DiaDent D.LUX+ User manual



Table of Contents
Table of Contents
1. Overview
2. Safety Instructions Device Storage
3. Package Contents
4. Product Description
5. Components Names
6. Operation Method
6.1 Preparation before use
6.2 Instruction for use and operation sequence
7. Hand-Piece Function and Set-up Method
7.1 Hand-piece function
7.2 Feature settings
7.3 Battery replacement
8. Charger Features and Cautions
8.1 Charger functions
8.2 Precautions during charging
9. Cleaning and Maintenance
9.1 Cleaning and maintenance
10. Disposal
11. Item Codes
12. Troubleshooting
13. Warranty
14. Technical Specification
15. Symbols
FCC Warning and EMC Information
Contact Information
5
6
8
9
9
10
11
11
12
12
12
12
13
14
15
15
16
16
17
18
19~21
21
ENGLISH

Overview
Overview

1. If the Light Probe is cracked or damaged, stop using it immediately. Being a plastic product
that may enter into mouth, the Light Probe can be hazardous.
2. Do not subject the handpieces to strong shocks or vibrations. Make sure that the Light Probe
is inserted securely and has not been pulled out before use (otherwise, it may cause
malfunctions or low light levels).
3. Never look directly at the output light. Otherwise, visual impairment may result.
4. As irradiating with strong light (in MAX mode) has a risk of burns, be careful when using it
and never use it continuously.
2. Safety Instructions
Safety Instructions
Safety Instructions
To reduce the risk of injury, always follow manufacturer’s cautions listed below as well as
other warnings and precautions throughout this manual.
CAUTION
Caution(s) indicates information, which could result in serious injury to the patient and/or the
operator and damage to the device, if not followed properly.
WARNING
1. The electromagnetic waves or other medical equipment in hospitals and clinics can cause
malfunction of this device.
2. Do not use this device on patients with pacemakers.
3. To reduce the risk of electric shock, do not try to open, disassemble or repair the hand-piece,
battery pack, or the charger base. Use only the charger and battery pack supplied by DiaDent.
4. Do not modify this equipment without authorization of the manufacturer.
5. Do not use this device in the presence of explosive and flammable materials to prevent
a possible explosion.
6. Use components/parts supplied by DiaDent only. Use of components/parts
manufactured by any other brand may cause damage or loss of performance and will thus void
the warranty.
GENERAL CAUTION
1.This device should be used exclusively by trained and certified dentists or endodontic specialists
for designated purposes described in this manual. If the device is used by any other users or
for purposes other than the designated purposes in this manual, the manufacturer bears
no liability due to injury from the incorrect usage.
2. Practice the technique and be familiar with the device before using on patients.
3. To prevent any possible injury, do not allow patients to handle or touch this device.
4. This product is not waterproof. Do not immerse the hand-piece, battery pack and charger base
in liquid of any kind.
5. For safe use, do not operate continuously in any mode and follow the specified duty cycle.
Application Notes

This product has been tested according to the IEC 60601-1-2 and is designed to provide reasonable
protection against harmful interference to general operation and installation of medical devices.
The device also has been designed and modified for professional health care enviroment
This product generates, uses, and can radiate radiofrequency energy and thus cause interference
to other devices nearby. In case of harmful interference, user is recommended to try one or more
of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation distance between the interference-causing equipment and receiver.
- Consult a local dealer or the manufacturer.
- The battery of the device has passed both battery transportation test (UN DOT.38.3) and battery
stability test (IEC 62133)
Device Storage
ⓛ Store the device in a safe place with the following conditions to avoid device malfunction
or fire:
- Dry place where there is no moisture.
- Avoid leaving the device in a place with direct sunlight, or air with dust, salt, or sulfur.
- Atmospheric Pressure 70 ~ 106kPa / Temperature -10 ~ 40°C / Humidity 30 ~ 85%
② The device should be cleaned after each use.
③ Should there be any defects or malfunctions, stop using the device immediately and call
the customer service team.
④ Please turn off the device when it is not in use. In case the battery pack will not be used
for an extended period of time, unplug the adapter and make sure the battery pack is NOT
at a completely discharged state before storing the device.
Safety Instructions
5. Use disposable sheath to prevent contamination of the Light Probe and dispose it immediately
after use, as it is for single use only.
6. When moving the device, remove the hand-piece from the charger and move the hand-piece
with the light protector and the charger separately and safely. The light protector should be
attached to the hand-piece so it will not roll and lose balance when placed on a flat surface.
WARNING :
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the , including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Safety Instructions
If the device is not operated properly, it might be a malfunction due to static electricity. In this case,
press the power button for more than 6 seconds to turn it off and on.

