CEFLA I-MMr User manual

97050840
Rev. 03
2022-10
MICROMOTOR
OPERATOR'S MANUAL
I-MMr • i-MMr L • i-MMs
EN
.

I-MMr • i-MMr L • i-MMs
2
EN
Contents
1. DESCRIPTION................................................................................................................................................................................................. 3
1.1. DESCRIPTION OF THE DEVICE................................................................................................................................................................3
1.2. INDICATIONS FOR USE ............................................................................................................................................................................3
1.3. IMPORTANT WARNINGS...........................................................................................................................................................................3
1.4. SYMBOLS...................................................................................................................................................................................................4
2. TECHNICAL DATA.......................................................................................................................................................................................... 5
2.1. MODE OF USE...........................................................................................................................................................................................6
2.2. ENVIRONMENTAL CONDITIONS FOR USE..............................................................................................................................................6
2.3. ENVIRONMENTAL CONDITIONS PERMITTED FOR TRANSPORT AND WAREHOUSING......................................................................6
2.4. APPLIED PARTS........................................................................................................................................................................................6
3. OPERATION OF THE DEVICE........................................................................................................................................................................ 6
3.1. CONNECTION TO THE POWER SUPPLY INSTRUMENT TUBING...........................................................................................................6
3.2. CONNECTION AND DISCONNECTION OF THE HANDPIECE..................................................................................................................6
3.3. OPERATING PARAMETERS......................................................................................................................................................................6
3.4. ILLUMINATION DEVICE.............................................................................................................................................................................6
4. REPROCESSING ............................................................................................................................................................................................ 7
4.1. CLEANING..................................................................................................................................................................................................7
4.2. STERILISATION.........................................................................................................................................................................................8
4.3. PRODUCTS TO BE USED..........................................................................................................................................................................8
5. MAINTENANCE............................................................................................................................................................................................... 9
5.1. REPAIRS AND OVERHAULS.....................................................................................................................................................................9
5.2. TROUBLESHOOTING ................................................................................................................................................................................9
6. SCRAPPING.................................................................................................................................................................................................. 10
7. GUARANTEE CONDITIONS ......................................................................................................................................................................... 10
8. DISPOSAL..................................................................................................................................................................................................... 10
9. MANUFACTURER......................................................................................................................................................................................... 10

I-MMr • i-MMr L • i-MMs
EN
3
1. DESCRIPTION
1.1. DESCRIPTION OF THE DEVICE
Name
Description
Light
Especially suitable for
Supplied as
95520443
i-MMs
MICROMOTOR
Yes
Implant & Endodontic
Non-Sterile and
Re-usable
95520452
i-MMr
MICROMOTOR
No
Prosthetic & Restorative
Non-Sterile and
Re-usable
95520435
i-MMr L
MICROMOTOR
Yes
Prosthetic & Restorative
Non-Sterile and
Re-usable
1.2. INDICATIONS FOR USE
The CEFLA Dental Micromotors are brushless electric micromotors controlled by a control unit inside CEFLA Dental Units.
They are intended to be connected with an ISO-type handpiece attachment: straight or contra-angle of equal, gear-reducing, or gear increasing speed.
They are intended for professional use in dental surgery such as: preventive dentistry, restorative applications, endodontic treatment, prosthetic
applications and implantology practices.
Read carefully the user manual before using Micromotor.
The use of the device must respect the instructions provided.
1.3. IMPORTANT WARNINGS
For the correct interpretation of the indications in this manual the Italian text is binding.
•The device is not suitable for use with blends of inflammable anaesthetics and oxygen or nitrous oxide.
•The instrument is provided in a non-sterile state. Before use, proceed with the applicable reprocessing procedure as detailed in this manual.
•Use of electrically-powered devices may interfere with the operation of active implantable devices such as pacemakers or other active devices. In
case of doubt regarding the treatment of patients with such devices, consult a cardiology expert or another competent medical institute.
•During use of the device dust and fragments from the patient’s oral cavity or from the same device may be projected into the environment (organic
and inorganic dust and metal dust, fragments of the device or its tips and potentially infected biological material).
•Protect the patient, when possible, using a dental dam.
•Instruct the patient to breathe through the nose in cases where the dam is not applicable.
•Medical personnel must put on suitable personal protection equipment.
•Suitably cool the surgical field during use.
•Use only handpieces and counterangles approved and legally marketed in your country.
•Use only burrs (or other similar tools) approved and legally marketed in your country and made with biocompatible materials complying with ISO
10993-1 and whose dimensions comply with EN ISO1797-1 standards.
•Use only lubricant approved and legally marketed in your country for dental handpieces.
•Before each use, verify the proper locking of the handpiece onto the micro-motor and of the burr onto the handpiece chuck.
•Do not use damaged or worn handpieces.
•Make sure that during operation of the motor there is full flow of cooling air. In the absence of cooling stop using the device and call authorised
technical service assistance.
•In case of visible damage, the emission of unusual noises and/or vibrations or if overheating is noticed, do not use the device and call authorised
technical service assistance.
•LED radiation: do not stare into beam or view directly with optical instruments.
The manufacturer will not be held responsible regarding safety, reliability or performance of the device if:
•The essential requirements for the location, detailed in the Dental Unit User Manual are not respected;
•The assembly, the additions, the adjustments, the calibrations and/or the repairs are not performed by authorised technical service personnel;
•If modifications, tampering, incorrect maintenance have been performed, or incompatible and/or non-original spare parts have been used on the
device;
•The device is not used in compliance with the use instructions and its intended use.
CONTRAINDICATIONS
There are no contraindications related to the use of micromotors because they are not intended to get in contact with the patient.

