Cairn MULTICAM User manual

Quick Start
MULTICAM
VERSION: AUG 2017 (SN 7072 - PRESENT)
The MultiCam is a powerful and stable device for pixel aligning images captured using different wavelengths, polarization
states, fields of view or other imaging modalities onto up to four camera sensors. It is designed to meet the most demanding
requirements for the sub-pixel mapping of Super Resolution imaging and is compatible with most scientific C mount and F
mount cameras.
For the purpose of this Quick Start Guide, please refer to the MultiCam diagram on Page 3. All words in bold blue text are
labelled on this diagram.
1. Connect the MultiCam to your existing 1x camera c-mount.
2. If using an inverted microscope, three support jacks are normally supplied to support the weight of the unit. Position
the support jacks in the jacking point on each output. The height can be adjusted and locked in position with the blue
locking collar.
3. Remove the C-mount camera tube and protective dust cap. The clamping screw may need to be loosened (using the
2.5mm allen key provided)
4. Screw each camera onto the C-mount camera tube and replace on each arm of the MultiCam with the calibration ring
still in place.
A) Mounting to your microscope
+44(0)1795 590140 www.cairn-research.co.uk
B) Mounting and positioning four cameras
1. Identify the transmitted port (1) on the MultiCam diagram
2. View a live image on the corresponding camera (ideally of a graticule using transmitted light) on the transmitted port.
For initial setup, it is essential to be able to view all four edges of the rectangular aperture, therefore it is not advisable
to use a fluorescent sample at this stage.
3. Close the rectangular aperture until in view on the image and rotate the camera until square.
4. Lock off the clamping screw
5. Insert the calibration cubes (each containing a 50% mirror) into the three combiner cubes by removing the magnetic
door covers. Magnets will ensure the cube locates correctly.
6. Repeat steps 1 to 4 on the remaining camera ports.
7. If the sample focus on any of the other cameras is not precisely matching the focus of the transmitted camera, an
adjustment is provided (to allow for any slight differences between positioning of the camera sensor). Remove the
calibration ring and move the camera in Z until focused. Lock off the clamping screw once again.
C) Centre the transmitted image
1. With the rectangular aperture still in view, centre the transmitted port (1) (using the 1.5mm allen key provided) by
adjusting the transmitted H&V controls. It’s useful to use a centred square or central cross-hair as a reference point
in your imaging software (if available).

+44(0)1795 590140 www.cairn-research.co.uk
For the remaining images, an additional method of XY alignment on the three combiner cubes allows fine adjustment of the
position of each image for precise pixel alignment in your imaging software.
D) Centre the remaining images
Often MultiCam units are supplied with populated filter cubes, however (if required), these can be installed using the guidance
notes below. Please note for all our emission splitter range we only recommend Chroma Ultra-Flat 2mm dichroic mirrors to
minimise distortion. Please let us know if you require any filter assistance.
E) Filter cube assembly
1. Identify port (2)* and combiner cube (1)
2. With the rectangular aperture still in view, centre the image on port (2) (using the 1.5mm allen key provided) by
adjusting the H&V controls on combiner cube (1)
3. Similarly, centre the camera on port (3) with the H&V controls on combiner cube (2)
4. Finally, centre port (4) with the H&V controls on combiner cube (3)
5. Note all the reflected images will be mirrored and will need to be flipped in your imaging software if overlaid.
6. Replace the calibration cubes with populated filter cubes for multi-channel imaging (see section E).
7. Open the rectangular aperture to slightly larger than the camera(s) field of view to minimize scattered light.
* MultiCam units may also be supplied with three output ports as an option, with upgrade potential for a forth camera. If so, port
2 will be absent, however, alignment steps for all other ports remain the same.
In addition to the two positions in the filter cube, there is also an auxiliary holder on each camera output port to accommodate
interference (bandpass) filters, long pass filters, ND filters, rotating polarisers, cylindrical lenses or Z-plane shifting lenses. The
holders supplied with the unit have an angled thread (indicated by angled line on the top) to avoid back reflections from filter
surfaces, however if using lenses or a polariser, then a flat holder is required. These can be purchased separately.
F) Auxiliary component mounts
Please consult filter manufacturers instructions regarding orientation of the filter.
Installing Dichroic Mirrors into the Cairn Filter Cube Installing Filters into the Cairn Filter Cube
Remove the 4 screws and
seperate the two halves. Replace the top half and
resecure the screws.
Place the dichroic, followed by
spring clip in
the recess
Remove the dichroic spring clip
CAIRN
www.cairn-research.co.uk
CAIRN
www.cairn-research.co.uk
CAIRN
www.cairn-research.co.uk
Turn the locking
ring anticlockwise
to remove it.
Carefully place
the recess.
Replace the locking
ring and rotate it
clockwise to secure.