DOSER Messtechnik DS5-AL User manual

Moisture meter type DS5AL
for absolute moisture measurement in waste paper balls,
wooden chips, pellets, etc.
MOISTURE METER
STICKING ELECTRODE
page 1version 072009
DS5-AL Operating instruction
Duly use:
The electronic moisture meter DS5-AL is used to quickly
determine the moisture in materials
material temperature range: +5 - +40°C
working temperature range: +5 - +40°C
storage temperature range: -20 - +70°C
. An absolute average
moisture content is measured.
materials: wastepaper balls, wooden chips, pellets
measuring range: 0 - 100 % water content

3
2
1
•field lines
‚material to be measured
ƒplate capacitor
Ser. Nr.
029871
made in germany
DS5
MODEMODE
OKOK
ON
OFF
ON
OFF
Auswahl der
Materialgruppe: 1
page 2
Moisture meter type DS5-AL
The measuring electrode of the meter is pushed into the paper or
cardboard piles and a high frequented field is able to pass through the
material. Taking in account the material setting, the percentage water
content is determined. The result can be read off immediately.
Method of operation:
Measuring principle:
The meter works in accordance with the principle of an opened plate
capacitor. The capacity of the capacitor depends on the material-
(dielectric)-constant of the material in between the plates. Compared with
air (e = 1) , for example water has a very high dielectric-constant (e = 80).
rr
The water content of a wet material can therefore be determined by
determining the dielectric constant of this material.
Safety tips:
- follow the operating instructions
- only use the meter as directed (see page
- keep the meter away from live and current carrying parts
- keep the meter dry
- avoid impacts
- protect the meter from heat
- keep the meter dry and try to prevent dirt from entering the case
- protect the meter from electrostatic discharge.
- the meter must be repaired or serviced only by qualified specialists
Damages caused by failure to follow the above mentioned Safety tips
are not covered by the warranty!
2)
•
‚
‚
ƒ
Measurement:
Insert the sensor between into the material (1).
The distance to the border (2)
3) should be at least 15 cm.
The result is displayed immediately
has to be at least 5 cm.
The sticking depth (
Turning on:
- the sensor must be dry and clean
- push "ON/OFF" button, until it shows "on” in the display and hold the
meter in the air so that the automatic zero point measurement /
correction can be carried out correctly
- release "ON/OFF" button the meter is ready for use
DS5
DS5
ONON
MODEMODE
MODEMODE
OKOK OKOK
ON
OFF
ON
OFF ON
OFF
ON
OFF
6.4%6.4%
Material: 4
Material adjustment:
- select the best fitting material number
- turn the meter on and then push the “MODE” button
- with the arrow buttons the corresponding material number can be
adjusted
- finally the "OK" button has to be pushed

DS3DS3
code: 00code: 00
MODEMODE
OKOK
ON
OFF
ON
OFF
Storing of Moisture Values:
If the storage is activated, by pushing the OK-button up to 250 values can
be stored. If the storage parameter is set to “on single” always one value is
stored, if the parameter is set to “on auto” every value > 0 is stored
automatically until the storage is full or the OK button is pressed again. The
number of measurements per second (mps) can be adjusted between 1 and
10. The storage procedure will be signed with an “stor” on the botom line at
the right side.
Restorage of the Stored Values:
By pushing the "OK" and the “arrow down” button, the stored values can
be restored. The result is shown in 2 colums. On the left average
value(AV), Minimum value (MI) and maximum value(MA) and on the right
side the stored values, they can be scrolled with the arrow buttons.
With our optional PC-interface cable and software DMI-Controller the
stored values can be transfered to a PC for storing them into a file or
showing them on a graph.
DS3
MAT
OK
ON
OFF
SPEICHER:
MA: 8,4
AV: 6,3
MI: 6,1
01: 7,8
02: 7,0
03: 6,1
04: 8,4
DS3
MAT
OK
ON
OFF
6.4%6.4%
F=
Holzgruppe: 4
HOLD
DS3
MODEMODE
OKOK
ON
OFF
ON
OFF
DOSER MESSTECHNIK
87629 Füssen
Tel: 08362 9159402
Fax: 08362 9159407
Cancellation of the Stored Values:
After changing the material adjustment or with code 14, the stored values
will be cancelled, which is shown as “delete storage” on the display.
page 3
Moisture meter type DS5-AL
Parameter Settings:
With code numbers parameter settings can be changed. Press both arrow
buttons, the display shows “code:00“, adjust the required code with the
arrow buttons and confirm it with OK. The following parameters are
changeable:
code 11: storage adjustments (OFF, ON single, ON auto)
ON single: single values will be stored by pressing the
OK-button
ON auto: All values > 0 will be stored automatically after
pressing OK until the storage is filled up or until the OK-button
is pressed again.
code 12: number of measurements per second (mps)
range 0 - 10
code 13: number of measurements until automatic switching off (loops)
range 0 - 999, loops = 0: no automatic swiching off
code 14: delete storage, through changing the material setting the
storage will be deleted too.
code 15: select language
code 21: attenuation ( brake), range 0 - 99
code 22: automatic maximum value measuring (automax)
no automatic maximum measurement (float)
code 95: toggle between material moisture and calibration
measurements (display of input values)
code 97: starts the calibration procedure
Turn Off the Meter:
- push "ON/OFF" button until the display shows the service address, this
keeps in the display as long as the “ON/OFF”-button is pressed
- release "ON/OFF" button: the meter turns off
If no measurement is carried out for an adjustable time,
the meter turns off automatically.

