Contech Brooks Sho-Rate Series User manual

Sho-Rate Series Models
1350G/1355G (Sizes 2-6)
Glass Tube Variable Area Flow Meters
Installation & Operation Manual
Sho-Rate
Model 1350G
(No Valve)
Sho-Rate
Model 1350G
with optional
integrated
flow controller
Sho-Rate
Model 1350G
with optional
needle valve
Sho-Rate
Model 1355G
(No Valve)
Sho-Rate
Model 1355G
with optional
integrated
flow controller
Sho-Rate
Model 1355G
with optional
needle valve
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

for local sales office contact information. Save this instruction manual for future reference.
this warning can result in serious personal injury and / or damage to the equipment.
result in serious personal injury and / or damage to the equipment.
Connect all products to the proper electrical and pressure sources.
and outlet connections. If no leaks are present, bring the system up to the operating pressure.
maintenance is being performed by qualified persons.
Essential Instructions
Read before proceeding!
European Pressure Equipment Directive (PED)
Compatibility (EMC directive 2014/30/EU).
The cable screen should be connected to the metal shell or gland and shielded at both ends over 360 Degrees.
The shield should be terminated to an earth ground.
European Electromagnetic Compatibility (EMC)
during the removal, installation or other handling of internal circuit boards or devices.
removed or adjusted.
Comments
devices. Damaged components, even though they appear to function properly, exhibit early failure.
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Contents
Section 1 Introduction
Description .................................................................................................................................................... 1
Specifications ................................................................................................................................................ 1
Optional Equipment ....................................................................................................................................... 4
Section 2 Installation
General...................................................................................................................................................... 5
Receipt of Equipment ................................................................................................................................. 5
Unpacking ................................................................................................................................................. 5
Return Shipment ........................................................................................................................................ 5
Recommended Storage Practice ................................................................................................................ 5
Installation ................................................................................................................................................. 6
Section 3 Operation
Operation ................................................................................................................................................. 12
Section 4 Bistable Inductive Limit Switches
Description .............................................................................................................................................. . 13
Technical Data Limit Switch ...................................................................................................................... 13
Hazardous Area for Size 2 Tube ...............................................................................................................13
Hazardous Area for Size 6 Tubes................................................................................................................13
Power Supply/relay Unit................................................................................................................................14
Maximum Operating Temperature................................................................................................................14
Recommended Spares.................................................................................................................................14
Pressure Equipment Directive (PED)97/23/EC.............................................................................................14
Section 5 Alarm Settings
Alarm Settings...........................................................................................................................................15
Section 6 Maintenance
Disassembly and Cleaning....................................................................................................................... 16
Reassembly Procedure..............................................................................................................................17
Section 7 Part List
