ANDalyze AND1100 User manual



Document:AND‐Prod‐1100‐8‐2013

AND1100Fluorimeter
forWaterTesting
UserManual

August,2013Edition2


©ANDalyze,Inc.,2011.Allrightsreserved.PrintedinUS


AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page2of16

TableofContents

2SafetyInformation……………………………………………….…………………………………………………………………3
2.1UseofHazardInformation
2.2Precautionarylabels
2.3ChemicalandBiologicalSafety
3GeneralInformation…………………..…………………………………………………………………………………………4
3.1OverviewofProduct
3.2GeneralFeatures
3.3AnalyteSensors
3.4OperatingEnvironment&Storage
3.5WaterTestingGuidelines
4Set‐UpandOperatingProcedures…………………………………………………………………………………………6
4.1ChargingtheBattery
4.2HomeTestingScreen
4.3SensorScreen
4.4SiteScreen
4.5StartOption
4.6MenuScreen
4.7FirmwareUpdate
4.8InstraCommLite
5SiteSet‐UpandManagement.……………………………………………………………………………………………….11
5.1NewSiteSet‐Up
5.2On‐SiteCalibration
5.3SelectingaSite
5.4DeletingaSite
5.5Re‐CalibratingaSite
6Maintenance……………………………………………………………………………………………………………………….…13
7CustomerServiceContactInformation…………………………………………………………………………………..14
8ConsumablesandReplacementItems……………………………………………………………………………………14
9TechnicalSpecificationforAND1100Fluorimeter………………………………………………………………….15
10LimitedWarranty…………………………………………………………………………………………………………………16
AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page3of16

2SafetyInformation
Pleasereadthisentiremanualbeforeunpacking,settinguporoperatingthisequipment.Payattentiontoall
danger,warningandcautionstatements.Failuretodosocouldresultinseriousinjurytotheoperatoror
damagetotheequipment.Makesurethattheprotectionprovidedbythisequipmentisnotimpaired,donot
useorinstallthisequipmentinanymannerotherthanthatspecifiedinthismanual.Iftheequipmentisused
inamannernotspecifiedbythemanufacturer,theprotectionprovidedbythemanufacturermaybeimpaired.

2.1UseofHazardInformation
DANGER:
Indicatesapotentiallyorimminentlyhazardoussituationwhich,ifnotavoided,couldresultindeathor
seriousinjury.
CAUTION:
Indicatesapotentiallyhazardoussituationthatmayresultinminorormoderateinjury.
Important:
Indicatesasituationwhich,ifnotavoided,maycausedamagetotheinstrument.Informationthatrequires
specialemphasis.
Note:
Informationthatsupplementspointsinthemaintext.

2.2PrecautionaryLabels

Readalllabelsandtagsattachedtotheinstrument.Personalinjuryordamagetotheinstrumentcouldoccurif
notobserved.Asymbol,ifnotedontheinstrument,willbeincludedwithadangerorcautionstatementinthe
manual.Thissymbol,ifnotedontheinstrument,referencestheinstructionmanualforoperationand/orsafety
information.ElectricalequipmentmarkedwiththissymbolmaynotbedisposedofinEuropeanpublicdisposal
systemsafter12Augustof2005.InconformitywithEuropeanlocalandnationalregulations(EUDirective
2002/96/EC),Europeanelectricalequipmentusersmustnowreturnoldorend‐oflifeequipmenttothe
manufacturerfordisposalatnochargetotheuser.

Note:Forreturnforrecycling,pleasecontacttheequipmentproducerorsupplierforinstructionsonhowto
returnend‐of‐lifeequipment,producer‐suppliedelectricalaccessoriesandallauxiliaryitemsforproper
disposal.


AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page4of16

2.3ChemicalandBiologicalSafety

DANGER:PotentialChemical/BiologicalExposureHazards.Handlingchemicalsamples,standardsandreagents
canbedangerous.Usersofthisproductareadvisedtofamiliarizethemselveswithsafetyproceduresandthe
correctuseofchemicals,andtocarefullyreadallrelevantMaterialSafetyDataSheets.

‐Normaloperationofthisinstrumentmayinvolvetheuseofhazardouschemicalsorbiologicallyharmful
samples.

‐Theusermustobserveallcautionaryinformationprintedontheoriginalsolutioncontainersandsafety
datasheetpriortotheiruse.

‐Allwastesolutionsmustbedisposedinaccordancewithlocalandnationallaw.

