manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Clements
  6. •
  7. Measuring Instrument
  8. •
  9. Clements SUC89851 User manual

Clements SUC89851 User manual

INSTRUCTION
MANUAL
INSTRUCTION MANUAL
VENTURI NOZZLE
FOR HIGH FLOW / HIGH VACUUM
PUBLICATION CODE: MO093
EDITION: n° 1
DATE: February 2021
Sponsor:
ICU Medical Australia Pty Ltd
Unit U, 10 – 16 South St, Rydalmere
NSW 2116, Australia
Tel: +61 2 9466 5300 -www.clements.net.au
FLOW METER S.p.A.
Via del Lino, 6 – 24040 LEVATE (Bergamo) – Italy
Tel. +39/035/594047 – Fax +39/035/594821
e-mail: info@flowmeter.it - www.flowmeter.it
0546
Contents
Important ................................................. 2
Application............................................... 3
Controls and connections ........................ 3
Installation and use..................................3
Cleaning and disposal .............................4
Technical features ...................................4
Important
CAUTION: INDICATION OF DANGER
Attention: Important indication
General information
•
Read this manual carefully before using
the venturi nozzle
for high flow/ high
vacuum.
•
AFTER UNPACKING, CONNECT THE DEVICE
AS DESCRIBED IN THE SECTION
“INSTALLATION AND USE”.
•
BEFORE EACH USE CARR
Y OUT THE
OPERATIONS DESCRIBED
IN THE SECTION
“INSTALLATION AND USE”.
•
INSTALLATIONS THAT ARE NOT ENVISAGED
BY THIS MANUAL MAY REDUCE THE SAFETY
LEVEL OF THE APPLIANCE.
•
BEFORE EACH USE, ENS
URE THAT THE
DIRECT SUCTION NOZZLE FOR HIGH FLOW /
HIGH VACUUM, IS DIRECTLY OR INDIRECTLY
CONNECTED TO A COLLE
CTION VESSEL
FITTED WITH AN OVERF
LOW VALVE AND
ANTIBACTERIAL FILTER
, AND THAT THE
CONNECTOR HOSES USED CONFORM WITH
THE STANDARD EN ISO 10079-3.
•
The company will not
accept any
responsibility if the instructions in this
manual are not observed, if original spare
parts and/or authorized technicians are
not used.
Connections
•
CONNECT THE UNIT AND CHECK THE SEALS
BETWEEN THE COMPONENTS AS DESCRIBED
IN “START-UP PROCEDURE” SECTION.
FAILURE TO MAKE THES
E CONTROLS MAY
COMPROMISE THE SAFET
Y AND
FUNCTIONING OF THE DEVICE.
•
OILS AND GREASES MUS
T NOT COME IN
CONTACT WITH THE EQUIPMENT.
•
The vacuum supply connector is positioned
on the back
part of the device and may be
identified by referring to the drawing on page
3 of this manual (pos. 4).
Operation
•
THE VENTURI NOZZLE FOR HIGH FLOW / HIGH
VACUUM, MUST BE USED CAREFULLY AND ONLY
BY PERSONNEL WHO ARE
FAMILIAR WITH THE
CONSEQUENCES OF THIS THERAPY.
•
Disconnect the vacuum supply when the
unit is not in use.
Service
•
All the modifications and repairs must only
be performed by personnel authorized by
the Manufacturer
or by hospital technicians
approved by the same company.
•
Original spare
parts must always be used
for the maintenance operations.
•
Check the direct suction nozzle every day
or in accordance with hospital routine.
•
For periodical updating reasons, the device
configuration can be subjected to changes.
Therefore, the Manufacturer
guarantees
spare parts to be available for at least 5
years from the manufacturing date.
•
Any modifications to the device must be
approved by the Manufacturer
, and carried
out in accordance with the procedures
prescribed.
The device has been
designed and
manufactured to satisfy the safety
requirements of the following standards:
EN ISO 10079-3
Page 2
Application
Venturi nozzles for high flow / high vacuum are used where simple On-Off suction control is required.
The Nozzles have a simple and robust
construction and are designed for applications where direct
suction is instantly required. The dual-scale vacuum gauge for controlling the line depression is accurate
and easy to read. The non-return valve protects the suction system from blowbacks. Venturi nozzles for
high flow / high vacuum are suitable for connection to most of terminal units available on the market.
Controls and connections
LEGEND
1 - Silicone protection sleeve (optional)
2 – Controlling vacuum gauge (optional)
3 – Non-return ball valve
4 – Specific gas quick coupling for suction supply
(VAC-IN)
5 – Chrome-plated metal body
6 – Sintered bronze filter
7 – Vacuum supply hose connector (VAC-OUT)
Installation and use
Check the functioning of the unit every day or in accordance
with the hospital routine. The connection procedures for the
venturi nozzle is described below.
•Connect the specific gas quick coupling for suction
supply VAC IN (4), to the rapid supply outlet of the
centralised hospital plant;
•Connect a suitable flexible hose to the vacuum supply
hose connector VAC OUT (7)
(e.g. silicone hose
autoclavable at 134 °C for 18 minutes), pressing onto
the end;
•Connect a suitable type of collection
container for
aspirated fluid to the other end of the hose.
The procedure below must be followed to make the appliance
operational:
•Check the suction on the control vacuum gauge (2), after
the block of the hose connector used for the suction. The
system is then ready for use.
After use, the procedure below must be followed:
•Disconnect the vacuum supply (4) from the main supply
outlet of the centralised hospital plant;
•Remove the collection container by disconnecting it from
the suction nozzle and detaching the PATIENT circuit.
Page 3
Installation and use
WARNING
FOR THE CORRECT USE OF THE COLLECTION CONTAINER, READ THE INSTRUCTIONS IN THE USER
MANUAL SUPPLIED WITH THE DEVICE.
THE LIQUIDS COLLECTED MUST ONLY BE DRAINED OUT IN AREAS ASSIGNED TO THE DISPOSAL OF
HOSPITAL WASTE AND/OR STRICTLY FOLLOWING THE INSTRUCTIONS OF THE AUTHORITIES FOR THE
TREATMENT OF THESE PRODUCTS.
Cleaning and disposal
•
Clean the device every day or in accordance with the hospital routine
•
Clean or replace the sintered bronze filter (6) if necessary.
•
Carefully clean all the external surfaces of the device using a soft damp cotton cloth rinsed in neutral detergent
diluted 10% in water.
CAUTION
DO NOT USE SOLVENTS OR ABRASIVE PRODUCTS FOR CLEANING: THESE CAN SERIOUSLY DAMAGE
THE SURFACES OF THE EQUIPMENT AND THE PLASTIC PARTS.
•
DO NOT IMMERSE THE UNIT IN DISINFECTANT;
•
DO NOT PLACE THE UNIT IN AN AUTOCLAVE;
•
DO NOT USE INFLAMMABLE PRODUCTS.
IF THE EQUIPMENT IS ACCIDENTALLY CONTAMINATED BY SUCTIONED LIQUIDS, IT MUST BE ISOLATED
AND THE NECESSARY CLEANING AND DISINFECTION OPERATIONS MUST BE
CARRIED OUT BY
QUALIFIED PERSONNEL.
IF NECESSARY, STERILISE THE VENTURI NOZZLE WITH A VALIDATED
ETHYLENE OXIDE
STERILIZATION CYCLE (E.G. 14% EO 12 H. / 50 KPA / 37 °C).
CAUTION FOR DISPOSAL
IF THE DEVICE IS DAMAGED AND CANNOT BE REPAIRED, DISPOSE AND DISCARD IT ACCORDING TO
THE INTERNAL PROCEDURES OF THE HOSPITAL. DO NOT RECYCLE OR REUSE ANY COMPONENT.
Technical features
Sales description ..................................................
Dimensions (mm)..................................................
Weight (g) .............................................................
Scale of the vacuum gauge...................................
Maximum suction flow...........................................
Maximum supply depression.................................
Vacuum supply connection ...................................
Vacuum connector................................................
Storage conditions ................................................
Operating conditions .............................................
Direct suction nozzle
Height:
195
Width:
68
Depth: 48
400
From 0 to -100 kPa graduated at 5 kPa
From 0 to -760 mm Hg graduated at 20 mm Hg
60 L/min ± 5 L/min at -98 kPa
-98 kPa
Thread M8x1 F
Hose connector Ø 10÷12mm
-40 °C ± 2 °C / +60 °C ± 5 °C and 40% to 70% RH
+5 °C ± 2 °C / +35 °C ± 5 °C
Page 4

