Cardinal Health IVAC 571 Operating manual

IVAC®Variable Pressure
Volumetric Pump
Models 571 & 572
Technical Service Manual

IVAC® Volumetric Pump (Models 571 & 572) 2/104 1000SM00018 Issue 2
Authorised EU Representive:
Cardinal Health*,
The Crescent,
Jays Close,
Basingstoke,
Hampshire
RG22 4BS,
United Kingdom
*Formerly ALARIS Medical UK Limited
t
Cardinal Health**,
10221 Wateridge Circle,
San Diego, CA 92121,
USA.
** Formerly ALARIS Medical Systems, Inc.
IVAC®is a registered trademark of
Cardinal Health, Inc. or one of its subsidiaries
All other trademarks belong to their respective owners.
© 1994 - 2008 Cardinal Health, Inc or one of its subsidiaries. All rights reserved.
This manual has been prepared for use by qualified service personnel only.
Cardinal Health cannot accept any liability for any breakdown or deterioration in
performance of parts or equipment resulting from unauthorised repair or modification.

Contents
Chapter 1 Introduction & Start Up 4
Chapter 2 Configuration & Calibration 13
Chapter 3 Routine Maintenance 22
Chapter 4 Troubleshooting 37
Chapter 5 Circuit Descriptions 43
Chapter 6 Spare Parts Replacement Procedures 47
Appendix A Specifications 83
Appendix B Spare Parts Listing 89
Appendix C Configured Options 97
Appendix D Disposal 99
Appendix E Service Contacts 101
Appendix F Document History 103

Chapter 1
Introduction and Start Up
In this chapter
Introduction 5
General Precautions 6
Views of Models 571 and 572 6
Controls and Indicators 8
Loading the IV Infusion Set 9
Starting the Pump 9
Programming 10
Operating on Battery Power 10
Secondary Infusion (Model 572) 11
Flow Sensor 12
Pole Clamp Accessories 12

IVAC® Volumetric Pump (Models 571 & 572) 5/104 1000SM00018 Issue 2
Introduction and Start Up
The IVAC® Volumetric Pump - Models 571 and 572 automatically regulates the infusion rate of IV solutions using a linear
peristaltic, volume displacement mechanism to regulate fluid flow at the prescribed rate. The pump can be operated with
a Flow Sensor and is compact and robust enough for most patient situations.
The IVAC® 571 Volumetric Pump operates in primary infusion mode only.
The IVAC® 572 Volumetric Pump has a primary and secondary infusion mode.
Product Familiarity
Ensure that you are fully familiar with the pump by carefully studying the Directions for Use (DFU) prior to operation and
prior to attempting any repairs or servicing. As part of continuous improvement, product enhancements and changes are
introduced from time to time.
Purpose of this Manual
This Technical Service Manual describes how to set up, test and maintain the following IVAC® Volumetric Pumps:
Model 571
Model 572
This manual is intended for use by personnel experienced in medical equipment testing and maintenance procedures.
Conventions Used in this Manual
BOLD Used for Display messages/values, self-test codes, controls and indicators referenced in this manual, for
example, CAL RATE, test code 11, ON/OFF switch.
'Single quotes' Used to indicate cross-references made to another section of this manual. For example, see
Chapter 2, 'Configuration and Calibration'.
underline Used to indicate a link to another section within this manual.
Italics Used to refer to other documents or manuals. For example, refer to the relevant Directions for Use
(DFU) for further information. Also used for emphasis, for example, ...if the gap still measures less
than...
Wherever this symbol is shown a Hints & Tips note is found. These notes provide useful advice or
information that may help to perform the task more effectively.
Wherever this symbol is shown a Toolbox note is found. These notes highlight an aspect of test
or maintenance that is important to know about. A typical example is drawing attention to a
software upgrade that you should check has been installed.
Introduction

