car MLB242 User manual

MLB242 Analysis Instrument
User Guide
Instrument for the analytical Quality Control of several sorts of foodstu.
For In-Vitro-Analysis only.
Similar to
CDR FoodLab Junior, CDR WineLab Junior, CDR BeerLab Junior,
CDR CiderLab Junior, CDR OxiTester Junior
Distributor:
Maufacturer: CDR s.r.l, Firenze, Italy
Development - Manufacturing - Sales – Technical Service


Manufacturer:
cdR srl
Via degli Artigiani, 6
50020 Ginestra Fiorentina - FI, Italy
Phone: +39-055-871431 www.cdr-mediared.it
Fax: +39-055-8714322 www.cdrfoodlab.com
P/Number:
Manual history
Revision Edition from Author cdR reader Doc size
1st edition 16.06.2016 J. Schenne sales team A5
Document made by ©
Ing.-Buero fuer Bio-Prozessanalytik
In case of missing information or descriptions that could cause misunderstandings,
please get in touch with me to solve it. Changes may be made only by the author; the
same applies to PDF documents!
Copyright:
The software for the MLB242 analyzer is the User and Instrument Control Software for
the MLB242 analyzer. The MLB242 analyzer software is the intellectual property of cdR
srl Italy. Intellectual property rights shall remain with cdR srl.
You are entitled to use the MLB242 analyzer software and the printed accompanying
material at your place of work only.
WindowsTM is a trademark of Microsoft company!


5
MLB242
User’s Guide MLB242-V1
Content
! Common Safety Issues......................................................................7
! 1 Hazards and Precautions......................................................................7
! 2 Maintenance and Hygiene...................................................................10
! 3 First-Aid Measures ..............................................................................10
! 4 Repairs..................................................................................................10
! 5 Safety of the product........................................................................... 11
! 6 Norms and standards used for the design and testing ................... 11
! 7 Conformity of the product ..................................................................12
1 General information .........................................................................13
1.1 Intended use.........................................................................................13
1.2 System description..............................................................................13
1.3 The MLB242 analyzer .........................................................................14
1.4 Storage of data.....................................................................................16
1.5 CDR Test cuvette .................................................................................16
1.6 CDR Reagent kit...................................................................................17
1.7 Pipettes.................................................................................................17
1.8 Testing ..................................................................................................17
1.9 Transportation and storage ................................................................19
1.10 Receipt of the analyzer........................................................................19
2 Installation ........................................................................................20
3 Operation ..........................................................................................22
3.1 Operation procedure ...........................................................................22
3.2 Switching on the analyzer...................................................................22
3.3 Preparation of reagents and samples................................................22
3.4 Test selection and execution..............................................................22
3.4.1 Cancellation of a test .................................................................25
3.4.2 Storing of Reagent Blank value ..................................................25
3.4.3 Exceeded measuring range .......................................................26
3.5 Management of results........................................................................26
3.6 Standardisation of tests......................................................................26
3.7 Settings Menu ......................................................................................27
3.7.1 Language ...................................................................................27
3.7.2 Date/Time ...................................................................................27
3.7.3 Laboratory name ........................................................................28

6
MLB242
User’s Guide MLB242-V1
3.7.4 Test Methods ..............................................................................28
3.7.5 Audio ..........................................................................................28
3.7.6 Brightness ..................................................................................29
3.7.7 Info .............................................................................................29
3.7.8 Export Status .............................................................................29
3.7.9 Reading Test...............................................................................30
3.7.10 Extra ...........................................................................................30
3.8 Performed Tests (Data Base) .....................................................30
4 Error messages ................................................................................32
5 Maintenance......................................................................................33
5.1 Recommended actions and precautions...........................................33
5.2 Software update ..................................................................................33
6 Appendix ...........................................................................................34
6.1 Additional products for analysis........................................................34
6.2 Spare parts ...........................................................................................34
6.3 Declaration of conformity ...................................................................35
6.4 Technical data ......................................................................................36
6.5 Data transfers.......................................................................................37
6.6 Disposal................................................................................................38
6.6.1 Disposal of instrument ................................................................38
6.6.2 Disposal of cuvettes and reagents .............................................38
6.6.3 Disposal of packaging material ..................................................38

