Cella BW Series User manual

INSTRUCTION MANUAL NI-222E
Rev. 2 12/09
BOURDON TUBE PRESSURE SWITCHES
All data, statements and recommendations supplied with this manual are based on information believed by us to be
reliable. As the conditions of effective use are beyond our control, our products are sold under the condition that the
user himself evaluates such conditions before following our recommendations for the purpose or use foreseen by him.
WEATHERPROOF: SERIES BW EXPLOSIONPROOF: SERIES BAX
A
B
B
A
A pressure attachment
B cable entry A pressure attachment
B cable entry
WEIGHT 2 kg dimensions in mm WEIGHT 3,7 kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings.
CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the
indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS
ARE CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the
case.
CONTENTS:
1 - GENERAL NOTES
2 - OPERATING PRINCIPLE
3 - IDENTIFICATION PLATE AND MARKINGS
4 - SET POINT REGULATION
5 - SET POINT CALIBRATION
6 - MOUNTING AND CONNECTIONS
7 - INSTRUMENT PLUMBING
8 - PUTTING INTO OPERATION
9 - FUNCTIONAL VERIFICATION
10 - TROUBLESHOOTING
11 - STOPPING AND DISMOUNTING
12 - DEMOLITION
SAFETY INSTRUCTIONS FOR USE IN HAZARDOUS ATMOSPHERES.
RECOMMENDATIONS FOR PRESSURE SWITCH SAFE USE.
The present document is the property of ETTORE CELLA SPA and may not be reproduced in any form, nor used for
any purpose other than that for which it is supplied.

INSTRUCTION MANUAL NI-222E
Rev. 2 12/09
3 - IDENTIFICATION PLATE AND MARKINGS1 - GENERAL NOTES
The instrument is fitted with a metal plate bearing all its
functional characteristics and – in case of explosionproof
execution (Series BAX) – also the markings prescribed
by standard EN-60079-0 and EN-61241-0. Fig. 2 shows
the plate mounted on explosionproof instruments.
1.1 FOREWORD
The wrong choice of a series or a model, as well as the
incorrect installation, lead to malfunction and reduce
instrument life.
Failure to abide by the indications given in this manual
can cause damage to the instrument, the environment
and persons.
Fig. 1 - Explosionproof instruments plate
1 Notified body that issued the type certificate and number of
said certificate.
2 CE marking and identification number of the notified body
responsible for production surveillance.
3 Apparatus classification according to ATEX 94/9 CE direc-
tive.
4 Type of protection and ambient temperature limits of opera-
tion.
1.2 ALLOWED OVERRANGE
Pressure exceeding the working range can be occa-
sionally tolerated provided they remain within the limits
stated in the instrument features (vacuum or proof pres-
sure). Continuous pressures exceeding the working
range can be applied to the instrument provided they are
clearly stated in the instrument features. The current and
voltage values stated in the technical specifications and
ratings must not be exceeded. Transitory overranges
can have a destructive effect on the switch.
1.3 MECHANICAL VIBRATIONS
Can generally lead to the wearing of some parts of the
instrument or cause false actuation. It is therefore rec-
ommended that the instrument be installed in a place
where there are no vibrations. In cases where this is
impossible it is a good idea to take measures to lessen
the effects (elastic supports, installation with the peg of
the microswitch positioned at right angles to the vibration
plane, etc.).
4 - SET POINT REGULATION
Each microswitch is independent and can be regulated
by means of a screw (for adjustment) in such a way that
it is released when the pressure reaches (increasing or
decreasing) the desired value (set point).
1.4 TEMPERATURE
Due to the temperature of both the environment and the
process fluid, the temperature of the instrument could
exceed the allowed limits (normally from -20° to +85°C).
Therefore, in case it does, suitable measures (protection
against heat radiation, fluid separators, cooling coils,
heated lockers) must be taken.
The instrument is usually supplied with the switches set
at the setting range value nearest to zero (factory cali-
bration).
The instrument is supplied with an adhesive rating plate
showing the set point calibration value. With factory
calibration the values are not indicated on the ratings as
these are temporary and will be modified with the defini-
tive values.
2 - OPERATING PRINCIPLE
The pressure, acting on the sensitive element (Bourdon
tube), determines its elastic deformation which is used to
activate one or two electric microswitches regulated at
set point values. The microswitches are of the rapid
release type with automatic rearm. When the pressure
moves away from the set values, returning towards the
normal values, the switch is rearmed. The differential
gap (difference between the set point value and the reset
value) can be set or adjustable (letter R in the contact
codes).
Prior to installation the instrument must be calibrated
and the definitive calibration values written on the adhe-
sive rating plate using a suitable indelible ink pen.
If the instrument has been ordered with a specific cali-
bration, it is a good rule to check the calibration values
marked on the relevant adhesive label, prior to installa-
tion.
The position of the adjustment screws is given in figure
The effect that the direction of rotation of the adjustment
screws has is described on the adhesive plate.
Fig. 2 - Electrical connections and adjustment screws
One contact instruments Two contact instruments Microswitch electrical circuit
State of the contacts at at-
mospheric pressure
1 - Microswitch set point calibration screws
3 - Terminal block
4 - Electrical connection identification plate
1 - Microswitch 2 set point calibration screws
2 - Microswitch 1 set point calibration screws
3 - Terminal block
4 - Electrical connection identification plate
Designation of the contacts:
C - common
NA - Normally open
NC - Normally closed

