the menu window M11 is for the entering of pipe outer diameter. The display will go
to the Mll menu window after the user presses
(2) Pressing
A/+
and ♦/-I keys. Each time of the
ENT is the ENTER key for any inputting or selections.
The
The
The
MENU
ON
OFF
MENUI
1
1
A/+
key pressing will proceed to the
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

-
10
-
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter
An optimum location would be defined as a straight pipe length full of liquid that is to be measured.
The piping can be in vertical or horizontal position. The following table shows
Piping Configuration
and
Transducer Position
Upstream
Dimension
Downstream
Dimension
Lup
x Diameters
Ldn
x Diameters
L up
----0.--
1
Ldn
—
g
m
10D
5D
i
Lup
1
Ldn
1OD
5D
----.—
i
Lup Ldn
1OD
5D
i
Lup
—
I
Ldn
12D
5D
I
I
Q
Lup
Ldn
I
M.Eript-,
20D
5D
i
1_,,,,,
_....__
Ldn
.
,-,...
20D
5D
E><
i
Lup
1
Ldn
30D
5D
1
/
YI
Examples of optimum locations.
Principles to selection of an optimum location
(1)
Install the transducers on a longer length of the straight pipe. The longer the better, and make
sure that the pipe is completely full of liquid.
(2)
Make sure that the temperature on the location does not exceed the range for the transducers.
Generally speaking, the closer to the room temperature, the better.
(3)
Take the pipe fouling into consideration. Select a straight length of a relatively newer pipe. If
the condition is not satisfying, consider the fouling thickness as part of the liner for a better
result.
(4) Some pipes have a kind of plastic liner, and between the outer pipe and the liner there may be a
certain thickness difference that will prevent the ultrasonic waves from direct traveling. Such
conditions will make the measurement very difficult. Whenever possible, try to avoid this kind
of pipes. If impossible, try our plug-in transducers that are installed permanently on the pipe by
drilling holes on the pipe while liquid is running inside.
§2.7 Transducers Installation
The transducers used by the
Q-FP 201
eries ultrasonic flow meter are made of piezoelectric
crystals both for transmitting and receiving ultrasonic signals through the wall of liquid piping
system. The measurement is realized by measuring the traveling time difference of the ultrasonic
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044

-11-
signals. Since the difference is very small, the spacing and the alignment of the transducers are
critical factors to the accuracy of the measurement and the performance of the system. Meticulous
care should be taken for the installation of the transducers.
Steps to the installation of the transducers
(1)
Locate an optimum position where the straight pipe length is sufficient, and where pipes are
in a favorable condition, e.g., newer pipes with no rust and ease of operation.
(2)
Clean any dust and rust. For a better result, polishing the pipe with a sander is strongly
recommended.
(
3
)
Apply adequate coupler to the spot where the transducers are to be installed and leave no gap
between the pipe surface and the transducers.
Extra care should be taken to avoid any sand or dust particles left between the pipe outer surface
and the transducers.
To avoid gas bubbles inside the upper part of the pipe, the transducers should be installed
horizontally by the side of the pipe.
§2.7.1 Transducers Spacing
The spacing value shown on menu window M25 refers to the distance of inner spacing between the
two transducers. The actual transducers spacing should be as close as possible to the spacing value.
§2.7.2 V-method Installation
V-method installation is the most widely mode for daily measurement with pipe inner diameters
ranging from 15 mm to 400 mm. It is also called reflective mode.
±45`
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

§2.7.3 Z-method Installation
Z-method is commonly used when the pipe diameter is above 200mm.
041111•1Nk
.4;"
Upstream Transducer
/I\
+45°1
\
I
I
-
\
Transducers
Spacing
- Downstream Transducer
§2.7.4 W-method Installation
W-method is usually used on plastic pipes with a diameter from 15mm to 50mm.
a
Az
-
Upstream Transducer
-
12
-
=45'
1
-
+45'
§2.7.5 N-method Installation
Rarely used method.
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

