manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Cardinal Health
  6. •
  7. Water Pump
  8. •
  9. Cardinal Health Kangaroo Joey Enteral User manual

Cardinal Health Kangaroo Joey Enteral User manual

Cardinal Health™ Kangaroo™ Joey
Enteral Feeding Pump
Quick Reference Guide
Intermittent
operation
Symptom Probable cause Corrections
Hold error Pump in Holding mode
for more than 10 minutes
Press
}
Continue to return to Holding screen, or press
and hold the Power button to turn o. Several options
are available from the Holding screen, such as
}
Run and
}
Adjust Settings.
Flow error Occlusion in pump-patient
line (downstream occlusion)
Check for and correct the occlusion in the line
between pump and patient. Replace Pump Set if
error can’t be resolved.
Feed error
Empty bag or occlusion in
bag-pump line
(upstream occlusion)
Check for empty bag and rell, or check for and correct
the occlusion in the line between the pump and feed bag.
Replace Pump Set if error can’t be resolved. Check the
valve pocket in the Pump Set loading area for dampness
or dirt.
Pump Set
use >24 Hrs
Pump Set in use beyond
recommended length of time
Informational warning message blinks on upper left of
screen. It’s recommended that the Pump Set be replaced.
Pump Set
dislodged
Pump Set not properly
loaded to pump
Check black ring retainer (MISTIC) and valve; reload Pump
Set and restart pump to continue. Replace Pump Set if
error can’t be resolved.
Battery low Battery charge too low Immediately plug pump into AC outlet to recharge battery
and to provide power for pump.
Rotor error
Pump Set tubing not
properly loaded on rotor,
or other unusual rotor
operating condition
Check that the tubing is loaded on the rotor and
that it’s not damaged, torn, etc. Reload tubing or
replace Pump Set.
System error
General error caused by
many factors; see Manual
for error list
Power down pump and attempt restart to clear error.
Call customer service with screen code if error
can’t be resolved.
Below are some of the probable causes of alarm conditions and how to correct them.
For additional operating instructions and
information, see the Operating Manual for
the Kangaroo™Joey Enteral Feeding Pump
“Intermittent Operation.”
Programming and Menu
Selection buttons
Power button
Status LEDs
Red: Error
Yellow: Hold/Pause
Green: Normal operation
Battery operation
Unplugging the pump from the AC line will automatically put the pump on battery power.
To recharge the battery, plug the pump into an AC wall outlet. The pump will automatically
begin charging. If AC power is interrupted, a built-in rechargeable battery will run the pump.
General instructions
1. Fill or spike feeding container and water bag prior to setting up pump.
2. Turn on pump by pressing Power button in lower right-hand corner.
3. To achieve proper accuracy, the uid line in the feeding set bag must be 6 in.
above the top of the feeding pump when initiating the feeding cycle.
4. Select
}
Keep Settings or
}
Clear Settings.
For day-to-day use with the same feeding regimen, select
}
Keep Settings.
To place the Kangaroo™ Joey Pump in Intermittent Mode
1. Select
}
More and then
}
Continuous/Intermittent.
Select
}
Intermittent Mode and then press
}
Done.
2. Load the feeding set per diagram on the feeding pump.* Don’t overstretch the tubing.
Priming the pump
1. Press
}
Prime Pump to access the pump priming options.
2. Press
}
Auto Prime to automatically prime the pump set. The pump will quickly prime
the feed line and stop the formula before reaching the end of the line.
3. Press and hold
}
Hold to Prime Feed to prime the feed line down to the connector
at the end of the pump set.
4. Press
}
Done.
Setting the bolus feed rate
1. Press
}
Adjust Bolus. Select
}
Bolus Rate.
Use the buttons on the left to set the rate from 1 to 400 mL per hour. Press
}
Enter.
2. Select
}
Number of Boluses.
Use the buttons on the left to set the number of bolus feeds; 99 maximum. Press
}
Enter.
3. Select
}
Volume per Bolus.
Use the buttons on the left to set the volume per bolus; 3,000 maximum. Press
}
Enter.
4. Select
}
Bolus Interval.
Use the buttons on the left to set the feed interval; 24 maximum. Press
}
Enter.
5. Press
}
Done.
Setting the ush interval/rate (option available only if a ushing set is loaded)
1. Select
}
Adjust Bolus, then
}
Flush Volume to set the volume of water administered per ush cycle
from 10 to 500 in increments of 1 mL. Press
}
Enter.
2. Select
}
Flush Interval to dene the time interval between the start of each ushing cycle
from 1 to 24 hours in increments of 1 hour. Press
}
Enter.
3. Press
}
Done.
4. Press
}
Run. A small drop will scroll down the screen when running.
Pump sets should not be reused after 24 hours of initial usage.
Quick Reference Guide: Intermittent operation
Cardinal Health™ Kangaroo™ Joey Enteral Feeding Pump
* For additional operating instructions and information, see the Operating
Manual for the Kangaroo™Joey Enteral Feeding Pump “Intermittent Operation.”
© 2020 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal
Health LOGO and KANGAROO are trademarks of Cardinal Health and may
be registered in the US and/or in other countries. All other marks are the
property of their respective owners. Lit. No. 2MS20-1146357 (04/2020)

