Chino TP-L0260EN User manual

Please store this instruction
manual in a designated place for
futur
e reference.
INST.No.TPL-02-5A
TP
TPTP
TP
-
L
LL
L0260
02600260
0260EN
ENEN
EN
Instruction Manual
CHINO
INSTRUCTION
S
Compact
Thermal Image Sensor


I
Custom Net ork Cable
(With a crossing cable and an RJ-45 terminal) 2.5m: 1 piece
* For direct connection to PC
(Refer to Page 6 for details.)
Custom Po er/Alarm Output Cable
(With O-tip terminals) 2.5m: 1 piece
Introduction
Thank you for purchasing a compact thermal image sensor. In order to use this product
safely, please read this instruction manual completely before use and confirm the correct
handling and instructions. In addition, please retain this instruction manual for future
reference.
Before using
Please confirm the contents of packing. If something is missing, please contact to your nearest distributor.
.
Thermal image sensor
Carl plugs for mounting
the universal head
into concrete: 3 pieces
Connector cap: 1 piece
Lens cap: 1 piece
Universal head: 1 piece
Scre s for the universal
head: 3 pieces
Mounting scre : 1 piece
This is attached to the sensor
Scre size
1/4
-
20UNC
Scre size
1/4-20UNC
CD-ROM: 1 piece
Application soft are
Instruction manuals (For the
sensor and the application
soft are
Quick manual: 1 sheet

II
This product corresponds to Items 10 (2) and 10 (4) of Export allo able items listed in
Appendix No. 1 of the Export Trade Control Ordinance in Japan.
When this product is exported or brought out of Japan, the permission by the Japanese
government is required under this control ordinance.
About this instruction manual
Under absolutely no circumstances may the contents of this instruction manual, in part or
in hole, be transcribed or copied ithout permission.
The contents in this instruction manual are subject to change ithout notice in future.
The figures in this instruction manual may be emphasized, simplified or omitted.
Every effort has been made to ensure that the details of this manual are accurate.
Ho ever, should any errors be found or important information be omitted, please inform
your nearest distributor.
“Microsoft” and “Windo s” are either trademarks or registered trademarks of Microsoft
Corporation, USA.
The company names and brand names used in this manual are the trademarks or
registered trademarks of respective companies.
About exemption from responsibility
Unless other ise specified in the guarantee clauses, e do not offer any guarantee about
this product.
We shall not be liable to a customer or a third party for any damages or indirect damages
by using this product or by unpredictable defects of the product.

III
Safety precautions
The safety precautions sho n in this manual indicate the important contents about safety.
Please be sure to understand and follo these precautions.
In this manual, in order to use this product safely, the precautions are described ith the
follo ing indications and marks.
This indicates a potentially hazardous
situation that, if not avoided, ill result in
death or serious injury.
This indicates a potentially hazardous
situation that, if not avoided, may result in
minor or moderate injury or cause property
damage.
[Safety precautions]
Do not operate this product in a place here flammable gas or explosive gas exists. It is
extreely dangerous to use this product under such environment.
In order to prevent an electric shock, make sure that the po er source is turned off before
connecting any iring.
● Stop using this product immediately, turn off the power source and contact to
your nearest distributor if it is broken or there is smoke/abnormal odor from this
product. Otherwise, it may cause fire.
● Do not repair or modify this product. If it is modified, operation of this product ill not
be guaranteed. It may also cause fire/electric shock.
● Avoid the use of this product in the follo ing places.
A place here ambient temperature varies idely
A place here humidity is high
A place here rain/ ater is splashed
A place here there are dusts like sand or dirt
A place here it is subjected to exposed in scorching heat or direct sunlight
A place here it s subjected to radiation
A place here strong electric circuit exists beside this product
A place here there is any inductive interference
A place here there is mechanical vibration/shock
● If this product is ired ith rong polarity, it may cause trouble or fire.
● Do not aim the sensor to a high temperature object like sun, or the imaging element
may be damaged and it may cause trouble on this product.
● In order to use this product safely, refer to "requirements" and handling methods
mentioned in this manual or it may cause damage or malfunction to this product or
cause damage also to other equipments.
● Optical components have special coating. Be careful that these parts become easily
dirty hen they are de -condensed. In addition, the infrared transmission is
deteriorated under the de condensation and a clear image is not provided.
Warning
Caution
Warning
Caution

