Ash OMNI 3 Setup guide

OMNI 3 Quick Start User Guide Rev 3.1
1 | P a g e
OMNI 3
Quick Start User Guide
Remove the OMNI 3 Camera Head and all cables from packaging.
1. Screw the cable as indicated with the M4 screw provided. Ensure that the
cable is pulled upward during assembly.
2. When screwed in place, bend the cable so that it feeds outward.
3. Ensure that the bottom ridge * under the screw has not been covered as it
could stop the OMNI 3 Arm bracket from attaching correctly.
STEP 1: ESD Grounding Cable Assembly (Optional)
1
2
3
*
STEP 2: Mouse & Keyboard Set-Up
2
1
2
3
4
1. Remove USB dongle from the underside of the mouse.
2. Ensure mouse is switched to on. Ensure keyboard is switched to on.
3. Remove USB dongle from the underside of the keyboard located in the
batteries compartment.
4. Remove yellow “arrow” tab in the keyboard.
Please remove standard +5 lens before installation of +10 plan1x, +25 or +50 lens.
Also ensure the newly installed lens is selected in System settings and save to Startup
Settings.

OMNI 3 Quick Start User Guide Rev 3.1
2 | P a g e
STEP 3: Arm Attachment to Stand
6
4
5
1. Attach Arm Receiver (1) to the stand using the 4 screws provided. Please ensure to have
the plated key* towards the bottom.
2. Hook the Arm Bracket (2) onto the Arm Receiver (1) using the 4 Hooks.
3. Please ensure that the rear hooks are pointing downwards. Tighten the two Thumb
Screws (3) on both sides of the back of the arm bracket to secure.
4. Route all cables (except HDMI cable) through the hole on the underside of the Arm
Bracket.
2
1
4
3
STEP 4: Peripherals and head Attachment to Stand
*Plated Key
3
2
1
1. Plug power cable into 24v DC power jack.
2. Plug External Keypad (if included; Preset optional accessory) into rear USB port.
3. Plug HDMI Cable into HDMI Connector port at the rear of OMNI 3 and connect to monitor.
4. Insert dongle from mouse and keyboard into USB ports at side of OMNI 3 (optionally the
keyboard dongle can be inserted in a rear USB port to free up a side port).
5. Now hook the OMNI 3 on to the hooks of the Arm Receiver by slotting the 4 hooks into
the unit.
6. Tighten the front 2 Thumb Screws on both sides to secure the OMNI 3.

OMNI 3 Quick Start User Guide Rev 3.1
3 | P a g e
OMNI 3 Software Setup
STEP 5: Setting the Correct Height for Measurement
1. Press and hold the power button on the OMNI 3 to switch on.
2. Click on the Settings icon on the main toolbar on the right-hand side of your
screen.
3. In System Settings, ensure the correct screen size and Lens Dioptre that you are
currently using are selected.
4. The user must set the Focal Plane of the lens first to suit the height of the
object.
To set the Focal Plane:
a) Place the unit in Overview Mode by clicking the overview icon.
b) Move the object with the area of interest under the crosshair on screen.
c) Exit Overview Mode by clicking the Overview icon again.
d) Click the Zoom In button or use the front panel dial to zoom to Full
Magnification.
e) Manually adjust the height of the OMNI 3 by moving the focus block up and
down on the stand until the object is in focus.
f) Now you are ready to measure at your desired focal plane!
Refer to user manual for full details
3
1
Note: Once you have achieved your Focal Plane DO NOT adjust the height of the object or the
height of the OMNI 3 if you wish to continue measuring at this focal distance.

OMNI 3 Quick Start User Guide Rev 3.1
4 | P a g e
1. Click on the Settings icon on the right-hand side of your screen.
2. In Camera Settings, adjust exposure and lighting setting to suit test sample
inspection conditions e.g.
Ringlight Settings:
•Select which Ringlight LED quadrant to turn on/off.
•Set the intensity value of the LEDS.
•Toggle all LEDS on/off.
Exposure Settings:
•Auto Exposure Mode: Gain and Iris settings are automatically adjusted but
brightness of the image can be adjusted.
•Manual Exposure Mode: Gain, Iris and Brightness settings can be manually
adjusted.
•Higher Gain values will increase noise on the image.
•Lower Gain values will reduce the amount of light entering the camera but
improve Focus Depth.
*Refer to user manual for full details
STEP 6: Camera & System Settings
2

OMNI 3 Quick Start User Guide Rev 3.1
5 | P a g e
●Where possible, the OMNI 3 should be kept in a clean environment away from dirt, dust, oil and
debris which could affect performance or degrade the system’s mechanical and electronic parts.
If a clean environment is not available, the OMNI 3 should be kept as clean and protected as is
possible. In harsh environments, cover the OMNI 3 with plastic when not in use.
●External surfaces can be wiped down with a clean, water moistened lint-free cloth (not wet).
Never wipe down with acetone or other harsh solvents that may damage surfaces.
●Cleaning Optics: Do not touch lens surfaces with your fingertips, as the resulting fingerprints
may destroy optical coatings over time. Only clean optical surfaces with proper cleaning
supplies, and only when necessary. If the lens is covered with loose dust, first try blowing off
this dust using a can of optical grade (oil-free) canned compressed air. Be careful not to shake
the can, or propellant may blow onto the lens. As an alternative, use a lens brush to gently wipe
off the dust. If the lens is soiled with greasy deposits which cannot be blown or brushed off, use
alcohol based commercial lens cleaner and a lens tissue or a lens cloth. Apply the lens cleaner
generously to dissolve grease, and then blot off the lens cleaner and dissolved grease using
minimum motion. Avoid rubbing the lens, since hard particles from deposits or the lens cloth
could scratch the optical coatings. Where lens contamination is unavoidable, consider adding
●Cleaning Mechanical parts: To ensure a long, trouble-free service life, wipe down the system
regularly to remove any dust or dirt from the stand. Should the focus block seize or require
tightening, please contact your Ash representative.
●The ringlight can be removed to gain access to the +5 Lens for cleaning by pressing the release
button at the back and pulling the ringlight down away from the OMNI 3 head.
•Indoor use only
•Exercise caution when lifting, handling or moving the Omni 3.
•Exercise caution when installing or removing accessories, ensuring a central and stable body
position, thus preventing any possible strains or injury.
•Ensure all screws and fixings are properly secured.
•Disconnect all power sources prior to moving or working on equipment.
•Do not stare at operating lights as it may be harmful to the eyes.
Maintenance and Care
General Safety

OMNI 3 Quick Start User Guide Rev 3.1
6 | P a g e
To prevent electrical shock or accidental short circuits leading to component damage, follow the points
outlined below.
•Ensure proper grounding of AC adapters.
•Do not operate with open housing, both from a safety and warranty perspective (there are no
user-serviceable parts or fuses in the system).
•Do not operate in environments of excessive humidity and keep away from liquids.
•Operate away from metal filings and similar debris.
•Should any power cords become damaged, switch off, disconnect and replace before using the
unit again.
•Do not position the equipment so that it is difficult to operate the disconnecting device, mains
plugs in this case.
•Please be aware of protection impairment if used in a manner not specified by the
manufacturer.
*You will find your full OMNI 3 User Manual on the supplied USB.