7
MLB242
User’s Guide MLB242-V1
! Common Safety Issues
Please read the Instruction Manual entirely prior operating your analysis system. In
order to ensure a high level of performance, all warnings and references to technical
safety in this document must be followed.
! 1 Hazards and Precautions
The cautions and safety regulations in this Instruction Manual meet international classi-
cations:
Danger - Electrical hazards
Warns of a risk of injury or of a risk to life (for example by electrical
shock).
Danger – Mechanical hazards
Warns of a risk of injury or of a risk to life.
Warns of a risk of injury or of the analyser being severely damaged.
Danger - Biohazard
Warns of a risk of biological material like blood or other human or ani-
mal specimen. Risk of infection! Potential infectious area/material!
Danger - Toxic substances
Warns of a risk of a serious intoxication!
Skin irritation
Inammable substance
Introduces rules to be observed
The following safety issues have to be observed at all times:

8
MLB242
User’s Guide MLB242-V1
Electrical safety
- Check that the operating voltage is set correctly before you
connect the device to the main power supply.
- To connect the analyzer to the main power supply, use only
sockets which are grounded to avoid the risk of an electrical
shock.
- Use only grounded extension power cords. The used power
cord has to correspond to the country specic regulations.
- Never intentionally disconnect the grounding contacts.
- There is the risk of electrical shock if the protective conduc-
tor is interrupted within or outside the analyzer, and/or the
grounded contacts have been disconnected from the line.
- Never remove protective guards or secured components since
you could expose electrically live parts in this way.
- Electrical connection contacts (plugs, sockets, etc.) can be
electrically live.
- Even after an analyzer has been switched o, components
(e.g. capacitors) can be under voltage as the result of an elec-
trical charge.
- All current carrying parts are sources of danger for an electri-
cal shock.
- Surfaces (oors, work tables) have not to be moist when you
are working with any electrical device.
- Carry out only the maintenance work and/or the replacement
of parts described in this document.
- Unauthorized work on the analyzer can lead to the guarantee
obligation becoming null and void with necessary expensive
service work to correct it.
- All work, which requires the analyzer to be opened, have to be
carried out by an authorized technician who is familiar with the
risks related thereto.

9
MLB242
User’s Guide MLB242-V1
Mechanical
Safety
Analyzer is operating
- Never open or unscrew analyzer’s casing parts while it is
switched ON. There is a risk of injury due to moving parts as
fans, pumps, motors or any other mechanical movements.
Biological Ma-
terial
Risk of Infection
- Wear gloves in all cases if there is a risk of infection.
- Avoid any direct contact with samples that are potentially in-
fectious. On cdR analyzers normally the reagents used togeth-
er with sample materials to test are not potential infectious.
In case of a contamination, immediately use a suitable skin
disinfection solution. Ask your doctor for aid.
- If sample material or reagent is spilled onto the analyzer, wipe
it o immediately and refer to the chapter !2 Maintenance and
Hygiene.
- Do not open a waste bottle as long as the analyzer is in pro-
cess.
Reagents and
Controls
- Observe the suggestions in the package inserts for an exact
use of the reagents and quality controls. Note that reagents
and quality control material can be biological material!
- Don’t use Reagents, Controls and other Liquids after expira-
tion date!
- Avoid any liquid penetration into the analyzer.
Restrictions Samples, Reagents and Controls
- For consumables no guarantee can be provided for any re-
sistance against organic solvents. For this reason, do not use
any organic solvents unless such solvents are expressly indi-
cated.
- Do not use any other rinsing or cleaning solutions as recom-
mended by the manufacturer or representative.
- Waste liquid has to be disposed in compliance with the legis-
lation.

