Checkpoint LDM100 User manual

User Manual
LDM100 Coating Thickness Gauge
Checkpoint Instruments
ul. Klecińska 125
54-413 Wrocław
Polska
kontakt@cp-instruments.pl
www.cp-instruments.pl

USER MANUAL
2
Contents
1Introduction.............................................................................................................. 3
2Safety Information................................................................................................... 3
3 Specification............................................................................................................ 4
4 System Description................................................................................................. 5
4.1 User interface and buttons.......................................................................................................... 5
4.2 Line starting ................................................................................................................................ 6
4.3 Measurement modes .................................................................................................................. 6
4.3.1 Single measurement....................................................................................................................... 6
4.3.2 Continuous measurement .............................................................................................................. 8
4.4 Measuring range ........................................................................................................................ 8
4.5 Battery indicator .......................................................................................................................... 8
4.6 Selection of the measuring unit ................................................................................................... 9
4.7 Language selection...................................................................................................................... 9
4.8 Alarm function ........................................................................................................................... 10
4.8.1 Switch on / off the alarm................................................................................................................ 10
4.9 Stored values.............................................................................................................................. 11
4.10 Accuracy measurement ............................................................................................................. 11
4.11 Calibration.................................................................................................................................. 12
4.11.1 Zero-point calibration.................................................................................................................... 12
4.11.2 Multi-point calibration...................................................................................................................... 12
4.12 Factory setting............................................................................................................................ 13
5 Disposal.................................................................................................................. 14
6 Contact ................................................................................................................... 14

USER MANUAL
3
1 Introduction
Thank you for purchasing a CP Instruments Coating Thickness Gauge.
This hand-held device for coating thickness measurement with a separate probe is a highly intelligent and
accurate measuring device with which layer thicknesses can be measured quickly and accurately on
almost all metal surfaces. The instrument not only displays the film thickness but also automatically
identifies the metal base (Fe stands for magnetic metals such as iron and steel, nFe denotes non-
magnetic metals such as aluminum, metal alloys and non-magnetic steel). With the device different
coatings can be measured:
non-magnetic coatings, ceramics, enamels, plastics, rubber coatings;
magnetic base materials such as iron or steel, non-ferrous metals such as nickel and
chromium;
corrosion protection coatings in the chemical and oil industries;
non-conductive paints;
plastic coatings and anodised coatings on non-magnetic conductive devices such as aircraft,
household appliances, vehicles, completely alloyed doors and windows, as well as other
aluminum products;
electrically conductive coating material.
Specifications and prices of all CP Instruments devices can be found on our website:
http://www.cp-instruments.pl. You will also find our contact details there. Our staff will gladly help you
with any questions.
2 Safety information
Please read this user manual carefully and completely before using the device for the first time.
The device may only be used by carefully trained personnel.
-This meter may only be used in the manner described in this manual. Use of the meter elsewhere may
result in dangerous situations.
-The device must not be exposed to extreme temperatures, direct sunlight, extreme humidity
or moisture.
-The device housing may only be opened by qualified personnel of the CP Instruments.
-The measuring device must never be placed on the user interface (eg on the keyboard side of a table).
-Never use the meter with wet hands.
-No technical modifications may be made to the device.
-The device should only be cleaned with a damp cloth. Please do not use abrasives or solvent-based
cleaning agents.
-The device may only be used with the accessories offered by CP-Instruments or equivalent.
-Furthermore, this meter must not be used if the environmental conditions (Temperature, humidity ...)
are not within the limits specified in the specification.
-The measuring device must not be used in an explosive atmosphere.
-Before each use, please check the meter by measuring a known size.
-The limits specified in the specification for the measured quantities may under no circumstances
be exceeded.
-Before starting the measurement, always check whether the correct measuring range has been set.
-If the safety instructions are ignored, it can damage the appliance and injury to the operator come.
This user manual is published by CP-Instruments without any warranty.
We expressly refer to our general warranty conditions, which can be found in our terms and conditions.
If you have any questions, please contact CP-Instruments.