‐Thetypeofprotectiveequipmentmustbeselectedaccordingtotheconcentrationandamountofthe
dangeroussubstanceatthespecificworkplace.


3GeneralInformation

3.1OverviewofProduct

TheAND1100™Fluorimeterenablesfieldtestingtobedoneintwoshortsteps,
eliminatingorsignificantlyreducingtestexpenditures,complexityandwaittimes
experiencedintraditionalheavymetalswatertesting.TheFluorimeterusesan
extremelysensitivemeasurementtechniquetodeterminethelevelsofheavymetal
contaminants.ANDalyze’sproprietaryCatalyticDNAsensorsuseaDNAzymereaction
thatfluorescesinthepresenceofatargetcontaminantsubstancesuchaslead.The
fluorescenceofthereactionismeasuredtodeterminetheconcentrationofthe
targetheavymetalandisreportedinpartsperbillion(ppb).Usingtheproducttotest
formetalsisasimple,quantitativetestthatallowsfordetectionatorbelowUSEPA
standardsindrinkingwater.Thetestisperformedbytakinga1milliliterwater
sample,injectingitthroughthesensor,andintotheAND1100fluorimeter.This
sampleisthenautomaticallyanalyzedandreportsresultsinlessthantwominutes.


AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page5of16

3.2GeneralFeatures

‐FluorescenceBasedSensing‐Reactionproducesquantitativefluorescencebasedresults.
‐SensorKit‐Eachdisposablecolor‐codedsensorisdesignedforaspecificheavymetaltesttarget.Kits
providesamplingtubes,syringesandcuvettes.
‐Rugged&ConvenientConstruction‐Tough,lightweight,impact‐resistantshellprotectselectronicsand
sensors.
‐KeyPad‐Quick,simplemenunavigationandonebuttonpushforsampleanalysis.
‐DataCapture&Reporting‐Timeanddatestampedwithsamplenumber.Storesover240,000testresults
andover75sitelocations.DatadownloadablethroughaUSBcable.
‐WaterandDustResistant‐Enclosurerating‐IP54
‐Portable‐BatteryoperatedandrechargeablethroughUSBcableortraditionaladapter.

Note:Seefulltechnicalspecificationsinsection9.

3.3AnalyteSensors

Tohelpdifferentiatethematerialsspecifictoeachanalyte,thepackaginghasbeenindividuallylabeledandthe
sensorhousingsandbuffercapshavebeencolorcoded.ForExample:

Green–Lead(Pb)Orange–Uranium(U)Blue–Copper(Cu)Gray–Mercury(Hg)Etc.

Note:Insomecasesthesamesensorisusedformultiplerangesbutdifferentbuffersand/ormethodswill
accompanythesensorinthekit.

3.4OperatingEnvironmentandStorage

Important:Thefollowingconditionsarenecessarytoensurecorrectinstrumentoperationandaccurate
results:

‐Placetheinstrumentfirmlyonaflatandevensurfaceduringoperation.
‐Maintainanambienttemperatureof10to40ºC(50to104ºF)forproperfluorimeteroperation.Pleasesee
themanualforeachspecificmetalforsensorspecifictemperatureconditions.
‐Therelativehumidityshouldbelessthan80%;moistureshouldnotcondenseontheinstrument.
‐Donotoperateorstoretheinstrumentinextremelydusty,damporwetlocations.
‐Keepthesurfaceoftheinstrument,thecellcompartmentandallaccessoriescleananddry.
‐Sensorsarestableupto1yearifstoredat<23˚C(74˚F),<50%R.H.awayfromdirectsunlight.Forbest
results,storeintherefrigerator(4˚C/39˚F).
‐Buffersolutionsarestableupto6monthsifstoredat<23˚C(74˚F)butifstoredintherefrigerator
(4˚C/39˚F),canbeusedforupto1year.
‐Sensorsandbuffersmustbebroughttoroomtemperaturebeforeuse.
AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page6of16

3.5WaterTestingGuidelines

Note:Thisisadrinkingwatertestkit.Ifyouaretestingothermatricessuchassurfacewater,groundwateror
industrialwater;contactANDalyzecustomerserviceforadditionalapplicationnotes.

