Condec UMC222 User manual

65443
UMC222
Digital Weight Indicator
Installation Manual


i
Contents
1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Front Panel Keypad and Annunciators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.1.1 Gross/Tare/Net Weighing Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1.2 USA or Standard Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.0 Installation and Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1 Unpacking and Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Enclosure Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Cable Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.4 Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4.1 AC Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 Board Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6 Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.7 Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.8 Load Cell Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.9 Serial Port Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.9.1 Serial I/O Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.0 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1 Setup Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.1 Direct Full Menu Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.2 Direct Cal Entry Only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.3 Front Panel Setup Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Main Menu Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Digital Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3.1 Parameter Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3.2 Configuration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.4 Serial Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.0 Options Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1 Option 1–Enables/Disables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1.1 Time and Date–Daylight Savings–Parameter 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1.2 5 Point Linearity Enable–Parameter 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1.3 Totalizer Operation–Parameter 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.1.4 Front Panel Keyboard Entry–Parameter 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.1.5 Count Feature–Parameter 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.2 Option 2–Smart Serial I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.2.1 To Create a File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.2 To Edit a File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.3 Option 3–Special Filtering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.3.1 First Stage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.3.2 Second Stage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.3.3 Third Stage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.4 Option 4 – Ticket Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Copyright © 2000 Condec. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
October 2000

ii
UMC222 Installation Manual
4.5 Option Test – Test Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.0 Auto Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.1 1 TyPE – Capacity Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.2 2 Ser – Serial I/O. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3 3 Opt – Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.0 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.0 Normal Weighing Mode Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.1 Display Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.1.1 Display Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.1.2 EPROM Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.2 LED Annunciators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.3 Function Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.4 Gross/Tare/Net Weighing Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.4.1 Display Mode on Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.4.2 ZERO Key Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.4.3 TARE Key Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.4.4 Overload and Underrange Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.4.5 Lb/Kg Conversion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.4.6 Auto Averaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.5 Serial Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.5.1 Serial Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.5.2 Demand Mode versus Continuous Data Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.0 Optional and Advanced Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.1.1 Demand Output Serial Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.1.2 Continuous Default Output Serial Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.1.3 Multiple Continuous Data Field Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.2 Battery Option – PN 55260. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8.2.1 Optional Battery Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
8.3 Panel Mount Kit–PN 55262. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
8.4 Tilt Stand Kit–PN 55257 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
8.5 Serial Expansion Board–PN 55255 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
9.0 Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9.1 ASCII Character Chart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9.2 Parameter Control Code Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
9.3 Smart Serial Count . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
9.4 Display and Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
9.5 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Introduction
1
About This Manual
The UMC222 indicator represents the latest in
state-of-the-art microprocessor technology specifically
applied to the weighing marketplace. This manual
provides information on installation, calibration,
configuration, and operation of the UMC222 digital
weight indicator.
Authorized distributors and their
employees can view or download this
manual from the Condec distributor
site at
www.4condec.com
.
1.0 Introduction
The UMC222 is a single-channel digital weight indicator designed to operate in a wide variety of scale and
weighing applications. The UMC222 digital weight indicator is ideally suited for platform or bench scale
applications requiring exceptional performance but can also be utilized for tank and truck scales. The indicator is
housed in a NEMA 4X stainless steel sealed case. The standard unit is equipped with a tilt stand base for tabletop or
wall mounting applications. The indicator front panel consists of a 5-button keypad, six seven-segment display
digits, and six LED annunciators (see Figure 1-1 on page 2).
Features of the UMC222 include:
• Full front panel digital configuration and calibration.
• Zero and span temperature compensation to ensure compliance with NTEP temperature range requirements (–10 to
40°C).
• Nonvolatile memory stores data for calibration, temperature compensation, configuration parameters, auto or fixed
tare values, PAZ and AZM values, and setpoint values.
• 10,000 displayed graduations in legal-for-trade applications; 80,000 displayed graduation in Non-Legal-For-Trade,
process weighing applications.
NOTE:
Use of more than 20,000 graduations may cause undesirable display instability in some applications.
• Analog sensitivity to 0.3
µ
V/grad at 20,000 graduations.
• Fifteen updates per second, with selectable digital averaging and auto averaging.
• Excitation for four 350
Ω
load cells or eight 700
Ω
load cells at 5VDC.
• Time and date print selection.