IVAC® Volumetric Pump (Models 571 & 572) 6/104 1000SM00018 Issue 2
Introduction and Start Up
General Precautions
Prior to using this pump, carefully read the Operating Precautions described in the Directions for Use
(DFU).
This pump contains static-sensitive components. Observe strict precautions for the protection of
static sensitive components when attempting to repair and service the pump.
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise
care to locate the pump away from any such hazardous sources.
Dangerous Voltage. An electrical shock hazard exists if the casing of the pump is opened or
removed. Refer all servicing to qualified service personnel.
MThis pump is protected against the effects of radio frequency emissions and is designed to be
fail safe if extremely high levels of interference are encountered. Should false alarm conditions
be encountered, either remove the source of the interference or regulate the infusion by another
appropriate means.
)6 #
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise
suspected to have been damaged, remove it from service for inspection by qualified service
personnel.
dWhen connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be
used. If the integrity of the external protective conductor in the installation or its arrangement is in
doubt, the pump should be operated from the battery.
Views of Models 571 and 572
Front View
Switch Panel
(Model 571)
Numeric
Display
Information Display
Door Latch
Switch Panel
(Model 572)
Battery and
AC power
Indicators
Primary/Secondary/CRIS
Infusion Switches
(Model 572 only)

IVAC® Volumetric Pump (Models 571 & 572) 7/104 1000SM00018 Issue 2
Introduction and Start Up
Views of Models 571 and 572 (continued)
Rear View
Open Door View
053(
4/
2%3%4
&,/7
3%.3/2
PE Connector
AC Power
Connector
Flow Sensor
Connector
Test Connector
Cover
Pressure Unit
Selector Switch
Reset Switch
Flow Sensor Storage
Lower Tubing Guide Posts
Upper Tubing Retainer
Air-In-Line Detector
Pressure Sensing Disc
Retainer
Pressure Transducer
Tubing Pincher
Pumping Mechanism
Vacuum Plunger

IVAC® Volumetric Pump (Models 571 & 572) 8/104 1000SM00018 Issue 2
Introduction and Start Up
Controls and Indicators
UP/DOWN
Increases or decreases infusion rate and pressure and volume values. Three
columns for faster/slower increase or decrease:
Left = 100's, Centre = 10's, Right = units (or in micro mode Left = 10's,
Centre = units, Right = fractions).
TOT
VOL
INF
CLEAR Clears total volume infused when both switches are held down together.
TOT
VOL
INF
TOTAL VOLUME INFUSED Displays total volume infused value (primary + secondary volume infused).
Sets the Information Display to continuously monitor volume infused.
LIMIT
PRES
PRESSURE LIMIT Displays or changes the occlusion pressure alarm level limit.
READ
PRES
READ PRESSURE Displays the IV infusion site pressure.
Sets the Information Display to continuously monitor infusion pressure.
PRI
SEC
PRIMARY/SECONDARY
(Model 572 only). Switches between PRIMARY and SECONDARY infusion
modes. Allows displays of VTBI or rate setting of one mode while operating
in another.
CRIS
CRIS On Model 572: only used in Self-test mode.
On Model 571: hidden switch, only used in Self-test mode.
R A T E
RATE Changes the infusion rate.
VOL
TO BE
INF
VOLUME TO BE INFUSED
(VTBI)
Sets the value of Volume To Be Infused. Displays volume remaining to be
infused.
RUN
HOLD
RUN/HOLD Starts and stops pump infusion. Silences/cancels alarms and advances
start-up instruction messages.
ON
OFF
ON/OFF Turns the pump on and off.
AC POWER INDICATOR When illuminated, indicates the pump is connected to an AC power supply
and the battery is being charged.
BATTERY INDICATOR Flashes to indicate the pump is operating on battery power.
0With a Flow Sensor in use, an illuminated ("0") in the Information Display window indicates a drop is
detected by the flow sensor in the drip chamber.
Flashing Displays When the pump is operating on battery power, both Displays flash on/off.
Front Panel
The English switch panel controls and indicators are described below. For information on overlays in other languages,
refer to the relevant DFU. See also Appendix B, 'Spare Parts Listing' for switch panel part numbers.
Rear Panel
PRESSURE UNIT SELECTOR
SWITCH
Sets the displayed pressure units. Millimetres of mercury (mmHg) or
Centimeters of water (cmH20).
053(
4/
2%3%4
RESET SWITCH Resets the circuit breaker.