Package Contents
Package Contents
3. Package Contents
Before removing the device from the box, check the items listed below and make sure you have
everything. If any item is missing or damaged, contact your sales representative immediately.
Contents:
•1x D-Lux+ Hand-piece
•1x D-Lux+ Charger
•1x Light Probe
•1x Light Protector
•1x C-Battery (Included in the hand-piece)
•1x Adapter
•1x Power Cord (1.5m or 1.8m)
•1x Disposable Sheaths (200 ea/Box)
•1x User Manual
Adapter Information
Manufacturer : EDAC Power Electronics Co., Ltd.
Model : EM1024A2
Input : 100-240V~, 1.0-0.5A, 50-60Hz
Output : 6V 2A
WARNING
Use of accessories, transducers and cables other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.

Product Description
, the cordless LED visible light curing unit, has a stylish design and excellent
cooling performance. Having overcome the overheating problem (which may deteriorate LED
performance and even damage the unit) caused while using the visible light curing unit, it is
capable of reliably producing high power. With the application of dual wavelength LED, it is
capable of polymerizing various composite resins. Ergonomic design makes it convenient to
use. Built-in rechargeable battery, wireless battery charging method and absence of contact
terminal between the handpiece and the charger prevent the recharging part from contamination
and contact failure.
4. Product Description
5. Components Names
Various built-in protection circuits (overcharge and over-discharge protection circuits)
enable the device to be used safely. Battery level shown on the display window allows
operator to check the battery level status and to maintain a constant light output.
<Charger>
Charging / Intensity Indicator
(Charging / Light Intensity Status)
Built-in LED
Intensity Sensor
<Hand-piece>
Light Probe
Light Protector
Operation
Button
Display Window
Mode / Power
Button
Product Description

Operation Method
Operation Method
6.2 Instruction for use and operation sequence
1) Connect the adapter (power supply) of the charger to the power cord and then insert the
plug into the outlet.
2) Connect the charging jack to the charger.
3) Mount the handpiece on the charger to begin charging. The charging indicator on the charger
turns orange when the unit is charging and turns green when it is fully charged. It takes about
8 hours for the hand-piece to fully charge for the first time.
4) Insert the Light Probe into the hand-piece.
5) Attach the Light Protector to the hand-piece by pushing the Light Protector all the way down
the hand-piece until it cannot move down further.
6) Before using the unit, cover the Light Probe with the Disposable Sheath.
7) Press and hold the Mode (Power) button for about 2 seconds to turn on the power.
8) Press the Mode (Power) button for about 1 second to change the mode in the order of STD,
SFT, HIG, ORT and MAX in the display window. Press the Mode (Power) button briefly for
less than 1 second to change the set-up time
(For Mode and Time, see Section7.1 Hand-piece Function P.11)
9) Press the Operation button to perform light irradiation for the set time and press the button
once again to stop the light irradiation. The light irradiation progress time is displayed
numerically in the display window.
6.1 Preparation before Use
1) Be sure to read the instruction manual thoroughly before use.
2) Make sure the device is not damaged or defective before use.
3) Check the battery status before use.
4) Check each component such as hand-piece, battery, light probe, etc. before use.
5) Make sure all components are cleaned and disinfected before use.
6) Cover the Light Probe with disposable sheath before use.
6. Operation Method
* This is the Light Probe mounting part. It is removable.
Never look at the light output section without Light Protector.
It may cause visual impairment.
<Display> <Light Probe mounting part>
Light Probe
Battery Indicator Mode Operation Time (S)
The light probe is replaceable. To replace it, pull it out from the hand-piece.
Replace the Light Probe by inserting it into the hand-piece.
* Replace the Light Probe
*Make sure the light probe is as close to the area to be treated as possible. Be careful not to
shine the light onto close-by teeth and gums.

