Aquatronica ACQ210-D User manual

IsACQ210D-0 Rev. 250907-2 1 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
High Range Conductivity
Interface (Density)
ACQ210-D
Instruction Manual

IsACQ210D-2 Rev. 250907-2
2 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
Table of Contents
• General Information............................................................................ Page. 3
• Pack contents...................................................................................... Page. 3
• Connection Diagram ............................................................................ Page. 3
•Connection to System.......................................................................... Page. 3
•Displaying the read value .................................................................... Page. 4
•Density Menu ...................................................................................... Page. 4
•Disconnect.......................................................................................... Page. 10
•Suggestions for an accurate reading .................................................. Page. 11
•Waste diposal legislation.................................................................... Page. 12
•Warranty ............................................................................................ Page. 13
•Declaration of conformity ................................................................... Page. 14
•Conversion table ................................................................................ Page. 15

IsACQ210D-0 Rev. 250907-2 3 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
Water density (or conductivity) is one of the parameters fundamental for the survival of
sh and invertebrates in an aquarium.
The amount of salt in the water can be established in various ways with different units of
measure. Electronically, however, electric conductivity can be measured once and through
a series of mathematical calculations the same quantity can be expressed with different
measurement units.
Aquatronica customers can therefore view the saltwater density in four different ways,
as they prefer.
So it is possible to choose which of the following values will be displayed:
Conductivity: Representing the content of mineral salts dissolved in water. It can be
measured electrically and is expressed in mS/cm (milliSiemens per centimeter) or µS/cm
(microSiemens per centimeter).
Salinity: Representing the amount of salts dissolved in a water sample. It can be expressed
in parts per thousand (ppt or PSU).
Density: Representing the ratio between the mass and volume of a liquid. Usually ex-
pressed for water in g/l (grams per liter).
Specic Gravity: Also called relative density, it is a dimensionless value representing the ra-
tio between the density of a liquid and the density of the same volume of water at 4°C.
A table on the back shows these values in their relative units of measure.
Aquatronica’s High-Range Conductivity Interface connects an Aquatronica high con-
ductivity probe to the “Aquarium Controller” system to measure and control the density
(conductivity) in aquariums.
English
Pack contents
You will nd:
One high conductivity probe connection interface.•
One BUS cable for connecting the interface to the power unit.•
Connection to “Aquarium Controller” system
Connect the conductivity probe’s connector to the interface (ACQ210-D).1.
2. Connect the ACQ210-D interface to the power unit (or HUB) using the provided BUS
cable
NOTE: Insert the correct end of the connector into the power unit; inserting it in
the other direction can seriously damage the equipment.
HIGH CONDUCTIVITY
PROBE
Connection diagram
BUS CABLE
ACQ210-D

IsACQ210D-0 Rev. 250907-2 4 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
(Fig. 1)
New device connected
S01
Density
(Fig. 2)
Mon 11/06/07 15:05
PU01 A B C D E F G H
Density 1.0232
After connection, the control unit will display a Plug-In
screen (Fig. 1), where a name can be assigned to the
connected sensor.
It may take several seconds for the control unit to
recognize the connected interface.
NOTE: The sensor’s name can be changed using
the control unit’s keypad.
If more than one of these sensors is connected, the user
may assign different names for each one to facilitate
menu navigation.
Displaying the read value
After the probe has been connected through the ap-
propriate interface, the values read by the probe will
be displayed on the main screen.
If several sensors were connected, their values can be
checked in order by pressing the and keys.
Density Menu
Once the probe and interface are connected, the “Den-
sity” menu will appear in the “Main Menu”, where all
of its settings can be programmed. All sensor menus
have the same structure in order to make them more
intuitive and simple.
In this menu it is possible to change the name of the
connected sensor, calibrate it in order to obtain a more
accurate reading, enable an acoustic or visual alarm,
read the conductivity (density) of the previous day,
week or month (selected in the “Settings” menu) and
enable/disable the outputs based on the read value.
Change name
This option modies the name given to the sensor
(Fig. 4).
To use this option, proceed as follows:
Main screen Main Menu Density Change Name.
• Select the letter to insert using the keys and move
within the word using the keys. When nished,
press “Enter”.
English
(Fig. 3)
Density
Change Name
Programs
Data Record
Alarm
Calibrate Sensor
Measurement Unit
(Fig. 4)
Dens_