IVAC® Volumetric Pump (Models 571 & 572) 9/104 1000SM00018 Issue 2
Introduction and Start Up
With Flow Sensor
1. Load the primed IV infusion set.
2. Connect the flow sensor to the drip chamber
ensuring the flow sensor is plugged into the pump.
3. Press the ON/OFF switch to power on the pump.
4. Use the UP/DOWN switches (JK) to set the rate.
5. Press the RUN/HOLD switch to confirm.
6. Press VOL TO BE INF (if required) then use the
JK switches to set VTBI.
7. Press the RUN/HOLD switch twice to start infusion.
Without Flow Sensor
1. Load the primed IV infusion set.
2. Press the ON/OFF switch to power on the pump.
3. Use the UP/DOWN switches (JK) to set the rate.
4. Press the RUN/HOLD switch to confirm.
5. Use the JK switches to set VTBI.
6. Press the RUN/HOLD switch twice to start infusion.
Starting the Pump
Loading the IV Infusion Set
1. Prime the IV infusion set slowly ensuring all air is removed from the pressure disc then close the roller clamp.
2. Open the door.
3. Load the primed IV infusion set (see Figures 1-1 and 1-2):
Insert the set tubing into the upper tubing retainer.
Grasp the pressure sensor disc tabs between thumb and forefinger with the flat side (membrane) toward the
pump
Hook the pressure sensor disc under the disc retainer then press up and in to properly seat the disc
Place the tubing between the lower tubing guide posts.
4. Close and latch the door.
5. Open the roller clamp. Check drip chamber to ensure no fluid flows.
6. Connect to test equipment as required.
Follow the Directions for Use supplied with the individual IV infusion set
Only use IVAC® "52" Series IV infusion sets
Figure 1-1 Figure 1-2
Pressure Sensing
Disc Retainer
Air-In-Line
Detector
Upper Tubing
Retainer
Tubing Pincher
Lower Tubing
Guide Posts

IVAC® Volumetric Pump (Models 571 & 572) 10/104 1000SM00018 Issue 2
Introduction and Start Up
Changing the infusion rate
1. Press the RUN/HOLD switch to place the pump on hold.
2. Press the RATE switch.
3. Use the JK switches to set the new rate.
4. Restart the pump at the new rate by pressing the RUN/
HOLD switch.
Clearing the total Volume Infused
1. Press the RUN/HOLD switch to place the pump on hold.
2. Press and hold the TOT VOL INF switch and the K
switch directly above simultaneously for 2 seconds until
display shows 000.0.
3. Resume infusion by pressing the RUN/HOLD switch.
Changing the Volume To Be Infused
1. Press the RUN/HOLD switch to place the pump on hold.
2. Press the VOL TO BE INF switch.
3. Set new volume by pressing the JK switches. OFF can
also be selected when a flow sensor is in use, see Notes
below.
4. Restart the pump by pressing the RUN/HOLD switch.
Notes:
1) Without a flow sensor in use, a VTBI value must be entered,
otherwise, the pump will be unable to operate.
2) With a flow sensor in use, a VTBI value isn't required and
OFF can be selected if required.
Programming
Adjusting the maximum pressure limit
(Occlusion pressure alarm level)
1. Simultaneously press and hold the LIMIT PRES
switch whilst using the JK switches to adjust
the pressure limit.
2. Release the LIMIT PRES switch.
Note: Pressure value will be displayed in mmHg or in
cmH20 depending on unit selected. See 'Pressure
Unit Selector'.
Reading the pressure
1. Press and hold the READ PRES switch. Wait at least
10 seconds to allow reading to stabilize.
Note: Pressure value will be displayed in mmHg or
in cmH20 depending on unit selected. See 'Pressure
Unit Selector'.
Selecting alternating display of total
volume infused
1. Press and release the TOT VOL INF switch three
times within 2 seconds while the pump is infusing or
on hold.
On Model 572, the display will alternate between
primary or secondary infusion and total volume
infused.
On Model 571, the display will continuously
show the total volume infused.
Selecting alternating display of infusion
pressure
1. Press and release the READ PRES switch three times
within 2 seconds while the pump is infusing or on
hold.
On Model 572, the display will alternate between
primary or secondary infusion and infusion
pressure.
On Model 571, the display will continuously
show the infusion pressure.
Operating on Battery Power
The pump operates on battery power when it is disconnected from the AC power. The battery power indicator, the
Information and Numeric Displays will flash whenever the pump is on battery power. In the event of a power failure, the
pump will automatically continue to operate on battery power.
Two alarms indicate the condition of the pump's battery:
LOW BATT alternating with the selected volume or pressure status. This indicates that approximately 30 minutes of
operating time remains on battery power
LOW BATT (constant). This indicates that the battery is discharged. Connect the pump to an AC power supply to
recharge the battery.
Note: The pump's battery is designed for limited duration use. Wherever possible, the pump should be used connected
to an AC power supply. If the pump is to be taken out of service for an extended period, it is good practice to charge the
battery periodically to ensure full capacity.