7.1 Hand-piece function
1) Press the Mode (Power) button for about 1 second to select the mode (5 different modes).
2) STD (Standard) mode irradiates with power of 900 mW/㎠ (±10%) for the selected time
(5, 10, 15 or 20 seconds).
3) SFT (Soft Start) mode starts irradiating slowly and gradually reaches a light intensity of
1,300 mW/㎠ (±10%) for the selected time (10 or 15 seconds).
4) HIG (High Power) mode irradiates with power of 1,300 mW/㎠ (±10%) for the selected time
(5, 10 or 15 seconds).
5) ORT (Orthodontic) mode irradiates with power of 1,800 mW/㎠ (±10%) for the selected time
(3, 4, 5 seconds or P). P (Pulse) : When the light is in ORT P (Pulse Mode), the light turns on
for 2 seconds and turns off for 1 second eight times.
6) MAX (Max Power) mode irradiates with power of 2,400 mW/㎠ (±10%) for the selected time
(1, 2 or 3 seconds).
7) Pressing the Operation Button during light irradiation stops the light irradiation immediately.
8) A flashing battery indicator shown on the Display Window while the unit is in use means that
it is running out of capacity and the low battery level may affect performance, limiting normal
operation. In order to restore normal function, it must be recharged immediately.
9) The device enters sleep mode after 5 minutes of inactivity with a buzzer sound and is
automatically turned off after 60 minutes.
7. Hand-piece Function and Set-up Method
Hand-piece Function and Set-up Method
Hand-piece Function and Set-up Method
10) Make sure to discard the Disposable Sheath after use.
11) After using the handpiece, please press and hold the Mode (Power) button for more than
2 seconds to turn off the power.
For safety, do not continuously operate in any mode and follow the specified duty cycle.

8. Charger Features and Cautions
2) Pressing and holding the Mode (Power) button for about 1 second changes the mode in
the order of STD, SFT, HIG, ORT and MAX in the display window. (Refer to the description
under Section7.1 Hand-piece Function P.11)
3) Press the Operation Button to perform light irradiation for the set time and press the button
once again to stop the light irradiation. The light irradiation progress time is displayed
numerically on the display window.
4) To stop the light while the light is in operation, press the Operation Button.
8.1 Charger functions
1) The charger consists of a Charging/Intensity Indicator, and a LED Intensity Sensor capable
of measuring the status of the light intensity.
2) Even with the charger connected to the power supply, the Charging/Intensity Indicator will
not be lit if the hand-piece has not been mounted (hereafter, the Charging/Intensity Indicator
shall be indicated as just the Indicator)
3) When the hand-piece is mounted on the charger, the indicator turns orange when charging
and turns green when the battery is fully charged.
4) If the Indicator is not lit, it means the hand-piece is not properly positioned in the charger
base. In this case, the hand-piece should be repositioned and reseated into the charger base.
1) If the hand-piece display window is not turned on, press the Power Button to turn on the
power.
7.2 Feature settings
7.3 Battery replacement
The battery pack comes inside the hand-piece.
- Do not touch the patient when changing the battery.
1) Pick out and remove the two gray silicone plugs with an appropriate tool. The silicone plugs
are located on back of the hand-piece on the battery cover.
2) Remove the screw on the battery cover at the back of the hand-piece by turning it
counterclockwise with a + screwdriver.
3) Remove the battery cover by sliding it down and out.
4) Remove the battery by pulling it out.
5) Insert the new battery pack in the proper direction (the terminal pin has to face upwards).
6) Cover the battery cover.
7) Tighten the screws in clockwise using a screwdriver
8) Insert the two gray silicone plugs back into the screw hole.
Hand-piece Function and Set-up Method
Hand-piece Function and Set-up Method