-13-
§2.8 Installation Checkup
Through the checkup of the installation, one can: check the receiving signal strength, the signal
quality Q value, the traveling time difference of the signals, the estimated liquid speed, the
measured traveling time of the signals and the calculated traveling time ratio. Therefore, optimum
measurement result and longer running time of the instrument can be achieved.
§2.8.1 Signal Strength
Signal strength indicates the amplitude of receiving ultrasonic signals by a 3-digit number. [000]
means there is no signal detected and [999] refers to the maximum signal strength that can be
received.
Although the instrument works well if the signal strength ranges from 500 to 999, stronger signal
strength should be pursued, because a stronger signal means a better result. The following methods
are recommended to obtain stronger signals:
(1)
Relocate a more favorable location, if the current location is not good enough for a stable and
reliable flow reading, or if the signal strength is lower than 700.
(2)
Try to polish the outer surface of the pipe, and apply more coupler to increase the signal
strength.
(3) Adjust the transducers both vertically and horizontally while checking the varying signal
strength, stop at the highest position, and then check the transducers spacing to make sure the
transducers spacing is the same as what the M25 shows.
§2.8.2 Signal Quality
Signal quality is indicated as the Q value in the instrument. A higher Q value would mean a higher
Signal and Noise Ratio (short for SNR), and accordingly a higher degree of accuracy would be
achieved. Under normal pipe condition, the Q value is in the range 600-900, the higher the better.
Causes for a lower Q value could be:
(1)
Interference of other instruments and devices such as a powerful transverter working nearby.
Try to relocate the flow meter to a new place where the interference can be reduced.
(2)
Bad sonic coupling for the transducers with the pipe. Try to apply more coupler or clean the
surface, etc.
(3)
Pipes are difficult to be measured. Relocation is recommended.
§2.8.3 Total Transit Time and Delta Time
The numbers displayed on menu window M93 are called total transit time and delta time
respectively. They are the primitive data for the instrument to calculate the flow rate inside the pipe.
So the flow rate indication will vary accordingly with the total time and delta time.
The total transit time should remain stable or vary little.
If the delta time fluctuates higher than 20%, it means there are certain kinds of problems with the
transducer installation.
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

-
14
-
§2.8.4 Time Ratio between the Measured Total Transit Time and the Calculated
Time
This ratio would be used to check the transducer installation. If the pipe parameters are entered
correctly and the transducers are installed properly, the value for this ratio should be in the range of
100±3. If this range is exceeded, the user should check:
(1) If the pipe parameters are correctly entered.
(2)
If the actual spacing of the transducers is right and the same as what the window M25 shows.
(3)
If the transducers are installed properly in the right directions.
(4)
If the mounting location is good and if the pipe has changed shape or if there is too much
fouling inside the pipes
Other poor
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

-15-
3. Menu Window Details
§3.1 Menu Windows Arrangement
MOO—M09 windows for the display of the flow rate, velocity, date time, totalizers, battery voltage
and estimated working hours for the battery.
M10-1\429 windows for entering the pipe parameter.
M30—M38 windows for flow rate unit selections and totalizer unit selections.
M40—M49 windows for response time, zeroing, calibration and modification password setup.
M50—M53 windows for the built-in logger
M60-M78 windows for time-keeper initialization, version and ESN information viewing and
alarms.
M82 window for viewing date totalizer.
M90—M94 are diagnostic windows for a more accurate measurement.
M97—M99 are not windows but commands for the outputting of display copying and pipe
parameter setups.
M+0—M+8 are windows for some additional functions, including a scientific calculator, viewer on
records such as total working hours, turn-on and turn-off times, dates and times when
the flow meter has been turned on or turned off.
Other menu windows such as M88 have no functions, or functions were cancelled because they are
not applied to this version of the software.
The major reason why the menu windows are arranged in this way is that the software programmer
hopes that the menu window arrangement for this version can be compatibility with the previous
versions of the flow meter software. This will make it easier for the former version users with this
flow meter series.
§3.2 Menu Window Details
Menu
window
No.
Function
MOO
Display three positive negative net totalizers,
signal strength, signal quality
and working status
MOl
Display POS totalizer, flow rate, velocity, signal strength, signal quality and
working status
MO2
Display NEG totalizer, flow rate, velocity, signal strength, signal quality and
working status
M03
Display NET totalizer, flow rate, velocity, signal strength, signal quality and
working status
M04
Display date and time, flow rate, signal strength, signal quality and working
status
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

—17—
M24
Window for selecting the transducer mounting methods
Four methods can be selected:
(0) V-method
(1) Z-method
(2) N-method
(3) W-method
M25
Display the transducer mounting spacing
M26
Entry to store the parameter configuration into the internal NVRAM
M27
Entry to load one set of saved parameters
M28
Select YES or NO for the instrument to determine whether or not to hold (or
to keep) the last correct value when poor signal condition occurs. YES is the
default setup
M29
Enter a value ranging from 000 to 999.
0 is the default value
M30
Window for selecting unit system. Default value is 'Metric'. The change from
English to Metric or vice versa will not affect the unit for totalizers.
M31
Window for selecting flow rate that will be used by the instrument afterward.
Flow rate can be in
0. Cubic meter
short for
(
m3)
1.
Liter
(1)
2.
USA
gallon
(gal)
3.
Imperial Gallon
(igl)
4.
Million USA gallon
(mgl)
5.
Cubic feet
(cf)
6.
USA liquid barrel
(bal)
7.
Imperial liquid barrel
(ib)
8.
Oil barrel
(oh)
The flow unit in terms of time can be per day, per hour, per minute or per
second. So there are 36 different flow rate units in total for selection.
M32
Window for selecting the totaliziers' working unit
M33
Select totalizer multiplier
The multiplier ranges from 0.001 to 10000
M34
Turn on or turn off the NET totalizer
M35
Turn on or turn off the POS totalizer
M36
Turn on or turn off the NEG totalizer
M37
(1)
Totalizer reset
(2)
Restore the instrument to the default parameters as the manufacturer did
by pressing the dot key followed by the backspace key. Take care or make note
on the parameters before doing the restoration
M38
Press-a-key-to-run or to stop totalizer for easier calibration
M39
Operational interface language selection
in Chinese
and
English.
This
selection makes it possible that more than 2 billions of people on the world
can read the menu.
M40
Flow rate damper for a stable value. The input range is 0 to 999 seconds.
0 means there is no damping. Default value is 10 seconds
M41
Lower flow rate cut-off to avoid invalid accumulation.
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

