Dynatech DS6100 Service manual

+

UES
ae
dns
cot
oR
à
DS6100
Patient
Simulator
dh
DNI
QD
N
EVA
DA
PO
BOX
1925
(89702)
IN
C.
1
2000
ARROWHEAD
DRIVE
CARSON
CITY,
NEVADA
89706
Formerly
PHONE:
702.883.3400
Dynatech
Nevada
Inc.
FAX:
702.883.9541
Operating
and
Service
Manual
%
DYNATECH
NEVADA
World
Class
People
—
World
Class
Products

PRESTI
ANI
Copyright:
Dynatech
Nevada,
Inc.
agrees
to
a
limited
copyright
release
that
allows
you
to
reproduce
manuals
and
other
printed
materials
for
use
in
service
training
programs
and
other
technical
publications.
If
you
would
like
other
reproductions
or
distributions,
submit
a
written
request
to
Dynatech
Nevada
Inc.
Warranty
and
Product
Support:
This
instrument
was
thoroughly
tested
and
found
to
meet
our
manufacturing
specifications
when
it
was
shipped
from
the
factory.
Calibration
measurements
are
traceable
to
the
National
Institute
of
Standards
and
Technology
(NIST).
New
Instrument
Warranty
1
year
on
parts
and
labor
Physical
damage
is
not
covered
Repair
Instrument
Warranty
90
days
on
calibration,
labor,
and
parts
replacement
Y
Physical
damage
is
not
covered
i
Dynatech
Nevada
Inc.
warrants
that
all
products
are
to
be
free
from
defects
in
parts
and
labor
as
described
in
the
warranty
policy
unless
a
different
period
is
specified.
If
you
discov-
era
defect,
Dynatech
Nevada
will
either
modify
the
product
to
correct
the
defect,
replace
the
product
with
new
or
factory
reconditioned
products,
or
refund
the
amount
paid
for
the
defective
product.
The
standard
terms
and
conditions
of
sale
are
printed
on
the
reverse
side
of
the
Dynatech
Nevada
invoice.
You
may
obtain
additional
copies
of
the
warranty
and
conditions
of
sale
from
Dynatech
Nevada.
If
you
retum
an
instrument
to
Dynatech
Nevada
Inc.,
we
recommend
United
Parcel Service,
Federal
Express,
or
Air
Parcel
Post.
You
must
receive
prior
authorization
from
Dynatech
Nevada
Inc.
for
C.O.D.
shipments,
otherwise
you
will
pay
transportation
charges.
Intended
Use
Disclaimer:
The
DS6100
is
intended
to
be
used
as
a
simulation
device
only.
Recommended
applications
include
routine
preventive
maintenance
inspections,
sales
demonstrations,
or
training.
The
DS6100
is
not
to
be
used
as
the
primary
calibration
stan-
dard
or
the
certification
test
instrument
for
medical
devices.
Dynatech
Nevada
Inc.
2000
Arrowhead
Drive,
Carson
City,
NV
89706-0403
P.O.
Box
1925,
Carson
City,
Nevada
89702-1925
800-648-7952
702-883-3400
Fax:
702-883-9541
Part
Number:
9508-0213
Revision:
C
Printed
in
the
USA.
Table
of
Contents
INTRODUCTION
..
5
FRONT
PANEL
........
6
OPERATION
.
10
USING
THE
PENDANT
KEYPAD
hrs
10
KEYPADCONTROLS
.…………
20000
11
MODIFTER/CAL
MODE
...........ccccccocsscsscscsccscevssessccsecsesece
LT
SIZE
ADJUSTMENTS:
SPEC
FUNC
54
ue.
18
CALIBRATION
LEVELS:
SPECFUNC78,82,&$4.....................
20
HEART
RATE
ADJUSTMENT:
SPEC
FUNC
53
.....................
21
ARTIFACT
SIMULATION:
SPEC
FUNC
47,
48,
8:
49
............
21
RESPIRATION
.
ee
22
DEFIB
SEQUENCE
..
..
seveso
...
24
PERSONALITY
MODULES
sense
ee
26
INTERACTIVE
IABP
PERSONALITY
MODULES
..............06
28
CARDIAC
OUTPUT
PERSONALITY
MODULES
...................
30
CAPNOGRAPHY
PERSONALITY
MODULE
.........................
32
BP
CABLES
...
...
a
„33
WIRING
CHARTS
…
34
CONNECTOR
INTERFACE
..........ccscssscsssssssscssesseorseessrsses
25
BATTERY
CARE
ee
6
PARTS
LIST/SCHEMATICS
ses
37
Table
of
Figures
Fig
1:
12
Lead
ECG
Amplitudes...6
Fig
8:
Artifact
Modifier.....................