IV
When you use accessories connected to this product, follo the instructions relating
toregarding the safety in this manual. If you do not follo these instructions, safety of this
product and the system ill not be guaranteed.
Your extra consideration is required hen this product is used in an application that may
subject to life threatening to human or property damage. Pay extra attention to safety
measures, avoid usage under extreme environment and build a fail-proof system.
Precautions for use
This product is a precision instrument. Do not drop it or make a strong impact on it.
Do not use it in ater.
Keep this product a ay from equipments generating strong static electricity and
electromagnetic aves including TVs, micro ave ovens and ireless applications, or it may
cause malfunction or trouble on this product.
Keep this product a ay from equipment generating strong high frequency or surge voltage,
or it may cause malfunction or trouble on this product.
Use neutral detergent for cleaning of this product and ipe it off lightly by rubbing ith the
cloth squeezed firmly. Do not use chemicals or bleaches including benzene, thinner and
alcohol.
Do not use or place this product in a place subjected to temperature of lo er than -10°C or
higher than 50°C, de -condensation, or humidity of lo er than 10%RH or higher than
80%RH.
Do not use or place this product in a place subjected to direct sunshine, dusty, high
temperature/high humidity or corrosive atmosphere.
When any service including repairs is required, contact to your nearest distributor.
Do not apply too much force on the custom po er supply and Net ork cables. It may cause
damage of the connection hich may lead ater immersion into the product, and may cause
trouble on this product.

【準備・設定編】 Contents
■ Introduction ......................................................... I
■ Safety precautions .............................................. III
■ Precautions for use ............................................ IV
■ Outline of product
1. Configuration........................................... 1
2. Features ............................................... 1
3. Functions ............................................. 1
4. Model/specification.................................. 1
5. Basic configuration ……………………… 2
■ Names and functions ........................................... 4
■ Connection .......................................................... 5
■ Installation ........................................................... 7
■ Operation ............................................................. 8
■ Monitor lamp ........................................................ 8
■ Operation mode...................................................... 8
1. Monitor mode........................................... 8
2. Capture mode ...................................... 8
■ Alarm output ........................................................ 9
■ About reset s itch ................................................. 11
■ Specifications ...................................................... 12
■
Precautions (For maintaining measurement accuracy)
......... 14
■Troubleshooting ………........................................ 14
■ External dimensions
1. Thermal image sensor............................ 15
2. Universal head........................................ 15
3. Protective case....................................... 16

1
Outline of product
1. Configuration
The system consists of the thermal image sensor utilizing a thermopile array detector and the
application soft are for displaying thermal images converted from the temperature data from the
sensor..
The application soft are allo s to connect up to 4 thermal image sensors and display thermal
image from these sensors.
2. Features
Built-in thermopile array detector ith approximately 2256 pixels. Each pixel can measure
temperature.
Lo cost, compact size and light eight (150g) fixed-mount type thermal image sensor
Robust and dust/splash-proof construction IP-65
Data transmission to a PC via Ethernet (LAN), Thermal image display by the application soft are
2 built-in circuits of non-voltage contact outputs. The sensor can be used as a fixed-mount type
alarm unit.
(When an alarm is activated, 1 image can be stored in the sensor.)
3. Functions
Alarm can be set from the application soft are or custom-made commands. The sensor provides
non-voltage contact outputs hen the alarm condition is met.
Monitor mode ( hen the supplied application soft are is used)
The temperature data of the entire thermopile array can be obtained continuously from the PC.
Capture mode ( hen the supplied application soft are is not used)
The temperature data in a single ro of the thermopile array can be obtained at a time by a
command from a master unit (e.g. PC or PLC).
When the capture mode is used, a series of communication command is required.
For the communication commands, contact to your nearest distributor.
4. Model/specification
TP-L
□□ □□ □ □
Pixels 02: 2256
Vie angle 60: 60°
Output E: Alarm + LAN
Specification N: Standard specifications

DC po er supply
(DC12V 1.5A)
5. Basic configuration
5-1. Connection ith 1 Thermal Image Sensor to PC (Monitor mode)
5-2. Connection of multiple Thermal Image Sensors (Example: 2 units)
Supplied custom Net ork cable (Crossing cable)
Custom po er/alarm output cable
Po er voltage: 12/24VDC
(Min. 9 ~ Max. 30VDC)
(Refer to Page 12 for details.)
PC
Alarm setup or
monitor ith the
PC
Alarm contact
outputs from the
thermal image
sensor
LAN
Contact output
* If a HUB ithout automatic crossing/straight converting function is used, a straight
cable is needed. Connect a crossing/straight converting connector bet een the custom
Net ork cable and a HUB. (Refer to Page 3.)
Alarm setup or
monitor ith the
PC
Alarm contact outputs
from the thermal
image sensor
Alarm contact outputs
from the thermal
image sensor
Thermal image sensor
PC
Thermal image sensor
DC po er supply
(DC12V 1.5A)
DC po er supply
(DC12V 1.5A)
Contact output
Contact output
LAN LAN
HUB (*)
Monitor mode ( hen the supplied application soft are is used)
The temperature data of entire thermopile array can be obtained from the PC.
Thermal image sensor