2
UMC222 Installation Manual
1.1 Front Panel Keypad and Annunciators
Figure 1-1 shows the UMC222 front panel. The UMC222 display consists of six seven-segment display digits.
Table 1-1 lists the front panel keys and their functions.
Figure 1-1. UMC222 Front Panel
Panel Key Function
ZERO Provides push-button auto zero (PAZ) over ±1.9% or 100% full scale capacity. Operates only in Gross
weighing mode. USA: rezeroes the scale if the gross weight is within the band of zero selected in the Set
Up Mode (2%, 5%, 20% or Full Scale). The Zero key will rezero the scale only in Gross mode.
OIML: rezeroes the scale if the gross weight is within the band of zero selected in the Setup mode (2%,
5%, 20% or Full Scale).
The Zero key will rezero the scale in either Net or Gross mode. If the scale is operating in the Net mode
and a valid Zero is executed, the unit will zero both the zero and tare values and returns the indicator to
Gross mode automatically.
MODE USA: Selects Net/Gross mode and recalls total or Average Piece Weight.
OIML: Selects the desired weight data to be viewed (Gross Recall or Total Recall). If the indicator is in the
Net mode, when the MODE key is pressed, the Net indicator LED is off and the recalled gross weight data
is momentarily displayed. The indicator will then automatically return to the Net display mode.
TARE Provides push-button tare entry over 100% of scale capacity. Pressing TARE key switches to net mode
and enters tare. USA – When issued in the Gross mode, the indicator will acquire the tare value and
automatically shift to the Net display mode.
OIML - When issued in the Gross mode, the indicator will acquire the tare value and automatically shift to
the Net Display mode if the Tare acquisition is > zero by (+1/2 grad). If a Tare is issued within ±1/2 grad of
zero, the indicator is set to Gross mode and the Tare value is set to zero (Tare = 0). If a Tare acquisition is
negative or the weight data is in motion, the indicator remains in the present mode and no tare is
performed.
UNIT If the indicator has been set up to perform conversions between the base unit and the selected
conversion unit, when this key is pressed the weight data is toggled between its calibrated base weight
unit (i.e. lb, kg,) and the conversion unit.
PRINT Provides a manual print function if the unit is wired to serial printer or other data device. See Section 8.4
on page 35 for serial output specifications. Issues a demand print command to the serial output. This key
also performs an accumulation of the current displayed weight data when the Totalizer mode is active.
Table 1-1. Front Panel Key Functions
CAPACITY CLASS
III
PRINT
UNIT
TARE
MODE
ZERO
Zero
Motion
Net
lb
kg
Total
n
= 10,000

Introduction
3
Table 1-2 summarizes the front panel annunciator functions.
1.1.1 Gross/Tare/Net Weighing Operations
The UMC222 offers two methods of Gross/Tare/Net operations (Standard, European). By default the indicator is set
to operate Gross/Tare/Net data in the more familiar USA (std.) conventional format. The selection of Eur sets the
Gross/Tare/Net operation to operate per OIML requirements.
1.1.2 USA or Standard Operation
When the standard indicator is initially powered up the display automatically comes up in the last mode it was in
(Gross or Net mode) prior to a power down.
Annunciator Function
ZERO On when scale weight is within ±0.25 displayed graduations of zero. Used in gross or net weighing mode.
NET On when the indicator is in net weighing mode.
lb/kg lb or kg LED is lit to show the current displayed weight units.
MOTION On when scale is in motion.
TOTAL On when the display mode is recalling the accumulated weight data.
Table 1-2. Front Panel Annunciators

4
UMC222 Installation Manual
2.0 Installation and Wiring
This section describes the procedures for connecting load cell and serial communications cables to the UMC222
indicator. Instructions for battery and CPU board replacement are included, along with assembly drawings and parts
list for the service technician.
Use a wrist strap to ground yourself
and protect components from
electrostatic discharge (ESD) when
working inside the indicator.
2.1 Unpacking and Assembly
All indicators are configured and tested prior to
shipment to ensure that they are fully functional.
Immediately after unpacking, visually inspect the
UMC222 to ensure all components are included and
undamaged. The shipping carton should contain the
indicator with attached tilt stand, this manual, and a
parts kit. If any parts were damaged in shipment, notify
Condec and the shipper immediately.
The parts kit contains the following items.
2.2 Enclosure Disassembly
The indicator enclosure must be opened to connect the
load cell and communications cables.
The UMC222 has no on/off switch.
Before opening the unit, ensure the
power cord is disconnected from the
power outlet.
Ensure power to the indicator is disconnected then
remove the screws that hold the front bezel to the
enclosure body. Place the indicator face-down on an
antistatic work mat, then lift the front bezel away from
the enclosure. Set the enclosure aside.
2.3 Cable Connections
The UMC222 provides four cord grips for cabling into
the indicator: one for the power cord, one for the load
cell cable, and two for serial communications.
Figure 2-1. UMC222 backplate
RLWS Part Number Description
19433 Square adhesive foot
15337 Wing knob
55450 Knob gasket
Table 2-1. UMC222 Parts Kit
!Caution
Warning
POWER
TB1
Load Cell
Cord Grip
TB2
REAR VIEW
TB3