WaterSampling
‐Forbestresultsusefreshlycollectedsample(unpreserved)foranalysis.Werecommendthatyouusethe
samplewithin1hour(maximumof2hours)ofcollectiontominimizeanymetallosstothewallsofthe
samplecontainer.Thisisparticularlyimportantfortestingtraceleadlevels.
‐OncethesampleismixedwithANDalyzesamplebuffer,testwithin15minutes.
pHRange
‐TestshaveshownthatenvironmentalsamplespreservedinacidtoapH<2cannotusuallybebroughtto
anappropriatepHwhenmixedwiththeANDalyzebuffer.Thesesamplesmustbefirstneutralizedwith
NaOHtoasuitablepHbeforemixingwithANDalyzebuffer.PleasecontactANDalyzecustomerservicefor
instructionsrelatedtopre‐treatmentofhighlyacidicorhighlybasicsamples.

4Set‐UpandOperatingProcedures
4.1ChargingtheBattery

Important:Beforeusingtheinstrumentassurethatthefluorimeterdeviceischarged.

ToConfirmtheBatteryLevel:
1. TurnPowerONbypressingandHoldingtheON/OFFbutton.(See
HorizontalArrowonPhototoRight)

2. Oncedeviceisinitiated,confirmthebatteryindicatorstatusisgreen
(SeeVerticalArrowonPhototoRight).Thebatteryindicatorshouldnot
bered(asseeninthephototoright)

ChargingtheBattery:
1. ReleasetheRubberBootatthebottomofthemeter.

2. AttachtheUSBtoMINI‐BCabletotheoutletatthebottomofthe
deviceandtoaUSBpoweredsourcesuchasacomputeroraUSB
charger.(SeePhototoRight).

Note:Whenthemeterischarging,thebatteryindicatorwilldisplayyellow
linesacrossandaroundtheindicator.




AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page7of16

4.2HomeTestingScreen

Optionscanbeaccessedby:
‐Usingtheup/downbuttonsandpressingSELECTbuttonwhenthearrowistotheleftofthedesiredoption.
‐ORbypressingthebuttonlocatedjustbelowthescreenunderthedesiredoptionsatbottom.

SelectableOptions:
1) Sensor‐SelectswhichSensorwillbetested.

2) Site‐Usessitespecificcalibrationsforaccuratetests.

3) Start‐Tobeginatestoftheindicatedmetal.

4) Menu‐Tochangeorviewinternalsettings.



4.3SensorScreen



SelectingtheSensoroptionallowstheusertochangethesensor
whichthemeterwilldetect.

Onceselected,theColoroftheareasurroundingthenameof
thesensorandthephotoofthesensorattherightontheHome
Screenwillmatchthecolorofthesensoritselfusedfortesting.
4.4SiteScreen

SelectingtheSiteoptionallowstheusertochangeorcreatenewsites.

Sitesareimportantbecauseeachwatersamplecontainsdifferences
whichaffecttestingresults.Itisrecommendedthateachlocationis
calibrated.Sobyselectingasite,thiscalibrationcanbeusedrepeatedly.



Pleaseseesection5belowforfurtherinstructionsonsettingupandcalibratingsites.

4.5StartOption

DependingonyourOptionssettings,
pressingtheStartbuttonwilleither
startthetestinginstructionsOR,if
turnedOFF,willbeginthetesting
immediately(foradvancedusers).
1
2
34
AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page8of16

4.6MenuScreen



TheMenuScreencanbeaccessedbypressingthebuttonlocatedbelowthescreen
underMenu.

SelectableOptions:
1. Results
2. Settings
3. Options
4. Advanced
5. About
6. Date/Time
7. Help

OptionsareselectableusingthearrowkeysandpressingSelectbuttonorby
pressingthebuttonlocatedbelowthescreenundertheSelect.

Results
‐Displaystheresultsofthepreviousfourtests.
‐Scrollupordowntoviewadditionalresults




Settings



‐Units:Determinesunitsfordisplayingmetal
concentration.Youcanchangethistoppb,ppm,
µMornM.

‐Minppb:Atconcentrationslessthenminppb,
“BelowLimit”isdisplayed.

‐Maxppb:Atconcentrationsgreaterthanmax
ppb,“AboveMax”isdisplayed.

‐Decimals:Changesthenumberofdecimals
displayedintheresults.Undersome
circumstancesthedecimalswillbedefaulted.

Note:TheDefaultvaluesarethesettings
recommendedbythemanufacturer.
About
‐DisplaystheHardwareandFirmwareversions
ofthemeterincludingthemostrecentdate
thefirmwarewasupdated.