5) If the Indicator is not turned on after the hand-piece is positioned properly in the charger,
disconnect the adapter from the charger for a few seconds and then try to charge again.
If the charger still does not work, please contact your representative immediately. It might
be the problem of charger or battery pack.
6) The Indicator displayed in green means the charging is complete. After charging is complete,
the Indicator may flash in orange briefly to check the charging status.
7) During charging, the bottom part of the main unit and the top part of the charger may
generate heat, which is normal.
8) If the hand-piece is removed while charging, the charging indicator may turn off after a few
seconds, depending on the residual amount of magnetic field.
9) If an error occurs while charging (for example, indicator lights up in green at the beginning
of charging, etc.), remove the adapter‘s charging jack from the charger and then reconnect it.
10) By irradiating the light on the Light Intensity Measurement Unit, light intensity can be
verified whether it is within the normal range on the Indicator. Measure the light intensity
in STD mode with the adapter connected. See below for LED color indications:
Green : Normal Orange : Shortage
CAUTION
Measurement must be done within 5 seconds in STD mode. If the light intensity is measured
in other modes or over 5 seconds, it could cause serious damage to the sensor and may
cause malfunction due to heat. In severe case, the sensor part might be broken.
8.2 Precautions during charging
1) Please use only the power adapter supplied by DiaDent Group International. Using a third-
party adapter may result in the internal parts being damaged, causing malfunction or fire.
2) This charger adopts the magnetic induction wireless charging system. Please remove any
foreign substance, magnets, metal plates, and other objects from the silicon wireless
charging pad. The product may be damaged due to heat generation/overheating and this
may also cause a fire.
3) Magnets or metals being near the charger may cause operational errors. Keep such materials
as far away from the charger as possible.
4) Please keep the silicone wireless charging pad and the hand-piece as close as possible so
that there may not be any gap between them while maintaining their vertical status.
(Charging time may be extended if they are twisted or gap exists between them.)
5) Do not drop the battery or subject it to shock. It may cause leakage and damage to
the battery, and may also cause a fire.
6) In case the battery level is low (or is fully depleted), a lot of heat will be generated when
unit is recharging. This is normal and is caused by the large amount of charging current.
Charger Features and Cautions
Charger Features and Cautions

9.1. Cleaning and Maintenance
1) Cleaning the Hand-piece
- Remove the hand-piece from the charger.
- Clean the outer surface of the hand-piece using a soft cloth moistened with a pH-neutral
surface cleaner.
- Do not submerge the hand-piece in disinfectant liquids of any kind.
2) Cleaning the Charger
- Remove the hand-piece from the charger.
- Disconnect the charger from the power adapter and power supply.
- Clean the outer surface of the charger using a soft cloth moistened with a pH-neutral surface
cleaner.
- Do not submerge the charger in disinfectant liquids of any kind.
3) Light Probe
- Remove the Light Probe from the hand-piece.
- Remove all foreign substance from the Light Probe and wipe its surface with soft cloth
moistened with a pH-neutral cleaning solution. (Do Not Autoclave)
- Always clean it after use to maintain cleanliness.
9. Cleaning and Maintenance
Cleaning and Maintenance
Cleaning and Maintenance
1) Before Use
- Inspect any damages on hand-piece or parts
- Check if the light probe is broken or contaminated. If the light output part is contaminated,
clean it thoroughly. (Contamination is the reason for power loss)
2) Monthly Inspection
- Check and tighten all the loose connection components.
- Separate light probe from hand-piece and gently clean the inside of the light source and
plastic protector with dry cotton swap. Caution: Don't apply any excessive force to the light
source.
3) Yearly Inspection
- Extensive checkup for the safe and normal function of the device. Inspect all the connection
components on the hand-piece as well as the cable and power cords.
When not in use, disconnect the power adapter and power supply from the charger.
7) If you use an adapter that is not supplied by DiaDent, the buzzer sound may occur while in
use. In this case, please disconnect the adapter immediately from the charger and use the
recommended adapter (provided by DiaDent) to prevent malfunction and failure of
the charger.
8) To disconnect / unplug the electric power, you have to disassemble the power plug from
the adapter. Do not place the adapter in a place where it is difficult to unplug it.