Installation and Wiring
5
2.4 Power Connections
Standard units are powered by either 115 VAC or 230
VAC volts or by optional battery pack units.
2.4.1 AC Units
Units are powered by standard AC power. The AC
power cord must be plugged into a 3-prong polarized
AC wall socket.
VAC Conversion
The UMC222 can be converted from 115VAC to
230VAC. The following steps are necessary to
complete this conversion.
Before beginning, disconnect the
AC power source. Failure to do so
can result in injury or death.
1. Disconnect the power source from the unit.
2. Remove rear outer case of the UMC222.
3. Remove rear connector bracket and four
standoffs.
4. Remove the CPU board.
5. Remove the protective insulator panel from the
solder side of the CPU board.
6. Cut circuit traces between E3/E4 and E5/E6 as
indicated with a sharp instrument like a razor
blade or an x-acto knife. Refer to Figure 2-2.
Figure 2-2. Circuit Trace Setup
7. Add jumper, E4 to E6 using a properly insulated
wire with a minimum size of #22 AWG.
8. Replace the protective insulator panel.
9. Reassemble the unit, test, and label unit for
230VAC.
10. Change the power cord.
2.5 Board Removal
If you must remove the UMC222 CPU board, use the
following procedure:
1. Disconnect power to the indicator. Loosen cord
grips and remove enclosure as described in
Section 2.2.
2. Unplug all connections to the CPU board. See
Figure 2-1 on page 4 for connector locations.
3. Remove the four nuts from the corners of the
CPU board.
4. Remove the CPU board from the enclosure.
To replace the CPU board, reverse the above procedure.
Be sure to reinstall cable ties to secure all cables inside
the indicator enclosure. Replace enclosure and torque
screws.
2.6 Replacement Parts
Table 2-2 lists the replacement parts for the UMC222.
2.7 Instrument Setup
The UMC222 operates with the EPROM program
KDK 1921. To verify the program installed in the
indicator, turn on the indicator and observe the
displayed value at the EP prompt (see Figure 2-3). The
EP prompt displays the family, set, and version level of
the installed EPROM.
Figure 2-3. EPROM Display
To ensure that the UMC222 is in proper operating
condition, the indicator can be tested with a load cell
simulator. The input signal should be as close as
possible to the normal system millivolt value.
Figure 2-5 on page 6 shows the simulator-to-indicator
wiring connection in a six-wire configuration. See
Section 2.8 for more information.
NOTE:
Six-wire configuration requires that the +SEN
lead be shorted to +EXC and the –SEN lead be shorted
to –EXC at the simulator only.
Warning
E3 E4 E6 E5
Jumper E4 to E6
Cut between
E3 and E4 Cut between
E5 and E6
Circuit Side of CPU Board
Converting from 115V AC to 220V AC
Part Number Description
56649 Thumbscrew
64983 Membrane keypad
59784 Front panel
63926 CPU board w/ EPROM
55257 Tilt stand w/ wing knobs
15337 Wing knob
64745 ESD supressor board
Table 2-2. UMC222 Replacement Parts
EP–1.12
EPROM Set 1 Version 12

