2. Starting operations
- The pressure switch should be installed and operated only by authorized personel.
- Cover and bottom of the device form a function unit. Exchanging the parts can cause measuring errors
or malfunctionning. For wall mounting remove the four front cover screws and the front cover. Then
fasten the device with four screws to the wall and finally remount the front cover. To damp strong
vibrations shock mounts must be used.
- Connect the sensor to the (3-pin + PE) cube plug at the bottom side of the UAD 3 - V3.
- The electrical connection (supply, analog output switching contacts) must be carried out according to
the connection tables depicted on the top of the device by removing the cover cap and insert the cable
through the cable gland PG 13,5. If required, additional cable glands can be installed in the cover cap
by breaking out the perforated cavities.
- The electrical connection must be carried out in accordance with the VDE 0100 regulations. In order to
ensure trouble-free operation it is essential to connect the protective lead.
- If inductive loads (magnets, contactors, etc.) are connected to the switch relays, suitable protective
devices (varistors etc.) must be provided.
3. Operating elements
Dimensions (in mm)
Operating Instructions 4-fold Trip Amplifier UAD 3 - V3
1. Product description
Intended applications
- The trip amplifier is only to be connected to input signals according to the values on the type label at
the bottom side of the device.
Barksdale GmbH
Dorn-Assenheimer Strasse 27
D-61203 Reichelsheim / Germany
Tel.: +49 - 60 35 - 9 49-0
Fax: +49 - 60 35 - 9 49-111 and 9 49-113
e-mail: [email protected]
www.barksdale.de
UAD 3 - V3
12
Electrical connection
Sensor connection
(In combination with 2-wire circuit pin 2 is not required)
8-digit 14-segment display
(Display 1)
Cover screws
Sign
Trend arrows
Menu buttons
20-part bargraph
4 x ø4,5
Switch point display
SP1...SP4
8-digit 14-segment display
(Display 2)
Analog-GND (0 V)
+18 V DC...+30 V DC
Analog-GND (0 V)
Switch contacts
Earth
M4
4...20 mA (0...10 V)
0V
S1
S2S3S4
S1
S2S3S4
4...20 mA (0...10 V)
Analog output 2
Analog output 1
Current input 4...20 mA
(2-wire)
+Ub
---
Signal
PE
Current / voltage input
0...10 V, 0...20 mA
(3-wire)
+Ub
–Ub
Signal
PE
Plug DIN 43650
3-pin + PE
Analog input No. 1 and 2
Pin 1
Pin 2
Pin 3
Earth
100
50
3846
87
150
35
80
136
ø4,6
Item-Nr.: 923-1280
Rev. --, 26. 06. 2004
Specifications are subject to
changes without notice.
Drilling pattern
43
List of functions UAD 3 - V3
Measur. mode .... .... Display of the actually measured value and the measuring unit
MENU Display keylock
UNLK No keylock, all parameters can be adjusted
LOCK Keylock active, all parameters visibal but can not be changed
SP1...SP4 Switch point menu SP1..SP4
MODE STND Standard evaluation (rising/falling)
WIND Window technology
ERRO Error output
ON .... Switch-on value for SP1...SP4; if the ON-value is smaller
than the OFF-value, the switch evaluatiuon is falling
OFF .... Switch-off value SP1...SP4
LEV Inversion of the switching output
HLFS High-level-fail-safe (Normally Open function)
LLFS Low-level-fail-safe ((Normally Closed function)
DEL 0,0s..9,9 s Switch-on / switch-off delay for SP1...SP4 in seconds
SPSC DSP1 Switch point reacts on display values of display 1
DSP2 Switch point reacts on display values of display 2
ANO1...ANO2 Analog output menu
ANOP ON Analog output in operation
OFF Analog output switched off
AOZS .... Scale the analog output - start value (e.g. 0 bar = 4 mA)
AOFS Scale the analog output - end value (e.g. 400 bar = 20 mA)
DISP1...DISP2 Display menu
MODE STND Display channel 1 resp. channel 2
DIFF Display difference between channel 1 and ch. 2 (ch. 1 – ch. 2)
UTXT ABCD Enter the measuring unit as text
DAMP 0,0s..9,9 s Damping of the displayed measuring value in seconds
OFFS .... Measuring value -Offset, means shifting the display range
CUT .... Cut-Off, means signal-surpression within the cut-off range
BGZS .... Scale the bargraph - start value
BGFS .... Scale the bargraph - end value
PEAK Peakholding menu
MIN .... Display the peak value „Min“
CLRM NO no deletion
YES delete „Min“-value memory
MAX .... Display the peak value „Max“
CLRX NO no deletion
YES delete „Max“-value memory
CAL1...CAL2 Calibration menu
DECP 0000..0,000 Determination of the decimal-point for the measuring value
ZSCL .... Scale the measuring-value indication - start value
FSCL .... Scale the measuring-value indication - end value
CODE Calibration of the manufacturer
4. Operation
After the unit is switched on, the unit starts an automatic self-test.
The device is menu operated and configured by the three keys on the front.
With the „M“ key (= mode) you change between the operation / indicating level to the dialog values and the
adjusted / actual values. With the keys („⇑“ = up) and („⇓“= down) you change between the dialog values
in the menu or change the values / functions in the menus.
A change of any configuration starts always with the M-Mode and indicated by the flashing cursor. After a
change has been made with the up- resp. down-key the M-mode key must be pressed to confirm each
configuration; to set numbers, each digit has to be confirmed with the M-Mode before adjusting the next
one. By confirming the last digit the new configuration will be stored in the memory.
Pushing the down key at the end of the sub-menu the software will switch automatically to the main-menu.
If the dialog is not continued within two minutes the device automatically returns to the measuring mode.
5. Key lock
Activating the („⇑“ = up) and („⇓“ = down) keys together for more than 5 seconds will block any changings
in all menues; shown by „LOCK“ in the display.
In this mode, all configuration values can be checked only, but not changed.
Repeating this action will unlock the configuration menu and shown by „UNLK“ in the display.
6. Error handling
The internal self-check software will monitor the proper functioning of the unit. When any of the following
failures will occur, the flashing display will indicate the following text:
Main Menu Sub Menu Value Description
Display Error Cause
max Positive excess of the measuring range The measured value exceeds the max. of the range
Channel 1 (indicated in display 1)
positive excess of the measuring range
Channel 2 (indicated in display 2)
min Negative excess of the measuring range The measured value is lower than the min. of the range
Channel 1 (indicated in display 1)
negative excess of the measuring range
Channel 2 (indicated in display 2)
anao Failure of the analog output 1 Output loop is not closed or short circuited
(indicated in display 1)
Failure of the analog output 2
(indicated in display 2)
data Stored data failure (EEProm) (internal) Memory failure
prog Processor failure (internal) Microcontroller failure
cal Calibration error channel 1 (internal) Calibration values are wrong
(indicated in display 1)
Calibration error channel 2 (internal)
(indicated in display 2)
Attention:
At first the peak value
menu for channel 1 in
display 1 is indicated,
then for channel 2 in
display 2.
65
Measuring mode Menu Structure UAD 3 - V3
Indication display 2
Channel 2
Indication display 1
Channel 1
Manufacturer menu
Code error