Disposal
11. Item Codes
D-Lux+ Regular KIT
D-Lux+ Hand-piece
D-Lux+ Charger
D-Lux+ Light Probe
D-Lux+ C-battery (Battery Pack)
D-Lux+ Light Protector
D-Lux+ Power Supply (Adapter)
D-Lux+ Power Cord
D-Lux+ Disposable Sheath
4008-1001
4008-1110
4008-1120
4008-1130
4008-1140
4008-1150
4008-1160
4008-1161
4008-1170
ITEM ITEM CODE
1) Hand-piece and Charger disposal
- Separate the battery pack from the hand-piece before disposal. Dispose of electronic parts
as electronic waste. Dispose of all other parts according to the relevant electronic waste
disposal regulations.
2) Battery Pack disposal
- Dispose of the battery pack as chemical waste in accordance with all applicable local
regulations.
- Do not throw the disposed battery pack into any heat/fire.
- To prevent the battery from overheating due to short circuit, dispose the battery in a
discharge state.
- Do not disassemble, open or cut the battery.
- Do not expose the battery to heat or fire.
- Pay attention to the short circuit of the battery.
10. Disposal
4) Light Protector
- Remove the Light Protector from the hand-piece.
- Wipe the surface of the Light Protector with soft cloth moistened with a pH neutral surface
cleansing solution.
Disposal Item Codes
Disposal Item Codes
CAUTION
Potential electromagnetic waves or the effects of other medical devices used in hospitals
may cause the malfunction of the device or adversely affect of other device.

4. Display Window is not
on
•Check the charge status of the battery pack.
•Check if the display window (OLED) is damaged.
5. The output is
abnormally weak
•Make sure the Light Probe is inserted all the way into the hand-piece
•Check if the Light Probe part is contaminated.
•If the performance of the light source (LED) is compromised, the Light
Probe should be replaced.
6. Charger does not work
•Make sure to use the power adapter provided by the manufacturer.
•Make sure the power adapter is properly plugged in.
• Make sure the charger and power adapter are intact and damage free.
•Ensure that the hand-piece is placed properly and firmly on the charger base.
•If the battery, hand-piece, or charger becomes wet, disconnect from
the power cord and dry immediately. Do not use until completely dried.
•Replace the battery pack.
•If the charger still does not work, please contact your representative
immediately.
7. The battery is discharged
quickly
•Lithium-ion battery has a fixed service life, resulting in performance
deterioration over extended periods of use. If problems arise during
its usage, the battery should be replaced with a new one.
•It is strongly recommended to replace the battery every two years.
Suggested here are some common problems you might encounter and how to fix them.
For problems not listed here, please contact your dealer or the A/S center of our head office.
12. Troubleshooting
Troubleshooting
Troubleshooting
•Check the connection between the charger and power adapter.
• Check the connection between the battery and hand-piece.
•Charge the battery pack.
•Check if any button is pressed accidentally.
• Press and hold the power button for at least 7 seconds to turn off
the power and then turn it back on again.
•Remove the battery from the hand-piece and then replace it.
1. Device will not turn on
2. Hand-piece does not
work
•Check that battery is fully charged before use.
•Check the connection between the charger and power adapter.
•Replace the battery pack.
3. The hand-piece operates
abnormally
Problem Solution
13. Warranty
1) The manufacturer provides a one (1) year warranty effective from the date of purchase. Any defects or
errors that are due to faulty manufacturing or material will be repaired or replaced within this warranty
period. Consumables such as light probe, light protector, battery pack, Power cord, adapter, disposable
sheaths are excluded.

