MODEL CTI-1200-1
TENSION INDICATOR
INSTRUCTION MANUAL A-342
5
Copyright 2023 - Revised, Comptrol All Rights Reserved
In either case, the output signal is available on J6 between terminals 'Out+' and 'Out-". A terminal marked 'Shld'
is also provided for a shield connection for this cable.
Cable containing a shielded twisted pair will minimize noise pickup. For best interference rejection, strip the
cable shield back to within a few inches of where the cable enters the enclosures. Connect the drain wire from
the cable shield to the stud right next to the hole where it enters the box, not to the J6 “Shld”terminal. Do not
connect the other end of the shield. Tape it.
If the isolated output is being used, the downstream equipment should be connected while performing the initial
calibration procedure.
DISPLAY DECIMAL POINT
Comptrol sets the digital panel meter decimal point per customer order when it ships. Changing it requires a pencil
soldering iron with grounded tip, rosin-core solder, and solder wick. Switch off the main disconnect SW1 while working on
the circuitry
On the back of the digital panel meter is a printed circuit board, with several split pads. The three pads labeled (SG1,
SG2, and SG3) on the bottom right corner are used to set the decimal point placement as follows:
SG1 = 1.999 SG2 = 19.99 SG3 = 199.9
If all of the gaps are open, there will be no decimal point display (1999). To enable a specific decimal point, it is
necessary to close the gap in the respective pad with solder. When re-assigning decimal places, remember to open
previously closed solder gap using solder wick to remove the solder.
NOTE: Turn off power to the board and meter at SW1, before working on the circuitry.
PART IV- CALIBRATION
NOTES ON CALIBRATION
To calibrate the CTl-1200 a calibrated load must be applied on the measuring roll. In the case of tension
measurement, the easiest and most accurate way is referred to as a "pull test". One or more cables ropes, or straps
are threaded through the web path, and weights are hung from one end to produce the required load. Make sure that
THEY ARE CAPABLE OF SAFELY SUPPORTING THE LOAD.
The cable should follow the web path at least over the preceding roll, over the measuring roll, and over the
succeeding roll. It doesn't matter which end gets tied off and which gets the weights, except that there should be
no driven rolls between the measuring roll and the weights.
When using a single cable, position it at the center of the web, and apply a weight equal to the maximum web
tension. Note that this applies a point load to the weakest part of the roll. If the full-scale load has the potential to
deform the roll, use multiple cables & weights distributed symmetrically about the center of the measuring roll.
When using multiple cables, it is impossible to apply the weights to all cables simultaneously, so they will need to slip
to equalize tension. Place something soft between the cables and the rolls. This also protects the surface of the
roll.
INITIAL CALIBRATION
A static strap is provided for use while making internal adjustments to the CTl-1200. One end of the strap is clipped to the
enclosure and the other is attached to the wrist.