21
Fig
2:
Front
Control
Panel........
8
Fig
9:
Wiring
5uV/VmmHg
monitor
..
34
Fig
3:
Pendant
Keypad.................
9
Fig
10:
Wiring
40uV/WVmmHg
monitor
.
34
Fig
4:
Modifier/Cal
Mode
............
17
Figll:
ECGConnecor..............
35
Fig5:Sizeddjwsi..................
19
Fig
12:
Pressure
Connectors
............
35
Fig
6:
Calibration
Functions
........
20
Fig
13:
Connector
Pin
Assignments
..
35
Fig
7:
Heart
Rate
Adjustment
.......
21

DS
6100
Patient
Simulator
DS
6100
Operator's
Manual
Introduction
DS
6100
The
DS
6100
is
a
battery-operated,
six-channel
patient
simulator
and
defib
trainer
that
simultaneously
simulates
ECG/arrhythmia,
respiration,
as
well
as
arterial,
pulmonary
arterial,
and
right
atrial
pressures.
Standard
waveforms
and
sequences,
selected
via
keypad,
are
stored
in
microcomputer
memory
within
the
simulator.
In
the
standard
defib
training
mode,
you
can
interface
the
DS
6100
with
Arrhythmia
Anne
and
Chris
Clean
manikins.
Among
the
many
optional
personality
modules
available,
you'll find
functions
for
special
purpose
ECG,
Cardiac
Output
Dilution
Curves,
Capnography,
and
IABP.
SAFETY
Levels
of
signals
available
via
the
ouput
connectors
and
the
Personality
Module
connector
do
not
exceed
12
volts,
and,
as
such,
do
not
constitute
a
potential
danger
to
the
operator.
However,
if
calibration
or
other
service
of
the
DS
6100
is
required,
we
recommend
that
only
qualified
service
personnel
be
permitted
to
remove
the
front
panel
of
the
simulator.
DS
6100
Operating
and
Service
Manual
5

Front
Panel
The
following
front
panel
controls
allow
you
to
operate
the
DS
6100.
See
Figure
2
for
locations.
1.
ON/OFF
SWITCH:
Switches
power
on
and
off.
Battery
operated
unit
is
rechargeable
with
the
supplied
charger.
2.
POWER
INDICATOR:
Illuminates
when
power
is
ON.
3.
ECG
SNAP
CONNECTORS:
Outputs
low
level
12
lead
ECG.
Attach
your
standard
patient
ECG
cable
lead
wires
to
these
snap
connectors
to
display
the
simulated
ECG
waveforms
on
your
monitor
screen.
1=+.25
mv
(+/-5%)
|
V1
=-.40
141.0
V2
=-.90
HI
=
+.75
V3
=+.60
aVR
=
-.60
V4
=
+1.0
aVL
=
-.25
V5
=+1.5
aVF
=
+.90
Ve
=+1.0
FIGURE
1:
12
Lead
ECG
Amplitude
NOTE:
To
minimize
60
cycle
artifact,
it
is
important
that
a
reference
electrode
connection
is
used.
For
example,
most
5
lead
ECG
diagnostic
recorders
require
that
a
simulator's
REF
signal
be
connected
to
the
RL
input
of
the
patient
cable.
The
DS
6100
can
also
output
the
ECG
in
a
modified
chest
lead
configuration
(MCL2)
used
for
bedside
monitoring.
Simply
attach
the
LL(+)
red,
RA(-)
white,
and
LA(ref)
patient
cable
snaps
to
the
patient
simulator.
Refer
to
the
operator’s
manual
of
your
ECG
monitor
for
hook-up
instructions.
4,
PRESSURE
OUTPUT
CONNECTORS
Output
for
arterial,
PA,
and
RA
pressures.
Simulates
the
electrical
output
of
the
BP
transducer
that
would
be
used
with
the
patient
monitor.
The
DS
6100
generates
signals
that
are
compatible
with
either
5
or
40
microvolt
transducers.
6
DS
6100
Operating
and
Service
Manual
EFI
SE
Front
Panel
§.
AUX
INPUT/OUTPUT
CONNECTOR:
Cable
interconnect
for
optional
waveforms
such
as
capnography
and
IABP.
6.
DEFIB
CONNECTOR:
Input
for
Arrhythmia
Anne
or
Chris
Clean
manikins
equipped
with
defibrillation
option.
4
7.
MANUAL
DEFIB
SWITCH:
Manually
simulates
defibrillation.