10
MLB242
User’s Guide MLB242-V1
Quality Accuracy and Precision of the measured results
- In order to ensure a awless operation of the analyzer mea-
sure control samples and watch the function of the analyzer
closely.
- Faulty measurement results may result in an incorrect diagno-
sis or range danger for patient.
Fire and Explo-
sion Hazards
- Do not place any inammable or hazardous explosive material
close to the device.
- Electrical sparks could cause re or explosions.
Operator Quali-
cation
- The analyzer should only be operated by trained personnel.
Improper operation of the analyzer might cause inaccurate
measuring results!
! 2 Maintenance and Hygiene
- No organic acid based cleaning substances should be applied. Instead use clean-
er designed for cleaning and disinfecting laboratory analysers.
- Only use a dampened cloth to clean the analyser.
- Never spray or pour cleaning solution directly onto the analyser that may negative-
ly impact the analyzer’s functions signicantly.
- Keep the analyser clean and do not spill liquids onto it.
- In case liquids were spilled onto the analyser, immediately absorb liquid with a
suitable cloth.
- Contact your distributor if your control measurements do not produce the expected
results.
Cleaning and Disinfection
- Use a suitable surface disinfectant solution available for laboratory instruments.
- In urgent cases use Surgical Spirit 94% if no other more suitable solution is availa-
ble.
! 3 First-Aid Measures
- If a person swallowed any chemical solution may be while servicing, wash out
mouth with water.
- If provided person is unconscious, than call a physician.
- If inhaled, remove to fresh air. If breathing becomes dicult, than call a physician.
- In case of contact, immediately wash skin with soap and copious amount of water.
- In case of contact with eyes, ush with copious amounts of water for at least 15
minutes.
- Assure adequate ushing by separating the eyelids with ngers, than call a physi-
cian.
! 4 Repairs
Repairs to the analyser may only be carried out by trained personnel, and replacement

11
MLB242
User’s Guide MLB242-V1
parts have to comply with the analyser specications. In case of analyser problems
contact your representative.
! 5 Safety of the product
European Directives consulted during the design stage. The analyzer has been de-
signed and constructed following the health and safety requirements of the directives:
European Directives consulted during the design stage
Directive
Number
Title and implementation into Italian legislation
Directive
2014/30/EC
Directive
2006/95/EC
EMC Directive 2014/30/EC of the European Parliament and of the Coun-
cil of 26 February 2014 on the harmonisation of the laws of the Member
States relating to electromagnetic compatibility (recast) (all languages),
aligned to the New Legislative Framework, will be applicable from 20
April 2016.
The Directive 2006/95/EC of the European Parliament and of the Council
of 12 December 2006 on the harmonisation of the laws of Member States
relating to electrical equipment designed for use within certain voltage
limits (codied version) was published in the OJEU L 374 of 27.12.2006,
p. 10-19. It entered into force on the 20th day following that of its publica-
tion, i.e. on 16 January 2007. The old Directive 73/23/EEC was repealed
on the same date.
! 6 Norms and standards used for the design and testing
The MLB242 analyzer corresponds to the essential requisites of the norms:
EN 55022:2010 Radio disturbance characteristics - Limits and methods of meas-
urement.
EN 61326-
1:2013
Electrical equipment for measurement, control and laboratory use -
EMC requirements - Part 1: General requirements
EN 61010-1
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 1: General requirements (IEC
61010-1:2010 + Cor. :2011
ENV 61010-2-
081
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-081: Particular requirements for
automatic and semi-automatic laboratory equipment for analysis
and other purposes (IEC 61010-2-081:2001