3
DC po er supply
(DC12V 1.5A)
Connection example of HUB ( ithout automatic crossing/straight converting function)
5-3. Stand alone usage of the thermal image sensor
5-4. Connection to PLC (Capture mode)
When the capture mode is used, a series of communication commands are required.
For the communication commands, contact to your nearest distributor.
Custom Net ork cable (Crossing cable)
HUB
Crossing/straight converting connector
LAN cable (Straight cable)
PC
Alarm contact output
from the thermal
image sensor
For image confirmation and alarm settings only
LAN
Contact output
Contact output
Other alarming devices
PLC
Capture mode (
hen the supplied CHINO application soft are is not used)
The temperature data in a single ro of the thermopile array can be obtained
at a time
by a command from a master unit (
e.g. PC or
PLC
).
LAN
DC po er supply
(DC12V 1.5A)

4
(7)
Names and functions
(1) Monitor lamp
(2) Lens
(4) (5)
Po er/alarm output LAN connector
connector
(3) Mounting scre (6) Reset s itch
(Behind the scre )
No
Name Function
(1)
Monitor lamp The state of alarm and the condition of the sensor is indicated
ith colors of blue, red and purple. (Refer to Page 8.)
(2)
Lens For forming image of radiation energy from a measurement
object on the detecting element
(3)
Mounting scre For mounting the sensor on a tripod (1/4-20UNC)
(4)
Po er/alarm output
connector
For connecting custom po er/alarm output cable
(5)
I/F connector For connection of custom Net ork cable
(6)
Reset s itch To initialize the sensor to factory default (The s itch is located
behind the mounting scre .) (Refer to Page 11 for details.)
(7)
Mfg nameplate Serial No., MAC address, etc. are sho n.

5
[
] mark
Connection
● Connect the custom po er/alarm output cable and the custom Net ork cable to this product.
Connect them to align the marks of [] back of this product and [Arro ] on the cable connectors.
To avoid misconnections, the custom Net ork cable is designed not to connect to the
po er/alarm output connector of this product. Similarly, the custom po er/alarm output cable
cannot be connected to the LAN connector of this product.
The connections are a quick-disconnect locking type. Insert the plug until it clicks.
To unplug, hold the sliding part (that has the [Arro ] marking on) and pull it out ard.
If the “A” part is pulled, and the plug may be damaged.
.
● Plug the connector of the custom Net ork cable to the Ethernet port of a master unit (PC, etc.)
and connect the O-tip terminals of the custom po er/alarm output cable to the terminal block
of
12VDC Po er Supply.
Custom Net ork
cable LAN connector
Custom po er/alarm output
cable ring terminals
Insulate the O-tip alarm output
terminals if not used.
[Arro ] mark
LAN port
Terminal block
Plug
A

6
[Alarm 1, / AC current] [Alarm 1, / DC current]
(AC)
(Relay)
(DC)
(Relay)
Connector cap
[Details of custom po er/alarm output cable] O-tip terminal 0.5-3.5
Po er +
Po er -
Alarm 1 +
Alarm 1 -
Alarm 2 +
Alarm 2 -
[Example of contact output circuit]
Use the contact output ith a protection device (e.g. fuse) for preventing malfunction.
Max. output load voltage 400V (Peak AC)
Max. continuous load current 0.1A (Peak AC)
Peak load current 240mA (100ms/1 shot)
Output capacity (Per 1 contact)
Max. output loss 300mW
● If the custom Net ork cable is not used (When using the thermal image sensor stand alone)
Cover the supplied connector cap to the connector marked ith [LAN] back of this product.
If you don’t cover the connector ith the cap, the aterproofing performance cannot be
satisfied. (It may cause ater immersion.)
Red
P
+
Black
AL1
+
Orange
AL1
-
Pink
AL
2
+
Bright
green
AL
2
-
Blue
P
-
AL1 +
AL1 -
AL1 +
AL1 -
Po er
Po er

7
Installation
By referring the figure belo , install the thermal image sensor at the measuring distance that a
measuring object can be imaged at the center of screen as large as possible.
<Relation bet een a measuring surface and a measuring distance>
D (Vie idth) = 1.4 × L (Measuring distance)
For accurate measurements, 8 x 8 pixels or more for a target is recomended.
1.4m
Do not install the sensor here subjected to vibration/shock or inductive ambient.
1m 3m 5m
29mm
4.2m
88mm
146mm
7.0m
: Vie per pixel