IsACQ210D-2 Rev. 250907-2
5 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
Programs
This option is used to create programs based on the
Density value.
To set up a program, the following must be estab-
lished:
- A reference value, dened as the Density for the
tank, which the system will aim to maintain.
- A ∆Max value, dened as the tolerance that must
be observed with respect to values higher than the
reference value.
For example: If the reference is set to 1.0240 and the
∆ Max is 0.0010, the system will be enabled at values
higher than 1.0250 (see chart).
- A ∆Min value, dened as the tolerance that must
be observed with respect to values lower than the
reference value.
For example: If the reference is set to 1.024 and the
∆ Min is 0.0020, the system will be enabled at values
lower than 1.0220 (see chart).
To insert a program, proceed as follows:
Main screen Main Menu Density Programs.
• Select “Insert” using the keys and press
“Enter” (Fig. 5).
Insert
Here it is possible to choose a desired Conductivity
(density) value as well as set
a minimum and maximum tolerance (Ex. Fig. 6).
To set this program, proceed as follows:
Main screen Main Menu Density Programs In-
sert.
• Select the relative conductivity value using the
keys and set the desired value using the keys.
• Set the “∆MIN” and “∆MAX” tolerances; select the
desired parameters using the keys and modify their
values using the keys.
• Select “Sockets” using the keys to choose
how the outputs will function when the conductivity
level goes above or below the set values. Then press
“Enter”.
• Select the outputs to be controlled using the keys.
The selected output will blink on both lines.
The outputs on the upper line determine which devices
must be enabled/disabled when the conductivity (den-
sity) goes above the set maximum value (Ref. + ∆MAX);
the outputs on the upper line are set using the key.
(Fig. 5)
Density
Insert
English
(Fig. 6)
Programs
Ref. 1.0240
∆MIN: ∆MAX:
0.0020 0.0010
Sockets Conrm
1.0240
1.0220
1.0250
Density
Time
Ref.+∆Max
Key to symbols
Values for which the system is enabled
∆Max=0.0010
∆Min=0.0020
Ref.
Ref.-∆Min

IsACQ210D-0 Rev. 250907-2 6 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
The example in (Fig. 7) shows enabling of the socket
“C”: pressing the key once, the socket is highlighted
and the symbol D appears above it.
To disable a socket, press the key a second time;
in this case, only the symbol D will appear above the
selected socket.
The outputs on the lower line determine which devices
will be enabled/disabled when the conductivity goes
below the set minimum value (Ref. - ∆MIN); the outputs
on the lower line are set using the key.
The example in (Fig. 7) shows disabling of the socket
“C”; only the symbol D appears.
After the desired sockets have been set, press “Enter”
to return to the previous menu (Ex. Fig. 6).
The “Conrm” eld is automatically selected: by press-
ing “Enter” the inserted program is saved.
If different conductivity (density) values are desired
based on the time of day, these can be set graphi-
cally.
To program this function, proceed as follows:
Main screen Main menu Density Programs In-
sert.
• Using the keys, select the conductivity reference
parameter (Ref.). Using the keys, select “Curve”
(Ex. Fig. 8). Afterwards move to the /\/\/\/\ symbol
and press “Enter”.
A screen opens where the desired conductivity value can
be graphically modeled over 24 hours (Ex. Fig. 9).
• Using the keys, select the time of day (in 2-hour
intervals, lower left corner). Using the keys, modify
the conductivity value (lower right corner) for the se-
lected time. When nished, press the “Enter” key.
(Fig. 7)
Top-off
PU01 A B C D E F G H
PU01 A B C D E F G H
D
D
English
(Fig. 8)
Programs
Curve /\/\/\/\
∆MIN: ∆MAX:
0.0020 0.0010
Sockets Conrm
(Fig. 9)
18:00 1.0232

IsACQ210D-2 Rev. 250907-2
7 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
English
View/Mod/Del
In this menu the inserted programs can be viewed
(View), modied (Mod) or deleted (Del).
To use this function, proceed as follows:
• Access the program to be modied or deleted by
pressing “Enter” on the “View/Mod/Del” eld.
•Use the keys to view the desired program (Ex.
Fig. 8).
• Press the “Enter” key. The specic screen appears
on the display (Fig. 10).
• Select “Mod” using the keys to modify the pro-
gram or change the desired parameters. Then press
“Enter” to conrm the changes.
• Select “Delete Program” using the keys to
delete the program. The delete screen will appear (Fig.
11). Press “Enter” to delete or “Esc” to cancel.
Delete All (Fig. 12)
In this menu all of the programs inserted in the menu
can be deleted at one time. To use this function, pro-
ceed as follows:
• Select “Delete All” using the keys and press
“Enter”. The delete screen will appear. Press “Enter”
to delete or “Esc” to cancel.
(Fig. 10)
Do you want to
modify or delete?
Modify
Delete Program
(Fig. 12)
Do you want to
delete all of
the programs?
Enter: Conrm
Esc: Cancel
(Fig. 11)
Do you want to
delete this
program?
Enter: Conrm
Esc: Cancel