Note:Updatefirmwareifithasnotbeendone
recently.

DateandTime
‐Setthedateandtimetomaintainaccurate
history.EditusingthearrowkeysandSave



Help
‐DisplaysHow‐Tovideoshowingasample
calibrationorprovideslinktoviewvideos
online.
AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page9of16

Options

PowerSave:Thissetstheminuteswhenthedisplay
willturnblacktosavepower.Settablefrom1to5
minutes.

PowerOff:Thisistheminuteswhichtheunitwill
turnitselfoffafteritgoesintopowersavemode.
Settablefrom6to15minutes.

CuvetteRemoval:TurningthisONwillcausea
screentoappearaftereachtestremindingtheuser
toremovethecuvettefromthesamplechamber.

TestInstructions:IntheONposition,4‐step
instructionswilldisplayinorderwhenTestis
selectedfromthehomescreen.Afterthe4thstep,
thetestwillbegin.IntheOFFposition,thetesting
beginsimmediately.




SensorLot:TurningthisoptionONallowstheuserto
enterthelotnumberofthesensorsatthetimeofa
sitecalibration.Thishelpstoidentifywhenanewsite
calibrationisrequired(newlotbatch).

Temperature:TurningthisoptionONallowstheuser
toenterthetemperatureofthesamplewateratthe
timeofasitecalibration.Thishelpstoidentifywhena
newsitecalibrationisrequired(significantchangein
watertemperature).

TemperatureUnits:SwitchbetweenCelsiusand
Fahrenheitfortemperatureentryanddisplays.

DisplayContrast:Adjuststhebrightnessofthedisplay.
Highervalueindicatesabrighterdisplay.

Language:ChoosebetweenEnglish,Chinese,
Portuguese,SpanishandGermanlanguagesforthe
fluorimeterdisplays.
Advanced

UsercanViewandSetparametersforeachspecificsensor.Edit
valuesusingup/downkeysandSELECT.

Window:Determinesthenumberofpointsthatwillbeusedto
calculatetherateoffluorescenceincreaseovertime.Weusea
movingaveragemethodtocalculatemaximumrate.(Ifthevalueof
windowandsamplesissame,thentheaverageslopeiscalculated)

Samples:Totaldatapointstakenwhichdeterminestimeuptowhich
dataiscollected.Avalueof100isapproximately20seconds.Increase
samplevalue,toincreasedatacollectiontime.Generallyanincrease
insamplescanincreasesensitivityespeciallyinmatricessuchaswaste
water.







AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page10of16

Scale:Slopeofthelinethatcorrelatestherateoffluorescence
increasevs.concentration.Thisvalueisdeterminedbythe3‐
pointcalibration.Somesensorshaveadefaultscalethatis
automaticallycalculatedfromothersensor’sscales.Inthis
case,“Auto”willbeindicatedinthedefaultscalefield.Anew
3‐pointcalibrationwillsupersedethedefaultsettings.

Offset:Interceptofthelinethatcorrelatestherateof
fluorescenceincreasevs.concentration.Alsoautomatically
determinedbythe3‐pointcalibration.

Delay:Pointsfromthestartofthetestwhich
arenotusedforcalculations.Thisisused
whenthereisadelayedreactionincertain
sensors.

R2:R‐Squaredisastatisticaltermindicating
howgoodonetermisatpredictinganother.In
ourcase,thisisderivedfromthebestfitline
createdbythe3‐pointlabcalibration.Values
closerto1arebetter.
3‐PointLabCalibration

Usingtheleft/rightkeysinthe
Advancedmenu,Selectthe
Calibrateoption.



Whenall(3)solutionshavebeen
tested,thecalibrationresults
screenwillbedisplayed.Click
Applytosavetheresults.The
settingswillthenbeviewablein
theAdvancedscreen.


Note:Thisisforlaboratory
calibrationONLYand
requires(3)standardized
solutionsfrom
manufacturerorprepared
byalaboratorytechnician.

FactoryReset

TheResetoptionwillrestorethesettingofthat
specificsensortofactorysettings.

Usingtheleft/rightkeysintheAdvancedmenu,
SelecttheResetoption.



SelectResetontheWarningscreentocomplete
theprocess.

