
Page 4 of 8 █ © Brüel & Kjaer Vibro ●January 2018 ● C106833.002 / V01
Technical alterations reserved!
EN
3 Hints for safe operation of VIBROCONTROL 950/960
General:
Please carefully read the operating instructions prior to set-up of the device. Make sure that your
VIBROCONTROL 950/960 device is suitable for your application without any restrictions.
Improper use:
Any improper or non-intended use may lead to malfunctions of the VIBROCONTROL 950/960 device
or to unwanted effects in your application. If the VIBROCONTROL 9x0 is used in a way not described
in the relevant user manuals, function and protection may be impaired and serious personal damage,
death or serious, irreversible injuries may result.
EU-directives:
All versions of the VIBROCONTROL 950/960 device conform to the relevant regulations and EC
directives.
Installation and operation:
Installation, electrical connection, set-up, operation and maintenance of your VIBROCONTROL
950/960 device(s) must only be carried out by qualified/trained personnel (electrician) authorised by
the machine operator in accordance with local- and national regulations for the installation of electrical
equipment.
Changing the setup parameters:
Before applying a new set of setup parameters to the VIBROCONTROL 950/960 device, please make
sure that doing so cannot cause any damage to persons and/or machinery.
Connecting the sensor(s):
Please make sure to meet the safe extra-low voltage (SELV) criteria when any sensors are connected
to the VIBROCONTROL 950/960 device so that no dangerous contact voltages are applied to the
sensor and/or transferred to the device. The sensor and the power supply of the VIBROCONTROL
950/960 device are not galvanic isolated.
Sensor cable mounting:
To prevent negative effects on the functioning of the VIBROCONTROL 950/960 device caused by
noise voltages, please lay shielded sensor cables and load cables separately.
Ingress protection (IP):
The VIBROCONTROL 950/960 is ranked as IP20. The VIBROCONTROL 950/960 device must be
mounted in a control cabinet with an ingress protection of at least IP54. The control cabinet should be
installed in accordance with local- and national rules and regulations.
Mounting:
Mount the VIBROCONTROL 950/960 device on a 35 mm DIN rail inside the control cabinet. Mount the
device vertically but make sure to leave enough space between the unit and the top and/or bottom of
the control cabinet. Only this way the air circulation will be sufficient to avoid excessive heating of the
device.
Connecting to a power supply:
The VIBROCONTROL 950/960 device has a voltage tolerance of +24 VDC ±5 %
Before connecting the VIBROCONTROL 950/960 device to a +24 VDC supply voltage, please make
sure that all terminal blocks are completely inserted.
The external +24 VDC supply voltage must be generated and supplied according to the SELV
requirements.
Protect the +24 VDC supply voltage externally with max. 2A. The ground (GND) of the DC supply is
directly connected with the ground (GND) of the sensor supply, if any. The SELV criteria must
therefore be met for the DC supply (safety extra-low voltage, circuit electrically isolated from other
circuits, not grounded). If the DC circuit is to be grounded (e.g. due to national regulations), the