-
16
—
MO5
Display date and time, velocity, signal strength, signal quality and working
status
M06
Display the wave shape of the receiving signal
M07
Display the battery terminal voltage and its estimated lasting time
M08
Display the all the detailed working status, signal strength, signal quality
M09
Display today's total flow, velocity, signal strength, signal quality and working
status
M10
Window for entering the outer perimeter of the pipe
Mll
Window for entering the outer diameter of the pipe
0 to 6000mm is the allowed range of the value.
M12
Window for entering pipe wall thickness
M13
Window for entering the inner diameter of the pipe
M14
Window for selecting pipe material
Standard pipe materials (that the user need not know the speed ) include:
(0) carbon steel (1) stainless steel
(2) cast iron (3) ductile iron
(4) copper (5) PVC
(6) aluminum
(7) asbestos
(8) fiberglass
M15
Window for entering the pipe material speed only for non-standard pipe
materials
M16
Window for selecting the liner material, select none for pipes without any
liner
Standard liner materials that the user need not know the speed include:
(1) Tar Epoxy
(2) Rubber
(3) Mortar
(4) Polypropylene
(5) Polystryol (6)Polystyrene
(7) Polyester
(8) Polyethylene
(9) Ebonite
(10) Teflon
m17
Window for entering the liner material speed only for non-standard liner
materials
M18
Window for entering the liner thickness, if there is a liner
M19
Window for entering the ABS thickness of the inside wall of the pipe
M20
Window for selecting fluid type
For standard liquids that the user need not know the liquid speed include:
(0) Water
(1) Sea Water
(2) Kerosene
(3) Gasoline
(4) Fuel oil
(5) Crude Oil (6) Propane at -45C
(7) Butane at OC
(8)Other liquids
(9) Diesel Oil
(10)Caster Oil
(11)Peanut Oil
(12) #90 Gasoline (13) #93 Gasoline (14) Alcohol
(15) Hot water at 125C
M21
Window for entering the fluid sonic velocity only for non-standard liquids
M22
Window for entering the viscosity of the non-standard liquids
M23
Window for selecting the proper transducers
There are 14 different types of transducers for selection.
If the user-type-transducers are used, 4 user type wedge parameters, which
will be prompted by the software, should be entered following.
If the 7r type transducers are used, 37r type transducers and pipe parameters
should be entered following.
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

-18-
M42
Zero point setup under the condition when there is no liquid running inside
the pipe.
M43
Clear the zero point set by the user, and restore the zero point set by the
factory
M44
Set up a manual flow bias. Generally this value should be 0.
M45
Scale factor for the instrument. The default value is '1'.
Keep this value as '1', when no user calibration has been made.
M46
Network environment Identification Number. Any integer can be entered
except 13(ODH, carriage return), 10 (OAH, line feeding), 42 (2AH), 38, 65535.
Every set of the instrument in a network environment should have a unique
IDN. Please refer to the chapter for communication.
M47
System locker to avoid modification of the parameters
M48
Not used
M49
Communication tester
M50
"Option" selection for the built-in logger. It also functions as the switch of
logger
M51
Time setup for the data logger
M52
(1)
Data logging direction control. If `To RS-232' is selected, all the data
produced by the data logger will be transmitted out through the RS-232
interface
(2)
If `To buffer ' is selected, the data will be stored into the built-in logger
memory
(3)
Buffer transferring and buffer clearing
M53
Logger buffer viewer. It functions as a file editor. Use Dot, backspace UP and
DN keys to browse the buffer.
If the logger is ON, the viewer will automatically refresh once new data are
stored
M54 Not used
M55 Nod used
M56
Not used
M57
Not used
M58
Not used
M59
Not used
M60
99-year calendar. Press ENT for modification. Use the dot key to skip the
digits that need no adjusting.
M61
Display Version information and Electronic Serial Number (ESN) that are
unique for each Q-FP 201 series flow meter.
The users can employ the ESN for instrumentation management
M62
RS-232 setup. Baud rate can be 75 to 115200 bps
M63
Not used
M64 Not used
M65
Not used
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