I-MMr • i-MMr L • i-MMs
4
EN
1.4. SYMBOLS
WARNING!
Failure to observe may result in equipment damage or injury to the user and/or patient.
Symbol to call the attention to other information found in the equipment User Manual.
NOTE: Identifies information that is especially important for the user and/or the assistant.
Symbol corresponding to “TYPE B PART APPLIED” according to IEC 60601-1.
Indicates a protection grade against direct and indirect contacts.
Part that may be sterilised in an autoclave.
Medical Device compliant with Directive 93/42/EEC as amended (Directive 2007/47).
CE issued by IMQ (0051).
In compliance with reference standards IEC EN 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62304, ISO 15223-1, ISO 9687,
Directive 2002/96/ECC and internal symbol meaning.
Disposal symbol in accordance with Directive 2012/19/EU.
Device identification code.
Device serial number.
Manufacturer.
For use only by trained medical personnel.
Mark of conformity with technical regulations of Ukraine.
China RoHS mark with Environmental Protection Use Period.

I-MMr • i-MMr L • i-MMs
EN
5
2. TECHNICAL DATA
i-MMs
i-MMr
i-MMr L
Type
Brushless
Power supply
32Vdc ± 12,5%
Max absorption
6 A
4,5 A
Max Electrical power absorbed
170 W
120 W
Max speed
40.000 ± 10% revs/min
40.000 ± 10% revs/min
Min speed
100 ± 10% revs/min
100 ± 10% revs/min
Max torque
5,3 Ncm
3,3 Ncm
Cooling type
Forced Air
Working times
Intermittent: 5 min. ON, 5 min. OFF
Pneumatic supply
300 ± 20 kPa (3 ± 0,2 bar)
Water Supply
250 ± 20 kPa (2,5 ± 0,2 bar)
Consumption
Cooling air
~ 25 Nl/min
Air Spray
~ 5 Nl/min
Water Spray
~ 150 cc/min
Spray
Integrated
Classification
Class IIa
(Directive 93/42 EEC and subsequent additions and amendments)
Class II type B (IEC 60601-1)
Emissions classification
CISPR 11 Class A Group 1
(IEC 60601-1-2)
Handpiece Connection
ISO 3964 TYPE 2 for version without LED.
ISO 3964 TYPE 3 for version with LED compatible with INTRAmatic® Lux.
LED Light
5000 K - 20000 mlm
Applicable regulations
IEC EN 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62304, ISO 15223-1, ISO 9687, Directive
2002/96/ECC
Dimensions
47,5 mm
35 mm
Weight
79 g
66,5 g
Cooling fluid regulation
On dental equipment
For the correct operation of the device it must be connected to the specific power supply and electronic control circuits, designed by the manufacturer,
by way of suitable connection supply tubing.
The manufacturer promises to provide, upon request, wiring diagrams, component parts lists, calibration instructions or any other information that may
be needed by authorised technical service assistance personnel.
The manufacturer reserves the right to bring about modifications at any moment without notice.

