Amvex FM Series Operating instructions

Cleaning Instructions
Use a clean damp cloth with a mild cleaning solution to wipe outside of product.
DO NOT gas sterilize with ETO. DO NOT clean with pungent hydrocarbons.
CAUTION: DO NOT submerge Dial Flowmeter in any form of liquid. This will cause
damage and void any warranty on the product.
Troubleshooting
Contact your dealer or the technical support department at Amvex Corporation for assistance if the
Dial Flowmeter does not function.
Maintenance Prevention
Inspect the product before and after use for any damage and ease of operation.
WARNING: When changing connectors on the Dial Flowmeter for service or replacement,
never re-attach connectors of a different gas. Doing so may result in patient
injury or damage to the equipment.
NOTE: Disconnect Dial Flowmeter from gas supply before servicing.
WARRANTY
Amvex Corporation warrants its Medical Secondary Equipment to be free from defects in material
and workmanship for a period of Five (5) years from the date of shipment. Within the rst Twelve
(12) months from date of shipment Amvex will repair or replace any part which is proven to
be defective at Amvex’s cost. After the rst Twelve (12) months, Amvex will send the parts to
the customer free of charge, but the shipping and installation will be borne by the Customer.
This warranty is valid only when the product has been properly installed according to Amvex
specications, used in a normal manner and serviced according to factory recommendations. It
does not cover failures due to damage which occurs in shipments or failures which resulted from
accidents, misuse, abuse, neglect, mishandling, alteration, misapplication or damage that may be
attributable to acts of God.
AMVEX CORPORATION DOES NOT HONOR VERBAL STATEMENTS CONCERNING THE
WARRANTY.
The distributor and/or dealer is not sanctioned to create verbal warranties about the product described
in this agreement. Any statements will not be honored or be made part of the agreement of sale. This
document is the nal, complete and exclusive terms of the agreement.
THIS WARRANTY IS INCLUSIVE AND REPLACES ALL OTHER WARRANTIES.
Amvex Corporation shall not, under any circumstances, be liable for incidental or consequential
damages including, but not limited to, prot loss, loss of sales or injuries to person(s) or property.
Correction of non-compliances as noted above will result in completion of all liabilities of Amvex
Corporation whether based on agreement, neglect or otherwise.
Amvex Corporation reserves the right to stop manufacturing any product or change materials, designs
or specications without notice.
All claims for warranty must rst be approved by Amvex’s Service Department: (support@amvex.
com or 905-764-7736 Toll Free: 1-866-462-6839). A valid Return Goods Authorization (RGA)
number must be obtained from Amvex prior to commencement of any warranty claim.
!
FM-D-MANUAL 08/10
4
CAUTION: Federal (USA) law restricts this device to sell by or on the order of a physician.
Gas:
Oxygen: O
Med Air: A
FM - XX Y Y - XX
Color Code:
USA: U
ISO: I
Dial Flowmeter Type:
Liters per minute (LPM)
25 LPM: 25
15 LPM: 15
8 LPM: 08
4 LPM: 04
Inlet Connection:
Ohmeda: OH
Chemetron: CH
1/8” FNPT: F2
1/4” FNPT: F4
1/8” MNPT: M2
1/4” MNPT: M4
DISS Male: DM
DISS Handtight: DH
DISS Hex Nut: DN
Puritan: PB
Schrader: SH
Oxequip: XQ
Medstar: MS
British Male: BM
French Male: FM
German Male: GM
Australian Male: AM
Japanese Male: JM
AGA Male: AG
0413
1111 Lakeside Drive, Gurnee,IL 60031-4099
Tel: 1-800-448-0770| Fax:847-855-6300
www.ohiomedical.com
25B East Pearce St., Richmond Hill, ON L4B 2M9 Canada| Tel: 905-764-7736 | Fax:905.763-7743 | www.amvex.com
OPERATING AND MAINTENANCE MANUAL
DIAL FLOWMETER
Authorized Representative
in the European Union:
EC REP
Oxygen Care Ltd.
2 Holfeld Business Park
Kilmacanogue Co Wicklow
Ireland

ATTENTION: Keep the Dial Flowmeter in a clean area when not being used.
ATTENTION: Ensure that all connections are tightened and free of leaks prior to use.
Only use an oxygen-safe leak detector when testing for leaks.
Intended Use
The Dial Flowmeter is used to dispense a certain ow of the gas specied on the label accurately
under the direction of a health care professional.
This manual and all other labels and inserts are strictly for the ease of operators that have
proper tools and knowledge in the repair of this device, and properly trained Amvex
representatives. DO NOT change, alter or modify intended use of the product.
Specications
Storage Temperature: -40 ºF (-40 ºC) to 140 ºF (60 ºC)
Dial Flowmeters calibrated at the pressure indicated on the Flowmeter, 70 ºF (21 ºC), standard
atmospheric pressure. Specications are subject to change without prior notice.