IsACQ210D-0 Rev. 250907-2 8 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
Data Record
The Data Record graphically displays variations in the
conductivity (density) during the previous 24 hours with
a minimum interval of 30 minutes (Ex. Fig. 13).
To display the data, proceed as follows:
Main screen Main menu Density Data Record.
• Using the keys, select the maximum (MAX), mini-
mum (MIN) or current conductivity (density). Using the
keys, move within the chart to view the conductivity
of a given time. Press “Enter” when nished.
Alarm
A visual or acoustic alarm can be set to notify the user
if the conductivity level goes below or above the “Less
than” or “Greater than” values (Ex. Fig. 14).
If the conductivity value exceeds these limits, the con-
ductivity value on the main screen will blink if the alarm
is set to “No Sound”.
If the alarm is set to “Sound”, the value will blink, an
acoustic signal will sound and the icon will appear
on the main screen.
To program this function, proceed as follows:
Main screen Main menu Density Alarm.
• Select the desired option using the keys:
OFF = alarm disabled
Sound = alarm and acoustic signal enabled
No Sound = alarm enabled and acoustic signal disa-
bled
• Select “Less than” using the keys and then
“Greater than” and with the keys program the
desired conductivity to set the limits beyond which the
alarm will be enabled.
After programming the various settings, press “En-
ter”.
English
Current value
11/06 9:30 1.0232
(Fig. 13)
(Fig. 14)
Alarm
OFF
Less than: 1.0220
Greater than: 1.0260

IsACQ210D-2 Rev. 250907-2
9 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
(Fig. 17)
Set reference and
wait for adjustment
Read value 51.4 mS
Calib. value 52.0 mS
1/1
(Fig. 18)
Calibration OK
Press any key to
continue
English
(Fig. 19)
Measurement Unit
mS/cm - µS/cm
ppt - Psu
GS
g/l
(Fig. 15)
Calibrate Sensor
New
Reset
(Fig. 16)
Do you want to
return to default
calibration values?
Enter: Conrm
Esc: Cancel
Calibrate Sensor
This menu allows you to calibrate the density probe
through the interface. By using the keys , you may
choose whether to perform a new calibration by selec-
ting “New”, or to delete a previous one and reset the
interface to the default settings by selecting “Reset”
(Fig. 16). A calibration should be performed when the
interface is rst connected to the system.
Note: Before performing the calibration, the probe
must be rinsed using tap water, dried carefully and
inserted in the Aquatronica conductivity (density)
solution.
In order to correctly calibrate the sensor, proceed as
follows:
Main screen Main menu Density Calibrate Sensor
New.
1) Select the function “New” by using the keys
then press “Enter” (Fig 15).
2) Using the keys set the value of the standard
solution next to “Calib. Value” (Fig. 17).
3) Wait 10 minutes in order to allow the probe’s read
value to stabilize. The read value may stabilize on a
value that is slightly different than the reference.
NOTE: keep the probe as straight as possible inside
the solution container.
4) Once 10 minutes have elapsed, press “Enter”.
5) The controller will display the calibration result (Fig.
18); rinse the probe and insert in aquarium
Note: the calibration may be cancelled at any time
by pressing “Esc”. This will return the calibration
parameters to those of the last completed calibration.
Measurement Unit
The measurement unit of the conductivity read by the
control unit may be modied (Fig. 17).
To modify this parameter, proceed as follows:
Main screen Main menu Density Measurement
Unit.
• Select the measurement unit using the keys and
press “Enter”.
Note: See conversion table at the end of this
manual.