Other Cardinal Health Water Pump manuals

Cardinal Health Alaris GP Manual

Cardinal Health

Cardinal Health Alaris GP Manual

Cardinal Health Alaris 7130/7131 Manual

Cardinal Health

Cardinal Health Alaris 7130/7131 Manual

Cardinal Health IVAC 571 Operating manual

Cardinal Health

Cardinal Health IVAC 571 Operating manual

Cardinal Health IVAC PCAM Manual

Cardinal Health

Cardinal Health IVAC PCAM Manual

Cardinal Health Kangaroo ePump User manual

Cardinal Health

Cardinal Health Kangaroo ePump User manual

Popular Water Pump manuals by other brands

Grundfos AP30 Installation and operating instructions

Grundfos

Grundfos AP30 Installation and operating instructions

Hyundai power products HYSP850D user manual

Hyundai power products

Hyundai power products HYSP850D user manual

Lincoln 83167 installation instructions

Lincoln

Lincoln 83167 installation instructions

Graco Monark instruction manual

Graco

Graco Monark instruction manual

Husky PL490 Operator's manual

Husky

Husky PL490 Operator's manual

ITT Industries JASCO JSP03 manual

ITT Industries

ITT Industries JASCO JSP03 manual

MPS HR1200 user guide

MPS

MPS HR1200 user guide

Drieaz S262 Replacement

Drieaz

Drieaz S262 Replacement

North Ridge Pumps DM troubleshooting guide

North Ridge Pumps

North Ridge Pumps DM troubleshooting guide

Grundfos TPE 2 Series Installation and operating instructions

Grundfos

Grundfos TPE 2 Series Installation and operating instructions

PFT BOLERO operating manual

PFT

PFT BOLERO operating manual

Clarke BPT600 Operating & maintenance instructions

Clarke

Clarke BPT600 Operating & maintenance instructions

Piusi Delphin PRO AC Installation, use and maintenance guide

Piusi

Piusi Delphin PRO AC Installation, use and maintenance guide

Ozito PXGTPS-0182 instruction manual

Ozito

Ozito PXGTPS-0182 instruction manual

Power Team PE55TWP-E110 operating instructions

Power Team

Power Team PE55TWP-E110 operating instructions

SymCom PumpSaver 235 installation instructions

SymCom

SymCom PumpSaver 235 installation instructions

Liberty Pumps 2LE Series installation manual

Liberty Pumps

Liberty Pumps 2LE Series installation manual

Pentair FAIRBANKS NIJHUIS 5430 Installation, operation and maintenance manual

Pentair

Pentair FAIRBANKS NIJHUIS 5430 Installation, operation and maintenance manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.