General...................................................................................................................................................... 18
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section A Essential Instructions ....................................................................................................................... A-1
Warranty, Local Sales/Service Contact Information ...................................................................... Back Cover
Figures
2- 1
Typical Flowmeter Installation ..................................................................................................................7
22
Dimensions - Sho-Rate 1350G ................................................................................................................ 7
2- 3
Dimensions - Sho-Rate 1350G Panel Mount ............................................................................................ .8
2-4 Dimensions - Sho-Rate 1350G Bezel ....................................................................................................... .8
2-6
Dimensions - Sho-Rate 1355G Panel Mount ............................................................................................ . 9
2-7
Dimensions - Sho-Rate 1355G Bezel ....................................................................................................... 10
2-8
Dimensions - Model 1350G/1355G with Flow Controller .......................................................................... 11
5-1 Alarm settings ......................................................................................................................................... 15
Tables
1-1
Specifications for Sho-Rate Models 1350G and 1355G ........................................................................... 2
1-2
Capacities Rib Guide Tubes, Spherical Floats for use with 1350G Only ................................................... 3
1-3
Tube and Float Code - Seventh Digit in Model Code for Scale Inscription ................................................ 3
1-4
Tube and Float Code - Eighth and Ninth Digit in Model Code for Scale Configuration ............................... 3
1-5
Capacities - Rib Guide Tubes, Spherical Floats for use with 1355G Series Only ...................................... 4
7-1 Parts List - Sho-Rate Models 1350G and 1355G ..................................................................................... 20
5-2 Standard bistable alarm .......................................................................................................................... 15
5-3 Bistable low and high flow alarm ............................................................................................................. 15
7-1 Parts Drawing Sho-Rate Models 1350G and 1355G ................................................................................ 19
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 1 Introduction
Section 1: Introduction
Description
The Sho-Rate Flowmeters are variable area, glass tube, flow rate
indicating meters. The basic elements are a tapered glass metering tube
and a metering float. Features include quick and simple removal or
installation of the tube and float while the meter remains in the process
piping.
Specifications (Reference Tables 1-1 thru 1-5)
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 1 Introduction
Table 1-1 Specifications for Sho-Rate Models 1350G and 1355G
Performance
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 1 Introduction
Table 1-2 Capacities - Rib Guide Tubes, Spherical Floats for use with 1350G Series Only
Meter
Size
Tube
No.
Float
Material
Water Air
GPH LPH SCFH NLPH
2
R-2-65-A G
Glass
0.010
0.041
0.12
3.2
Sapphire
0.021
0.079
0.19
5.0
Stainless Steel
0.049
0.18
0.37
9.8
Carboloy
0.10
0.36
0.65
17
Tantalum
0.10
0.40
0.71
18
R-2-65-B G
Glass
0.014
0.06
0.16
4.4
Sapphire
0.028
0.10
0.25
6.7
Stainless Steel
0.07 0.25 0.48 12
Carboloy
0.12
0.48
0.80
21
Tantalum
0.14
0.53
0.87
22
R-2-65-C G
Glass
0.12
0.47
0.99
26
Sapphire
0.22
0.83
1.3
35
Stainless Steel
0.41
1.5
2.1
55
Carboloy
0.65
2.4
3.1
81
Tantalum
0.70
2.6
3.3
87
R-2-65-D G
Glass
0.68
2.5
3.9
100
Sapphire
0.99
3.7
5.1
130
Stainless Steel
1.6 6.3 7.9 200
Carboloy
2.5
9.5
11
290
Tantalum
2.7
10.0
12
310
6
R-6-65-A G
Glass
2.2
8.5
13
340
Sapphire
3.3
12
17
440
Stainless Steel
5.6
21
25
660
Carboloy
8.3 31 36 940
Tantalum
8.8
33
38
1000
R-6-65-B G
Glass
8.7
33
46
1200
Sapphire
12
47
59
1500
Stainless Steel
20 76 86 2200
Carboloy
29
100
110
3100
Tantalum
30 110 120 3300
Air flows are at 14.7 psia and 70 Degrees F/1.01 Bar and 21.1 Degrees C
Table 1-3 Tube and Float Code -
Seventh Digit in Model Code for Scale Configuration
Code Model 1350 Tube Model 1355 Tube
A
R-2-15-A G
B
R-2-15-B G
C
R-2-15-C G
D
R-2-15-D G
F
R-6-15-B G
G
R-2-65-A G R-2-15-AAAA G
H
R-2-65-B G
J
R-2-65-C G
K
R-2-65-D G
L
R-6-65-A G
M
R-6-65-B G
N No Tube No Tube
Table 1-4 Tube and Float Code -
Eighth and Ninth Digit in Model Code for Scale Configuration
Meter