INSTRUCTION MANUAL NI-222E
Rev. 2 12/09
5 - SET POINT CALIBRATION
In order to proceed with the calibration and the periodical
functional verification of the instrument a suitable cali-
bration circuit (fig. 5) and an adequate pressure source
is required.
5.1 PRELIMINARY OPERATIONS
Weatherproof pressure switches (BWX)
Remove the blocking device fixed to the side of the in-
strument case and the adjustment screw access plate
(Fig. 3). Remove the cover by rotating it in an anticlock-
wise direction.
Fig. 3 - Weatherproof pressure switch blocking device -
a - Plumbing wire
b - Plumbing
c - Blocking nut
d - Blocking bracket
e - Adjustment screw access
plate
Explosionproof pressure switches (BAX) (Fig. 4)
CAUTION: do not open the cover of explosionproof
pressure switches (Series BAX) when energized, in
explosive atmospheres.
Loosen the locking headless screw situated on the
cover using a 1,5 hexagonal key then unscrew the cover.
Remove the internal blocking device inserted on the
closure plugs and slide out the plugs.
Fig. 4 - Explosionproof pressure switch blocking device
5.2 CALIBRATION CIRCUIT AND OPERATIONS
Prepare the control circuit as indicated in Fig.5.
The warning lamps should be connected to contact 1 or
2 in the NO or NC position according to the required
contact action.
Connection of C and NO terminals
If the circuit is open at the working pressure, the
switch closes the circuit as the pressure increases
when the desired value is reached.
If the circuit is closed at the working pressure, the
switch opens the circuit as the pressure decreases
when the desired value is reached.
Connection of C and NC terminals
If the circuit is closed at the working pressure, the
switch opens the circuit as the pressure increases when
the desired value is reached.
If the circuit is open at the working pressure, the
switch closes the circuit as the pressure decreases
when the desired value is reached.
The test instrument should have a measurement range
approximately equal to or slightly wider than the pressure
switch range and should have an accuracy consistent
with the precision required to calibrate the set point.
The pressure switch must be mounted in the normal
installation position, i.e. with the pressure connection
pointing downwards.
Fig. 5 - Calibration circuit
PS - Pressure switch
CA - Test pressure gauge
V1 - Inlet valve
V2 - Inlet valve
P - Pressure source
Test fluid:
air for P 10 bar
water for P > 10 bar
Avoid forcing the elastic support of the microswitch by
hand or with tools. This could affect the instrument func-
tioning.
CAUTION: if the switch is of the kind with adjustable
dead band (letter R in the contact codes) before pro-
ceeding with the following operations it is necessary to
proceed with the adjustment of the dead band (see at-
tachment NI-705 for model BWXUR and NI-704 for
model BAXUR)
Increase the pressure in the circuit up to the desired set
point value for the first microswitch.
Use a wide bladed screwdriver, as indicated on the ad-
hesive plate, until the relative lamp turns on (or turns off).
- If the instrument is equipped with only one contact the
calibration is complete.
- If it is equipped with two contacts continue in the fol-
lowing manner.
Vary the pressure until the desired set point value for the
second microswitch is reached.
Act on the adjustment screw of the second contact.
Repeat former operations on the first contact, then op-
erations on the second contact, until the required set
point precision is obtained. This is necessary due to the
reciprocal influence which the microswitches have on the
sensitive element of the instrument.
Check the calibration values (varying the pressure in the
circuit accordingly) and record them on the adhesive
plate using a pen with indelible ink.
5.3 FINAL OPERATIONS
Disconnect the instrument from the calibration circuit
Weatherproof pressure switches (Series BWX)
Take the cover, ensure that the sealing gasket is cor-
rectly fitted into its seat, and insert the cover onto the
case, with the blocking gap positioned in correspon-
dence to the blocking bracket. Turn the cover clockwise
closing it tightly. Mount the adjustment screw access
plate, then the blocking device as in Fig. 3.
Explosionproof pressure switches (Series BAX).