6
UMC222 Installation Manual
2.8 Load Cell Wiring
Units are equipped with a six-wire load cell connection. Figure 2-4 shows the location of TB1 on the back of the
indicator.
Figure 2-4. Load Cell Connector Location
The following diagram illustrates the load cell hookup TB1 as shown in Figure 2-5.
Figure 2-5. Load Cell Wiring From Indicator
NOTE:
S2 and S3 jumpers are for setting the load cell excitation for either REM(ote) or LOC(al) sensing.
POWER
TB1
Load Cell
Cord Grip
TB2
REAR VIEW
TB3
S1
FR
DIR
SETUP
U2
S2
TB1
–EXC
–SENSE
–SIGNAL
+SIGNAL
+SENSE
+EXE
654321
LOAD CELL CONNECTOR
1 2
OPEN
S3
RF1
SHIELDED
A/D
CONFIGURATION
SETUP
SENSE
REM. LOC.
SENSE
REM. LOC.
Check load cell color
code for proper wiring
E13
E14

Installation and Wiring
7
2.9 Serial Port Wiring
The UMC222 serial port can be configured to communicate directly to a printer, remote display, or other device
using RS232. The standard UMC222 comes equipped with two RS232 serial communication ports (TB2 on the
back of the indicator). Port 1 is a bi-directional port that can be configured for either duplex or simplex operation.
Port 2 is for output only.
NOTE:
The serial communications capability of the
UMC222
is dependent on the product specifications of the
receiving device.
Access to this serial communication port is through the cord grip (TB2) located on the back of the UMC222
indicator (see Figure 2-6). Wiring is extended through this connector and wired inside the indicator. See the output
connector diagram in Figure 2-7 for connector and wire identification.
Figure 2-6. Serial Port Location
The standard serial port connection (TB2), is a 4-wire connector. Either a two position terminal block can be added
or a six position replacement for TB2 can be installed in the adjacent E points (E13 and E14) shown in Figure 2-7
which can be used for optional input wiring.
Figure 2-7. E13 and E14 Locations
Figure 2-8 illustrates a typical wiring pattern of a six position replacement.
Figure 2-8. Alternate Serial I/O wiring
POWER
TB1 TB2
REAR VIEW
TB3
TB2
Serial Port Location TB3
Serial Port Location
TB2
SIG COM
IN
OUT
OUT
4321
SERIAL
COMMUNICATION
E13
E14 E13 and 14
LOCATIONS
for optional remote
input
TB2
4321
SERIAL
COMMUNICATION
SIG COM
IN
OUT
OUT
P1
P2
RS232
E13
E14
IN
OUT
remote
input
wiring
i
for future
enhancements

8
UMC222 Installation Manual
2.9.1 Serial I/O Wiring
Serial Port 1 is a bi-directional (full duplex) port supporting active/passive RS232 serial communications. The
following table points out the TB2 wiring.
TB2 Serial I/O Wiring Table
Port TB2
Port 1 (P1) Transmit OUT TB2-2
Receive IN TB2-3
SIG COM TB2-4
Port 2 (P2) Transmit OUT TB2-1
SIG COM TB2-4
Table 2-3. Serial I/O Wiring

Configuration
9
3.0 Configuration
Prior to calibration, the UMC222 must be digitally
configured, or assigned a set of operating parameters.
To configure the UMC222 indicator, the indicator must
be placed in setup mode.
3.1 Setup Mode
To enter the setup mode, you must access the setup
switch which is located on the CPU board. Remove the
four large screws on the front of the indicator and open
the enclosure.
There are three methods for entering the setup mode:
•Direct Full Menu Entry
•Direct Cal Entry Only
•Front Panel Setup Mode
3.1.1 Direct Full Menu Entry
Set the setup rocker switch S1 position SW-1 (DIR) to
the closed position on the CPU board. This allows you
full access to all main menu entry functions.
Figure 3-1. Configuration Setup Switch
Once configuration and calibration (Section 3.0 and
Section 5.0) of the UMC222 is complete, open S1-1
and close up the unit.
3.1.2 Direct Cal Entry Only
Set the setup rocker switch S1 position SW-2 to the
closed position on the CPU board. This allows access to
the calibration menu options.
Once calibration (see Sections 5.0) of the UMC222 is
complete, open S1-1 and close up the unit.
3.1.3 Front Panel Setup Mode
Set the setup rocker switch S1 position SW-1 to the
closed position. Enable the front panel access in Option
1/parameter 5 Kbd, open S1, SW-1.
Press and hold the
MODE
and
UNIT
keys for
approximately three seconds which allows the serial
configuration option to appear.
Once serial configuration (see Section 3.4) of the
UMC222 is complete, select
DONE
, by scrolling with
the
ZERO
key in the main menu and press the ENTER
(Print) key to exit.
3.2 Main Menu Selection
Table 3-1 defines the functional operation of each key
on the front panel of the indicator when the unit is in the
set up mode.
Pressing the ZERO key on the indicator front panel
while in the setup mode allows you to advance through
the main menu selections. Those menu selections are
listed in Table 3-2 shown below.
S1
CAL
DIR
SETUP
OPEN
12
Key Function During Setup
Menu select. When pushed will return
operator to main menu.
Parameter/subparameter select
Parameter/subparameter data select
Recalls parameter name and decrements
parameter select
Used as an Enter key for numeric data entry
and calibration entry
Table 3-1. Front Panel Key Functions
Display Menu Selection
1. CAL Calibration
2. Conf Configuration
3. Ser Serial Setup
4. Auto Auto Setup
5. Opt1 Time and Date / Linear / Totalizer / Piece Count
6. Opt2 Smart Serial
7. Opt3 Adaptive Filtering
8. Opt4 Ticket Count
t. tESt Expanded Resolution / Input mV value / Serial
I/O port test
d. dOnE Exit, Setup (only in front panel key access)
Table 3-2. Main Menu Selections
ZERO
MODE
TARE
UNIT
PRINT

