American Marine PINPOINT II User manual

American Marine is proud to introduce the PINPOINT® Calcium
Monitor. The only truly affordable and accurate digital Calcium
measurement instrument designed exclusively for marine saltwater.
Powered by a 9-volt battery (not included). PINPOINT® Calcium
Monitor will operate for approximately 150 total hours before battery
replacement is indicated on the display.
NOTE: The Calcium Monitor unit is NOT waterproof and must be
operated on a dry surface. Liquid contact on the printed circuit board will
cause corrosion and void warranty.
•Probe Set-up
1)This probe is a solid state single piece, permanently filled model Ver III.
2)Remove the storage bottle that protects the probe tip by unscrewing
the bottle from the bottle cap. When not in use, probe should be stored in
the protective bottle without any fluid added to the bottle.
3) Be careful not to damage the flat PVC membrane on the probe tip.
This membrane should be kept as clean; free from algae, etc.
•Meter Set-up
1)All calibration fluids and tank water sample should be room
temperature for greatest accuracy.
2)Pour the 100 ppm Ca++ calibration fluid into a clean cup or storage
bottle.
3)Pour the 1,000 ppm Ca++ calibration fluid into a 2nd clean cup
4)Install a new 9-volt battery or the PINPOINT AC Adapter to the battery
compartment wires on the back of the meter.
5)Attach the BNC connector from the probe cable to the front of the
meter. Lock it into place with a 1/4 turn to the right.
•Calibration and Measurement
1)Activate meter on by pressing the ON/OFF/CAL button for 1 second.
2)Now press the ON/OFF/CAL button continuously for 3-4 seconds (you
will hear a beep) when “CAL” appears on the display.
3)The number 100 should appear on the display. If not; then press the
Range button to change the calibration value to 100.
4)Put the probe tip into the 100 ppm Ca++ Calibration Fluid. Wait until
the number 100 begins to flash on the display. This may take a minute or
so for a new probe. Let the flashing continue at least 3-5 minutes is
advised. Then press the Enter button to register the 100 ppm Ca++
calibration. The LCD should display the word “good”
5)Shake the probe dry. Press the Range button until 1000 is displayed.
6)Put the probe tip into the 1000 ppm Ca++ Calibration Fluid. Wait until
the number 1000 begins to flash on the display. This may take a minute
or two. Let the flashing continue for at least 3-5 minutes. Longer “flashing
time” will yield the most stable readings. Then press the Enter button to
register the 1000 ppm Ca++ calibration. The LCD should display the
word “good”.
7)Now press the ON/OFF/CAL button to read Calcium ppm.
8)Test your sample. Take a sample of water away from your system in a
clean cup and insert the probe to take a precision measurement. Allow
enough time for the reading to settle down.
9)Rinse the probe in RO or DI water, shake the probe dry and store it in
the protective bottle.
10)The meter will store the calibration even when the power is off.
•Calcium Monitoring
The accuracy of this and other electronic instruments is the greatest
immediately following a calibration process. Measure the calcium
frequently, if the value drifts, calibration should be performed to verify the
reading. The probe should not be totally submerged. Only the bottom
1 –2 inches of the probe tip should be in the water.
Troubleshooting and Important Points to Remember on back
PINPOINT®II
Calcium
Monitor ver. 1.03
User's Guide
American Marine Inc.
54 Danbury Road, Suite 172
Ridgefield, CT 06877 U.S.A.
Phone/FAX 914-763-5367
www.americanmarineusa.com
©Copyright 2007
American Marine Inc.
American Marine Inc
The world’s finest selection of electronic measurement and
control instruments designed and optimized for the
professional aquaculture industry.
PH
ORP/REDOX
Oxygen
Calcium (Marine)
Conductivity Freshwater Hardness
Wireless Temperature
Nitrate (Marine)
Salinity (Marine and Koi)
Complete line of all calibration materials and accessories
Important Points to Remember
• There are several possibilities for the meter to show “err” during the calibration process:
1-the probe tip is very dry (common with a new probe) and it should be hydrated by placing the tip into 1,000 ppm fluid
calibration fluid for and hour or so and then the 100 ppm calibration fluid for an hour. Then recalibrate.
2-The probe has reached the end of its useful life
3-The calibration fluids have become contaminated.
• At the end of the calibration process if the meter shows a ppm reading of zero; one or both of the calibration fluids may
be out of specification and are no longer accurate. Obtain new calibration fluids and recalibrate. Do not mix old
calibration fluids with new calibration fluids; discard the old set.
• Calcium Probe stability will be greatly improved when the probe tip is allowed to properly hydrate for 5-10 minutes in
any calibration fluid prior to calibration.
•PINPOINT Calcium Monitor is NOT waterproofand must be operated on a dry surface. Liquid contact with any of the
printed circuit boards will cause corrosion and void the warranty.
• Measurements above the instrument range will be indicated on the LCD as “HI”.
• The Calcium Probe tip should never be placed in RO or DI water for an extended period of time.
• If you observe an unusually high or low calcium ppm value, it is advisable to re-check the calibration.
• The Calcium Probe tip should be stored dry. The storage bottle can be used to prevent physical damage to the tip and
should not be filled with any fluid.
• The Calcium Meter will always store the last calibration even when the power is off.
• Replace batteries in the PINPOINT Calcium Monitor when the appropriate LCD battery icon appears.
•Display Range: 000.0 – 9,772 ppm Ca+
Resolution: 000.2 ppm
Accuracy: +/- 2 % of the reading after calibration
Warranty
PINPOINT®Calcium Monitor by American Marine Inc. is warranted to be free of defects in material and workmanship
for a period of 2 years from date of sale.
PINPOINT® Calcium Probe is warranted to be free of defects in material and workmanship for a period of 6 months
from the date of sale.
Positive proof of purchase is required for warranty claim. American Marine Inc. will not be liable for any costs of
removal, installation, transportation charges, or any other charges, which may result in connection with a warranty claim.
American Marine Inc. will not be liable for any damage or wear to products or livestock caused by abnormal operating
conditions, water damage, abuse, misuse, unauthorized alternation or repair or if the product was not installed in
accordance with the printed operating instructions. Any defective product must be sent freight prepaid with appropriate
documentation supporting the warranty claim. Replacement or repair will be at the discretion of American Marine Inc.
American Marine Inc.
54 Danbury Road, Suite 172
Ridgefield, CT 06877 U.S.A.
www.americanmarineusa.com