INSTRUCTION MANUAL NI-222E
Rev. 2 12/09
Insert the closure plugs of the adjustment screw access
holes, block them using the internal device and if nec-
essary seal them with plumbing. Screw on the cover and
block it using the headless screw with which it is
equipped (Fig. 4).
Replace the supplied protection caps on the pressure
attachment and cable conduit.
IMPORTANT: the protection caps should only be defini-
tively removed during the connection steps (see §6).
6 - MOUNTING AND CONNECTIONS
6.1 MOUNTING
Surface mount the instrument by means of the holes
provided, or pipe mount using the appropriate bracket
(see Fig. 9). The chosen position must be such that
vibrations, the possibility of shocks or temperature
changes are within tolerable limits. The above also ap-
plies to direct mounting. With gas or vapour process
fluid, the instrument must be positioned higher than the
pipe inlet (see Fig. 8). With a liquid process fluid, the
instrument can be positioned higher or lower, indiffer-
ently (see Fig. 7 e 8). ). In this case, during set point
calibration the negative or positive head must be taken
into account (distance h in fig.7 and 8).
6.2 PRESSURE CONNECTIONS
For a correct installation it is necessary to:
Mount a shut-off valve with drain (root valve) on the
process tube to allow the instrument to be excluded and
the connection tubing to be drained. It is recommended
that said valve has a capstan-blocking device aimed at
preventing it being activated casually and without au-
thorisation.
Mount a service valve near the instrument to permit
possible functional verification on site. It is recom-
mended that the service valve is closed with a plug to
prevent the outlet of the process fluid caused by the
incorrect use of said valve.
Mount a three-piece joint onto the threaded attachment
of the instrument to permit the easy mounting or removal
of the instrument itself.
Carry out the connection using a flexible tube in such a
way that variations in the temperature of the tube itself
do not force the instrument attachment.
Ensure that all the pressure connections are airtight. It is
important that there are no leakage in the circuit.
Close the root valve and the relative drain device.
Close the service valve using a safety plug.
6.3 ELECTRICAL CONNECTIONS
It is recommended to carry out the electrical connections
according to the applicable standards. In case of explo-
sionproof instruments (Series BAX) see also the Stan-
dard EN-60079-14 and EN-61241-14. If the electrical
connection is carried out in a protected tube, it shall be
made so that condensate is prevented from entering
instrument enclosure.
The arrangement shown in Fig. 7 or 8 is therefore rec-
ommended.
CAUTION: fittings used for the electrical
connection of the pressure switch Series BAX
(explosionproof) shall be certified according to
ATEX, and shall guarantee instrument degree of
protection (IP65). To ensure anti-loosening of lock
coupling or cable gland, we suggest to seal the
thread of connection to the case with an anaerobic
sealant. For example, you can use an anaerobic
sealant type as Loctite ® 542.
Check that there is no power in the lines.
Remove the cover and carry out the cabling and connec-
tions to the terminal block (see Fig. 2).
Flexible cables with a maximum section of 1.2 mm2
(16AWG) are recommended using the pre-insulated fork
thimbles. Do not touch the adjustment screws and do
not bend the elastic microswitch supports in order to
prevent the instrument calibration being altered.
Ensure that no deposits or wire ends remain inside the
case.
As soon as connection steps are completed, mount the
cover on and make sure it is tight and blocked See Fig. 3
and 4.
6.4 SPECIAL NOTE FOR INSTALLATION OF
CATEGORY 2 G D PRESSURE SWITCHES
Explosionproof pressure switches (Series BAX)
can be installed on processes requiring apparatus
of group II category 1 in an ambient requiring
apparatus of group II category 2 (see Fig. 6).
Fig. 6 - Installation of Group II Cat. 2 GD instruments
Gas (G) Dust (D)
7 - INSTRUMENT PLUMBING
Weatherproof pressure switches (Series BWX)
The plumbing, aimed as a guarantee against possible
tampering of the calibration and electrical connections,
can be carried out using a flexible steel wire (c) inserted
into the holes in the screw (a) and the bracket (e) pro-
vided for this purpose (see Fig. 3).
Explosionproof pressure switches (Series BAX)
Plumbing is not necessary as the cover is blocked with a
headless screw, the closure plugs of the adjustment
screw access holes are blocked by means of the internal
blocking device and the instrument does not have to be
opened when installed.