IsACQ210D-0 Rev. 250907-2 10 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
About
Provides information on the control unit’s rmware ver-
sion. To use this function, proceed as follows:
Main screen Main menu Density About.
Disconnect
If the density interface is disconnected, a message
will appear on the display (Fig. 21). Press “Enter” to
indicate that the message has been read.
On the main screen the “?” icon will appear next to
the name of the “Density” sensor and in the lower left
corner (Fig. 22).
If the conductivity interface is reconnected, the con-
trol unit will automatically begin displaying the read
value again.
To denitively eliminate the density sensor from the
system, after disconnecting it, proceed as follows:
Main screen Main menu Density Disconnect.
The “Data Record” and “Calibrate Sensor” functions
disappear from the “Density” menu (Fig. 23) and the
“Disconnect” function appears.
• Select this function using the keys and press
“Enter”.
• The disconnection screen will appear (Fig. 24). Press
“Enter” to disconnect or “Esc” to cancel.
(Fig. 21)
Device
Disconnected
S01: Density
(Fig. 22)
Mon 11/06/07 15:05
PU01 A B C D E F G H
Density ?.?
English
(Fig. 23)
Density
Change Name
Programs
Alarm
Measurement Unit
Disconnect
(Fig. 24)
Disconnect
Density
Enter: Conrm
Esc: Cancel
(Fig. 20)
Aquatronica
FW version: x.y
Press any key to
continue

IsACQ210D-2 Rev. 250907-2
11 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
Suggestions for an accurate reading of the Conductivity (density)
Precise readings depend greatly on proper maintenance of the connected sensor. Beyond
its intrinsic qualities, how the sensor is cared for is particularly important. This will, in
fact, provide reliable readings. Below is a list of some simple suggestions for optimum
conductivity (density) readings in aquariums:
- Handle the probe with care.
- If the probe is stored out of the water, it must be thoroughly dried to prevent oxidation
of the electrodes.
Periodically clean the probe (every week) with tap water, subsequently drying it thoroughly
to eliminate any deposits.
- Periodically calibrate the instrument (approximately every month as indicated
to page 24) to correct any reading imperfections due to probe wear.
- Replace the probe after a period of approximately 15 months.
- Do not install the interface in direct contact with wet or damp parts.
- Never use calibration solutions that have been left open or have expired.
- After the bottled solution has been opened, it must be used in 6 months (if properly
closed after use). After this period it must be replaced.
- Position the probe in the tank or sump away from strong currents to prevent excessive
oxygenation inside the latter.
- Do not immerse the probe completely in water. The cable’s seam must always be ap-
proximately 2 cm above the water.
IMPORTANT
For reliable conductivity readings, use only AQUATRONICA electrodes and cali-
bration solutions.
The use of other brands of electrodes could cause incorrect readings of the
instrument.
NOTE: in case of malfunctions or any doubts about the use of this interface, please
contact AQUATRONICA’S free Technical Assistance.
English

IsACQ210D-0 Rev. 250907-2 12 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
DISPOSAL OF ELECTRIC AND ELECTRONIC PARTS
Pursuant to Article 13 of Legislative Decree No. 151 of 25 July 2005, “Imple-
mentation of Directives 2002/95/CE, 2002/96/CE and 2003/108/CE,
regarding the reduction in use of dangerous substances in electrical
and electronic equipment, as well as waste disposal”:
Products bearing the barred dustbin symbol must be disposed of
separately from other waste. The user must therefore dispose of
the product in question at suitable recycling centers for electronic
and electro-technical waste, or he/she must turn over the used
product to the retailer when buying a new equivalent product, on
a one-to-one basis.
Separate waste collection allows used equipment
to be recycled, treated and disposed of without negative con-
sequences for the environment and health, and it allows the
materials in the equipment to be recycled. Illegal dumping of the
product by the user entails the administrative sanctions stated in
Legislative Decree No. 22/1997 (Article 50 et seq of Legislative
Decree No. 22/1997).
Separate collection of used products and packaging allows mate-
rials to be recycled and used again. Reuse of recycled materials
helps prevent environmental pollution and reduces the demand
for raw materials.
Local regulations may provide for the separate collection of
household appliances at municipal waste sites or retailers when
a new product is purchased.
English