Specification
Specification
14. Technical Specification
Product Name
Light Wavelength Range
Light Intensity
Wireless Charger
Battery
Frequency
Modulation Method
Charging Time
Duty Cycle
Operation time per full charge
Applied Part
Applied Part Maximum Tempeature
Dimensions
Power Supply
(Medical Adapter)
Weight
Hand-piece
Light Probe
Charger
Hand-piece
Light Probe
Light Protector
Input
Output
Charger
Operating
Conditions
Transportation
and Storage
Conditions
Ambient Pressure
Relative Humidity
Temperature ( C)
O
Ambient Pressure
Relative Humidity
Temperature ( C)
O
2) To obtain warranty coverage and to return the device to the manufacturer’s service facility, please
contact your sales representative.
3) Products returned for warranty coverage should be securely packed with all components and shipped
with postage prepaid. DiaDent does not accept shipments sent collect. DiaDent will be responsible for
the shipping of the replacement or repaired product.
4) Manufacturer disclaims liability and is not responsible for repair or replacement of product in the
case of:
•Noncompliance with this instruction manual
•Operating and/or handling the device without care
•Damage caused by natural disasters
•Incorrect diagnosis or poor result due to operator error
•Device damages resulting from repairs or modifications not approved/serviced by the manufacturer
30 ~ 85 %
10 ~ 30 ℃
70 ~ 106 (KPa)
30 ~ 75 %
-10 ~ 40 ℃
70 ~ 106 (KPa)
Li-ion 3.6V 2600mAh/Rechargeable
6V 2A
D-Lux+
405nm, 460nm
MAX 2,400mW/㎠
Approx 8 hours
Light Probe
MAX 3 sec x 700 times over
201 x 33 x 44mm (L x W x H)
77 x 15 x 15mm (L x W x H)
124 x 100 x 72mm (L x W x H)
119g
17g
10g
215g
100 ~ 240V~ , 50~60 Hz, 50VA
60 ℃
Operation Time : for each mode
Break Time : at least 60 sec.
WPC 1.2 Qi-compliant device (Dual-channel
Amplitude Shift Keying (ASK) demodulation)
110~205kHz
42 dB㎂/m
Max. Transmission Power

15. Symbols
Symbols
Solution
Symbol Description Symbol Description
Type B Applied Part Class II equipment
Do not Reuse
Subject to electrical/electronic
equipment waste regulations
Sitting Prohibited Temperature Limitation
Humidity Limitation Atmospheric Pressure Limitation
Symbol Description Symbol Description
Cautions indicate which, if not observed,
could result in damage to the user and/or
device.
Direct current
CE certificate: Product conformance with
the applicable European Union Directives
Authorized EU Representative
Manufacturer
Serial No.
Catalog Number
Manufacture Date
See instruction manual
Caution, hot Surface
Keep Dry Be careful to prevent burns
Refer to accompanying Documents
Refer Servicing to
Qualified personal only
Symbols
REF
The Restriction of the use of
certain Hazardous Substances
in electrical and electronic equipment

FCC Warning
FCC Warning
Caution: Any changes or modifications to this device not expressly approved by the party responsible
for compliance could void your authority to operate the equipment.
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.
This equipment should be installed and operated with minimum 20 cm between the radiator and your
body. This transmitter must not be collocated or operating in conjunction with any other antenna or
transmitter unless authorized to do so by the FCC.
■ IC Statements
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to
the following two conditions: (1) This device may not cause interference; and (2) This device must
accept any interference, including interference that may cause undesired operation of the device.
RF Exposure
The antenna (or antennas) must be installed so as to maintain at all times a distance minimum of
at least 20 cm between the radiation source (antenna) and any individual. This device may not be
installed or used in conjunction with any other antenna or transmitter.
■ Hereby, DiaDent Group International declares that our (RE type : D-Lux+ wireless charger) is
in compliance with RED 2014/53/EU. The full text of the EU DoC is available at the following
internet address http://www.diadenteurope.com/
■ FCC Warning
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or more of the
following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
•Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.