8.
PERSONALITY
MODULE
CONNECTOR:
Input
for
optional
modules.
Used
to
add
numerous
functions
and
waveforms
to
the
standard
set.
9.
KEYPAD
CONNECTOR:
Input
for
hand-held
keypad.
10.
KEYPAD
REST:
Storage
area
for
keypad.
11.
BATTERY
CHARGER
INPUT
CONNECTOR
(Located
on
inner
well)
Jack
used
to
recharge
battery.
Unit
can
be
operated
while
recharging.
DS
6100
Operating
and
Service
Manual
7
>
‘
RQ
RE
so
amet
ce
gg

Front
Panel
©
Tea
DS
6
100
nevacane
Patient
Simulat
2
| |
ㆍ
一
一
一
|
<
ECG
©
3
mーー
of
ART
vi
Gİ
PA
PE
+
№3
4*
9
|
|
一
个
一
RA
~
+
KEYPAD
ČT
İl
vs
>
5
Ve
PERSONALITY
©
MODULE
SIMULATE
6
DFE
DEFIB
N
7
NOTE:
The
Battery
Charger
Input
Connector
(#11)
is
located
on
the
inner
well.
FIGURE
2:
Front
Panel
Controls
8
DS
6100
Operating
and
Service
Manual
Keypad
DYNATECH
NEVADA
inc
(__)
Patient
Simulator
X
r
“
NSR
E
2AVBI
o
A-FU
4
JAVB
E
РМС1
CPLT
VTAC
y
da
FIGURE
3:
Pendant
Keypad
DS
6100
Operating
and
Service
Manual

Operation
USING
THE
PENDANT
KEYPAD
Direct
Functions:
Select
the
rhythm
desired
by
pressing
one
of
15
Direct
Function
keys.
The
following
pages
list
all
of
the
functions
and
code
numbers
available
in
the
standard
simulator
unit.
Example:
To
initiate
a
bradycardia
rhythm,
press
and
release
the
key
labeled
BRAD.
The
simulator
automatically
determines
the
proper
time
in
the
cardiac
cycle
to
begin
a
new
rhythm.
Special
Functions:
To
initiate
a
Special
Functions
rhythm,
press
and
release
the
red
SPEC
FUNC
key
followed
by
the
appropriate
two
digit
code,
found
in
the
menu
on
either
side
of
the
keypad.
The
simulator
displays
the
*
to
echo
the
Special
Function
key
selection.
You
must
enter
a
two
digit
special
function
code
whenever
the
*
prompt
is
displayed.
Example:
To
select
a
bigeminy
rhythm
press
the
red
SPEC
FUNC
key
followed
by
code
27.
You
are
free
to
select
any
function
in
any
order,
and
in
doing
so,
simulate
virtually
any
patient
condition
sequence.
Some
of
the
Special
Functions
are
actually
programmed
sequences
of
rhythms
or
events.
A
single
PAC
or
PVC
can
be
inserted
into
any
rhythm
by
pressing
the
PAC,
PVCI,
or
PVC2
key.
When
any
of
these
keys
are
pressed,
the
respective
PAC
or
PVC
will
be
inserted
into
the
next
cardiac
cycle.
Only
one
PAC,
PVCI,
or
PVC2
can
be
inserted
at
any
one
time.
The
8
character
LCD
display
echoes
key
entries
by
displaying
the
abbrevi-
ated
function
name.
10
DS
6100
Operating
and
Service
Manual
Operation
KEYPAD
CONTROLS
Code
Function
00
NORMAL
SINUS
-
Provides
a
convenient
way
to
return
to
normal
sinus
rhythm
at
a
rate
of
70
BPM.
01
SINUS
TACHYCARDIA
-
Rapid
regular
rhythm
with
normal
P
wave
and
heart
rate
of
140
BPM.
02
SINUS
BRADYCARDIA
-
Slow
regular
rhythm
with
normal
P
wave
and
heart
rate
of 35
BPM.
.
03
ASYSTOLE
-
No
beats
but
very
small
irregular
signal
remains.
04
ATRIAL
FLUTTER
-
Regular
ventricular
rhythm;
atrial
flutter
with
4:1
conduction.
05
ATRIAL
FIBRILLATION
-
Rapid
irregular
atrial
signal
with
no
real
P
waves.
Irregular
ventricular
rate.
06
2ND
DEGREE
AV
BLOCK
I
-
Wenckebach.
Irregular
rhythm
with
normal
P
wave.
PR
interval
lengthens
progres-
sively
until
a
dropped
beat
occurs
and
cycle
repeats.