Accuracy
Float
Material
Decal Scale Inscription
MM
Scale
Linear
Scale
Custom
Decal
Percent
Scale
Standard
(1350-10%)
(1355- 5%)
Glass 1A 2A 3A 4A
Stainless Steel 1B 2B 3B 4B
Sapphire 1C 2C 3C 4C
Carboloy 1D 2D 3D 4D
Tantalum 1E 2E 3E 4E
Calibrated
(1350-5%)
(1355- 2%
Glass 1G 2G 3G 4G
Stainless Steel 1H 2H 3H 4H
Sapphire 1J 2J 3J 4J
Carboloy 1K 2K 3K 4K
Tantalum 1L 2L 3L 4L
*Note: Code for No Float/No Scale = “ZZ”
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 1 Introduction
Table 1-5 Capacities - Rib Guide Tubes, Spherical Floats for use with 1355G Series Only
Meter Tube Float Maximum Flow Rate*
Size
2
No. Material Water (CC/Min)
Air
R-2-15-AAAA G
Glass
0.59
50 SCC/M
Sapphire
1.1
79 SCC/M
Stainless Steel
2.6
150 SCC/M
Carboloy
5.2
280 SCC/M
Tantalum
5.8
310 SCC/M
R-2-15-D G
Glass
5.5
370 SCC/M
Sapphire
10
520 SCC/M
Stainless Steel
20
830 SCC/M
Carboloy
34
1200 SCC/M
Tantalum
36
1300 SCC/M
R-2-15-A G
Glass
17
0.82 SLPM
Sapphire
26
1.0 SLPM
Stainless Steel
46
1.6 SLPM
Carboloy
70
2.4 SLPM
Tantalum
75
2.5 SLPM
R-2-15-B G
Glass
53
2.3 SLPM
Sapphire
80
3.0 SLPM
Stainless Steel
130
4.6 SLPM
Carboloy
200
6.7 SLPM
Tantalum
210
7.1 SLPM
R-2-15-C G
Glass
90
4.0 SLPM
Sapphire
130
5.2 SLPM
Stainless Steel
220
7.9 SLPM
Carboloy
340
11 SLPM
Tantalum
360
11 SLPM
6
R-6-15-A G
Glass
210
9.5 SLPM
Sapphire
320
12 SLPM
Stainless Steel
540
18 SLPM
Carboloy
790
25 SLPM
Tantalum
840
26 SLPM
R-6-15-B G
Glass
560
23 SLPM
Sapphire
820
29 SLPM
Stainless Steel
1300
43 SLPM
Carboloy
1900
60 SLPM
Tantalum
2000
63 SLPM
Air flows are at 14.7 psia and 70 Degrees F/1.01 Bar and 21.1 Degrees C
Optional Equipment
Standard Needle Valve
The standard needle valve can be supplied integrally mounted to the inlet
or outlet of the instrument. For more details on the needle valve go to our
website: BrooksInstrument.com, select Documentation, Precision Valves
& Flow Controllers, select Brooks-Line IV, CART, 8503 or 8504 valves.
Flow Contollers
Flow controllers can be supplied integrally mounted to the inlet or outlet of
the instrument. For the flow controller's complete instruction manual go to
our website: BrooksInstrument.com, select Documentation, Precision
Valves & Flow Controllers, select FC8800, or FC8900.
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 2 Installation
Section 2: Installation
General
This section provides installation instructions for the ®
Brooks Sho-Rate
Models 1350G/1355G Flowmeters. Refer to Section 1 of this manual for
connections.
Receipt of Equipment
When the equipment is received, the outside packing case should be
checked for damage incurred during shipment. If the packing case is
damaged, the local carrier should be notified at once regarding his liability.
A report should be submitted to the nearest Brooks Instrument location
listed on the Global Service Network page on our website:
BrooksInstrument.com/GlobalSupportCenters
Remove the envelope containing the packing list. Carefully remove the
instrument from the packing case. Make sure spare parts are not discarded
with the packing materials. Inspect for damaged or missing parts.
Unpacking
Carefully unpack the meter and inspect it for any damage that may have
occurred during shipment. The flowmeters are shipped completely
assembled and tested. It should not be necessary to tighten or adjust any
of the parts when it is received.
Return Shipment
Prior to returning any instrument to the factory for any reason, visit our
website for instructions on how to obtain a Return Materials Authorization
Number (RMA #) and complete a Decontamination Statement to
accompany it: . BrooksInstrument.com/Service All instruments returned to
Brooks also require a Material Safety Data Sheet (MSDS) for the fluid(s)
used in the instrument. Failure to provide this information will delay
processing of the instrument.
Instrument must have been purged in accordance with the following:
Recommended Storage Practice
If intermediate or long term storage is required for equipment, as supplied
by Brooks Instrument, it is recommended that said equipment be stored in
accordance with the following:
•
Within the original shipping container.
•
Stored in a sheltered area, preferably a warm, dry heated warehouse.
•
Ambient temperature: 70°F (21.0°C) nominal, 110°F (43°C) maximum,
45°F (7.1°C ) minimum.
•
Relative humidity 45% nominal, 60% maximum, 25% minimum.
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 2 Installation
Upon removal from storage, a visual inspection should be conducted to
verify the condition of equipment is “as received”. If the equipment has
been in storage for an excess of two (2) years or in conditions in excess of
those recommended, all pressure boundary seals should be replaced and
the device subject to a hydrostatic/pneumatic pressure test to 150% of
rated pressure.
Installation
The flowmeter should be mounted within 6° of true vertical. The inlet
connection to the flowmeter is in the bottom end fitting. The connections
are normally horizontal, female NPT. Be sure the piping is adequately