4.7FirmwareUpdate

ANDalyzeiscontinuallyaddingnewfeaturesandoptionstotheAND1100throughnewfirmwareupdates.To
updateyourAND1100firmware,pleasevisitandalyze.comandlogintoregisteryourdevice.Onceloggedin,
youwillseetheoptiontoupdateyourdevice.Followtheinstructions.






AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page11of16

4.8InstraCommLite

Thissoftwareallowsuserstolaunchandrecordlivesampleresults,downloadallsampleresultsfromthe
instrumentandperformrelatedconfigurationtasks.TheInstraCommapplicationwasdevelopedby
AutonomousInnovations,Inc.andacomplimentary"Lite"versionhasbeenmadeavailablewiththepurchase
ofANDalyze'sAND1100.TodownloadyourcomplimentaryversionofInstraCommLite,pleasevisit
andalyze.comandlogintoregisteryourdevice.Foradditionalinformationandoperatinginstructionssee
InstraCommLiteSolutionsNotes(Availableseparately).

5SiteSet‐UpandManagement
Thissectionprovidesinformationconcerningtheset‐upandmanagementofSites.Asdifferentsample
locationscontaindifferentwatercharacteristics(samplematrix)itisrecommendedtocreateanewsitefor
eachsamplearea.Thisallowstheusertotrackandcomparetheresultsateachlocationwhilealsonormalizes
theresultsthroughanon‐sitecalibration.

5.1 NewSiteSet‐Up

Sitesareimportantbecauseeachwatersamplecontainsdifferenceswhichaffecttestingresults.Itis
recommendedthateachlocationissitecalibrated.Sobycreating,calibratingandselectingthatsiteinthe
future,thiscalibrationcanbeusedrepeatedly.PleaseseethespecificTestingandCalibrationManualfor
detailedinstructions.

Tocreateanewsite:

1. Scrolltothe“Site”fieldand
press“Select”.
2. Usingtheup‐down
keys,select“NewSite”.
3. Enterthenameofthelocation(8
charactersmax)andpress“Save”.
  
4. IfSensorLotoptionis
turnedON,youwill
havetheoptionto
enterthelotofthe
sensors.
5. IftheTemperature
optionisturnedON,
youwillhavethe
optiontoenterthe
temperatureofthe
samplewater.
AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page12of16

5.2 On‐SiteCalibration

Whentestingwateratanewlocation,theinstrumenthastobecalibratedforaccuratereadings.Completing
theOn‐SiteCalibrationrequirestwoseparatesolutionsandtests.Attheendoftheprocess,thefluorimeteris
calibratedfortestingwateratthatsiteandalsoprovidestheleadtestresultsforthissite.Acalibrationmay
onlybeaccuratewhenoriginallyperformed.Donotrelyonacalibrationtobeaccurateoverlongtimeperiods
asenvironmentalsamplescanvarygreatly.

TheOn‐Sitecalibrationprocessadjustsforchangesinsamplematrix(suchasionicstrengthandpH)by
calculatingtherecoveryofaknownspikeinthatmatrix.Whenauserperformsanon‐sitecalibrationforLead,
forexample,thefluorimetersoftwarecalculatestheppbofleadinthesample(a)andtheppbofleadin
samplespikedwith25ppblead(b)usingthecalibrationcurvefromthe3pointcalibration(initiallycompleted
atfactoryorinlaboratory).Itthencalculatesanaccuracyfactorbasedon%recovery.Accuracyfactor=25/b‐a.
Thisfactorisstoredforaparticularsitethattheuserenters.Thesampleconcentrationdisplayedafteranon‐
sitecalibration=a*accuracyfactor.

Important:On‐Sitecalibrationmustbeperformediftestingisbeingdone:
‐Atanewlocationwhichhasnotbeenpreviouslysaved.
‐Toadjustforsignificantchangesinsamplingtemperature.
‐Whenbeginningtouseanewlotofsensors.

SeethespecificTestingandCalibrationManualfordetailedinstructions.

5.3 SelectingaSite

1. OntheHomescreenselecttheSiteoption.
2. Usingtheup‐downkeyshighlightthe
desiredsiteandpressSelect.
3. OntheSite/SensorDetails,chooseSelect.

5.4 DeletingaSite

1. WhenintheSite/SensorDetailsscreen,
usingtheleft‐rightkeys,scrolltoDeleteand
select.
2. OntheConfirmDeletescreen,selectDelete.


AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page13of16

5.5 Re‐CalibratingaSite

1. WhenintheSite/SensorDetails
screen,usingtheleft‐rightkeys,
scrolltoCalibrateandselect.
2. Followinstructiontocalibrate.