07
2ND
DEGREE
AV
BLOCK
II
-
Slow
regular
rhythm
with
normal
P
wave.
PR
interval
in
conduced
beats
is
normal;
3:1
block.
08
3RD
DEGREE
AV
BLOCK
-
Slow
regular
ventricular
rhythm;
regular
atrial
rhythm.
Independent
atrial
and
ventricular
rhythms.
09
PAC
INSERT
-
Premature
atrial
contraction.
A
single
PAC
is
inserted
into
the
current
rhythm
when
button
is
pushed.
10
PVCI
INSERT
-
Premature
ventricular
contraction.
A
single
PVC1
is
inserted
into
the
current
rhythm
when
button
is
pushed.
11
PVC2
INSERT
-
Premature
ventricular
contraction.
A
single
PVC
is
inserted
into
the
current
rhythm
when
button
is
pushed.
12
COUPLET
-
Irregular
rhythm
with
frequent
couplets
and
1
pair
ventricular
beats.
PVCs
are
premature;
compensatory
pause
follows
couplet.
13
VENTRICULAR
TACHYCARDIA
-
Irregular,
rapid
rhythm
with
irregular,
successive
PVCs.
No
P
waves.
DS
6100
Operating
and
Service
Manual
11

Operation
ee
o
14.
VENTRICULAR
FIBRILLATION
-
Irregular,
rapid,
chaotic
waveforms
with
no
QRS.
AT
ㆍ ㆍ ㆍ
15
ATRIAL
TACHYCARDIA
-
Rapid,
regular
rhythm
with
biphasic
P
wave.
16
ATRIAL
TACHYCARDIA/ABERRANCY
-
Rapid,
regular
rhythm
with
biphasic
P
wave
and
notched,
wide
QRS.
17
FREQUENT
PACS
-
Irregular
rhythm
with
frequent
pre-
mature
PACs
and
a
biphasic
P
wave.
18
PEDIATRIC
TACHYCARDIA
-
Rapid
regular
rhythm
with
tall
R
wave
and
narrow
QRS.
BLOCKS
19
1ST
DEGREE
AV
BLOCK
-
Normal
beats
except
with long
PR
interval
of
.25
sec.
20
BUNDLE
BRANCH
BLOCK
-
Wide
QRS
complexes.
Regular
rhythm
with
a
PR
interval
of
.16
sec.
JUNCTIONAL
22
ACCELERATED
JUNCTIONAL
-
Regular
rhythm
with
inverted
P
wave
and
short
PR
interval.
23
JUNCTIONAL
-
Slow
regular
rhythm
with
inverted
P
wave
and
short
PR
interval.
24
FREQUENT
PICs
-
Irregular
rhythm
with
frequent,
premature
PJCs
followed
by
a
pause.
Inverted
P
wave
and
short
PR
interval.
VENTRICULAR
26
IDIOVENTRICULAR
-
Slow
reg.
rhythm,
no
P
waves,
wide
QRS.
27
BIGEMINY
-
Irregular
rhythm
alternates
between
PVC
and
normal
beat.
28
UNIFOCAL
PVCs
-
Frequent,
premature
unifocal
PVCs
followed
by
pause.
PVC
has
wide
QRS.
12
DS
6100
Operating
and
Service
Manual
Operation
29
MULTIFOCAL
PVCs
-
Regular
rhythm
with
frequent
PVCs
that
alternate
between
type
1
and
type
2.
Pause
follows
PVCs.
30
TRIPLET
-
Frequent
runs
of
3
consecutive
PVCs.
PACEMAKER
31
VENTRICULAR
-
Pacer
artifact
precedes
QRS.
Regular
rhythm,
no
P
waves,
wide
QRS.
32
ATRIAL
-
Normal
rhythm,
pacer
artifact
precedes
P
wave.
33
AV
SEQUENTIAL
-
Prolonged
PR
interval.
Pacemaker
artifact
precedes
P
and
QRS
waves.
Wide
QRS.
34
SENSE
&
CAPTURE
FAIL
-
Frequent
loss
of
capture,
low
amplitude
escape
beat,
sensing
failure
of
escape
beat.
35
FAILURE
TO
CAPTURE
-
No
capture.
Pacer
artifact
amplitude
is
8mV
with
1
msec
duration.
MISC
ECG
38
CARDIAC
FAILURE
-
Intermittent
PVCs
and
couplet
followed
by
ventricular
tachycardia,
followed
by
ventricular
fibrillation,
followed
by
asystole.
39
ST
ELEVATION
-
Regular
rhythm
with
ST
Elevation
at
2
mm
80
msec
after
J
point.