10 UMC222 Installation Manual
Figure 3-2 provides a graphic representation of configuration parameters associated with the UMC222.
Figure 3-2. Configuration Menu
3.3 Digital Configuration
3.3.1 Parameter Overview
Table 3-3 on page 11 lists configuration parameters and describes their values. The following paragraphs give the
procedure for configuring the UMC222.
Do not lean the front bezel with the exposed board against the housing enclosure without placing a
static shield between exposed board and housing.
3.3.2 Configuration Procedure
1. Unscrew the four screws on the face plate bracket. Pull the face plate away from the enclosure.
2. Locate the setup switch S1 on the UMC222 CPU board (lower portion of board).
3. Close switch S1-1 by moving to the left (closed) position (Section 3.1). A prompt appears with a parameter
number and parameter name.
4. Use the ZERO key to scroll through the Main menu selections, see Table 3-2 on page 9. Select display 2
Conf.
5. Table 3-3 lists the configuration parameters available.
6. Scroll through the parameter selections by using the MODE key and select which parameter to change.
7. Use the TARE key to change data.
8. If numeric data entry (such as capacity), press the TARE key to activate the data entry mode. Press the TARE
key to select the digit to be modified.
9. Use the ZERO key to increment the digit and the MODE key to decrement the digit.
XXXXXXX XXXXXXXXXXXXXXXXXXXXX
CAP
capacity
entered value
.0001
.0005
.0002
.001
lb.
kg. 2
1
4
6
Off
3
1
5
TESTOPTION 4
OPTION 3
OPTION 1 OPTION 2SERIAL
CALIBR
CONFIG AUTO
RES
resolution BASE
cal. base unit FILT
digital filter ACTN
motion detect.
TARE
enable tare
.2
.005
.002
.01
.05
.02
.1
.5
Off
1
5
3
10
5
2
legal-for-trade
10
100
20
Off
Tare
Zero
Print
Off
kg
lb
Z TrK
zero band A per
zero range CONV
enable conver.
0-ON
factory default
IN
remote input GTN
regulatory
mode for tare
50
1
.5
2
10
5
20
100
12
8
10
Auto 2
16
Auto 1
Auto 3
50
10
20
Off
On
Std
Eur
!Caution

Configuration 11
10. Press the PRINT key to enter that value and exit out of the numeric data entry mode.
11. To abort the data entry mode, press the UNIT key.
When configuration is complete, set SW1-1 open (right) to return the unit to normal operating mode.
NOTE: Values marked in bold denote pre-programmed factory default settings.
Parameter Description Data Value Selection
1 Capacity - enter desired capacity NTEP to 10,000 (up to 999990 available).
2 Resolution - enter resolution .0001 to .1, .0002 to .2, .0005 to .5, 1, 2, 5, 10, 20, 50, 100
3 Base unit - unit of calibration lb or kg conversion
4 Digital filtering 1, 2, 4, 6, 8, 10, 12, 16, Auto 1, Auto 2, Auto 3
(1 least filtering, 16 most filtering, auto 1 least filtering-quick
response, auto 3 most filtering-slow response)
5 Motion (action) Off, 1, 3, 5, 10, 20, 50
6 Tare mode Off, on
7 Zero track Off, .5, 1, 3, 5, 10 graduations
8 Zero aperture 2, 5, 10, 20, 100 % of full scale. Only default is legal-for-trade
9 Conversion select - enable
conversion Off, lb, kg
10 Stored zero Stored –factory default
11 Remote input selections Off, zero, tArE, print.
Recommended factory installation.
This option requires a 2-position connector (part number 55391)
be soldered to the board for easy access).
12 Mode selection –G/T/N Std., Eur - gross/tare/net mode selection EUR=OIML
Table 3-3. Configuration Parameter Overview

