Martini Instruments pH 58 User manual

OPERATIONAL GUIDE
• Remove the electrode cap and immerse the
meter in MA9015 storage solution for 2 hours
to activate the electrode.
• Turn the meter on by pressing the ON/OFF
button. All the used segments on the LCD will
be visible for 1 second or as long as the
button is pressed.
• Immerse the electrode in the solution to be
tested. Stir gently and wait for the reading to
stabilize, i.e. the instability indicator (hour-
glass) on the LCD turns off.
• Select the pH or ORP (mV) mode by press-
ing the SET/HOLD button.
• The displayed pH value is automatically com-
pensated for temperature. The ORP (mV)
value will be displayed on the main LCD,
while the temperature will always be dis-
played on the secondary LCD.
•To freeze the display, while in measurement
mode, press and hold the SET/HOLD button.
The “HOLD” message appears on the sec-
ondary display and the reading will be fro-
zen on the LCD. Press any button to return
to the normal measurement mode.
•To turn the meter off, press the ON/OFF
USER MANUAL
pH 58
Pocket-size
pH/ORP/°C/°F Meter
ISTPH58R1 05/05
button. The “OFF” message will appear on
the secondary display. Release the button.
Note:
• Before taking any measurement, make sure
the meter is calibrated (the CAL tag is on).
• After use always turn the meter OFF, rinse
the electrode with water to minimize contami-
nation and store it with a few drops of storage
(MA9015) or pH7 (M10007) solution in the
protective cap.
NEVER USE DISTILLED OR DEIONIZED WA-
TER FOR STORAGE PURPOSE.
CALIBRATION PROCEDURE
The ORP (mV) range is factory calibrated.
Contact the nearest Service Center for
recalibration, if necessary.
For better accuracy, frequent calibration of the
tester is recommended. Calibration is also nec-
essary after electrode replacement, after testing
aggressive chemicals and where extreme ac-
curacy is required.
• Turn the meter on and select the pH mode
with the SET/HOLD button.
• Press and hold the ON/OFF/CAL button until
the “OFF” message on the secondary LCD
is substituted by “CAL”. Release the button.
• The instrument enters the calibration mode
by displaying “pH 7.01 USE” (or “pH 6.86
USE” if the NIST buffer set was selected).
• For a single-point calibration, immerse the
electrode in any buffer, i.e. pH 4.01, 7.01 (or
6.86), 10.01 (or 9.18).
www.milwaukeeinst.com
• The meter activates the automatic buffer rec-
ognition. If no valid buffer is detected, the
meter keeps the USE indication active for 12
seconds, and then replaces it with WRNG,
indicating that the sample being measured is
not a valid buffer. Otherwise, if a valid buffer
is detected, then its value is shown on the
primary display, and REC appears on the
secondary LCD.
• If the pH 7.01 (or pH 6.86) was used, press
the SET button to exit the Calibration mode
and the “OK 1” message will appear on the
display. The first calibration point is stored
and the meter returns to normal measure-
mentmode.
For better accuracy, it is always recom-
mended to perform a 2-point calibration.
• For a two-point calibration, immerse the elec-
trode in pH 7.01 (or pH 6.86) buffer solution.
• After the first point has been accepted, the
meter will then ask for the second buffer and
the message “pH 4.01 USE” will appear.
• Rinse the electrode and immerse it in the
second solution (pH 4.01, 10.01 or 9.18).
• If a valid buffer value is detected, the REC
message is displayed and the meter com-
pletes the calibration procedure. The LCD
shows the accepted value with the “OK 2”
message and the instrument returns to the
normal measurement mode. Otherwise, if no
validbuffer is detected, the meter displaysthe
WRNGmessage.
ELECTRODE REPLACEMENT
• Remove the protective cap and unscrew the
plastic ring on the top of the electrode.
• Pull out the MI58P electrode and replace it
with a new one.
• Make sure the gaskets are in place before
screwing back the ring.