40
ST
DEPRESSION
-
Regular
rhythm
with
ST
Depression
at
3
mm
80
msec
after
J
point.
41
AGONAL
-
No
P
waves;
wide
complexes
with
periods
of
asystole.
42
VENTRICULAR
ASYSTOLE
-
No
QRS;
regular
atrial
rhythm.
43
CONVERSION
-
Vfib
with
CPR
followed
by
Vfib
without
CPR
for
2
seconds
prior
to
defib.
Defib
artifact,
Brady
thythm
with
CPR,
Brady
rhythm
without
CPR,
followed
by
NSR.
DS
6100
Operating
and
Service
Manual
13

Operation
ECG
ARTIFACT
46
DEFIBRILLATION
-
Positive
and
negative
baseline
satura-
tion.
Artifact
is
automatically
followed
by
preselected
rhythm.
See
#59
below.
47
60
CYCLE
ARTIFACT
-
Artifact
amplitudes
0.1,
0.4,
and
0.7
mV
p-p.
Use
INC
and
DEC
keys,
or
eliminate
with
STD
key.
48
ECG/RESP
ARTIFACT
-
Artifact
amplitudes
0.4,
0.6,
and
1.0
mV
peak
to
peak.
Use
INC,
DEC
keys.
STD
will
eliminate.
49
MUSCLE
ARTIFACT
-
Artifact
amplitudes
0.4,
0.6,
and
1.0
mV
peak
to
peak.
Use
INC
&
DEC
keys.
STD
will
eliminate.
50
CPR
-
CPR
artifact
at
70/min.
ADJUST
53
ECG
RATE
ADJUST
-
Increase
or
decreases
rate
in
10%
increments.
STD
returns
to
standard
rate.
S4
SIZE
ADJUST
-
Increases
and
decreases
ECG,
BP,
and
respiration
waveforms,
or
returns
them
to
standard
size.
Use
after
selecting
desired
channel.
55
AUTO
TREND
-
Automatically
varies
NSR
heart
and
respiration
rates,
and
blood
pressures.
Vtach
episode
4
every
1
hour;
trend
repeats
every
2
hours.
DEFIB
CONVERSION
59
RHYTHM
SELECT
-
Permits
selection
of
post defib
rhythm,
which
is
then
automatically
inserted
after
defib
artifact.
60
DEFIB
DISABLE
-
Disables
auto
defib
sequence.
RESPIRATION
71
APNEA
-
Thoracic
impedance
is
simulated
via
ECG
output
snaps.
72
RESP
AT
10
BPM
73
RESP
AT
20
BPM
74
RESP
AT
40
BPM
75
RESP
AT
80
BPM
14
DS
6100
Operating
and
Service
Manual
m
με
μα
μα
μα
ya
m
μι
na
m
pe
m
pa
m
Operation
76
CVA
COINCIDENCE
-
ECG
and
respiration
are
synchro-
nized
and
timed
appropriately
for
cardiovascular
artifact.
Impedance
drops
immediately
after
each
R
wave.
77
PEDIATRIC
APNEA/BRADY
-
Intermittent
apnea
periods.
Heart
rate
drops
from
140
BPM
tp
70
BPM
each
apnea
period.
CALIBRATION
78
HEART
RATE
CALIBRATION
-
Calibrated
heart
rates
from
30
to
300
BPM.
81
|
LINEARITY/SPEED
-
2.5
Hz
triangular
ECG
waveform.
82
ECG
SIZE
CALIBRATION
-
Cal
pulse
amplitudes
from
.25
mV
to
2.5
mV.
83
ARRHYTHMIA
SEQUENCE
-
Automatic
sequence
of
arrhythmias
over
3
minute
period
of
time.
NSR,
Dropped
Beat,
AV
Pace,
V
Pace,
Atrial
Tach,
Couplet,
RBBB,
Triplet,
Bigeminy
(3
Foci),
VTach.
84
RESPIRATION
SIZE
CAL
-
Respiratory
depth
ranging from
0.25
to
2.5
ohm;
calibrated
for
RA-LA
lead.
BLOOD
PRESSURE
62
ZERO
PRESSURE
-
Zeroes
all
three
pressure
channels
as
well
as
auxiliary
CO2
channel.
63
PATENT
LINE
-
Outputs
a
30
mmHg
flush
waveform
on
PA
channel.
PA
waveform
follows
flush.
Trailing
edge
of
flush
and
PA
pressure
waveforms
demonstrate
ideal
frequency
characteristics
of
pressure
line.
64
RESONANT
LINE
-
Outputs
a
50
mmHg
flush
waveform
on
PA
channel.