I-MMr • i-MMr L • i-MMs
6
EN
2.1. MODE OF USE
The method of application of the device is intermittent. The operational cycle indicated respects the following times:
Type of Instrument
Operation (minutes)
Pause (minutes)
i-MMs
5
5
i-MMr
5
5
i-MMr L
5
5
2.2. ENVIRONMENTAL CONDITIONS FOR USE
•Ambient Temperature 10 ÷ 40 °C
•Relative Humidity 30 ÷ 75 %
•Atmospheric Pressure 700 ÷ 1060 hPa (700 ÷ 1060 mBar)
2.3. ENVIRONMENTAL CONDITIONS PERMITTED FOR TRANSPORT AND WAREHOUSING
•Ambient Temperature between -20 ÷ +70 °C
•Relative Humidity 10 ÷ 100 %
•Atmospheric Pressure 500 ÷ 1060 hPa (500 ÷ 1060 mBar)
2.4. APPLIED PARTS
The parts that, during standard use, necessarily come into contact with the patient in order for the device to carry out its functions correctly, are the
handle covers.
3. OPERATION OF THE DEVICE
3.1. CONNECTION TO THE POWER SUPPLY INSTRUMENT TUBING
Bring the motor next to the instrument tubing and rotate it, on the connector itself until the coupling point is found. Insert the motor completely and then
fully tighten the ring nut.
3.2. CONNECTION AND DISCONNECTION OF THE HANDPIECE
Handpieces complying with ISO 3964 and with the INTRAmatic®Lux. type may be used. Insert the handpiece on the coupling until the connection “click”
is heard.
Never mount a handpiece onto a motor that is running.
To detach the handpiece from the motor pull lightly.
3.3. OPERATING PARAMETERS
Operational modes, speed adjustment, inversion of the rotation direction, continous or reciprocating movements, torque adjustment, spray and LED
lighting activation (if available) are to be made through the Dental Unit command panel. Refer to the Dental Unit instruction manual for the setting of the
operational parameters.
3.4. ILLUMINATION DEVICE
In the micro-motors set up with LEDs for lighting of the surgical field, these light up automatically when the micro-motor is running.
After the motor stops, the LED remains lighted for a pre-set time. Lighting depends on settings made on the Dental Unit.

I-MMr • i-MMr L • i-MMs
EN
7
4. REPROCESSING
RACCOMANDATION: reprocess shall be performed within 30 minutes of the end of use.
The micromotors have been tested for 250 cycles (as required by EN ISO 14457).
Before sanitisation operations, the operator must detach the handpiece from the micromotor and the micromotor from the dental
units tubing.
Before reprocessing, visually inspect the micromotor. Reprocess only micromotors free of any unacceptable deterioration such as
corrosion and damages.
To maintain a level of hygienic security, at the end of each use and within a short time he must clean, disinfect and sterilize the
micromotor.
4.1. CLEANING
For cleaning follow the instructions below:
•Create a cleaning solution following the instructions for light organic debris.
Use a cleaner that is legally marketed and approved in your country (see paragraph PRODUCTS TO BE USED).
•Soak a lint-free cloth in the prepared detergent solution and tightly wrap the micromotor for 2 minutes.
Pay particular attention to have the cloth contact the front and rear connections of the device.
•Unwrap the device and dispose of the wipe.
•Soak a new lint-free cloth in the detergent solution and wipe the device by the following:
1 Wipe the external surface. Repeat this step using a fresh portion of the wipe if any visible soil remains.
2 Wipe the mating surface where the handpiece connects by wrapping the shaft housing with the wipe and rotating the device. Repeat this step with
a fresh portion of the wipe if visible soil remains.
3 Wipe the edge of the anterior mating surface. Press the wipe against the surface and rotate the device a few times both clockwise and
counterclockwise. Repeat this step with a fresh portion of the wipe if visible soil is observed.
Pay particular attention to the screw heads, recesses, and crevices of the device.
4 Wipe the edge of the rear mating surface of the device. Press the wipe against the surface of the device and rotate the device.
5 Repeat only step 4.1.
•Soak a new lint-free cloth in water (do not use saline solution) and wipe the micromotor by the following:
1 Wipe the external surface.
2 Wipe the mating surface between the handpiece and micromotor by wiping the stud and rotating the device a few times clockwise and
counterclockwise.
3 Wipe the edge of the anterior mating surface. Press the wipe against the surface to achieve good contact and rotate the device a few times both
clockwise and anti-clockwise.
4 Wipe the edge of the rear mating surface of the device. Press the wipe against the surface to have good contact and rotate the device a few
times.
•Using compressed air, dry the article.
•Using a new clean lint-free cloth, wipe the device to remove any excess of water or lubricant.
VISUAL INSPECTION.
At the end of the cleaning procedure, visually inspect the device under normal lighting conditions.
If any visible soil is present, repeat the cleaning procedure.