IVAC® Volumetric Pump (Models 571 & 572) 11/104 1000SM00018 Issue 2
Introduction and Start Up
Secondary Infusion (Model 572)
Starting the Pump with Secondary Infusion
1. Load the primed set. See 'Loading the IV Infusion Set' earlier in this
chapter for instructions.
2. Prepare the secondary infusion using a secondary solution container
and the IVAC® "52" Series back check valve set; lower the primary
container. See Figure 1-3 below.
Prime the secondary IV infusion set in accordance with the set
Directions for Use
Attach a secondary IV infusion set to the upper Y-site of the
primary IV infusion set
3. Press the ON/OFF switch to power on the pump. Pump always starts
up in PRIMARY mode.
4. Use the JK switches to set the primary rate.
5. Press VOL TO BE INF if required.
6. Set primary VTBI if required.
7. Press the SEC switch to place the pump into SECONDARY mode. The
pump will sound a four beep verification tone.
8. Use the JK switches to set the secondary rate.
9. Press the VOL TO BE INF switch.
10. Set secondary VTBI by using the JK switches.
11. Press RUN/HOLD to start secondary infusion.
WARNING:
Secondary infusion applications using a back check
valve set must have a VTBI setting equal to the
volume in the secondary container; this will require
consideration of such variables as factory overfill,
medication additions, priming volume, etc.
Underestimating the volume will cause remaining
secondary solution to be infused at the primary rate;
overestimating will result in primary solution being
infused at the secondary rate.
Introduction
This mode of operation supports automatic secondary infusions ("piggybacking") in the same pump channel. When
the secondary VTBI reaches zero, a transition tone will sound and the primary settings will automatically take effect.
Secondary mode can be used where a second, independent VTBI is required, and also when an automatic rate change is
required.
When the pump is programmed and delivering in the secondary mode, the primary infusion is temporarily stopped and
fluid is drawn from the secondary (higher) container. Delivery from the primary container resumes when the fluid level in
the secondary line is level with the fluid in the primary container.
Notes:
1) The primary fluid container must hang lower than the secondary fluid container to allow the secondary infusion to
run.
2) On completion of the secondary infusion the pump will automatically return to primary infusion.
When using a flow sensor it must be positioned on the
primary IV infusion set (as shown in Figure 1-4). Correct
placement of a flow sensor is essential for proper
operation.
Figure 1-4
Secondary Infusion Set Up
with Flow Sensor
Flow
Sensor
Figure 1-3
Secondary Infusion Set Up
Y-Site
Primary
container
Secondary
container
IV Infusion set
patient end
Back Check
valve

IVAC® Volumetric Pump (Models 571 & 572) 12/104 1000SM00018 Issue 2
Introduction and Start Up
Verifying a value in one mode while operating in another
1. Press and release the appropriate (current) mode switch. For example, SEC.
2. Within two seconds press the switch representing the information to be displayed. For example, PRI.
Changing the infusion manually to primary mode
1. Press the RUN/HOLD switch to set pump on hold.
2. Press the PRI switch; pump sounds a three beep verification tone and goes into PRIMARY mode.
Secondary Infusion (Model 572) continued
Additional Programming during Secondary Infusion
Flow Sensor
A flow sensor (part number 180) is used to detect an empty solution container. It is connected to the drip chamber and
when a drop is detected in the drip chamber an illuminated ("0") appears in the Information Display.
When not in use, the flow sensor can be stored on the pump handle.
Pole Clamp Accessories
The following pole clamp accessories allow the pump to be mounted on conventional IV poles:
Universal Pole Adapter (part number 6767)
Pole Adapter (part number 5767A)*
Pole Clamp Accessory (part number BC100A)*
* must be used in conjunction with each other.
Refer to Chapter 6, 'Spare Parts Replacement Procedures' for assembly information.