PA
waveform
follows
flush.
Trailing
edge
of
flush
and
PA
pressure
waveforms
demonstrate
resonant
frequency
characteristics
of
pressure
line.
DS
6100
Operating
and
Service
Manual
15

Operation
65
DAMPED
LINE
-
Ouputs
a
50
mmHg
flush
waveform
on
PA
channel.
PA
waveform
follows
flush.
Trailing
edge
of
flush
and
PA
pressure
waveforms
demonstrate
damped
frequency
characteristics
of
pressure
line.
|
66
CATHETER
WHIP
-
Pressure
artifact
caused
by
motion
of
catheter.
On
PA
channel.
ーー
|
67
PRES/RESP
ARTIFACT
-
Varies
level
of
respiration
artifact
on
pressures.
PROCEDURES
68
IABP-
Arterial
pressure
with
cardiac
assist
pressure
pulse.
2:1
augmentation.
Simulation
only;
not
interactive
with
the
IABP.
Output
on
arterial
channel.
69
PA
WEDGE
-
Outputs
a
PAW
waveform.
70
SWAN-GANZ
INSERTION
-
Output
on
PA
channel.
Right
atrium
to
right
ventricle;
right
ventricle
to
pulmonary
artery;
pulmonary
artery
to
wedged
pulmonary
artery;
wedged
pulmonary
artery
to
pulmonary
artery.
16
DS
6100
Operating
and
Service
Manual
Modifier/Cal
Mode
MODIFIER/CAL
MODE
The
bottom
row
of
the
keypad
has
three
dual
function
keys:
CPLT/up
arrow,
VTACH/down
arrow,
and
VFIB/STD.
These
keys
can
be
used
in
conjunction
with
the
SPEC
FUNC
keys
listed
in
Figure
4
to
increase,
decrease,
or
reset
the
standard
value
for
parameters
such
as
rate,
waveform
amplitude,
or
artifact
level.
When
selected,
these
functions
will
display
one
of
the
following
prompts:
INC,
DEC,
or
STD.
See
following
pages
for
operating
instructions.
INC
(Up
Arrow):
Increases
parameter
value.
Momentarily
displays
“at
max”
when
maximum
value
is
reached.
DEC
(Down
arrow):
Decreases
parameter
value.
Momentarily
displays
“at
min”
when
minimum
value
is
reached.
STD:
Resets
parameter
value.
All
of
the
remaining
keys
exit
the
Modifier/Cal
function,
retain
parameter
level,
and
simulate
rhythm
selected.
(i.e.
NSR,
Bradycardia,
etc.)
53
Rate
Adjust
54
Size
78
Heart
Rate
Cal
70
BPM
82
ECG
Size
Cal
84
Resp
Size
Cal
47
60
Cycle
Artifact
48
ECG/Resp
Artifact
49
ECG/Muscle
Artifact
FIGURE
4:
Special
Functions
used
in
the
Modifier/Cal
Mode
DS
6100
Operating
and
Service
Manual
17

Modifier/Cal
Mode
SIZE
ADJUSTMENTS:
SPEC
FUNC
54
The
Size
Adjust
function,
SPEC
FUNC
54,
allows
you
to
adjust
the
waveform
size
up
or
down
from
the
STD
level.
Figure
9
lists
waveform
amplitudes
for
each
channel.
To
adjust
the
size
of
a
waveform
channel:
STEP
1:
Select
SPEC
FUNC
54
(SIZE)
STEP
2:
DS
6100: Enter
desired
channel
number,
1
-
6.
ECG
=
channel
1,
ART
=
2,
PA
=3,
RA
=4,
AUX
=S5,
RESP
=
6
STEP
3:
Press
INC
to
increase,
DEC
to
decrease,
STD
to
reset
to
standard
level.
The
size
adjust
function
is
intended
for
inservice
and
demonstration
applications
when
size
adjustments
may
be
helpful
in
demonstrating
a
clinical
situation
i.e.
Increasing
the
PA
Wedge
pressure
to
simulate
LV
heart
failure.
NOTE:
Pressure
and
R
wave
values
are
defined
for
normal
sinus beats
only.
To
obtain
nonvarying
pressure
values,
select
SPEC
FUNC
67
and
decrease
minimum
level.
For
calibration
or
performance
check
applications,
use
SPEC
FUNC
78
(Heart
Rate
Cal),
SPEC
FUNC
81
(Linearity/Speed),
SPEC
FUNC
82
(ECG
Size
Cal),
and
SPEC
FUNC
84
(Resp
Size
Cal).