IsACQ210D-2 Rev. 250907-2
13 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
Dear Customer,
Thank you for the trust you have placed in AQUATRONICA by purchasing this product.
AQUATRONICA subjects all its products to severe quality tests. If, in spite of the controls,
the product should malfunction, we recommend you immediately contact the dealer/pro-
ducer for any checks or interventions necessary.
- General warranty regulations
AQUATRONICA guarantees that this product operates properly and that it is free of
manufacturing faults and defects. If the product becomes defective during the warranty
period, AQUATRONICA will repair or replace it. Defective parts will be replaced ex-works
AQUATRONICA’s plant (recipient responsible for shipping expenses). Accessories and
components not made by AQUATRONICA are covered by their manufacturers’ warranties
only. This is the only warranty offered by AQUATRONICA. All others are excluded. Unless
due to malice or gross negligence, AQUATRONICA is not responsible for injury to people
or damage to things resulting from product malfunctioning. This warranty is valid only for
those up to date with their payments.
- Conditions
The warranty will be valid for a period of 24 months from the date of purchase only upon
presentation of this certicate, which must bear the retailer’s stamp, the date of sale and
the product’s serial number, accompanied by the invoice or receipt issued by the retailer,
indicating the serial numbers of the installed products. AQUATRONICA
can refuse to recognize the warranty if this information is incomplete or has been tam-
pered with after purchase. The warranty will be valid only if, at the time of purchase, the
product was properly stored as it was originally packaged by AQUATRONICA, who alone
can ensure its origin and adequate protection.
- Warranty exclusions
This warranty does not cover:
a) periodic checks, maintenance, repairs, or replacement of parts due to normal wear
b) malfunctioning due to negligence, incorrect installation, improper use or non-observance
of the technical instructions and, in general, any malfunction not traceable to manufacturing
faults or defects of the product and therefore the responsibility of AQUATRONICA
c) products modied, repaired, replaced, installed or in any way tampered with by anyone
without the prior written authorization of AQUATRONICA
d) accidents originating from force majeure or other causes (e.g. water, re, lightning,
bad ventilation, etc.) not depending on AQUATRONICA
Products with manufacturing faults or defects recognizable through normal dili-
gence must not be resold or installed. The sole competent court for any disputes
concerning the interpretation or execution of this warranty is that of Reggio
Emilia (Italy).
Product code:
Serial number:
Date of purchase:
Retailer’s stamp
Day Month Year
Warranty

IsACQ210D-0 Rev. 250907-2 14 - 16
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy Tel.: +39 0522 494403
DECLARATION OF CONFORMITY
Standard of reference ISO/IEC Guide 22 and EN 45014
Number of conformity: 003-2006
Name of the manufacturer: Aquatronica division of A.E.B. srl
Address: via dell’Industria, 20
CorteTegge
42025 Cavriago (RE) Italy
DECLARES THAT THE ELECTRONIC UNITS
Code: ACQ012 (6 shuko plugs + I/O)
ACQ012 AUS (6 australian plugs + I/O)
ACQ012 F (6 french plugs + I/O)
ACQ012UK(6 english plugs + I/O)
ACQ012 ZA (6 South Africa plugs + I/O)
ACQ013 (4 shuko plugs + 4 F plugs + I/O)
ACQ013 CH (8 swiss plugs + I/O)
ACQ220 (PC interface)
ACQ200 (I/O interface)
ACQ210-RX (REDOX sonde interface)
ACQ210-PH (PH sonde interface)
ACQ210-TL (temperature and level sonde interface)
ACQ210-MS (conducibility sonde interface)
ACQ210-D (density sonde interface)
ACQ450 (dosing pump 4 modules)
ARE IN COMPLIANCE WITH THE FOLLOWING PRODUCT
SPECIFICATIONS:
FIELD Directive Description References Test
Result
EMC 89/336/EEC EMC directive Official Journal of the
European Union L139
May 23 1989
applied
LVD 73/23/EEC Low voltage
directive
Official Journal of the
European Union L077
March 26 1973
applied
THEREFORE THEY ARE IN COMPLIANCE WITH THE REQUISITES OF
THE CE MARK
The equipment was checked in a typical working configuration
Place of issue: Cavriago (RE) Italy Date of issue: 05/23/06
The A.E.B. srl legal representative
Paterlini Ivan
Declaration of Conformity