USER MANUAL
4
3 Specification
Measuring range
Resolution
Accuracy
Dimensions
Weight
Ambient temperature
0 … 1250 µm / 0 … 50 mil
0,1 µm / 0,1 mil
± (2 % + 2 µm) / ± (2 % + 0,1 mil)
166 x 68 x 30 mm
180 g (with batteries)
-10 … +50 °C
Note: Perform a measurement only when the probe has reached the specified ambient temperature.
Please do not observe the first measurement results if you are not sure whether the sensor has
already reached the specified ambient temperature.

USER MANUAL
5
4 System description
4.1 User interface and buttons
Press the short to turn on the device and long to turn off
the device again. The meter automatically turns off after 180
seconds of inactivity.
(Menu/Back: Press the long to get to the
briefly to return to the previousmain menu. Press
(Statistics/Height): Press the short in measuring
mode, to the statistics data.
Max / Min. / Average values displayed by the device.
This screen will be displayed in the next picture.
(Data/Enter): Press the short in measuring mode
to display the recorded data. This screen will be displayed in
the next picture.
(Calibration/Backlight): In measurement mode, press
and hold to turn on the backlight. Then you can
or
adjust brighteness via
Press the to go to the page for calibration.
The screen will be displayed in the next picture.
menu level.

the screen
4.3 Measurement modes
The coating thickness gauge has two different measurement modes, the single measurement (SI) and the
continuous measurement (CO). Long press to enter the main menu. There, select the menu item
"Options" and move on to Mode, where you press .
USER MANUAL
6
mode to clear the recorded data.
(Delete/Down): Press the
4.2 Line starting
Press the to turn on the instrument. A beep sounds and "Model no. PCE-CT 27FN" is displayed on
4.3.1 Single measurement
SI mode is the measurement mode of the single measurement:
Press the probe vertically against the surface to be measured. The unit will beep once and display the
coating thickness value when the readings have been taken.
Note: Do not pull the probe over the surface to be measured.
The screen will be shown in the next picture:
1025.0 µm
measured value
Average 605.3 µm the average of all recorded data
Min. 49,6 µm
the smallest measured value
Max. 1025.0 µm
the highest measured value
in measurement

USER MANUAL
7
standard deviation ±20,5 µm
(2 % + 2 µm)
Fe
The measured material is a magnetic metal such as steel
NFe
The measured material is a non-magnetic metal such as
aluminum

USER MANUAL
8
4.3.2 Continuous measurement
The CO mode is the measuring mode of continuous measurement:
Press the probe vertically against the surface to be measured. The instrument will beep continuously and
display the coating thickness reading when the readings have been taken.
Note: During the measurement, do not lift the probe off the surface to be tested until the measurement is
complete. The screen will be displayed in the next picture:
4.4 Measuring range
The device measures layer thicknesses between 0 μm and 1250 μm. If the thickness is outside this range,
"---" will appear in the display. The screen will be shown in the next picture:
4.5 Battery indicator
The device is powered by four AAA batteries. If the icon is green, the battery is fully charged. After
a period of use, the battery icon will become shorter, indicating the current capacity of the battery.
starts flashing, the battery is almost dead.When the color changes from green to red and
Note: Please recharge the batteries or replace them when they are nearly empty. Otherwise, this could
affect your measurement accuracy.

USER MANUAL
9
4.6 Selection of the measuring units
Press the long to get to the main menu. Then move to the menu item "Unit" via the main menu for
to select between μm and mil.
the options menu. Press the
4.7 Lanugage selection
Press the long to get to the main menu. Then move to the menu item "Language" via the main
to select between English (EN), Chinese (CN) and German (DE).
menu for the options menu. Press the

4.8 Alarm funtion
To adjust the alarm range, long-click on to get to the main menu. Then move to the menu option
"Alarm limit" via the main menu for the options menu.
USER MANUAL
10
Press the or (long for fast scroll) to adjust the range.
4.8.1 Switch on/off the alarm
Press the long to get to the main menu. Then move to the menu item "Alarm" via the main menu for
the options menu. Press to turn the alarm on or off. When the alarm is on, the unit displays the
following picture on the display:
If the measurement is outside the alarm limits, the unit will beep three times to signal this.