BATTERY REPLACEMENT
When the batteries become weak, the battery
symbol on the LCD will light up to advise that
only a few hours of working time is remaining.
The meter is also provided with BEPS (Battery
Error Prevention System), which avoids any
erroneous readings due to low battery level by
automaticallyswitching the meter off.
It is recommended to replace the batteries im-
mediately.
To replace the batteries unscrew the battery
compartment cap and replace all four 1.5V bat-
teries while paying attention to their polarity.
Make sure the gasket is in place before screw-
ing back the cap.
Batteries should only be replaced in a non-
hazardous area using the battery type specified
in this instruction manual.
SPECIFICATIONS
RANGE -2.00 to 16.00 pH
±1000 mV (ORP)
-5.0 to 60.0°C / 23.0 to 140.0°F
RESOLUTION 0.01 pH / 1 mV
0.1°C / 0.1°F
ACCURACY (@25°C) ±0.05 pH / ±2 mV
±0.5°C / ±1°F
TYPICAL EMC DEVIATION ±0.02 pH
±2 mV
±0.3°C / ±0.6°F
TEMPERATURE COMPENSATION
Automatic for pH, from -5 to 60°C
CALIBRATION, pH Automatic, 1 or 2 points
with 2 sets of memorized buffers
(pH 4.01, 7.01, 10.01 or 4.01, 6.86, 9.18)
CALIBRATION, ORP Factory calibrated
PROBE ReplaceableMI 58P
ENVIRONMENT -5 to 50°C; 100% RH max.
BATTERY TYPE 4 x 1.5V; IEC LR44, A76
BATTERY LIFE approx. 250 hours of use
AUTO-OFF after 8 minutes of non-use
DIMENSIONS 200 x dia 38 mm
WEIGHT 100 g
ACCESSORIES
MI58P Replaceable electrode for pH 58
M10004B pH 4.01 buffer, 20 ml sachet, 25 pcs.
M10007B pH 7.01 buffer, 20 ml sachet, 25 pcs.
M10010B pH 10.01 buffer, 20 ml sachet, 25
pcs.
MA9004 pH 4.01 buffer, 230 ml bottle
MA9006 pH 6.86 buffer, 230 ml bottle
MA9007 pH 7.01 buffer, 230 ml bottle
MA9009 pH 9.18 buffer, 230 ml bottle
MA9010 pH 10.01 buffer, 230 ml bottle
MA9015 Electrode storage solution, 230 ml
MA9016 Electrode cleaning solution, 230 ml
MA9020 ORP test solution (200/275 mV), 230
mlbottle
M10000B Electrode rinse solution, 20 ml sa-
chet, 25 pcs.
Note: When the calibration procedure is com-
pleted, the CAL tag is turned on.
•To quit the procedure and return to the last
calibration data, after entering the calibration
mode press the ON/OFF button.
The secondary LCD displays “ESC” for 1
second and then the meter returns to the
normal measurement mode.
•To reset to the default values and clear a
previous calibration, press the SET/HOLD
button after entering the calibration mode and
before the first point is accepted.
The secondary LCD displays “CLR” for 1
second, the meter resets to the default cali-
bration and the CAL tag on the LCD turns off.
SETUP
Setup mode allows the selection of temperature
(°C or °F) and pH buffer set for calibration. To
enter the Setup mode, press the ON/OFF but-
ton until “CAL” on the secondary LCD is re-
placed by “TEMP” and the current temperature
unit (e.g. TEMP °C). Then:
•for °C/°F selection: use the SET/HOLD but-
ton; then press the ON/OFF button once to
enterthe buffer set selection or twiceto return
to the normal measurement mode.
•to change the calibration buffer set: after
setting the temperature unit, press ON/OFF
once and select the buffer set (“pH 7.01
BUFF” or “pH 6.86 BUFF” for NIST) by
WARRANTY
This instrument is warranted against defects in mate-
rials and manufacturing for a period of two years
from the date of purchase. Probe is warranted for 6
months. If during this period the repair or replace-
ment of parts is required, where the damage is not
due to negligence or erroneous operation by the user,
please return the parts to either distributor or our
office and the repair will be done free of charge.
Damages due to accidents, misuse, tampering or
lack of prescribed maintenance are not covered.
Milwaukee/Martini instruments reserves the right to make
improvements in design, construction and appearance
of its products without advance notice.
using the SET/HOLD button. Press ON/OFF
to return to the normal measurement mode.

