To Prime the Administration Set
1. Use only MedSystem III®administration
sets.
2. Close roller clamp. Spike container. Fill drip
chamber to approximately half full.
3. Confirm that vent is in the correct position.
4. With cassette inverted, prime set slowly.
Use a syringe to aspirate air from
SmartSite® needle-free valve. Massage
pressure dome to expel all air.
To Insert the Cassette
Note: Pump must be turned ON prior to
cassette loading. Slide clamp must be in
the fully closed (out) position.
1. With tubing downward, insert cassette
upward and in at a 45° angle.
2. Press the slide clamp in until flush with
the cassette.
3. Gently pull tubing collar down and press
it into the recess below the cassette.
4. Three "beeps" indicate proper placement.
To Set Up a Primary Infusion
1. Select channel by pressing "A, Bor C".
2. Press SELECT to highlight the value you
wish to program (Rate or Volume
Remaining or to clear the Volume Infused.)
3. Use ↑, ↓, Fast ↑, or Fast ↓ to reach the
desired value, then press ENTER.
The highlight bar will automatically move
to the next field.
Note: The direction of the "Fast" key can be
changed by pressing the corresponding
↑or ↓arrow.
4. To recall a setting before Enter is pressed,
press MORE OPTIONS then RECALL.
5. Press START/STOP to begin or stop infusion.
6. VERIFY:
· flow of solution from the primary solution
container
· settings on the standard display are correct
· Infusion Status "Infusing" on the selected
channel.
To Change Rate or VTBI
1. Select channel by pressing "A, Bor C".
2. Change the rate using ↑, ↓, Fast ↑, or Fast
↓to reach the desired value, then press
ENTER.
3. Or press SELECT to move to the Volume
Remaining field. Change the VR using
↑, ↓, Fast ↑, or Fast ↓to reach the
desired value, then press ENTER.
Note: Infusion does not stop during changes.
New rate is in effect immediately upon
pressing the Enter key.
To Clear the Volume Infused
1. Select Channel by pressing "A, B, or C".
2. Press SELECT to move to the Volume
Infused (VI) field.
3. Press CLEAR to reset to zero.
Press ENTER.
Note: Information does not clear from the Total
Volume page.
OR
1. Press MORE OPTIONS, then TOTVOL.
Total Volume page appears.
2. Press CLRTOT. All values for each channel
will clear and flash. Press ENTER.
Note: Information will be cleared from all
channel pages, A, B, and C.
To Stop a Channel
Temporarily - "Stopped" Status
1. Select channel by pressing "A, Bor C".
2. Press "START/STOP".
3. Verify Infusion status "Stopped" on the
selected channel. If cassette remains in the
channel, a "CHANNEL NOT IN USE"
advisory will sound in two minutes.
Note: In the OR Device Modes, there is no "Stopped"
option, and "Standby" mode is reached directly
when "STOP/START" is pressed.