CONDUCTIVITY - DENSITY - SALINITY - SPECIFIC GRAVITY
Conductivity Density Salinit
y
Specific Gravity
(mS/cm) ( g/l) (ppt/PSU)
35,5 mS/cm 1.013,8
g
/l 22,4
pp
t/PSU 1,0164
36 mS/cm 1.014,1
g
/l 22,7
pp
t/PSU 1,0166
36,5 mS/cm 1.014,4
g
/l 23,0
pp
t/PSU 1,0169
37 mS/cm 1.014,6
g
/l 23,4
pp
t/PSU 1,0172
37,5 mS/cm 1.014,9
g
/l 23,8
pp
t/PSU 1,0174
38 mS/cm 1.015,1
g
/l 24,1
pp
t/PSU 1,0177
38,5 mS/cm 1.015,4
g
/l 24,5
pp
t/PSU 1,0179
39 mS/cm 1.015,7
g
/l 24,8
pp
t/PSU 1,0182
39,5 mS/cm 1.015,9
g
/l 25,2
pp
t/PSU 1,0185
40 mS/cm 1.016,2
g
/l 25,5
pp
t/PSU 1,0187
40,5 mS/cm 1.016,5
g
/l 25,9
pp
t/PSU 1,0190
41 mS/cm 1.016,7
g
/l 26,2
pp
t/PSU 1,0193
41,5 mS/cm 1.017,0
g
/l 26,6
pp
t/PSU 1,0195
42 mS/cm 1.017,3
g
/l 26,9
pp
t/PSU 1,0198
42,5 mS/cm 1.017,5
g
/l 27,3
pp
t/PSU 1,0201
43 mS/cm 1.017,8
g
/l 27,7
pp
t/PSU 1,0204
43,5 mS/cm 1.018,1
g
/l 28,0
pp
t/PSU 1,0206
44 mS/cm 1.018,4
g
/l 28,4
pp
t/PSU 1,0209
44,5 mS/cm 1.018,6
g
/l 28,7
pp
t/PSU 1,0212
45 mS/cm 1.018,9
g
/l 29,1
pp
t/PSU 1,0214
45,5 mS/cm 1.019,2
g
/l 29,5
pp
t/PSU 1,0217
46 mS/cm 1.019,4
g
/l 29,8
pp
t/PSU 1,0220
46,5 mS/cm 1.019,7
g
/l 30,2
pp
t/PSU 1,0223
47 mS/cm 1.020,0
g
/l 30,5
pp
t/PSU 1,0225
47,5 mS/cm 1.020,3
g
/l 30,9
pp
t/PSU 1,0228
48 mS/cm 1.020,5
g
/l 31,3
pp
t/PSU 1,0231
48,5 mS/cm 1.020,8
g
/l 31,6
pp
t/PSU 1,0234
49 mS/cm 1.021,1
g
/l 32,0
pp
t/PSU 1,0236
49,5 mS/cm 1.021,4
g
/l 32,4
pp
t/PSU 1,0239
50 mS/cm 1.021,6
g
/l 32,7
pp
t/PSU 1,0242
50,5 mS/cm 1.021,9
g
/l 33,1
pp
t/PSU 1,0245
51 mS/cm 1.022,2
g
/l 33,5
pp
t/PSU 1,0248
51,5 mS/cm 1.022,5
g
/l 33,8
pp
t/PSU 1,0250
52 mS/cm 1.022,8
g
/l 34,2
pp
t/PSU 1,0253
52,5 mS/cm 1.023,0
g
/l 34,6
pp
t/PSU 1,0256
53 mS/cm 1.023,3
g
/l 34,9
pp
t/PSU 1,0259
53,5 mS/cm 1.023,6
g
/l 35,3
pp
t/PSU 1,0262
54 mS/cm 1.023,9
g
/l 35,7
pp
t/PSU 1,0264
54,5 mS/cm 1.024,2
g
/l 36,1
pp
t/PSU 1,0267
55 mS/cm 1.024,4
g
/l 36,4
pp
t/PSU 1,0270
55,5 mS/cm 1.024,7
g
/l 36,8
pp
t/PSU 1,0273
56 mS/cm 1.025,0
g
/l 37,2
pp
t/PSU 1,0276
56,5 mS/cm 1.025,3
g
/l 37,6
pp
t/PSU 1,0278
57 mS/cm 1.025,6
g
/l 37,9
pp
t/PSU 1,0281
57,5 mS/cm 1.025,9
g
/l 38,3
pp
t/PSU 1,0284
58 mS/cm 1.026,1
g
/l 38,7
pp
t/PSU 1,0287
58,5 mS/cm 1.026,4
g
/l 39,1
pp
t/PSU 1,0290
59 mS/cm 1.026,7
g
/l 39,6
pp
t/PSU 1,0293
59,5 mS/cm 1.027,0
g
/l 39,8
pp
t/PSU 1,0296
60 mS/cm 1.027,3 g/l 40,2 ppt/PSU 1,0299
Conversion Table

A.E.B. Srl Division
Via dell’Industria, 20 - 42025 Cavriago (RE) Italy
Tel.: +39 0522 494403 Fax: +39 0522 494410
http:\\www.aquatronica.com
E-mail: [email protected]
Table of contents