I-MMr • i-MMr L • i-MMs
8
EN
4.2. STERILISATION
Before each autoclave cycle lubricate the micromotor with a quick squirt of a lubricant for dental handpieces legally marketed and approved in your
country (see paragraph PRODUCTS TO BE USED).
Spray the lubrificant into the central hole in the back side of the micromotor only.
Reconnect the motor onto the instrument tubing and activate it so as to expel the excess lubricant.
Dry the outside of the micromotor with gauze or cotton wool before proceeding on with sterilisation.
Set following parameter according to Type of sterilizer:
Type of Sterilizer (EU market)
Exposure time at 121 °C
Exposure time at 134 °C
Drying time
Dynamic-air-removal (e.g., prevacuum)
20 min
4 min
40 min
Type of Sterilizer (USA market)
Exposure time at 250° F
Exposure time at 270° F
Drying time
Gravity displacement
30 min
/
40 min
Dynamic-air-removal (e.g., prevacuum)
/
4 min
40 min
•For sterilization use a wrap legally marketed and approved in your country (see paragraph PRODUCTS TO BE USED).
•Do not use ultrasound cleaners.
•Do not immerse the handpiece in disinfectant or sanitizing solution.
•Never sterilise the handpiece in a dry heat steriliser.
•Do not leave the handpiece in the autoclave, but remove it immediately after the cycle.
•Periodically check the autoclave according to the requirements of the manufacturer. Temperatures higher than the indicated
limit may damage the micromotor.
4.3. PRODUCTS TO BE USED
The Manufacturer recommends to use the following products:
CEFLA s.c. does not guarantee proper device operation if products other than those indicated are used.
EU market
DETERGENT FLUIDS
•IC100 (ALPRO MEDICAL GMBH) or other CE marked detergent
LUBRICATING OILS
•Daily Oil PLUS –500 ml (CEFLA s.c.)
STERILIZATION WRAPS
•Use only CE marked sterilization wraps
USA market
DETERGENT FLUIDS
•Sani ProZyme™- 3,8l (Crosstex)
LUBRICATING OILS
•Statcare™- 500ml (SciCan)
STERILIZATION WRAPS
•H200 - (Halyard Health)
All the products listed above must be stored and used according to the instructions of their Manufacturers.