Popular Amplifier manuals by other brands

Kramer 6601 user manual

Kramer

Kramer 6601 user manual

Quicksilver SET Mono Amplifier operating instructions

Quicksilver

Quicksilver SET Mono Amplifier operating instructions

StewMac 64 REVERB UNIT Assembly instructions

StewMac

StewMac 64 REVERB UNIT Assembly instructions

Linear LT1222 manual

Linear

Linear LT1222 manual

QUAD 405-2 Instruction book

QUAD

QUAD 405-2 Instruction book

McIntosh MC225 owner's manual

McIntosh

McIntosh MC225 owner's manual

Analog Metric 12AU7 user manual

Analog Metric

Analog Metric 12AU7 user manual

JVC KS-A101J Operation manual

JVC

JVC KS-A101J Operation manual

Blaupunkt MA-5850 Service manual

Blaupunkt

Blaupunkt MA-5850 Service manual

CYP AU-D820S Operation manual

CYP

CYP AU-D820S Operation manual

Marantz 240 Service manual

Marantz

Marantz 240 Service manual

Planet Audio PSA4CB user manual

Planet Audio

Planet Audio PSA4CB user manual

Pioneer A-447 operating instructions

Pioneer

Pioneer A-447 operating instructions

Atlona AT-UHD-CAT-8 manual

Atlona

Atlona AT-UHD-CAT-8 manual

Merlaud AMB 80 D Operation & maintenance instructions

Merlaud

Merlaud AMB 80 D Operation & maintenance instructions

Bosch PLE-2MA120-EU Installation and user instructions

Bosch

Bosch PLE-2MA120-EU Installation and user instructions

McIntosh MC611 owner's manual

McIntosh

McIntosh MC611 owner's manual

diytube stereo 35 Instructions for assembly operation

diytube

diytube stereo 35 Instructions for assembly operation

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.