Other Cardinal Health Water Pump manuals

Cardinal Health Alaris 7130/7131 Manual

Cardinal Health

Cardinal Health Alaris 7130/7131 Manual

Cardinal Health IVAC 571 Operating manual

Cardinal Health

Cardinal Health IVAC 571 Operating manual

Cardinal Health Kangaroo ePump User manual

Cardinal Health

Cardinal Health Kangaroo ePump User manual

Cardinal Health IVAC PCAM Manual

Cardinal Health

Cardinal Health IVAC PCAM Manual

Cardinal Health Alaris GP Manual

Cardinal Health

Cardinal Health Alaris GP Manual

Popular Water Pump manuals by other brands

Watershed Innovations HYDRAPUMP SMART FLEX Instructional manual

Watershed Innovations

Watershed Innovations HYDRAPUMP SMART FLEX Instructional manual

Graco Modu-Flo AL-5M instructions

Graco

Graco Modu-Flo AL-5M instructions

Messner MultiSystem MPF 3000 operating instructions

Messner

Messner MultiSystem MPF 3000 operating instructions

Xylem Bell & Gossett WEHT0311M Installation, operation and maintenance instructions

Xylem

Xylem Bell & Gossett WEHT0311M Installation, operation and maintenance instructions

WilTec 50739 Operation manual

WilTec

WilTec 50739 Operation manual

Franklin Electric Little Giant 555702 HRK-360S instruction sheet

Franklin Electric

Franklin Electric Little Giant 555702 HRK-360S instruction sheet

Ingersoll-Rand PD02P Series Operator's manual

Ingersoll-Rand

Ingersoll-Rand PD02P Series Operator's manual

VS ZJ Series Operating	 instruction

VS

VS ZJ Series Operating instruction

Flotec FPZS50RP owner's manual

Flotec

Flotec FPZS50RP owner's manual

SKF Lincoln FlowMaster II User and maintenance instructions

SKF

SKF Lincoln FlowMaster II User and maintenance instructions

Xylem Lowara LSB Series Installation, operation and maintenance instructions

Xylem

Xylem Lowara LSB Series Installation, operation and maintenance instructions

Water Duro Pumps DCJ500 Operating & installation instructions

Water

Water Duro Pumps DCJ500 Operating & installation instructions

Action P490 Operating instructions & parts manual

Action

Action P490 Operating instructions & parts manual

Flo King Permacore Reusable Carbon Bag Disassembly. & Cleaning Instructions

Flo King

Flo King Permacore Reusable Carbon Bag Disassembly. & Cleaning Instructions

ARO ARO PD15P-X Operator's manual

ARO

ARO ARO PD15P-X Operator's manual

Pumptec 112V Series Operating instructions and parts manual

Pumptec

Pumptec 112V Series Operating instructions and parts manual

Virax 262070 user manual

Virax

Virax 262070 user manual

Neptun NPHW 5500 operating instructions

Neptun

Neptun NPHW 5500 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Cardinal Health™ Kangaroo™ Joey
Enteral Feeding Pump
Quick Reference Guide
Intermittent
operation
Symptom Probable cause Corrections
Hold error Pump in Holding mode
for more than 10 minutes
Press
}
Continue to return to Holding screen, or press
and hold the Power button to turn o. Several options
are available from the Holding screen, such as
}
Run and
}
Adjust Settings.
Flow error Occlusion in pump-patient
line (downstream occlusion)
Check for and correct the occlusion in the line
between pump and patient. Replace Pump Set if
error can’t be resolved.
Feed error
Empty bag or occlusion in
bag-pump line
(upstream occlusion)
Check for empty bag and rell, or check for and correct
the occlusion in the line between the pump and feed bag.
Replace Pump Set if error can’t be resolved. Check the
valve pocket in the Pump Set loading area for dampness
or dirt.
Pump Set
use >24 Hrs
Pump Set in use beyond
recommended length of time
Informational warning message blinks on upper left of
screen. It’s recommended that the Pump Set be replaced.
Pump Set
dislodged
Pump Set not properly
loaded to pump
Check black ring retainer (MISTIC) and valve; reload Pump
Set and restart pump to continue. Replace Pump Set if