MRI WARNING: This product contains magnetic, ferrous material that may affect the result
of an MRI, unless Dial owmeter is ordered as MRI conditional.
Operating Instructions
1. Turn Dial Flowmeter off by turning knob fully counter clockwise, until window shows “0”.
2. Inspect the Dial Flowmeter for damage. If any is found, do not use the Dial Flowmeter.
3. Connect the Dial Flowmeter to the supply pressure specied on the Dial Flowmeter.
WARNING: The accuracy may be affected if the temperature of the gas is different
than 70º F (21º C) and the supply pressure is different than that indicated
on the Dial Flowmeter.
WARNING: Connection to the gas source must be done by using only the
appropriately indexed tting.
4. Adjusting the Flow: To Decrease Flow: Turn the knob counter clockwise
To Increase Flow: Turn the knob clockwise
WARNING: Flow is only available at the listed increments. There is no ow between
increments.
WARNING: To avoid injury ALWAYS conrm ow requirement for patient prior to
dispensing.
5. Turn knob completely counter clockwise to achieve maximum Flood/ Flush ow. (Flood/Flush
Setting not available on all models)
IMPORTANT: SAFETY INSTRUCTIONS
This manual is to provide you with information on the Dial Flowmeter and to ensure
the proper use and safety for this product.
READ AND UNDERSTAND ALL THE SAFETY AND OPERATING INSTRUCTIONS
CONTAINED IN THIS BOOKLET BEFORE USING THIS PRODUCT.
IF YOU DO NOT UNDERSTAND THESE INSTRUCTIONS, OR HAVE ANY QUESTIONS,
CONTACT YOUR SUPERVISOR, DEALER OR THE MANUFACTURER BEFORE ATTEMPTING TO
USE THE APPARATUS.
WARNING: Indicates a potentially hazardous situation, if not avoided, could result in
death or serious injury.
ATTENTION: Indicates a potentially hazardous situation, if not avoided, could result in
minor or moderate injury.
CAUTION: Indicates a potentially hazardous situation, which, if not avoided could result
in property damage.
Receiving Inspection
Remove product from package and inspect for damage. Verify that the model received is in
working order. If product is damaged, or incorrect do not use. Contact your dealer, equipment
provider or manufacturer.
ATTENTION: It is very important to allow product to remain in original packaging for 24
hours to acclimatize to room temperature before use.
ATTENTION: Store the product in a sealed package to avoid environmental damage. The
operating and storage temperature for the Dial Flowmeter should reect typical
environmental conditions of a medical facility environment.
User Responsibility
WARNING: Service of this device should only be performed by properly trained
individuals.
This product performs as explained in this manual. This holds true as long as the assembly, use,
repair and maintenance are properly followed according to our instructions. Periodic review of this
device is recommended. If any damage or defects are present, the product should not be used.
This includes parts that may have been altered, contaminated, worn or missing. If any of the
above are noted, immediate repair / replacement is required. In compliance with the Amvex
Warranty, repair of this device is not to be performed by anyone other than an Amvex trained
professional and done in strict accordance to the written instructions provided by Amvex. If this
device is subject to improper maintenance, repair, use and or abuse leading to malfunction of the
device, replacement is the sole responsibility of the user.
WARNING: Operation of this device is not to be done if ammable anesthetics are
present due to the possibility of explosion caused by static charge.
TO ELIMINATE THE RISK OF EXPLOSION OR FIRE:
• Never attempt to attach a Dial Flowmeter directly to a cylinder.
• Never use grease, oil, organic lubricants or ammable materials on or near the Dial Flowmeter.
• Never smoke in an area where oxygen is being used.
• Never use any type of ame or ammable or explosive material near the Dial Flowmeter.
• Always follow CGA and ANSI standards for Dial Flowmeters and Medical Gas Products
(E-7) and Oxygen Handling (G-4).
!
!
!
!
!
23
!
Gas Scale Increments Accuracy
Averages
Oxygen/Air 0-15 LPM
Oxygen/Air 0-8 LPM
Oxygen/Air 0-25 LPM
Oxygen/Air 0-4 LPM .03, .06, .12, .25, .5,
.75, 1, 1.5, 2, 3, 4
+/- 0.25 LPM from 0-1 LPM,
+/- 10% of reading from 1.5
LPM and above
.5, 1, 1.5, 2, 2.5, 3, 4, 5,
6, 7, 8
.5, 1, 2, 3, 4, 5, 6, 8, 10,
12, 15
.5, 1, 2, 3, 4, 5, 6, 8, 10,
15, 25
!
!
+/- 0.25 LPM from 0-1 LPM,
+/- 10% of reading from 1.5
LPM and above
+/- 0.25 LPM from 0-1 LPM,
+/- 10% of reading from 1.5
LPM and above
+/- 20% from 0-1 LPM
+/- 10% above 1 LPM