12 UMC222 Installation Manual
3.4 Serial Configuration
The UMC222 has two serial ports. Both Serial Ports 1 and 2 are ASCII-compatible RS232 outputs. The serial
format is compatible with most printers, scoreboards, and other remote devices. Each output can be disabled, set for
print on demand mode, or set to output data continuously.
Table 3-4 shows the configuration selections for Parameter 3 which control the configuration of Ports 1 and 2.
Parameter
(Mode Key)
Subparameter
Selection
(Mode Key)
Data Value
Selection
(Unit Key) Explanation
1 P1 Off, On Port 1 enable, Smart Serial files 2.1 –2.4
2 baud 600, 1200, 2400,
4800, 9600,
19200
Port 1 baud rate
3 Par None, Even,
Odd, One Port 1 parity
4 dUPL Off, On, Adr Port 1 duplex enable as standard or address (RS485)
4.1 Chk None, or, CkS Duplex check select
4.2 res None, Echo, Ack Duplex response format
4.3 Adr Adr xx (01) Address (1 to 98)
5 P1- Print, Cont, none Serial type selection. If Par 4 Dupl = adr only none selectable.
5.1 dat Gtn, DISP Port 1 G/T/N or display data
5.2 EOL CrLF, Cr Port 1 End of line cr/lf or cr
5.3 SOt None, StX Port 1 Start of test
5.4 EOt None, Eot, FF, LF Port 1 End of text
5.5 del Off,1 Sec,2 Sec,
3 Sec, 4 Sec Port 1 Print delay after Cr
Smart Serial: Code 700 delays serial xmit for # of seconds.
6P2 Off, Print, Cont Port 2 enable
Smart serial: Files 2.5 –2.8
6.1dat Gtn, DISP Port 2 G/T/N or display data
6.2EOL CRLF, Cr Port 2 End of line cr/lf or cr
6.3SOt None, StX Port 2 Start of test
6.4EOt None, Eot, FF, LF Port 2 End of text
6.5delt Off,1 Sec,2 Sec,
3 Sec, 4 Sec Port 2 Print delay after CR
Smart serial Code 700 delays serial xmit for # of seconds
7 baud 600, 1200, 2400,
4800, 9600,
19200
Port 2 baud rate if port 1 not set for DU
8 Par None, Even,
Odd, One Port 2 parity if port 1 not set for DU
Table 3-4. Serial Configuration Prompts

Configuration 13
All serial characters in the data format are in ASCII and consist of the following:
Table 3-5. Data Formats
The transmission of serial data can be initiated in either demand mode, continuous mode, or RS232. Refer to
Section 7 for advanced detailed information on serial configuration options.
Data Formats
1 Start Bit
7 Data Bits
1 Parity Bit (Even Parity)
2 Stop Bits