Popular Measuring Instrument manuals by other brands

Diesse VES-MATIC EASY Service manual

Diesse

Diesse VES-MATIC EASY Service manual

Mirion Technologies UltraRadiac Plus manual

Mirion Technologies

Mirion Technologies UltraRadiac Plus manual

Ryobi E49IR01 Operator's manual

Ryobi

Ryobi E49IR01 Operator's manual

CREASLIN HSV3261 user manual

CREASLIN

CREASLIN HSV3261 user manual

Ballantine 340 instruction manual

Ballantine

Ballantine 340 instruction manual

Kobelt 7178 Series Owner's installation, operation and maintenance manual

Kobelt

Kobelt 7178 Series Owner's installation, operation and maintenance manual

ABB TTF200 operating instructions

ABB

ABB TTF200 operating instructions

CEM FC-35 user manual

CEM

CEM FC-35 user manual

Brooks Instrument GT 1305 Installation and operation manual

Brooks Instrument

Brooks Instrument GT 1305 Installation and operation manual

Speedrite Fault Finder Handset user manual

Speedrite

Speedrite Fault Finder Handset user manual

ABERLINK Axiom Too installation manual

ABERLINK

ABERLINK Axiom Too installation manual

Pitney Bowes MAILSTATION 2 Quick install guide

Pitney Bowes

Pitney Bowes MAILSTATION 2 Quick install guide

mru OPTIMA 7 BIOGAS user manual

mru

mru OPTIMA 7 BIOGAS user manual

Motorola R2010D Operator's manual

Motorola

Motorola R2010D Operator's manual

ATAGO PAL-38S instruction manual

ATAGO

ATAGO PAL-38S instruction manual

Agilent Technologies E8362C Service guide

Agilent Technologies

Agilent Technologies E8362C Service guide

Thermokon FT-RDF18+ quick start guide

Thermokon

Thermokon FT-RDF18+ quick start guide

WAGNER BI 2200 instruction manual

WAGNER

WAGNER BI 2200 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.