6Maintenance

CAUTION:Alwaysdisconnectpowerfromtheinstrumentbeforeattemptinganycleaningoperations.

Important:Undernocircumstancesshouldtheinstrument,displayortheaccessoriesbecleanedwithsolvents
suchaswhitespirit,acetone,etc.

Fluorimeter

‐Cleantheenclosure,samplecellcompartmentsandallaccessorieswithasoftdampcloth.Amildsoap
solutioncanalsobeused.Donotgetexcesswaterinthesamplecellcompartments.Donotinsertabrush
orsharpobjectintosamplecellcompartmenttoavoiddamagingthemechanicalcomponents.Drythe
cleanedpartscarefullywithasoftcottoncloth.

Display

‐Takecarenottoscratchthedisplay.Donottouchthescreenwithballpointpens,pencilsorsimilarpointed
objects.
‐Cleanthedisplaywithasoft,lint‐freeandoil‐freecottoncloth.
‐Dilutedwindowcleanerliquidcanalsobeused.

SampleChamber

‐Splashesorspillsonandintheinstrumentshouldbecleanedupimmediately.Removeanyliquidinsidethe
samplechamberbyusinganon‐lintswab. Donottiptheinstrumenttoemptyliquidfromthesample
chamber.

Sensors

‐Neverreusesensors!Oncethesensorbagisexposedtoair,itshouldbeusedimmediately(within30
minutes)asexposuretohumidity,airand/orheatcanaffectitseffectiveness.Disposeusedsensorandall
otherusedconsumablessuchassampletubes,syringes,plasticpipettesaftereachanalysis.

Pipettes

‐The100µLautomaticpipettecanbereusedasitusesdisposablepipettetips.Pipettetipsshouldbe
disposedofimmediatelyafteruse.
AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page14of16

7CustomerServiceContactInformation
ContactusbyEmail:
[email protected]

ByTelephone:
8883880818(TollFreeinUS)or+18574536740

ByFax:
+18573861277

BusinessServiceHours:
9:00amto5:00pmEasternStandardTime(USA)MondaythroughFriday

CompanyAddress:
ANDalyzeInc.
800BoylstonStreet
PrudentialTower,Floor16
Boston,MA02199USA

Website:
www.andalyze.com

8ConsumablesandReplacementItems
‐FluorimeterKit
oCapableofmeasuringmultiplemetals.
oKitIncludes:Fluorimeter
USBtoMINI‐BCable
100µLFixedVolumePipetteandTips
pHTestStrips
‐SensorKits
oEquipmentfor(25)Testsand/orCalibrations
KitIncludes:(25)SensorBagswithSensor&Cuvette
(25)SampleTubes(withbuffer)
(25)1mLSyringes
(25)DisposableTransferPipettes(NotincludedinHighRangeCopper)
8mLAnalyteStandardSolution
InstructionManuals
MaterialSafetyDataSheets(MSDS)

AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page15of16

9TechnicalSpecificationforAND1100Fluorimeter
TheAND1100Fluorimeterisaplatformcentrichand‐helddeviceforheavymetalstestingfordrinkingwater
andindustrialwatersupplies.Itusesanextremelysensitivemeasurementtechniquetodeterminethelevelsof
heavymetalcontamination.
InstrumentSpecifications:
LightSource:LightEmittingDiode
Detector:PhotomultiplierTube
Filters: Excitation485nm
Emission535nm
Certifications:
CEMarked:ComplieswiththefollowingEuropeanUnionDirectives:
LowVoltageElectricalEquipmentDirective2006/95/EC
ElectromagneticCompatibility(EMC)Directive2004/108/EC
PowerSpecifications
Power: BatteryOperated
PowerRecharging:viaUSBcable
Powerusage:Thebatterylastsover225testsonafullcharge(at45seconds/test).
MechanicalSpecifications

Dimensions:Width:3.6inches/9.14cm
Depth:8.0inches/20.3cm
Height:2.25inches/5.72cm
Weight:1.25lb/565g
IPratingIP54(water&dustresistant)
Dataspecifications
Interface:USB2.0
Data: Downloadedincsvformat

Note:Allspecificationsaresubjecttochangewithoutnotice.