12
MLB242
User’s Guide MLB242-V1
EN 60950-1
European Standard CEI EN 60950-1 (CEI 74-2) “Information tech-
nology equipment – Safety. Part 1: General Requirement.” VI° Ed.
File n°8683 (February 2007) , with amendment A11 (CEI 74-2;V2)
Technical le 10285 (March 2010), A1 (CEI 74-2;V2) Technical le
12566 E (October 2012), A12 (CEI 74-2;V3) Technical le 12567 E
(October 2012) and A2 (CEI 74-2;V4) Technical le 13635 (June
2014).
ETSI EN 300 328 V 1.7.1 (2006) “electromagnetic compatibility
and Radio spectrum Matters (ERM); wideband transmission sys-
tem; Data transmission equipment operating in the 2,4 GHz ISM
band and using wide band modulation techniques; Harmonized
EN covering essential requirements under article 3.2 of the R&TTE
Directive.
! 7 Conformity of the product
The instrument has the CE-mark in accordance with the Directive 89/336/CEE and
successive modications and integrations. Each modication that alters the design and
construction characteristics of the instrument and not expressly authorised by cdR S.r.l.
will cancel the conformity and consequently the right of use.
Use of the instrument that is not described by this document is considered. cdR S.r.l.
will not be held responsible for damage caused by not respecting the
technical specications described herein and by the improper use of the
instrument or biological products. See chapter 6.3 for the Declaration of
conformity.

13
MLB242
User’s Guide MLB242-V1
1 General information
1.1 Intended use
The MLB242 analyzer is made for the analysis on several sorts of food using cdR’s
dedicated test kits together with special test application instructions for use on produc-
tion lines or special laboratories. All analytical tests provided by the MLB242 analyzer
are well calibrated after valid international standards.
1.2 System description
MLB242 analyzer is a standalone, compact instrument designed and made after latest
technology. Its main part is the spectrophotometer with 4 channels and 2 beams each
of dierent wavelength. Free cells can also be used for incubation purposes. It is
temperature-controlled at 37.0°C ±1), for pre-warming of test cuvettes. The analyzer is
calibrated and controlled automatically after the power ON where also other ele-
mentary function checks will be done. Possible malfunction will be shown on display
automatically after detection.
The analyzer is equipped with a colour LCD touch screen. All operations take place by
using a nger or a special touch-screen pen which is to put on a symbol or message.
The analytical tests are internally stored along with the associated calibration param-
eters. Test results will be stored with a serial or positive identication in a data base
where it can be called up. A lithium battery ensures a loss of all data. The language and
other basic functions can be set up in Settings Menu. Following language selection can
be done: Italian, English, German, French and Spanish.
CdR test cuvettes 10 or 20 ps., pre-led with reagent, are packed in an aluminium bag
often together with enzyme vials or R1a reagent vials.
The analytical tests are mainly run in 2-phases. In the rst phase, a dened amount of
sample material is pipetted into each pre-lled cuvette to R1 reagent. After incubation,
a blank value is measured and stored. For colour handling, a reagent R2 must add to
the cuvette content and after the second incubation the colour change can be read.
The results are shown on the display and automatically memorised.
The data transfer to a host computer can be done by the USB interface. After connec-
tion a temporary drive will be installed for the transfer of data.
The analyzer must be powered by a LPS - Limited Power Source - that automatically
adapts to the existing voltage from 115V to 230V. The secondary voltage is 24Vdc +/-
10%, 3 A. This LPS is in accordance with Clause 2.5 of EN 60950-1.
Useful hint
In this Manual a bullet “•” draws your attention to an instruction.
Example:•Pressenter-keytoconrm...