USER MANUAL
11
4.9 Stored values
Press the to display the stored data and press or to display the last 50 stored values.
To clear the data, press Then move to the menu item "Data deletion" via the main menu for the
options menu. Then you can choose between "single deletion" (deletion of individual data) and
"total deletion" (deletion of all values).
4.10 Accuracy measurement
The user can control the accuracy of the device according to the given reference standards. The device is
supplied with standard plastic films. These can be used to test the measurement accuracy of the
instrument with the help of supplied calibration blocks. They can also be used to protect the sensor on
rough or hot surfaces. The measured value should be within the accuracy range specified in the operating
instructions. For example, if the accuracy is specified as ± 2% + 2 μm, the measurement result should be
between 47 and 53 μm when the Coating Thickness Gauge is used to measure a 50 μm film thickness
plastic film. Otherwise, the device must be calibrated.

USER MANUAL
12
4.11 Calibration
The instrument has been factory calibrated and a built-in self-test function performed before the
measurement. Therefore, it usually only needs to be tested before the measurement if the measured
value is 0 when measuring an uncoated metal. If this is not the case, a zero point calibration should be
performed.
4.11.1 Zero-point calibration
Press the and select "Zero Point Calibration" by briefly pressing and "000" will flash on the
screen. Then measure an uncoated substrate. When "0" is displayed on the screen, the zero calibration
is successful. After this process, the instrument can usually be reused for accurate coating thickness
measurements. Nevertheless, due to abnormal base materials or extreme environmental conditions,
additive errors may occur. If this is the case, you can use the standard plastic films to perform a multipoint
calibration.
4.11.2 Multipoint calibration
Use the instrument to place standard plastic films on a calibration block if the value found is outside the
accuracy limits of this manual. Lift the probe ("Pt1 ~ μm" flashes in the screen to "), press or
for fast scrolling) to set the value of the actual thickness. Repeat this with other plastic films. The use of
two of these standard plastic films is better for calibration than just one. For example, you can use a thin
and, alternatively, a thick foil. If the coatings to be measured are relatively close to the thickness of the
other, it is sufficient to use only a film of similar thickness for the calibration.

USER MANUAL
13
4.12 Factory settings
If zeroing with standard plastic films does not work, it may be necessary to reinitialize the device. First
press briefly select "Reset", then "Complete Reset" and confirm with
"Complete reset" will flash in the display and indicate together with a beep that the initialization is
completed. To ensure high measurement accuracy, it is recommended to perform a zero point calibration
after initialization. All previously saved values and settings are deleted after initialization and the device is
reset to factory settings.
Note: When measuring the coating thickness of alloys such as aluminum or stainless steel alloy, the
instrument must be calibrated as follows:
1. Turn on the coating thickness gauge, initialize it to factory settings, and turn it off.
2. Turn it back on and calibrate it to 6 standard slides (if accuracy was affected, keep the unit away
from any material during the entire calibration process).

USER MANUAL
14
5 Disposal
NOTE according to the battery regulation (BattV)
Batteries must not be disposed of with household waste: The end user is legally obliged to return. Used
batteries can be returned, among other things, to designated collection points or CP-Instruments.
Collection point according to BattV:
Checkpoint Instruments
ul. Klecińska 125 bud.BETA
54-413 Wrocław
For the implementation of the ElektroG (take back and disposal of WEEE) we take back our equipment.
They are either recycled by us or disposed of by a recycling company according to legal requirements.
6 Contact
For questions about our product range or the measuring device please contact CP-Instruments:
Postal:
Checkpoint Instruments
ul. Klecińska 125 bud. BETA
54-413 Wrocław
Phone:
+49 (71)796 57 54
WEEE-Reg.-Nr.DE69278128
Table of contents
Other Checkpoint Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PRECISION DIGITAL
PRECISION DIGITAL ProtEX-RTP PD6830 instruction manual

E-Mon
E-Mon Class 3200 installation instructions

dymax
dymax ACCU-CAL 150 user guide

NIVELCO
NIVELCO PiloTREK W-100 User and programming manual

Endress+Hauser
Endress+Hauser Deltabar S FMD77 operating instructions

LG
LG RKS-7000Q owner's manual