18
DS
6100
Operating
and
Service
Manual
EEE
EEE
ES
EE
NEN
Modifier/Cal
Mode
DS
6100
SIZE
ADJUST
LEVEL
-
SPEC
FUNC
54
CH/PARAMETER
|-3
-2
-1
STD
|
+1
+2
+3
+4
1
ECG
R
WAVE
25
1.50
1.75
|
Imv
[1.25
[1.50
[1.75
|
2.0
AMPLITUDE
(mV)
À
2
ART
SYS/DIAS
116/62
[123/66 [130/70
|
137/73
1143/76
[150/80
[157/84 [164/88
MEAN
(mmHg)
80 85 90 95
100
1105
{110
|
115
3
PA
SYS/DIAS
14/6
|21/9
127/12133/15
|
41/19]
48/22]
55/25
61/28
MEAN
(mmHg)
8
13
17
22.127
32
37
|
42
4
RA
MEAN
(mmHg)
4
le
8
10
112
[14
|
16
5
AUX
(CO,
CO2)
Not
Adjustable
6
RESP(RA-LA)
(ohms)
25
|.50
|175
|1.0
|125
|
1.50
|1.75
|
2.0
FIGURE
5.
Size
Adjust
Special
Function
DS
6100
Operating
and
Service
Manual
19

Modifier/Cal
Mode
CALIBRATION
FUNCTION
LEVELS:
SPEC
FUNC
78, 82,
&
84
Special
function
keys
78,
82,
and
84
allow
you
to
adjust
calibration
levels
for
heart
rate,
linearity/speed,
ECG
size,
and
respiration
size
as
follows:
ADJUSTMENT
LEVELS
-4-3
-2
-1
[STD[+1
+2
+3
+4
+5
+6
+7
+8
+9
+10
+11
+12
SPEC
FUNC
78:
HRCAL
30
40
50
60|70
180
90
100
121
142
163
183
203 227
250
268
300
SPEC
FUNC
82:
ECG/SIZE/CAL
25 50
75)
1
11.25
1.5
1.75
20
25
mV
SPEC
FUNC
84:
RESP
SIZE
CAL
25
50
75]
1
[1.25
15
1.75
20
2.5
Ohm
FIGURE
6:
Calibration
Functions
20
DS
6100
Operating
and
Service
Manual
Modifier/Cal
Mode
HEART
RATE
ADJUSTMENT:
SPEC
FUNC
53
The
Rate
Adjust
function,
SPEC
FUNC
53,
offers
you
the
option
of
adjusting
rates
up
or
down
in
10%
increments.
Figure
7
lists
the
heart
rates
which
can
be
obtained
for
the
various
rhythms.
On
Power-up,
each
rhythm
has
the
standard
heart
rate
(STD)
listed
below.
For
performance
check
applications,
the
Heart
Rate
Calibration,
SPEC
FUNC
78
(ECG
RATE),
provides
calibrated
heart
rates
from
30
to
300
BPM.
See
Fig
6.
RHYTHM
-40%
-30% -20%
-10%
STD
+10%
+20%
+30%
Bradycarda
21
24
28
31
35
38
42
45
Ventricular
21
24 28
31
35
38
42
45
Junctional
24
28
32
36
40
44
48
52
Normal
Sinus
42
49
56
63
70
77
84
91
Sinus
Tach
84 98
112
126
140
154
168 182
Atrial
Tach
90
105
120
135
150
165
180
195
Atrial
Fib
60
70
80
90
100
110
‘120
130
FIGURE
7:
Heart
Rates
Obtained
via
Rate
Adjust
Spec
Func
53
ARTIFACT
SIMULATION:
SPEC
FUNC
47,
48,
&
49
Various
types
of
artifact
can
be
superimposed
on
the
ECG.
The
type
and
level
of
artifact
is
selected
via
the
Special
Function
keys
47,
48,
and
49.
To
enable
artifact,
select
the
desired
function,
increase
or
decrease
the
level
of
artifact
via
the
INC,
DEC
keys.
The
STD
key
returns
the
simulator
to
the
standard
level
of
artifact.
FUNC
o
TYPE
STD
+]
+2
+3
47
ECG
60
Hz
0
그
.4
-7
mV
48
ECG
RESP
0
4
.6
1.0
mV
49
ECG
MUSCLE
|
0
4
6
1.0mV
FIGURE
8:
Artifact
Modifier
Functions
DS
6100
Operating
and
Service
Manual
21

Respiration
RESPIRATION
The
DS
6100
provides
a
varying
thoracic
impedance
signal
via
all
10
ECG
electrode
output
snaps.