14 UMC222 Installation Manual
4.0 Options Configuration
The UMC222 offers a selection of features that are available in the configuration of the indicator. The options setup
mode allows the operator to expand the capabilities of the indicator. The available features are:
•Option 1 - Enables/Disables
•Option 2 - Smart Serial I/O
•Option 3 - Special Filtering
•Option 4 - Ticket Counter
•Option t - Test Mode
To access options configuration, go into the setup mode
and scroll to the options menu. Close switches SW1
(Figure 4-1). If the option mode has been enabled,
selections are available.
Figure 4-1. Closed Configuration Switches
In the option mode, primary function keys (ZERO
MODE, etc.) operate as secondary function keys
(Figure 4-2).
Figure 4-2. Front Panel Keys With Alternate Functions in
Options Menu
Use the ZERO key to select options 1 through 4 and test
menu. Push the MODE key to access the sub
parameters.
Some options, once enabled, may have an option
submenu available. Use the MODE key to access the
option submenu. To make parameter data selections in
the option submenu, use the TARE key. Display
prompting is provided in each case.
Key Functions Summary
Use the list below as a quick reference when selecting
and configuring the options.
Figure 4-3. Quick Key Functions for Options
S1
CAL
DIR
SETUP
OPEN
12
CAPACITY CLASS III
PRINT
UNIT
TARE
MODE
ZERO
Zero
Motion
Net
lb
kg
Total
n
= 10,000
MENU
SELECT ENTER
PARAMETER
RECALL
DECREMENT
PARAMETER/
SUB-PARAMETER
SELECT
SCROLLS
THRU
DATA
SELECT
ZERO
MODE
TARE
UNIT
PRINT
ZERO –Increments option number (menu select)
MODE –Parameter and subparameter select
TARE –Data select
UNITS –Recalls parameter name and
decrements parameter select
PRINT –Enter key for numeric data entry and for
calibration entry

Options Configuration 15
4.1 Option 1–Enables/Disables
Option 1 enables or disables various parameters that enhance the indicator’s functionality. Those parameters that
can be enabled or disabled include:
•Time and date (parameter 2)
•Five point linearization for span calibration (parameter 3)
•Ten digit totalizer (parameter 4)
•Front panel keyboard access (parameter 5)
•Count feature (parameter 6)
Option 1 configuration parameters are shown in Table 4-1 and the menu structure is graphically illustrated in
Figure 4-4 on page 16.
Table 4-1. Option 1 Configuration Prompts
Parameter
(mode key) Subparameter
Selection
Subparameter
Data Selection
(tare key) Explanation
1 OP.1 Off, On Expanded RAM option enable
2 time & date Off, On Time and date enable. Smart serial codes 400 to 402 brings
time/date into the serial data.
2.1 dLS Std, dLS Daylight savings enable
2.2 Hr 12, 24 12 or 24 hour selection for print
2.3 A–Pt AM, PM Subparameter of 2.2 hr - 12
2.4 ti hh.mm.ss Time recall/setting (setting in 24 hour mode)
2.5 Eur On, oFF Date print format enabled for day/mo/yr. dd/mm/yr
2.6 dat xx.xx.xx Date recall/setting (setting in mo/day/yr) mm/dd/yy
2.7 pos On, Under,
Above Location of printed time/date
Above, on, or under the weight information
2.8 type Letter, No. Print letters (May 1, 2000) or numbers (05/01/00)
3 Lin Off, On 5 point linearity enable
4 Tot -
net
gross
Off, On Total enable. Smart serial codes 230 to 233 to print total
mode data
4.1 bnd Off, 1 to 9, 10,
20, 30, 40, 50 Total zero band
5 kbd Off, On Off, On Front panel keyboard access to configuration of
non-legal-for-trade options (time and date, serial I/O setup
and ticket counter)
6 Cnt Off, On Off, On Enable count mode. Smart serial codes 234-237 and 240 to
243 to print count mode data

16 UMC222 Installation Manual
Option 1 Menu
Figure 4-4. Option 1 Menu
XXXXXXX XXXXXXXXXXXXXXXXXXXXX TESTOPTION 4
OPTION 3
OPTION 1 OPTION 2SERIAL
CALIBR
CONFIG AUTO
OFF ON
Off
On dLs
Std. 12
24
TIME/DATE DLS HR A –PT
Am
Pm
TI EUR DATE POS
of date
LIN
linearity TOT
TYPE BND
total band KBD
keyboard access
CNT
count mode
Letter
No
Off
On
On
Off
Off
1
Off
On
Off
On
hh.mm.ss On
Off
mm/dd/yy On
Under
Above
2
3
4
5
6
7
8
9
10
20
30
40
50
Table of contents
Other Condec Accessories manuals
Popular Accessories manuals by other brands

Waterpik
Waterpik Complete Care CC-01 quick start guide

HomeMatic
HomeMatic HmIP-STHO-A Mounting instruction and operating manual

Wenglor
Wenglor P1KL Series operating instructions

OXFORD CRYOSYSTEMS
OXFORD CRYOSYSTEMS PheniX Operation & instruction manual

Oppermann Regelgeräte
Oppermann Regelgeräte CO2-W-LC manual

Sonoff
Sonoff S55 Series Quick installation guide