Product features and function
3. Spec
- To setup the spec. for the FLAT mode.
1. Go Back : To exit the ‘Spec’ menu.
2. Tolerance : To setup the tolerance of the tilted direction display line.
3. Model : 10 memory spaces are given to save the users’ specification
setting. And user could recall them later.
4. X : The width of the setting points.
5. Y : The length of the setting points.
6. Pitch : The pitch of the screwed legs at the home appliances or other
machines.
7. Angle : It is set by 360 deg. normally.
4. Calibration
- To calibrate the absolute zero. Please follow the steps carefully.
If not, it can not be guaranteed the angle data.
Fig.1
Fig.2
Step 1 [XY Calibration]
Put the inclinometer on the surface as like Fig.1 and choose
the ‘XY 1st’. Then you might see ‘XY 1st Please Touch Screen’.
If the inclinometer is settled well on the surface, touch the screen
and wait until the inclinometer gathers the data and averages them.
Put the inclinometer as like Fig.2 and choose the ‘XY Rev’.
And touch the screen if the inclinometer is settled.
Step 2 [X Calibration]
Put the inclinometer on the surface as like Fig.3 and choose
the ‘X 1st’. Then you might see ‘X 1st Please Touch Screen’.
If the inclinometer is settled well on the surface, touch the screen.
Put the inclinometer as like Fig.4 and choose the ‘X Rev’.
And touch the screen if the inclinometer is settled.
Fig.3
Fig.4
* [Y Calibration]
User doesn’t have to do the Y calibration because the Y calibration
is done automatically if the X calibration is done.
* [Default]
If the calibration is not done well, user can choose the ‘default’.
Then IM-2DT will recall the zero setting values at the factory.