CAL
OFSt Calibration offset - the calibration
offset is a single point adjustment
which can be usedto alter the
calibration scaling values across
the entire measuring range without
affecting the calibration slope.
This method can be used instead of
performing a two point calibration
when a constant measurement
error is found to exist across the
entire range. To perform a calibration
offset press the ^and vbuttons
simultaneously at the CAL OFSt
function. A "live" reading from the
input will be seen,make a note of
this reading. Press the Fbutton, the message SCLE will now
be seen followed by the last scale value inmemory. Use the ^or vbutton to adjust
the scale value to the required display value for that input.
For example if the "live" input reading was50 and the required display value for this
input was 70 then adjust the SCLE value to 70.
ZERO
RNGE Zero range - the zero range function allows a limit value to be set (in engineering
units) above which the display will not zero i.e. if a zero operation is attempted via
the Pbutton, remote input or set zero function when thedisplay value is greater
than the zero range setting the display will refuse to zero and give a ZERO RNGE
Err message (note that the CAL OFSt function is also affected by the ZERO
RNGE setting). For example if the zero range setting is 10 the instrument will only
respond to a zero operation if the display reading at the time is between -10 and 10.
If the zero rangefunction is not required it can be set to OFF by pressing the ^and
vbuttons simultaneously at this function. When switched off the instrument can be
zeroed no matter what the display value.
Note that the instrument keeps track of the value being zeroed at each operation,
when the total amount zeroed from repeated operations becomes greater than the
zero range value the instrument will reject the zero operation and a ZERO RNGE
Err message will be seen. To allow a zero operation beyond this point either the
ZERO RNGE function value will need to be raised or a new zero reference point
introduced via the CAL ZERO function.
If repeated zero operations are required the ZERO RNGE function should be set to
OFF or alternatively the tARE operation could be considered.
CAL
ZERO Calibration zero - the calibration zero function is used following a calibration via
CAL1 and CAL2. A calibration zero operation at this time ensures that the display
zero and the ZERO RNGE reference zero are at the same point after a calibration.
After a calibrationthe calibration zero can also be used to select a zero point other
than the display zero as the reference for the ZERO RNGE function. For example if
the CAL ZERO operation is carried out with a display reading of 500 and a ZERO
RNGE reading of 10 the zero range function will allow the display to zero only if the
current display reading is between 490 and 510. To perform a calibration zero press
the ^and vbuttons simultaneously at the CAL ZERO function, a live reading will
be seen, press the Fbutton, the message CAL ZERO End should now be seen
indicating that the instrument has acceptedthe zero point. Although the display
reading will not change as a result of the calibration zero the input value on the
display at the timeof the operation will be the new zero reference point for the ZERO
RNGE function.
USER
En 4 4mA input scaling without a live input.
This scaling method can be used with 4-20mA inputs only. The instrument can be
scaled for a 4-20mA input without a live input i.e. this is an alternative method to the
CAL1 and CAL2 method of scaling. To perform the first point (USER En 4) scaling
simply press the ^and vbuttons simultaneously when the USER En4 function
has been reached. The display will now indicate a value. Use the ^or vbutton to
change this valueto the display value required for a 4mA input.
Page 20 of 30 PM4AMMAN-1.4-0
Input
offset
offset