-19-
M66
Not used
M67
Input the frequency range for the frequency output. The biggest range is
OHz-9999Hz. Default value is 1-1001 Hz
M68
Enter a flow rate value that corresponds to lower frequency
M69
Enter a flow Rate value that corresponds to higher frequency
M70
LCD display backlight control. The entered value indicates how many seconds
the backlight will be on with every key pressing.
M71
LCD contrast control. The LCD will become darker when a small value is
entered.
M72
Working timer. It can be cleared by pressing ENT key, and then select YES.
M73
Enter Lower Flow Rate value that will trigger the #1 Alarm. There are two
virtual alarms in the system. By "virtual" we mean that the user must redirect
the output of the alarms by setuping the output hardware in M78 and M77
M74
Enter the higher flow rate value that will trigger the #1 Alarm.
M75
Enter the lower flow rate value that will trigger the #2 Alarm.
M76
Enter the higher flow rate value that will trigger the #2 Alarm.
M77
Buzzer setup.
If a proper input source is selected, the buzzer will beep when the trigger
event occurs
M78
OCT (Open Collect Transistor Output) setup
By selecting a proper input source, the OCT hardware will close when the
trigger event occurs
M79
Not used
M80
Work as a keypad and display for another handheld set by RS-232 connected
with the handset
M81
Not used
M82
Date totalizer
M83 Not used
M84
Not used
M85
Not used
M86
Not used
M87
Not used
M88
Not used
M89
Not used
M90
Display signal strength, signal quality, time ratio on the upper right corner.
M91
Displays the Time Ratio between the Measured Total Transit Time and the
Calculated time. If the pipe parameters are entered correctly and the
transducers are properly installed, the ratio value should be in the range of
100±3%. Otherwise the entered parameters and the transducer installation
should be checked.
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
ADDS INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

M92
Displays the estimated fluid sound velocity. If this value has an obvious
difference with the actual fluid sound speed, pipe parameters entered and the
transducer installation should be checked again.
M93
Displays total transit time and delta time(transit time difference)
M94
Displays the Reynolds number and the pipe factor used by the flow rate
program.
M95
M96
Not used
Not used
M97
Command to record the pipe parameters entered by the user either to the
built-in data logger or to RS-232C serial interface
M98
Command to record the diagnostic information either to the built-in data
logger or to RS-232C serial interface
M99
Command to copy the current display either to the built-in data logger or to
RS-232C serial interface
M+0
Browse the 64 recorded instrument power-on and power-off date and time
with the flow rate at the time of power on and off
M+1
Displays the total working time of the instrument
M+2
Displays the last power-off date and time
M+3
Displays the last power-off flow rate
M+4
Displays the times of instrument powered on(the instrument has been
powered on)
M+5
A scientific calculator for the convenience of field working.
All the values are in single accuracy.
The drawback is that the user can't operate it by direct key-pressing
M+6
Not used
M+7
Not used
M+8
Not used
M+9
Not used
M-0
Entry to hardware adjusting windows only for the manufacturer
-
20
-
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter

4. How To
§4.1 How to judge if the instrument works properly
When 'R' is displayed in the lowest right corner of LCD display, the instrument is working properly
generally speaking.
If an
flashes on that place, there could be poor signal received. Please refer to the chapters on
diagnosis.
If an T is displayed, it means that there is no signal detected.
If a `.1' is displayed, it means that the hardware of this instrument could be out of order. Refer to the
chapter on diagnosis.
§4.2 How to judge the liquid flowing direction
(1) Make sure that the instrument works properly
(2)
Check the flow rate for the indication. If the displayed value is POSITIVE, the direction of
the flow will be from the RED transducers to the BLUE transducers; if the displayed value is
NEGATIVE, the direction will be from the BLUE transducers to the RED transducers;
§4.3 How to change between units systems
Use menu window M30 for the selection of unit system in English or Metric system.
§4.4 How to select a required flow rate unit
Use menu window M31 to select the flow unit first and then the timing unit.
§4.5 How to use the totalizer multiplier
Use window M33 to select a proper totalizer. Make sure that the totalizer pulse is appropriately
speeded. It should not be too fast and neither too slow. A speed of producing a pulse in several
seconds or minutes is preferable.
If the totalizer multiplier is too small, there can be a loss of accumulation pulse because the output
device can output only one pulse in a measurement period (500milliseconds)
-21-
If the totalizer multiplier is too large, the output pulse will be too fewer for the devices that are
connected with the instrument for a quicker response.
www.automat.com.ar aw@automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111)4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter
Table of contents