supported to prevent undue strain on the meter.
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 2 Installation
Figure 2-1 Typical Flowmeter Installation
It is recommended that a final leak test of the system plumbing and meter
be performed before subjecting it to process fluid.
(See "4-2 Reassembly Procedure" on p. 4-2, step 6)
Figure 2-2 Dimensions - Sho-Rate 1350G
HORIZONTAL
LINE
VERTICAL
LINE D
FLOWMETER
B
FLOWMETER B
C
A D A
E E
C
A - Inlet Valve B - Outlet Valve C - Bypass Valve
D - Control Valve E - Drain Valve
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Figure 2-3 Dimensions - Sho-Rate 1350G Panel Mount
Figure 2-4 Dimensions - Sho-Rate 1350G Bezel
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Figure 2-6 Dimensions - Sho-Rate 1355G Panel Mount
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Figure 2-7 Dimensions - Sho-Rate 1355G Bezel
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Figure 2-8 Dimensions - Model 1350G/1355G with Flow Controller
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Section 3: Operation
After the flowmeter has been installed in the flow system, it is ready for
operation. An optional built-in needle control valve may be provided to control
the flow through the flowmeter. These control valves are designed
for fine control. Excessive tightening may damage the valve seat and limit its
effectiveness as a control valve. If tight shut-off is required, it is
recommended that a separate shut-off valve should be installed in the line
immediately before the flowmeter.
Flow indication is measured using the center of the spherical float as the
reference point.
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Description
T he Brooks bistable limit switches are inductive ring initiators designed to be
slipped over the glass tube. The switch operates as a coil that is inductively
a ctuated by a 316 Stainless Steel or Carboloy ball float. The limit switches can
be set to any desired limit values, by sliding the switch along the metering
tube. The Brooks ring initiators create a highly sensitive, stable latching
signalling on high or low flows, or deviations from a controlled flow.
Technical data limit switch
Hazardous area for size 2 tube
Hazardous area for size 6 tubes
Switch type Inductive
Function bistable (latching)
Repeatability < 0.5 mm
Input voltage 8Vdc nominal
Current consumption < 1 mA (direction A to B)
>3 mA (direction B to A)
Max. float speed 10 m/s
Ambient temperature 0°C to 40°C
Protection to IEC 60529 IP67
Cable specification 2 conductor , 2m PVC-covered
Conductor cross section 2
0,14 mm (#23 AWG)
Housing material Crastin (PBT)
Ring initiator RC 10-14-N3 (size 2 tube)
PTB 99ATEX2128 X II 2 G EExia IIC T6
EMC to EN 60947-5-2
In compliance with EN 50227
Self inductance Limax = 70 µH
Self Capacitance Cimax = 90 nF
Ring initiator RC 15-14-N3 (size 6 tube)
PTB 99ATEX2128 X II 2 G EExia IIC T6
EMC to EN 60947-5-2
In compliance with EN 50227
Self inductance Limax = 120 µH
Self Capacitance Cimax = 90 nF
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Power supply/relay unit
P roximity switches may only be connected to a power supply/relay unit
Maximum operating temperature
certified EEx ia IIB or EEx ia IIC, EEx ib IIB or EEx ib IIC. The switch adopts
the same classification as the power supply/relay unit. The power supply/relay
unit must fulfill NAMUR requirement, EN 50227 (DIN 19234).
The maximum fluid temperature for the Sho-Rate series with the inductive
switches is 70°C.
Recommended spares
Pressure Equipment Directive (PED)97/23/EC
Flow meter complies under Sound Engineering Practices (SEP)
Model Size Description Part number
Sho-Rate Size 2 Bistable alarm incl. mounting parts 273-G-006-ZZZ
Sho-Rate Size 6 Bistable alarm incl. mounting parts 273-G-007-ZZZ
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Alarm Settings
T he minimum setting distance of a switch and a fitting is 8 mm, and the
minimum setting distance between 2 switches is approx. 30 mm, see figure 1.
The standard alarm function is a bistable (latching) function, see figure 2.
O nce the ball float moves inside the inductive switch, the alarm function is
activated, it remains activited, even if the float continues to move towards the
alarm zone, thus leaving the inductive switch. The relay will drop out as soon
as the float crosses the switch from the opposite direction, and moves back the
alarm from the alarm zone into the normal operating range. The actual float
position above or below the alarm switch is precisely indicated.
Figure 3 displays low and high flow alarm settings.
The intrinsically safe power supply/relay unit from Pepperl+Fuchs (KF series)
gives the customer the ability to switch over the alarm function in "Normally
open" or "Normally closed".
Figure 5-2: Standard bistable alarm
Figure 5-3: Bistable low and high flow alarm
Figure 5-1: Alarm settings
Section 5: Alarm Settings
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP

Disassembly and Cleaning
It is recommended the user periodically inspect the tube and float, and
clean if necessary. Dirt or foreign materials adhering to the tube and float
may cause inaccuracy and sticking of the float. The metering tube
(Borosilicate glass) and related parts may be cleaned with any solvent
which does not attack glass. To disassemble use the following procedures:
1.
Remove the front shield.
2.
Loosen the jack screw by turning it counterclockwise with a 5/32” hex
wrench. The tube may now be removed from the meter housing.
3.
On meter sizes 1 through 6, the tube, float and float stops may be
cleaned as an assembly or may be diassembled for cleaning. Using a
small hook, remove either Teflon® float stop from the metering tube and
remove the float. Be careful not to chip the tube.
4.
Packing seats may now be removed.
5.
The needle control valve assembly may be removed by turning the valve
body counterclockwise. The valve seat, stem and packing then may be
removed easily from the valve body for cleaning or replacement.
Section 6: Maintenance
Telefone: (11) 5035-0920 - E-mail: [email protected]
Av. Dr. Lino de Moraes Leme, 1.094 - Vila Paulista - São Paulo/SP
This manual suits for next models
2
Table of contents
Other Contech Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

DIALAB
DIALAB DTN-410 K user manual

Siemens
Siemens SITRANS FC430 operating instructions

Tektronix
Tektronix TriMode P7700 Series Installation and safety instructions

ADA INSTRUMENTS
ADA INSTRUMENTS ProDigit MICRO operating manual

DeLaval
DeLaval MM15 Preventive maintenance

Aimco
Aimco AUDITOR ATDA-8000 Series user guide