Pump Setup Reglo Digital Rev B 09-09-06
Part 2 – Tubing connectors
The ferrule forms a tight seal as it
is deformed when being screwed
into the flow module. Place the
ferrule and connector of the tube
as shown before attaching them.
Make sure to cut the PTFE tubing
at right angles. It is recommended
to use the tube cutting device
included in the Installation set.
Note that the arrows on the top of
the module indicate the direction
of the flow for correct sample
temperature equilibration.
Use the brown adapter (the perifit)
to connect the PTFE tubing and
black threaded nut to the soft
pump tubing (expanding the
narrow bore of the pump tubing
with a pipette tip may help). If 1.5-
2mL tubes are used, the white lids
can be used to guide the tubing.
Note that the lid needs to be
Place the pump and sample holder
so that the samples are drawn
through the flow module. This
setup will reduce the need to clean
the pump tubing and prevents
leakage.
For details, please check the Explorer Operator Manual.
In
Out