Popular Measuring Instrument manuals by other brands

Labino Apollo 1.0 operating instructions

Labino

Labino Apollo 1.0 operating instructions

ALZA Professor DP-85 Directions for use

ALZA

ALZA Professor DP-85 Directions for use

Entes EMK-01 manual

Entes

Entes EMK-01 manual

Master Meter Allegro installation guide

Master Meter

Master Meter Allegro installation guide

Ditel ALPHA-D Quick installation guide

Ditel

Ditel ALPHA-D Quick installation guide

PlantCare Plantomat PRO Instructions for use

PlantCare

PlantCare Plantomat PRO Instructions for use

CASTLE SONIK S operating manual

CASTLE

CASTLE SONIK S operating manual

Messtechnik MiniAir64 Operating	 instruction

Messtechnik

Messtechnik MiniAir64 Operating instruction

Hanna Instruments HI 3221 instruction manual

Hanna Instruments

Hanna Instruments HI 3221 instruction manual

Awair Glow C AWAIR0PD2 quick start guide

Awair

Awair Glow C AWAIR0PD2 quick start guide

palintest Micro 600 manual

palintest

palintest Micro 600 manual

WYLER Clinotronic S quick reference

WYLER

WYLER Clinotronic S quick reference

Honeywell BW RigRat quick start guide

Honeywell

Honeywell BW RigRat quick start guide

Fluidwell F110-A manual

Fluidwell

Fluidwell F110-A manual

BASETech Porty II operating instructions

BASETech

BASETech Porty II operating instructions

Lamtec LT3 user manual

Lamtec

Lamtec LT3 user manual

Gemco Semelex Safetimeter 1999 Series operating manual

Gemco

Gemco Semelex Safetimeter 1999 Series operating manual

Thermo Scientific Orion 7070iX Installation and operation

Thermo Scientific

Thermo Scientific Orion 7070iX Installation and operation

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.