14
MLB242
User’s Guide MLB242-V1
1.3 The MLB242 analyzer
Ports at back site for power supply and data transfers.
4 Reading cells, with dierent wavelengths.
Operating cell is indicated by a blue shining control LED.
Cells not used for testing can be used for incubation pur-
poses.
Power ON: press button short
Power OFF: press button for 3 seconds
4,3” TFT-Colour Display with touch-function.
Figure 1 MLB242 Modules
TOUCH function: is an active area on display expressed as a symbol that can be used
to carry out a function.
Operating parts
Colour touchscreen display
Top line with:
Analyzer name, HOME button
Optional: Battery loading status.
Date/Time
Main part
Menus for Settings, Tests
And Results (DATABASE)
Figure 2 Main Dialogue
Measuring cells
After test selection a blue illuminated LED shows the reading cell that must be used for
chosen test. Automatically the cell will be checked, if it fails the operator will be warned
and the specic test is inhibited for further use. As a new innovation every measuring
cell is equipped with an Insertion Tube Sensor at the button to control whether a cell
is occupied by a cuvette or not. (see
chapter 3.7.10)
During reading don’t touch cuvette or
other parts and avoid any vibrations!
Figure 3 Measuring cells

15
MLB242
User’s Guide MLB242-V1
Beeper
A beep will be heard:
- after power on/o
- after touch of function on screen
- at end of incubation cycle
- if an error occurs
Connections
At rear site following parts are available:
Figure 4 connections and its symbols
Power switch ON/OFF
24V DC port for external power unit (internal pin is +)
USB port type B for technical service and for the data transfer to computer.
Connection to the host PC.
To upload analytical data, connect an USB-cable between analyser and USB-port of
the pc. See chapter 6.5.
Power requirements
The analyzer must be powered by an external power supply of secondary 24Vdc +/-
10%; 3 A with limited power supply LPS after Clause 2.5 of EN 60950-1, branded with
European Quality Marks for Safety like: TUV, VDE, SEMKO, NEMEKO etc. Use the
power adapter supplied with the analyser to meet important safety regulations.
Normally a standard power cable is supplied with the adapter with a protective ground
contact. Check if the plug will t the power outlet (convenience outlet) you have. If you
need a common adapter use one that meets a standard regulation for safety.

16
MLB242
User’s Guide MLB242-V1
Instrument nameplate
The instruments nameplate displays symbols for safety and identication data.
Address of manufacturer, CE mark
Symbols for attention and wasting
Type: MLB242 and
Serial number:
Year of production:
Power requirement: 24V, 3A AC
Figure 5 Nameplate
Important:
- Never remove the nameplate from the original position chosen by the manufacturer.
- Do not modify or falsify the data displayed on the nameplate.
- Do not clean the nameplate with abrasive products.
1.4 Storage of data
Analyser data, test specic data and results are stored in the memory as long as they
will be overwritten or deleted by user interaction.
1.5 CDR Test cuvette
For opening or closing of the cuvette, it is much recommended to place the cuvette into
a free incubation cell and turn then the cup on or o. This procedure keeps the optical
window clean.
Pre-lled test cuvette
1 cup that closes the cuvette
2 triangular mark that indicates the reagent lling level of 1ml
3 optical window, keep it clean, don’t touch it with ngers!
Figure 6 CDR test cuvette

17
MLB242
User’s Guide MLB242-V1
1.6 CDR Reagent kit
A test kit contains 10 or 20 pre-lled cuvettes in a welded aluminium bag.
Important information on plate:
Test name: e.g. Acidity
REF: 300xxx (order code)
Instrument name
Number of test cuvettes inside
LOT: reagent lot number
EXP: expiration date
Store at: storing temperature
Link for method download
Symbols should be observed!
Figure 7 Bag of pre-lled test cuvettes
MSDS: For all cdR test kits Material Safety Data Sheets are available on cdR’s website
downloadable only in English language: http://resources.cdrfoodlab.com/msds.php
For testing, transportation, storing and wasting you should know any dangers or consequences!
Symbol explanation:
Expiration date followed by date e.g. 07/2014
Storing condition followed by temperature e.g. 15 to 25°C
For symbols for safety see chapter ! 1 Hazards and Precautions.
1.7 Pipettes
For a maximum of accuracy it is much recommended to use especially for the displace-
ment of doughy sample material a positive displacement one. Ask your distributor for
more information. Operate with pipette according to manufacturer’s description! Know
that a pipette is an instrument that needs after a certain time of use a special mainte-
nance, otherwise results could be wrong!
MiniPet: types of dierent volumes can be in the accessories. It is not of highest accu-
racy and should be used only as described in the test application instructions.
1.8 Testing
For testing go as described in the specic test application instructions available in
Menu Settings and Test Methods. For quick information about the test procedure,
touch the question mark that is visible in Main Dialogue.
All following illustrations are examples!