INSTRUCTION MANUAL NI-222E
Rev. 2 12/09
8 - PUTTING INTO OPERATION If this is not the case it is necessary to stop operation,
dismount by means of the three-piece joints and carry
out the verification in a test room.
As the signal transmitted by the instrument is used in a
complex system, it is necessary that the means of put-
ting it into operation are established by those in charge
of the plant. CAUTION: do not open the cover of explosionproof
pressure switches (Series BAX) when energized, in
explosive atmospheres.
The instrument comes into operation as soon as the root
valve is opened. Any possible drainage of the connection
tubing can be carried out by removing the safety plug
and opening the service valve with the necessary
caution.
Verification consists in checking the calibration value
and possibly regulating the adjustment screw (see §4 ).
In case of explosionproof instruments (Series BAX),
inspections of the electrical installation are to be carried
out also according to customer procedures and at least
in accordance with Standard EN-60079-17 and EN-
61241-17.
In case of explosionproof instruments (Series BAX),
initial inspections are to be carried out according to cus-
tomer procedures and at least in accordance with Stan-
dard EN-60079-17 and EN-61241-17.
The explosionproof instruments (series BAX) installed in
explosive atmospheres for the combustible dust pres-
ence, must be periodically cleaned up externally in order
to avoid dust accumulating.
9 - FUNCTIONAL VERIFICATION
This will be carried out according to the Client’s control
procedures. Series BWX instruments can be verified on
the plant if installed as illustrated in Fig. 7 and 8.
The instruments Series BAX may be checked on site
only if apparatus suitable for explosives atmosphere are
used and provided that the electric line is not
energized.
10 - TROUBLESHOOTING
IMPORTANT NOTE: operations involving replacement of essential components must be carried out at our work-
shop, especially for instruments with explosionproof certificate; this is to guarantee the user the total and cor-
rect restoration of the product original characteristics.
MALFUNCTION PROBABLE CAUSE REMEDY
Set point shift
Permanent deformation of the sensitive
element due to fatigue or non-tolerated over-
ranges.
Variation of the elastic features of the sensi-
tive element due to its chemical corrosion.
Recalibrate or replace the sensitive element.
Recalibrate or replace the sensitive element
with another made of a suitable material. If
necessary apply a fluid separator.
Slow response
Clogged or obstructed connection line.
Root valve partially closed.
Too viscous fluid.
Check and clean line.
Open valve.
Provide instrument with suitable fluid separator.
No actuation or
undue actuation
Root valve closed.
Microswitch contacts damaged.
Loosened electrical joints.
Interrupted or short-circuited electrical line.
Open the valve
Replace the Microswitch.
Check all electrical joints.
Check the conditions of the electrical line.
Undue actuation Accidental shocks or excessive mechanical
vibrations. Modify the mounting.
Remove the instrument cover and disconnect the electri-
cal cables from the terminal block and earth screws.
11 - STOPPING AND DISMOUNTING
Before proceeding with these operations ensure that the
plant or machines have been put into the conditions
foreseen to allow these operations.
Remove the screws fixing the case to the panel (or pipe)
and remove the instrument, taking care to slide the elec-
trical conductors out from the case.
With reference to figures 7 and 8 Replace the instrument cover. Insulate and protect the
conductors remaining on the plant. Temporarily plug the
tubing (4).
Remove the power supply (signal) from the electrical
line.
Close the root valve (6) and open the drain.
Remove the plug (2), open the valve (3) and wait until
the process fluid has drained from the tubing through the
drain.
Do not dispose of the process fluid into the environment,
if this can cause pollution or damage to people.
Unscrew the three-piece joint (8).
CAUTION: do not open the cover of explosionproof
pressure switches (Series BAX) when energized, in
explosive atmospheres.
Unscrew the three-piece joint (10) (electrical cable tub-
ing).
In case of explosionproof instruments (Series BAX) it is
recommended to follow - at least – the standard EN-
60079-17 and EN-61241-17 for the withdrawal from
service of electrical apparatus.
12 - DEMOLITION
The instruments are mainly made of stainless steel and
aluminium and therefore, once the electrical parts have
been dismounted and the parts coming into contact with
fluids which could be harmful to people or the environ-
ment have been properly dealt with, they can be
scrapped.