AND1100Fluorimeter
AND1100FluorimeterforWaterTestingJuly1,2013


ANDalyzeInc.800BoylstonStreet,PrudentialTower,Floor16,Boston,MA02199USATel.+1857.453.6740www.andalyze.com

Page16of16

ANDalyzeInc.
2109SOakStreet,Suite102
Champaign,IL61820USA

Email:[email protected]
Telephone:+18883880818or+12173280045
10 LimitedWarrantyforFluorimeterDevice

ANDalyzeprovidesthefollowinglimitedwarrantyfortheANDalyzeAND1100Fluorimeterdevice(the
“FluorimeterDevice”).ANDalyzewarrantsthattheFluorimeterDevice(includingthesoftwareusedonthe
FluorimeterDevice),whenusedinaccordancewiththeuserdocumentation,willoperateinallmaterial
respectsinconformitywiththespecificationsstatedintheuserdocumentationforaperiodofninety(90)days
fromthedateofyourreceipt(the“WarrantyPeriod”).Ifitdoesnot,yoursoleremedyandANDalyze’stotal
liabilityforsuchmaterialnonconformityintheFluorimeterDevicewillbe,atANDalyze’soptionanddiscretion,
torepairorreplacetheFluorimeterDeviceatANDalyze’sexpenseortorefundthepurchaseprice(butnotany
taxes,exportorshippingfees)andsubjecttothelimitationsinTheforegoingremedyissubjecttothe
LimitationofLiabilityasdescribedwherein.Toqualify,youmustnotifyANDalyzeduringtheWarrantyPeriod
ofanyproblemsthatyouexperiencewiththeFluorimeterDevice.ANDalyzewillhavenoliabilityforany
nonconformityofwhichyoufailtonotifyANDalyzepriortotheexpirationoftheWarrantyPeriod.This
warrantydoesnotapplyto(i)FluorimeterDevicewhichhasbeenusedinamannerotherthanasauthorized
undertheseTermsandthedocumentationprovidedwiththeFluorimeterDevice(includingtheproduct
brochureandfluorimeterspecifications);(ii)anysoftwareontheFluorimeterDevicethathasbeenmodifiedby
youoranypartyotherthanANDalyzeorwhichhasbeenimproperlyinstalledtotheextentsuchmodification
orimproperinstallationcausedthebreachofwarranty;(iii)failurescausedbyaccident,neglect,failureto
maintainasuitableoperatingenvironment,tampering,oranyothereventotherthanordinaryuse.





Popular Measuring Instrument manuals by other brands

Vortex BLUEONE BWO 155ZM Original operating instructions

Vortex

Vortex BLUEONE BWO 155ZM Original operating instructions

Power Probe BASIC instruction manual

Power Probe

Power Probe BASIC instruction manual

smartheart 19-006 instruction manual

smartheart

smartheart 19-006 instruction manual

SWI L500S Installation and operation manual

SWI

SWI L500S Installation and operation manual

Innova OBD2 owner's manual

Innova

Innova OBD2 owner's manual

Rototherm AM60 instruction manual

Rototherm

Rototherm AM60 instruction manual

Bosch PLR 25 Original instructions

Bosch

Bosch PLR 25 Original instructions

Northern Design PowerRail300 operating guide

Northern Design

Northern Design PowerRail300 operating guide

Anton Paar Lovis 2000 M/ME Standard operating procedure

Anton Paar

Anton Paar Lovis 2000 M/ME Standard operating procedure

jenway 6305 operating manual

jenway

jenway 6305 operating manual

KERN ABP Set up and operating manual

KERN

KERN ABP Set up and operating manual

AFRISO DIM 20 instruction manual

AFRISO

AFRISO DIM 20 instruction manual

Endress+Hauser Mycom CLM 152 operating manual

Endress+Hauser

Endress+Hauser Mycom CLM 152 operating manual

Mitsubishi Electric ME96NSR Operator's manual

Mitsubishi Electric

Mitsubishi Electric ME96NSR Operator's manual

Ametek Z-tron III Installation and operating istructions

Ametek

Ametek Z-tron III Installation and operating istructions

Anemoment TriSonica Mini user manual

Anemoment

Anemoment TriSonica Mini user manual

Landis+Gyr ULTRAHEAT 50 Series Glossary

Landis+Gyr

Landis+Gyr ULTRAHEAT 50 Series Glossary

UE Systems Ultraprobe 15000 Touch instruction manual

UE Systems

UE Systems Ultraprobe 15000 Touch instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.