8
Operation
● Install the supplied application soft are to your personal computer.
(For the details, refer to the instruction manual for the application soft are.)
● For more accurate measurements, ait 30 minutes after po er-on until this product is stabilized.
● Use this product ith noiseless and stable po er source.
Monitor lamp
The state of product is indicated by 3 colors on the LED indicator; blue, purple and red.
Item Color Specification
Mode is not set (Factory Default) Blue Lit once for 3 seconds
Alarm values are not set Blue Lit once for 3 seconds
Alarm being monitored Blue Lit continuously
Set to Captured Mode Purple
Lit continuously
Reboot is required after all settings have
completed Purple
Lit for 3 seconds t ice
Alarm is activated Red Lit continuously
Abnormal internal memory Red Lit once for 3 seconds
* Refer to Page 14 Troubleshooting.
Abnormal
State Abnormal internal temperature Red Lit once for 3 seconds
* Refer to Page 14 Troubleshooting.
Operation mode
1. Monitor mode: When CHINO application soft are is used
The temperature data is sent continuously to a master unit (e.g. PC) by a command from the
soft are.(The frame rate of the thermal image on the soft are is 3fps ithout any Alarm activated.)
The alarm set on the sensor can provide a non-voltage contact output hile sending the
temperature data.(The frame rate goes do n to 1fps or less ith the alarm set on the sensor.)
2. Capture mode: Without CHINO application soft are
The temperature data in a single ro of the thermopile array can be obtained at a time by a
command from a master unit.
When Capture mode is used, a series of communication commands are required.
For the communication commands, contact to your nearest distributor.

9
Alarm output
1. Specifications
Item Specification
Alarm area 1 area
Alarm set value 2 values in a specified area
Setting resolution 1 degree (°C/ºF)
Alarm activation condition When consecutive 4 frames have satisfied the alarm
condition
Alarm reset condition When the measured temperature returns from the
alarm set value in 4 images continuously
Output Control Auto Alarm Reset/Manual Alarm Reset
Resetting condition of alarm
output manual reset
Rebooting of the thermal image sensor
Command from a PC
Alarm Output Non-voltage contact output
Number of outputs 2 points [Alarm name: AL1 (alarm 1), AL2 (alarm 2)]
Output logic Output logic selectable (N.O/N.C)
Alarm condition Selectable (high-limit, lo -limit)
Alarm condition hen po er is off
Opened
2. Image storage
The image data can be stored hen alarm is activated.
Item Content
Number of storage
1 image
An image can be stored by AL1 or AL2 (to be
specified).
Storage medium Built-in EEPROM
Storage data Temperature data ( ithout time stamp)
Storage timing
Every time hen the alarm is activated/ hen the first
alarm is activated only
Selectable by a PC (Refer to Page 10.)
Reading of stored data Can be retrieved from a PC
Erasing of stored data Automatic erasing or erasing by a command ith the
alarm setting from a PC

10
警報有
警報有警報有
警報有 警報無
警報無警報無
警報無
警報状態
警報状態警報状態
警報状態
警報出力
警報出力警報出力
警報出力
警報出力
警報出力警報出力
警報出力
画像保存
画像保存画像保存
画像保存
画像保存
画像保存画像保存
画像保存
警報状態
警報状態警報状態
警報状態が
がが
が連続
連続連続
連続で
でで
で発生
発生発生
発生した
したした
した場合
場合場合
場合は
はは
は、
、、
、発
発発
発
生時
生時生時
生時の
のの
の最初
最初最初
最初の
のの
の画像
画像画像
画像を
をを
を保存
保存保存
保存する
するする
する
画像上書
画像上書画像上書
画像上書き
きき
き
警報有
警報有警報有
警報有
熱画像
熱画像熱画像
熱画像センサ
センサセンサ
センサの
のの
の電源再投入
電源再投入電源再投入
電源再投入又
又又
又は
はは
は上
上上
上
位
位位
位機器
機器機器
機器からの
からのからの
からのコマンド
コマンドコマンド
コマンドまで
までまで
まで受信
受信受信
受信まで
までまで
まで
保持
保持保持
保持する
するする
する
自動復帰
自動復帰自動復帰
自動復帰
保持
保持保持
保持
● Alarm output and storing image
熱画像
熱画像熱画像
熱画像センサ
センサセンサ
センサの
のの
の電源再投入又
電源再投入又電源再投入又
電源再投入又は
はは
は
アプリケーションソフト
アプリケーションソフトアプリケーションソフト
アプリケーションソフトからの
からのからの
からの
警報解除
警報解除警報解除
警報解除まで
までまで
まで保持
保持保持
保持する
するする
する
Alarm state
Auto Reset
Contact output
Image saving
ON
OFF
ON
Image is over ritten.
Manual Reset
Contact output
Image saving
The image is saved only at the time
of hen alarming state turns from
Off to On.
Contact output state ill be held until
the sensor restarts or you reset the
alarm by clicking [Alarm Reset] button
in Alarm Setup Windo .