IVAC® Volumetric Pump (Models 571 & 572) 14/104 1000SM00018 Issue 2
Configuration & Calibration
Self-test mode is used to access a number of configuration and test routines which are designed to test and verify many
of the pump functions, defaults and calibrations without requiring internal inspection. Configuration settings are also
accessed via self-test mode.
Each test is accessed by selecting a Test ID number as shown in the table below.
Self-Test Mode
Test ID Description
001 Software Revision Display
002 Configuration Status Display
003 Flash Display Test
004 Segment Display Test
005 Sequence Display Test
006 EEPROM Test and Data Display
007 A/D Channels Test
008 Rate Calibration Display/Set
009 * Configuration Rate Display/Set
010 Maximum Rate Display/Set
011 * VTBI Configuration Display/Set
012 Pressure Display
013 Pressure Calibration Set
014 Maximum Occlusion Pressure Limit Display/Set
015 Default Occlusion Pressure Limit Display.Set
016 Alarm History Display/Clear
017 Switch Test
018 AC Power Run Time Display/Reset
019 Battery Run Time Display/Reset
020 Secondary Mode Completion Tone Display/Set
021 Air-In-Line Test
022 I/O Port Test
023 Language Display/Set
024 CIM ID Display/Set
025 Alarm Tone Display/Set
026* AIL Alarm Threshold Display/Set
027 System Vacuum Display
028 Enter Operational Setup Mode
029 Burn-In Test
030 Not used
031* Infusion Mode Configuration Display/Set
* A mode select jumper (part number 132350) must be
used to change settings, see 'Mode Select Jumper' for
further details.
Exiting Self-Test Mode
Self-test mode can be exited at any time:
Turn the pump off by pressing the ON/OFF switch
OR
Select test 028 then press the RUN/HOLD switch
to enter the start-up mode of operation
Entering Self-Test Mode
1. Press the RUN/HOLD and ON/OFF switches
simultaneously until the display turns on. Initially,
the current software revision level is displayed.
2. Press the RUN/HOLD switch. The Information
Display will read: TEST ID and the Numeric
Display will read: 002 which means the pump is
ready to start test 002.
Note: The display may bypass the display of the
software revision level and advance directly to
TEST ID 002, depending on how long the RUN/
HOLD switch is held down.
3. Use the JK switches to select the Test ID
required.
Test Execution
1. After selecting the Test ID number (see above),
press the RUN/HOLD switch to start the test.
2. To exit a test, press the RUN/HOLD switch.
The three columns of JK switches
allow faster/slower increments:
Left = 100's
Centre = 10's
Right = Units

IVAC® Volumetric Pump (Models 571 & 572) 15/104 1000SM00018 Issue 2
Configuration & Calibration
In order to perform certain tests and to change any pump settings, a mode select jumper must be installed.
Procedure:
1. With the pump off, remove the test connector cover from the back of the rear case.
2. Connect the mode select jumper (part number 132350) to the test connector, orientating the jumper so that CR1 is at
the top.
3. Place the pump in self-test mode then select and carry out the test(s) as required.
4. When the test is complete, switch the pump off and remove the mode select jumper from the test connector.
5. Refit the test connector cover.
Mode Select Jumper
Configuration Settings
Test ID Setting Action/Data Displayed
009 Rate Configuration
RATE CFG
Display and/or set the current rate range.
STANDARD = rate range of 001 to 999 ml/h
ALL RATE = rate range of 0.1 to 999.9 ml/h. Note: Use this setting where
fractional flow rates are required.
To change the setting:
Install the mode select jumper. See 'Mode Select Jumper'.
Press the CRIS (lower hidden switch on Model 571) and VOL TO BE INF
switches together and wait for a click.
Press the Junit (right column) switch to select STANDARD or the K unit
switch to select ALL RATE.
010 Maximum Rate
MAX RATE
Display and/or set the maximum infusion rate.
999 = set to STANDARD rate configuration (ml/h). This rate cannot be changed.
0.1 and 999.9 = set to ALL RATE configuration (ml/h).
Note: Rate configuration is set in Test 009.
To change the maximum rate setting (for ALL RATE configuration):
Press and hold the PRI (upper hidden switch on Model 571) while using the
J K switches to adjust the value.
011 VTBI Configuration
VTBI CFG
Display and/or set the Volume To Be Infused (VTBI) limit. VTBI limit ranges are:
MACRO = 1 to 9999 ml, MICRO = 0.1 to 999.9 ml.
If set to ALL RATE configuration (see Test 009), VTBI CFG can be set to MACRO
or MICRO.
If set to STANDARD rate configuration, VTBI CFG is set to MACRO and is not
adjustable.
To change the VTBI limit (for ALL RATE configuration):
Install the mode select jumper. See 'Mode Select Jumper'.
Press the CRIS (lower hidden switch on Model 571) and VOL TO BE INF
switches together and wait for a click.
Press the Junit switch to select MACRO or the K unit switch to select
MICRO.
014
Maximum Occlusion
Pressure
MAXIMUM
Display and/or set the maximum occlusion pressure limit. Pressure values are
displayed in MMHG or CMH20depending on the pressure units selected. See
'Selecting Pressure Units' for further information.
This pressure limit setting determines the upper limit of the pumps' variable
pressure range thereby fixing the upper limit when the LIMIT PRES switch is
used during normal operation.
To change the setting:
Press and hold the PRI (upper hidden switch on Model 571) while using the
J K switches to adjust the value.
Press the RUN/HOLD switch to save the new value.
Enter Self-Test mode. See 'Entering Self-Test Mode' in previous section for instructions.
Note: For default settings refer to Appendix C 'Configured Options'.