INSTRUCTION MANUAL NI-222E
Rev. 2 12/09
ETTORE CELLA SPA Viale de Gasperi, 48 - Casella Postale (P.O. Box) 96 - I 20010 Bareggio (MILANO) ITALY
WEATHERPROOF EXPLOSIONPROOF
Fig. 7 - Example of connections - Fig. 8 - Example of connections -
Fig. 9 - Example of mounting -
Montaggio a parete
Montaggio su staffa per tubo 2" (palina verticale)
Montaggio su staffa per tubo 2" (palina orizzontale)
NOTE With gas or vapour
process fluid, the instrument
must be positioned higher than
the pipe inlet (see Fig. 8). With
a liquid process fluid, the in-
strument can be positioned
higher or lower, indifferently
(see Fig. 7 and 8). In this case,
during set point calibration the
negative or positive head
must be taken into account.
(Distance h in Fig.7 and 8).
LEGEND
1 - Fitting
2 - Drain plug
3 - Service valve
4 - Piping
5 - Three piece fitting
6 - Root valve with drain
7 - Process piping
8 - Three piece fitting
9 - “T” fitting
10 - Three piece fitting
11- Curve
12 - Blocking joint
13 - M6 screws (No. 4)
14 - Bracket for 2” pipe
15 - Vertical pipe
16 - Horizontal pipe
Surface mounting
2” pipe mounting (vertical pipe)
2” pipe mounting (horizontal pipe)
This manual suits for next models
1
Table of contents
Other Cella Switch manuals
Popular Switch manuals by other brands

Symbol
Symbol WS5000 Series installation guide

TRENDnet
TRENDnet TE-910plus Specifications

Heatmiser
Heatmiser neoHub user manual

Transition Networks
Transition Networks SM24-1000SFP-AH installation guide

Fujitsu
Fujitsu XG700 - Switch user guide

Weidmuller
Weidmuller IE-SW-BL05T-5TX Hardware installation guide