error can’t be resolved.
Battery low Battery charge too low Immediately plug pump into AC outlet to recharge battery
and to provide power for pump.
Rotor error
Pump Set tubing not
properly loaded on rotor,
or other unusual rotor
operating condition
Check that the tubing is loaded on the rotor and
that it’s not damaged, torn, etc. Reload tubing or
replace Pump Set.
System error
General error caused by
many factors; see Manual
for error list
Power down pump and attempt restart to clear error.
Call customer service with screen code if error
can’t be resolved.
Below are some of the probable causes of alarm conditions and how to correct them.
For additional operating instructions and
information, see the Operating Manual for
the Kangaroo™Joey Enteral Feeding Pump
“Intermittent Operation.”
Programming and Menu
Selection buttons
Power button
Status LEDs
Red: Error
Yellow: Hold/Pause
Green: Normal operation
Battery operation
Unplugging the pump from the AC line will automatically put the pump on battery power.
To recharge the battery, plug the pump into an AC wall outlet. The pump will automatically
begin charging. If AC power is interrupted, a built-in rechargeable battery will run the pump.
General instructions
1. Fill or spike feeding container and water bag prior to setting up pump.
2. Turn on pump by pressing Power button in lower right-hand corner.
3. To achieve proper accuracy, the uid line in the feeding set bag must be 6 in.
above the top of the feeding pump when initiating the feeding cycle.
4. Select
}
Keep Settings or
}
Clear Settings.
For day-to-day use with the same feeding regimen, select
}
Keep Settings.
To place the Kangaroo™ Joey Pump in Intermittent Mode
1. Select
}
More and then
}
Continuous/Intermittent.
Select
}
Intermittent Mode and then press
}
Done.
2. Load the feeding set per diagram on the feeding pump.* Don’t overstretch the tubing.
Priming the pump
1. Press
}
Prime Pump to access the pump priming options.
2. Press
}
Auto Prime to automatically prime the pump set. The pump will quickly prime
the feed line and stop the formula before reaching the end of the line.
3. Press and hold
}
Hold to Prime Feed to prime the feed line down to the connector
at the end of the pump set.
4. Press
}
Done.
Setting the bolus feed rate
1. Press
}
Adjust Bolus. Select
}
Bolus Rate.
Use the buttons on the left to set the rate from 1 to 400 mL per hour. Press
}
Enter.
2. Select
}
Number of Boluses.
Use the buttons on the left to set the number of bolus feeds; 99 maximum. Press
}
Enter.
3. Select
}
Volume per Bolus.
Use the buttons on the left to set the volume per bolus; 3,000 maximum. Press
}
Enter.
4. Select
}
Bolus Interval.
Use the buttons on the left to set the feed interval; 24 maximum. Press
}
Enter.
5. Press
}
Done.
Setting the ush interval/rate (option available only if a ushing set is loaded)
1. Select
}
Adjust Bolus, then
}
Flush Volume to set the volume of water administered per ush cycle
from 10 to 500 in increments of 1 mL. Press
}
Enter.
2. Select
}
Flush Interval to dene the time interval between the start of each ushing cycle
from 1 to 24 hours in increments of 1 hour. Press
}
Enter.
3. Press
}
Done.
4. Press
}
Run. A small drop will scroll down the screen when running.
Pump sets should not be reused after 24 hours of initial usage.
Quick Reference Guide: Intermittent operation
Cardinal Health™ Kangaroo™ Joey Enteral Feeding Pump
* For additional operating instructions and information, see the Operating
Manual for the Kangaroo™Joey Enteral Feeding Pump “Intermittent Operation.”
© 2020 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal
Health LOGO and KANGAROO are trademarks of Cardinal Health and may
be registered in the US and/or in other countries. All other marks are the
property of their respective owners. Lit. No. 2MS20-1146357 (04/2020)