EMC (Electromagnetic Compatibility) Information
AC Mains
AC Mains
Immunity to
Proximity Fields
from RF wireless
Communications
Equipment
EN 61000-4-3:2006+A2:2010
IEC 61000-4-3:2006+A2:2010 Enclosure
Table 9 in
IEC 60601-1-2:
2014
Electrical Fast Transient/
Burst Immunity
EN 61000-4-4:2012
IEC 61000-4-4:2012
± 2 kV, 100 kHz repetition
frequency
Surge Immunity EN 61000-4-5:2014
IEC 61000-4-5:2014
Line to Line
± 0.5 kV, ± 1 kV
Line to Ground
± 0.5 kV, ± 1 kV, ± 2 kV
AC Mains
Immunity to
Conducted
Disturbances
Induced by RF fields
EN 61000-4-6:2014
IEC 61000-4-6:2013
EN 61000-4-8:2010
IEC 61000-4-8:2009
3 V
0.15-80 MHz
6 V in ISM bands
Between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Phenomenon Basic EMC
standard or testmethod Port tested Test level/requirement
Mains terminal
disturbance voltage
EN 55011:2016+A1:2017
CISPR11:2015+A1:2016 AC Mains Group1, Class A
Radiated disturbance EN 55011:2016+A1:2017
CISPR11:2015+A1:2016 Enclosure Group1, Class A
Harmonic Current
Emission
EN 61000-3-2:2014
IEC 61000-3-2:2014 AC Mains Class A
Voltage change,
Voltage fluctuations
and Flicker Emission
EN 61000-3-3:2013
IEC 61000-3-3:2013 AC Mains
Pst: 1
Plt: 0.65
Tmax:0.5
dmax: 4%
dc: 3.3%
Electrostatic Discharge
Immunity
EN 61000-4-2:2009
IEC 61000-4-2:2008 Enclosure
± 8 kV/Contact
± 2, ± 4, ± 8,
± 15 kV/Air
Radiated RF
Electromagnetic
Field Immunity
EN 61000-4-3:2006+A2:2010
IEC 61000-4-3:2006+A2:2010 Enclosure
3 V/m
80 MHz-2.7 GHz
80% AM at 1 kHz
EMC Information
EMC Information

Phenomenon Basic EMC
standard or testmethod Port tested Test level/requirement
EMC Information
NOTE :
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
EMC (Electromagnetic Compatibility) Information
EN 61000-4-11:2004
IEC 61000-4-11:2004
AC Mains
AC Mains
Power Frequency
Magnetic Field
Immunity
EN 61000-4-11:2004
IEC 61000-4-11:2004 Enclosure 30 A/m
50 Hz & 60 Hz
Voltage dips EN 61000-4-11:2004
IEC 61000-4-11:2004
0 % UT : 0.5 cycle
At 0°,45°,90°,135°,180°
225°,270°and 315°
0 % UT; 1 cycle
and
70 % UT; 25/30
cycles
Single phase: at 0°
Voltage
interruptions
0 % UT; 250/300 cycle
EMC Information
Table of contents
Other DiaDent Dental Equipment manuals
Popular Dental Equipment manuals by other brands

Nevadent
Nevadent SMD 24 A1 Operating instructions and safety instructions

OrthoAccel Technologies
OrthoAccel Technologies AcceleDent OPTIMA Directions for use

Gendex
Gendex GXS-700 user manual

bredent
bredent Polylux 2 Instructions for use

Poldent
Poldent Endostar Instructions for use

Ancar
Ancar SD-300 Touch Expert user manual