IVAC® Volumetric Pump (Models 571 & 572) 16/104 1000SM00018 Issue 2
Configuration & Calibration
Test ID Setting Action/Data Displayed
015
Default Occlusion
Pressure
DEFAULT
Display and/or set the default pressure limit.
Pressure values are displayed in MMHG or CMH20depending on the pressure
units selected. See 'Selecting Pressure Units' for further information.
This setting determines the occlusion pressure limit at power up.
To change the setting:
Press and hold the PRI (upper hidden switch on Model 571) while using the
J K switches to adjust the value.
Press the RUN/HOLD switch to save the new value.
020
Secondary Mode
Completion Tone
(Model 572 only)
Display and/or set the audible tone that sounds when the SECONDARY infusion
finishes and switches to PRIMARY infusion.
1 CLICK = one-click tone
4 BEEPS = four-beeps tone
To change the setting:
Press the Junit switch to select 4 BEEPS or the K unit switch to select 1
CLICK.
023 Language
Display and/or set the language to be displayed. Languages are ENGLISH,
DUTCH, FRENCH, GERMAN, ITALIAN, SPANISH or SWEDISH.
To change this setting:
Press the J K unit switches to scroll through list and select language.
024 CIM ID
CIM ID
Setting is not in use. OFF = CIM not in use, 001 to 999 = CIM ID number
Important: Ensure the setting is OFF.
To change setting to OFF:
Press and hold the VOL TO BE INF switch while using the J K switches to
select OFF.
Press the RUN/HOLD switch to save the new setting.
025 Alarm Tone
Display and/or select the alarm tone volume level.
Approximate decibel volumes are:
HI TONE = 90 db
MED TONE = 80 db
LOW TONE = 70 db
To change the setting:
Press the Junit switch to select alarm tone.
WARNING: Take any necessary precautions against prolonged exposure to the
HI TONE alarm as this may result in hearing damage.
026 AIL Alarm Threshold
UL AIR
Display and/or set the air bubble length maximum value. This value is the
nominal air bubble threshold in microlitres used for air-in-line detection. Value
must be between 50 and 150 microlitres.
To change the setting:
Install the mode select jumper. See 'Mode Select Jumper'.
Press the CRIS (lower hidden switch on Model 571) and VOL TO BE INF
switches together and wait for a click.
Press the J K switches to enter the value (in 10 µl increments).
031 Infusion Mode
INFM CFG
Display and/or set the pump infusion mode.
PRI ONLY = PRIMARY infusion mode only (Model 571).
SEC/CRIS = PRIMARY and SECONDARY infusions modes. (Model 572).
To change the setting:
Install the mode select jumper. See 'Mode Select Jumper'.
Press the CRIS (hidden switch on Model 571) and VOL TO BE INF switches
together and wait for a click. Press the J unit switch to select PRI ONLY or
the K unit switch to select SEC/CRIS.
Note: Model 571 must be set to PRI ONLY.
Configuration Settings (continued)