I-MMr • i-MMr L • i-MMs
EN
9
5. MAINTENANCE
At least once a week lubricate the O-Ring seals found in the handpiece coupling with silicone lubricating grease. Wearing disposable gloves, apply a
film of grease on the fingertips and lubricate with the fingers.
If liquids that potentially leave incrustations are used to supply the spray, it is recommended that the circuit be flushed well with water at the end of each
use.
The manufacturer requires an annual check or overhaul of the device by authorised service assistance personnel.
5.1. REPAIRS AND OVERHAULS
Installation, overhaul, calibration and repair of the device must be performed exclusively by technical service assistance personnel authorised by the
manufacturer.
5.2. TROUBLESHOOTING
FAULTS
PROBABLE CAUSES
INTERVENTION SUGGESTED
The motor does not turn.
Motor blocked.
Request technical service assistance.
Blocked handpiece.
Lubricate the handpiece.
Send the handpiece to technical service
assistance.
Electronic control board broken.
Request technical service assistance.
Instrument tubing broken.
Request technical service assistance.
The motor runs but its speed is not adjustable.
Adjustment command broken.
Request technical service assistance.
Electronic control board malfunction.
Request technical service assistance.
The motor functions sporadically.
Instrument tubing defective.
Substitute the instrument tubing.
Bearings damaged.
Request technical service assistance.
The motor does not provide sufficient torque.
Torque set to its minimum.
Recover the proper adjustment.
Endodontics mode selected.
Recover the proper adjustment.
Power supply / motor malfunction.
Request technical service assistance.
The motor overheats.
Insufficient cooling air.
Request technical service assistance.
Handpiece with strong friction.
Lubricate the handpiece.
Send the handpiece to technical service
assistance.
The motor turns regularly.
The burr does not turn.
Handpiece not connected properly.
Slide the handpiece out and reinsert it
properly.
Handpiece broken.
Send the handpiece to technical service
assistance.
Defective transmission coupling.
Request technical service assistance.
The handpiece rotates on its coupling.
Blocked handpiece.
Send the handpiece to technical service
assistance.
The handpiece does not remain inserted on
the micro-motor.
Handpiece damaged.
Substitute the handpiece.
Motor hook ring damaged.
Request technical service assistance.
Water leakage at the handpiece.
O-ring seals worn or damaged.
Substitute the O-rings.
Water leakage at the instrument tubing
coupling.
O-ring seals worn or damaged.
Substitute the O-rings.

I-MMr • i-MMr L • i-MMs
10
EN
6. SCRAPPING
Scrapping the device must be carried out with respect for the laws in force for electrical and electronic equipment, according to the individual national
laws.
The materials used for the manufacture are not harmful for human beings or animals that come into contact with or are exposed to them.
7. GUARANTEE CONDITIONS
The manufacturer provides the end user with a 12-month warranty, starting from the date of installation, covering all operational faults, defects of
materials or manufacture.
In case of justified complaints the manufacturer or the Authorised Repair Centre will proceed with the repair or substitution of the product free of charge.
To be able to make use of the free repair or substitution, it is an indispensable condition that, shipped together with the device, there is a proof of user's
purchase document for the same device, which must include clearly legible references to the product, its serial number and date of purchase.
The guarantee expires when any damages and/or their consequences may be attributed to unsuitable procedures or modifications of the product
performed by third parties not authorised by the manufacturer: that is, if non-original spare parts or components are used.
The recognition of any other claims from any origin whatsoever, in particular any requests for indemnification for damages or interest are excluded.
The manufacturer may not be held responsible for damages, or injuries and their respective consequences derived:
•From excessive wear.
•From connection of the instrument not compliant with the CE regulations.
•From improper tampering or maintenance performed by unauthorised personnel.
•From use of non-original accessories or spare parts.
•From the lack of observance of the instructions for use, for assembly or maintenance, or from improper use of the product.
•From unusual chemical, electrical or electronic influxes.
•From faulty connections (air water, electricity).
The guarantee does not cover the conductors such as the flexible “fibre optics”, nor any other elements made of synthetic materials.
8. DISPOSAL
Dispose of the device and accessories in compliance with local regulations.
Abide by national recycling laws and the hospital's current recycling procedures in order to ensure proper disposal of the Micromotor device.
For more information on recycling, contact the Minister of the Environment or local authorities.
Do not dispose of this product as household waste. The directive does not apply to contaminated products.
9. MANUFACTURER
Manufactured by
CEFLA s.c.
Headquarters
Via Selice Provinciale, 23/a - 40026 Imola (BO) Italy
Tel. +39/0542 653111 Fax +39 0542 653344
Plant
Via Bicocca, 14/c - 40026 Imola (BO) Italy
Tel. +39 0542 653441 Fax. +39 0542 653555


www.cefla.com
This manual suits for next models
2
Table of contents
Popular Engine manuals by other brands

AvioMac
AvioMac 70cc Twin instruction manual

Briggs & Stratton
Briggs & Stratton Twin Cylinder L-Head Repair manual

Briggs & Stratton
Briggs & Stratton Vanguard 470000 Series Operating and maintenance instructions

Kohler
Kohler SDMO Technic 6500 A AVR UK C5 Instruction and maintenance manual

Eaton
Eaton X70 Repair manual

Danfoss
Danfoss Series 20 Service manual and repair instructions