Other Martini Instruments Measuring Instrument manuals

Martini Instruments Mi 105 User manual

Martini Instruments

Martini Instruments Mi 105 User manual

Martini Instruments Mi 106 User manual

Martini Instruments

Martini Instruments Mi 106 User manual

Martini Instruments MI 408 User manual

Martini Instruments

Martini Instruments MI 408 User manual

Martini Instruments Mi 160 User manual

Martini Instruments

Martini Instruments Mi 160 User manual

Martini Instruments MI 405 User manual

Martini Instruments

Martini Instruments MI 405 User manual

Martini Instruments MI 404 User manual

Martini Instruments

Martini Instruments MI 404 User manual

Martini Instruments MI 805 User manual

Martini Instruments

Martini Instruments MI 805 User manual

Martini Instruments MI 170 User manual

Martini Instruments

Martini Instruments MI 170 User manual

Martini Instruments pH 56 User manual

Martini Instruments

Martini Instruments pH 56 User manual

Martini Instruments MI 411 User manual

Martini Instruments

Martini Instruments MI 411 User manual

Martini Instruments MI 406 User manual

Martini Instruments

Martini Instruments MI 406 User manual

Martini Instruments MI 413 User manual

Martini Instruments

Martini Instruments MI 413 User manual

Martini Instruments MI 605 User manual

Martini Instruments

Martini Instruments MI 605 User manual

Martini Instruments pH 55 User manual

Martini Instruments

Martini Instruments pH 55 User manual

Martini Instruments MI 415 User manual

Martini Instruments

Martini Instruments MI 415 User manual

Martini Instruments MI 806 User manual

Martini Instruments

Martini Instruments MI 806 User manual

Martini Instruments MI 306 User manual

Martini Instruments

Martini Instruments MI 306 User manual

Martini Instruments MI 415 User manual

Martini Instruments

Martini Instruments MI 415 User manual

Popular Measuring Instrument manuals by other brands

Somogyi Elektronic Home MVT 808ACE instruction manual

Somogyi Elektronic

Somogyi Elektronic Home MVT 808ACE instruction manual

Stanley Tabs 25222 User instructions

Stanley

Stanley Tabs 25222 User instructions

LaMotte Tracer 1766 manual

LaMotte

LaMotte Tracer 1766 manual

Tektronix P6209 instruction manual

Tektronix

Tektronix P6209 instruction manual

Siemens SICAM Q200 7KG97 manual

Siemens

Siemens SICAM Q200 7KG97 manual

Keysight N9010AK-MTU Installation note

Keysight

Keysight N9010AK-MTU Installation note

Vaisala SHM40 user guide

Vaisala

Vaisala SHM40 user guide

Galvanic Applied Sciences 943-TGS-CE Operation manual

Galvanic Applied Sciences

Galvanic Applied Sciences 943-TGS-CE Operation manual

Nieaf-Smitt EazyAmpPLUS manual

Nieaf-Smitt

Nieaf-Smitt EazyAmpPLUS manual

Agilent Technologies Cary 60 user guide

Agilent Technologies

Agilent Technologies Cary 60 user guide

ABB Relion RES670 Operator's manual

ABB

ABB Relion RES670 Operator's manual

HENNING MSM12 V3 manual

HENNING

HENNING MSM12 V3 manual

Ambient Weather WS-25 user manual

Ambient Weather

Ambient Weather WS-25 user manual

TSI Instruments AIRASSURE 8144 Operation and maintenance manual

TSI Instruments

TSI Instruments AIRASSURE 8144 Operation and maintenance manual

Tektronix 1L10 instruction manual

Tektronix

Tektronix 1L10 instruction manual

AIR SYSTEMS INTERNATIONAL AQT-02 operating manual

AIR SYSTEMS INTERNATIONAL

AIR SYSTEMS INTERNATIONAL AQT-02 operating manual

Ahlborn ALMEMO 2390-5 operating instructions

Ahlborn

Ahlborn ALMEMO 2390-5 operating instructions

Inepro PRO1 Series manual

Inepro

Inepro PRO1 Series manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.