IVAC® Volumetric Pump (Models 571 & 572) 17/104 1000SM00018 Issue 2
Configuration & Calibration
Calibration Procedures
Equipment required:
50ml syringe
Pressure gauge
"52" Series IV infusion set
Procedure:
1. Remove the battery cover from the pump to
allow access to the Accessory Interface Board.
See Chapter 6, 'Spare Parts Procedures' for
instructions.
2. Prepare a modified IV infusion set as follows:
Completely remove the clear membrane
cover from the white pressure sensing disc.
Ensure no membrane remains as this could
compromise the integrity of the vacuum
Attach the air-filled IV infusion set to the
syringe. Use the luer fitting on the distal end
of the set or cut approximately 1 inch (2.5
cm) off the distal end and press the tubing
firmly into the tip of the syringe. Ensure a
good seal is formed between the two devices
Splice a T-fitting into the IV infusion set
tubing approximately 1 foot (30 cm) up from
the syringe attachment. Connect the pressure
gauge to the T-fitting port
Push the syringe plunger in until it stops
3. Load the modified IV infusion set into the pump
and close the door. Ensure that the set tubing is
not kinked.
4. Enter self-test mode and select test 027. See 'Entering
Self-Test Mode' for instructions.
5. Press the RUN/HOLD switch to enter the Vacuum Display mode and display VAC MMHG.
6. Pull the syringe plunger out until the pressure gauge reads -95 mmHg and immediately clamp off tubing below the
pump to hold the vacuum stable.
7. Adjust the vacuum detect potentiometer R45 on the Accessory Interface Board until the numeric display on the
pump reads -95 ± 5 mmHg.
Note: R45 is the upper potentiometer on the Accessory Interface Board.
8. Perform the 'Vacuum Retention Verification' procedure as detailed in
Chapter 3, 'Routine Maintenance'.
Vacuum System Calibration (Test 27)
Syringe
Pressure
Gauge
Proximal
End of Set
Selecting Pressure Units
The pressure units can be configured to be displayed as millimetres of mercury (mmHg) or
as centimeters of water (cmH20).
To select the pressure units:
Set the Pressure Unit Selector switch on the rear panel to the required position. Pressure Unit
Selector Switch
Do not adjust the lower
potentiometer, R4, this
is pre-adjusted during
manufacture.

IVAC® Volumetric Pump (Models 571 & 572) 18/104 1000SM00018 Issue 2
Configuration & Calibration
Calibration Procedures (continued)
Dry Set Pressure Calibration (Test 13)
Equipment required:
Test jumper cable (142109)
Pressure gauge
"52" Series IV infusion set (dry set)
Procedure:
1. Remove the chassis from the body. See Chapter 6, 'Spare Parts Procedures' for instructions.
2. Attach the jumper of the test jumper cable between connector J7 on the Logic Board and connector P7 on the
Accessory Interface Board.
3. Connect the battery pack to the battery connector.
4. Set up equipment as shown below.
Note: Set the pressure unit selector switch (on the back panel of the pump) to the same units as the air pressure
source.
5. Enter self-test mode, press the RUN/HOLD switch and select test 12. Press the RUN/HOLD switch again to display the
pressure MMHG (or CMH20).
Ensure the pump is connected to the AC power supply for approximately 30 minutes before continuing with this
calibration.
6. Enter self-test mode, press the RUN/HOLD switch and select test 13. Press the RUN/HOLD switch again to display
PRES CAL.
7. With the door open and no set loaded, adjust the zero potentiometer (R20 on the Analog Board) so the pump reads
000 ±1 mmHg (or 000 ±1.4 cmH20).
8. Install an IV infusion set, close the door and apply a test pressure of 000 mmHg (000 cmH2O) to the distal end of the
set. Record the reading displayed as the PI offset value.
9. Apply a test pressure until the meter reads 400 mmHg (544 cmH2O) and adjust the span potentiometer (R23 on the
Analog Board) so the pump reads 400 mmHg ±2 (544 cmH20 ±10 ) plus the offset value, PI, recorded in the previous
Step.
10. Return the test pressure to 000 mmHg (000 cmH2O), bleed off any air from the IV infusion set then remove the IV
infusion set.
Note: If the set is removed when still under pressure may result in damage to the pressure sensing disc film.
11. Leave the door open and reverify that the pump reads 000 ±1 mmHg (000 ±1.4 cmH2O). If not, repeat the procedure
from Step 7, until no further adjustment of the zero and span potentiometers is required.
6
Reservoir
Pressure
Gauge
Air
Pressure Source
Proximal
End of Set