18
MLB242
User’s Guide MLB242-V1
Touch ? help to open the rst appli-
cation dialogue.
This dialogue asks you to place the
cuvette with sample 1 into the indicat-
ed measuring cell for blank reading.
Figure 8 Blank reading
Figure 9 Help dialogue Figure 10 Help dialogue
For every test you can have a look of a number of illustrations that explain the test
procedure entirely.
Touch:
- for next dialogue
- for previous dialogue
- to leave a dialogue, Exit
Reagent indications
While reading test applications you will be confronted by dierent reagents terms. Un-
derstand the following description:
R1 basically all reagents pre-lled in test cuvettes are called as R1, also in bulk ware
where the reagent must be pipetted into empty test cuvettes.
R1a is a reagent that activates the R1.
R2 is a reagent that causes an optical change depending to a concentration.
R3 is a reagent that causes an optical change depending to a concentration.

19
MLB242
User’s Guide MLB242-V1
1.9 Transportation and storage
Packaging, movement and transportation
The analyzer and its accessories are packed together in a cardboard box. To transport
it, use its cardboard box to protect it against atmospheric inuences. Please note that
the carton does not provide sucient protection against rain, sun, snow and wind.
Storage conditions
The analyzer in its carton must be stored in a closed environment within the tempera-
ture range +5°C to +30°C (41°F to 86°F) and a relative non-condensed humidity of
20% to 90%. The storage place must not be exposed to contaminating agents such as
humidity, dust, acids, corrosive gases, salt, smoke etc. Do not stack the cartons more
than ve high.
Conservation and/or disposal of the packaging
The manufacturer recommends conserving the cardboard box in case of future trans-
portation purposes. The disposal of the packaging does not present any particular dan-
gerous aspects to persons, animals or materials. Before disposal of the cardboard box,
consult the Council Directive 94/62/CE on packaging and the disposal of packaging.
1.10 Receipt of the analyzer
Each analyzer is checked following specic procedures on instruments by cdR prior
shipment. However, damage could occur during transportation and will not be the
responsibility of cdR. Eectuate a visual control of the packaging and the apparatus to
check that there is no visual damage due to knocks and bangs. If damage is evident,
call with your sales representative to nd a solution.
Control whether all parts are available by Packing List.

20
MLB242
User’s Guide MLB242-V1
2 Installation
For the installation of the analyzer proceed as follows:
A. Place MLB242 on a plane surface in such a position that it is not exposed by me-
chanical vibrations or extreme temperature uctuations or excessive light.
B. Connect the supplied external power adapter (LPS: Limited Power Supply) to the
port at the rear of the instrument 24 Vdc, 3A (Figure 4) using the appropriate power
cord to the mains (110V to 230V). In case the adapter is dierent from the one included
with the instrument, check that the electrical and safety characteristics (24 Vdc, 3A) are
equivalent (CE mark) and that the polarity of the connectors is respected Figure 4).
C. After power ON the analyser, the photometer optics will be controlled together with
other functional parts. A malfunction will be indicated on display.
After a while the following information is shown:
Example!
Figure 11 appears for seconds
Followed automatically by next dialogue:
Figure 12 a, b: appears only if tube insertion sensor is o
Test applications are loaded into the working memory for quick access.
After warming up a Warning message appears:
• Make sure that the reading cells are empty!
• Make sure that no cell is occupied by a cuvette and conrm by touching the green
check mark!
According to the customer needs, the analyser can be congured for Oils & Fats, Wine,
Beer or other food products.
Similar analyzers have their own test application prole like the FoodLabJunior or
Table of contents