11
Reset
s itch
Reset s itch
(behind the scre )
Remove the mounting
scre .
■ About reset s itch
In case of an error, etc. at the LAN environment settings, the set values can be reset to the default
values as shipped from factory by removing the mounting scre and then pushing the reset s itch.
Turn the po er on hile pushing the reset s itch for 1 second continuously. Resetting to the
default values becomes effective (The violet monitor lamp flashes 2 times.) By re-po er-on,
the set values are reset to the default values at the shipment from factory. Usually do not push
the reset s itch.
[Factory Default settings]
Items Specifications
IP address 192.168.1.254
Subnet mask 255.255.255.0
Gate ay address 0.0.0.0 (Invalidation)
Sensor number 1
UDP port address 50828 (fixed) *1
Emissivity 1.00 (all pixels)
Alarm setting Alarm invalid
*1 If you use UDP port on your PC and the UDP port address that you are using is
overlapping ith this address, change the UDP address that you are using.
[MEMO]

1
■ Specifications
Items Specifications
Detecting element Thermopile array ith 48 x 47 pixels
Measurement avelength Center avelength 10µm
Measurement vie angle 60° x 60°
Temperature measurement
range -20°C to 300°C
Temperature resolution 0.5°C (at 100°C black body)
Accuracy ratings
±2% of measured value or ±3
°C, hichever is greater
(Ambient temperature 25 ± 2
°C)
Focus Fixed focus
Flame rate 3fps (1fps or slo er if alarm function is enabled)
Emissivity correction 0.10 ~ 1.00 (0.01 increment)
Radius resolution 21.8mrad
Working temperature range -10°C to 50°C
Working humidity range 10 to 80%RH (no de condensation)
Storage temperature range -20°C to 60°C
Housing
IP65 ( ith custom
cable connection or connector cap, and
mounting scre mounted)
External dimensions W62 x H62 x D59 (mounting scre not included)
Weight About 150g (sensor unit)
Material
Polycarbonate Resin Color: Black
Conforming standard Influence ithin 3% of measurement range under the
environment of CE (EN61326-1/Class B)
mounting scre size 1/4-20UNC
Po er supply 12 to 24VDC (Min. 9 to Max. 30VDC)
Current consumption *1 Max. 2.5VA at 12VDC
Inrush current *1
Max. 1A at 12VDC
(When po er supply S8VM-01512 made in Omron is
used)
Connection Custom-connector connection
Connector name [DC IN/IN/OUT]
*1 The inrush current should be considered for the selection of po er source.
Thermal image specifications
General specifications
Po er specifications

13
Items Specifications
Number of contact output 2 Outputs
Kind of contact No-voltage contact output
Cable terminal names AL1, AL2
Max. output load voltage 400V (Peak AC)
Max. continuous load current 0.1A (Peak AC)
Peak load current 240mA (100ms/1 shot)
Contact Rating (per
contact)
Max. output loss 300mW
Connection Custom-connector connection
Connector name [DC IN/IN/OUT]
Interface standard 10BASE-T/100BASE-TX
Data/Protocol
Data communication by UDP
Measured temperature data → To a master unit (PC, etc.)
Setting data ← From a master unit (PC, etc.)
Setting data IP address/alarm setting information, etc.
DHCP
Not supported
Thermal image sensor IP
address
192.168.1.254 (default at shipment)
Thermal image sensor
subnet mask 255.255.255.0
Default gate ay 0 or FFFF (Not set)
Connection Custom-connector connection
Connector name [I/F]
■ Accessory (sold separately)
Protective case: Model: TP-ZCC1 (Refer to Extrernal Dimension.)
■ Minimum requirements for the CHINO application soft are
Compatible PC: DOS/V computers ith PC/AT compatible
* The LAN port is required.
* Display resolution: 800 x 600 pixels or more recommended
* .NET Frame ork 2.0 or later has been installed and runs correctly.
Compatible OS: Windo s 2000 (SP4 or later)/XP/Vista
* Windo s XP or later recommended
* .NET Frame ork 2.0 or later required
Contact output specifications
Ethernet specifications
This manual suits for next models
1