cal. low:
cal. high:
calibration:
NP= 028
NP= 028
NP= 028<
UM= 031
UM= 031
UM= 031
F= 248
F= 223<
F= 000
MV= 202
MV= 199
MV= 000
Basic Calibration:
start calibration procedure with code 97
1. zero point
The zero point only can be tested, not adjusted. On the display the zero
point (NP) and the actual measuring value (UM) are shown. At the test
the sensors have to be clean and dry! The instrument has to be hold free
into the air. Than wait till NP and UM are nearly equal. The value should
be between 10 and 390. Is UM <1 or > 400, then no zero point
calculation is working, in this case the instrument needs to be repaired.
With the OK-button to the next calibtration step.
2. measuring in test module PE50R at high amplification
As shown on the picture, the instrument has to be sticked into the test
module.With the arrow buttons the factor (F) can be changed till the
display shows MV=200.
With OK-button to the next step.
3. measuring in test module PE50R at low amplification
With the arrow buttons the factor (F) can be changed till the display
shows MV=200.
With OK-button the changed factors will be stored.
The factors can be changed in the range 150 - 255. If this is not enough, the
instrument needs to be repaired.
page 4
Moisture meter type DS5-AL
Oven Drying Method:
The oven drying method is the most accurate way to measure the material
moisture in paper NORM EN20 287 or DIN ISO 287or wood DIN 52183.
We recommend this for testing and calibrating of our moisture meters.
Short description:
- for measuring the weights we recommend a balance with an measuring
range of 200g and an accuracy of 0,01g
- For drying you need an oven with adjustable temperatures of 105°C
- Take a sample from the waste paper, avoid border parts. The sample
should be at least 20g
- It is very important to take the first weight of the sample immediately, as
air humidity may change the moisture content.
Name of the first weight: wet weight (WW)
- The probe must be dried in the oven until the weight is constant,
the maximum drying temperature: 105 °C (ISO 287),
- the name of the dry weight is DW
- moisture calculation:
DIN 52183: MOISTURE = 100% * (WW - DW) / DW
DIN ISO 287: MOISTURE = 100% * (WW - DW) / WW

Customer Specific Linearisation Curves:
It is possible to store 10 special linearisation curves with up to 8 free
programmable points each. The values in this storage will stay in the
memory during replacement of the battery. The adjustemt of the calibration
points can be done with the PC-software DMI-Controller, or they can be
changed through the kexpad as decribed below.
Moisture meter type DS5-AL
page 5
The microcontroller in the meter takes the analog measurement value with
automatic zero point correction.
The resulting value is our input value.
You can display the input value with code 95. 1:0000
According to a rectangular coordinate system, the input are assign in the X-
direction
and the moisture result are assign in the Y-direction.
According to this, X-values are input values ,
Y-values are the assigned moisture values.
X-values (input values) can appear in the area 0-4095,
this input values ( can be dedicated to moisture values (Y-values)
The moisure values have to be written in %-values * 10
(for example: for the moisuture value 12,3 % ==> Y-value = 123).
For chaging the calibration points you need to put in code 40 to 49 for
curve 0 to 9 (e.g. code 43 for curve no. 3).
The values signed with “<” can be changed with the arrow keys.
Confirm the changed value with “ok”, then you can change the next value.
Calibration sample:
Special curve 3:
1: 0000 0000
2: 0043< 0000 till the input of 43 the display shows 0,0%
3: 0200 0152 200 is the calibration value with our testmodule, with
the testmodule the instrument will show 15,”%
4: 1430 0500 500 =50% is the maximum value
5: 5000 0500 input 1430 to 5000 the display stays at 50,0%
6: 0000 0000 points 6 to 10 not used
Y-values)
7: 0000 0000
8: 0000 0000
9: 0000 0000
10: 0000 0000
The maximum value for the input value is bewteen 4000 and 5000,
depending on the basic calibration factor.
With code 31, the defoult calibration curves can be selected
With code 30, the special curves can be selected.
Input 1:
0000
Special curve 3:
1: 0000 0000
3: 0200 0152
2: 0043< 0000

DOSER Messtechnik GmbH & Co.KG
Kemptener Str. 73
87629 Füssen
Germany
T: +49 8362 9159 402
F: +49 8362 9159 407
page 6
Moisture meter type DS5-AL
Optional extras:
- test module type PE50R for testing and calibrating
- PC-Software DMI-Controller to adjust parameter settings and to transfer
measured values for analysys and documentation
- special calibration of the moisture meter
Our operating instructions are intended for guidance and to provide information on our products and their uses.
They should not be taken to imply special characteristics or suitability for any specific purpose, other than those
stated.
We constantly work to improve our products and reserve the right to alter our products and operating instructions
without advanced notification.
Battery:
The meter works with a commercial 9V block battery. The capacity is
continuously controlled. If the battery is running low,
the display shows “BAT”
Change battery:
- open the battery box for example with a small screw driver
- take out the battery
- insert new battery, observing the correct polarity
- close the battery box
Attention!
In accordance with battery legislation, all used batteries
must be disposed off in special battery collecting bins.
The disposal of old or used batteries as part of normal
waste is not allowed!
Ser. Nr.
009871
made in germany
DS3DS3
6.4%6.4%
MODEMODE
OK
ON
OFF
M=
material: 1
BAT
Table of contents