4 
Type of Sensing : Induction coil sensing 
 for AC current. 
Temperature Coefficient : 0.2x(Spec Acc'y)/ 
 ℃, < 18℃or > 28℃ 
Maximum Output Impedance : 75Ω. 
Maximum Jaw Opening : 30mm. 
Maximum Conductor Size : 29mm diameter. 
Size : 72mm (W) x 148mm (L) x 36mm(D) 
 W/O cable. 
Weight : 250grams. 
Indoor use. 
Maximum Altitude : 2000 meter. 
Accessories : Manual , Carrying Case. 
 
Precautions and Preparations for 
measurement 
 
1. Do not apply the voltage to the output plugs. 
 
2. Do not use or store this instrument in a 
 high temperature or high humidity 
 environment and do not store the unit in 
 direct sunlight. 
5 
3. Do not measure current before the unit 
 is not combined with DMM. 
4. If the instrument is used near noise gen- 
 erating equipment , be aware that output 
 voltage may become unstable or get more 
 errors. 
 
THIS INSTRUMENT MUST NOT BE 
 USED ON UNINSULATED CONDUCT- 
 ORS AT A VOLTAGE GREATER THAN 
 250V ac/dc. 
 
Electriacl Specification 
Accuracy is ±(%reading + number of 
Ampere) at 23℃±5 ℃, less than 75% R.H. 
Range Measure Output Accuracy 
400A 
3Amp 3mV 
30Amp 30mV 
350Amp 350mV 
400Amp 400mV ±(3.2% + 1A) 
 50 ~ 60Hz 
±(1.9% + 0.5A) 
 50 ~ 60Hz