IVAC® Volumetric Pump (Models 571 & 572) 19/104 1000SM00018 Issue 2
Configuration & Calibration
Dry Set Pressure Calibration (Test 13) continued
12. Press the Junit and RATE switches simultaneously for a minimum of one second to set the 000 mmHg (000 cmH2O)
calibration point.
13. Reload the IV infusion set and apply test pressure until the meter reads 400 mmHg (544 cmH2O). When the pressure
reading is stable, press the TOT VOL INF and the J100 switches simultaneously for a minimum of one second to set
the 400 mmHg (544 cmH2O) calibration point.
14. Press the RUN/HOLD switch to return to test 13 and turn the pump OFF. Remove the IV infusion set (first bleeding off
any pressure).
15. Disconnect the battery pack from the battery connector.
16. Remove the test jumper cable and reassemble the pump.
17. Perform the 'Dry Set Pressure Verification' procedure as detailed in Chapter 3, 'Routine Maintenance'.
Calibration Procedures (continued)
Both Steps 12 and 13 must be performed to correctly set the calibration parameters. Performing
only one step will result in the message RECONFIG REQD when the pump returns to operation.

IVAC® Volumetric Pump (Models 571 & 572) 20/104 1000SM00018 Issue 2
Configuration & Calibration
Calibration Procedures (continued)
Rate Accuracy Calibration (Test 8)
There are two types of rate accuracy associated with the pump, these are System Rate Accuracy and Pump Rate Accuracy.
System Rate Accuracy is the rate accuracy of the pump when used with a standard "52" Series infusion set. Due to slight
variations in between sets, the system accuracy of the pump is ± 5%. Pump Rate Accuracy is the absolute rate accuracy of
an individual pump which can only be determined with the use of a characterised set.
The following procedure describes how to calibrate the Pump Rate Accuracy. For instructions on how to verify Pump Rate
Accuracy, see 'Rate Accuracy Verification' in Chapter 3, 'Routine Maintenance'.
Equipment required:
Characterized set (191806)
Stopwatch
Class A burette (50 ml) with 0.1 ml
graduations
Preparation:
1. Set up the equipment as shown.
The vertical distance from the
fluid level of the container to
the pump pressure transducer
should be 34 ±4 inches
The burette zero line should
be within ± 2 inches from the
pump pressure transducer.
Verify that the inside of the
wall is free of droplets
Ensure fittings are secure and
there is no leakage through
the 3-way stopcock
Ensure that the air has been
purged from the characterised
set and all equipment
2. Close the set roller clamp.
3. Plug the pump into the AC mains
and turn the pump ON.
Characterised Sets (part number 191806)
Must be used for rate accuracy calibration
Sets are labeled with a serial number, a nominal volume and an expiry date (180 days from the date
the set was characterised)
Handle sets with care at all times. Sets must not be pressurised. Prior to loading or removing from
the pump, manually open the auto-pinch-off mechanism. Do not stretch the set excessively
Record the number of times each set is used in a log book. A set must not be used for more than 20
rate runs.
Burette
Water Source
Burette Clamp
Table
or Bench
Tubing
Set Tubing
Used Fluid
Recptacle
Three-way
Stopcock
Pressure Transducer
(level with)
34 ± 4 inches (86 ± 10 cm)
Equipment Stand
This manual suits for next models
1
Table of contents
Other Cardinal Health Water Pump manuals
Popular Water Pump manuals by other brands

Einhell Royal
Einhell Royal HWA 1100 NIRO operating instructions

Liberty Pumps
Liberty Pumps Builder S30 Series installation manual

TELLARINI POMPE
TELLARINI POMPE EEM Series User and maintenance manual

Wilo
Wilo DrainLift S Series Installation and operating instructions

Hitachi
Hitachi CHROMASTER 5160 instruction manual

Pfeiffer Vacuum
Pfeiffer Vacuum HILOBE 1004 operating instructions