Several
respiration
rates
from
apnea
to
80
BPM
can
be
selected.
Once
a
respiration
rate
has
been
selected,
the
respiration
rate
remains
fixed
during
subsequent
rhythm
selections
or
until
a
new
respiration
rate
is
selected.
The
amplitude
of
the
thoracic
impedance
signal
(R)
can
be
varied
using
Size
Adjust,
SPEC
FUNC
54
and
selecting
Channel
#6.
Use
the
INC,
DEC,
and
STD
keys
to
vary
the
impedance
amplitude.
Respira-
tion
amplitudes
will
differ
depending
upon
the
lead
that
is
monitored.
Figures
5
&
6
list
amplitudes
for
the
RA-LA
lead.
NOTE:
For
calibration
applications,
use
the
Respiration
Size
Cal
Function,
SPEC
FUNC
84
(Figure
6).
This
function
displays
impedance
amplitudes.
Pressure
waveforms
are
modulated
by
the
respiration
waveforms
to
simulate
respiration
artifact.
The
respiration
artifact
level
can
be
varied
using
Pressure/Respiration
Artifact,
SPEC
FUNC
67.
Cardiovascular
artifact
or
ECG/respiration
coincidence
can
be
simu-
lated
by
selecting
the
CVA
Coincidence,
SPEC
FUNC
76.
The
impedance
signal
is
synchronized
with
the
ECG.
Impedance
decreases
after
each
R
wave,
simulating
a
decrease
in
impedance
due
to
an
increase
in
pulmonary
blood
volume
during
systole.
22
DS
6100
Operating
and
Service
Manual
Defib
Training
DEFIBRILLATION
TRAINING
In
the
Defib
Training
Mode,
you
can
interface
the
DS
6100
with
the
Laerdal
“Arrhythmia
Anne”
and
the
Armstrong
“Chris
Clean”
mani-
kins
for
realistic
defibrillation
and
ECG
paddle
monitoring
training.
Follow
the
guidelines
provided
by
your
training
manikin
for
proper
cable
connection.
STEP
1:
Connect
the
torso
skin
cable
into
the
Defib
Training
Connec-
tor.
Turn
the
DS
6100
on.
STEP
2:
Select
life
threatening
arrhythmia.
i.e.
VFIB.
STEP
3:
To
enable
the
automatic
defib
conversion
feature
and
select
the
post
defib
rhythm:
a.)
Press
and
release
the
red
Special
Function
key
b.)
Enter
code
number
59
to
initiate
the
Defib
Rhythm
Select
function.
c.)
Select
the
desired
post-defib
rhythm
by
pressing
a
direct
function
key,
or
by
pressing
the
red
Special
Function
key
followed
by
the
two
digit
code
number
of
the
desired
rhythm.
Ifa
post
defib
rhythm
is
not
programmed,
the
defib
artifact
will
be
followed
by
asystole
until
another
rhythm
is
manually
selected.
STEP
4:
To
initiate
the
defibrillation
sequence:
Defibs
equipped
with
apex/sternum
paddles
-
F
irmly
place
the
paddles
on
the
electrodes
located
on
the
manikin
chest
skin
or
torso
cover.
Charge
the
defib
to
25
joules
for
Arrhythmia
Anne
or
up
to
400
joules
for
Chris
Clean,
and
defibrillate.
DS
6100
Operating
and
Service
Manual
23

Defib
Training
Defibs
equipped
with
disposable
electrodes
-
Plug
in
adapters
and
connect
cables
from
defib
to
adapter
snaps.
Charge
defib
to
desired
energy
level
and
discharge
the
defibrillator.
CAUTION:
See
the
Arrhythmia
Anne
and
“Chris
Clean”
instruction
manuals
for
the
proper
procedures
and
use
of
the
defibrillator
on
the
manikin.
Under
no
circumstances
should
a
defibrillator
be
discharged
directly
into
the
DS
6100.
After
defibrillation
of
the
manikin,
the
DS
6100
will
simulate
defibrillation
artifact
followed
by
your
preselected
rhythm.
The
programmed
post
defibrillation
rhythm
can
be
changed
at
any
time
by
repeating
STEP
2.
STEP
5:
To
repeat
a
sequence,
select
another
life
threatening
rhythm
and
defibrillate.
The
defib
conversion
feature
can
be
disabled
at
any
time
by
selecting
SPEC
FUNC
60
(NO
DEFIB).
MANUALLY
INITIATING
A
DEFIB
SEQUENCE
Use
the
defib
switch
on
the
DS
6100”s
front
panel,
or
SPEC
FUNC
46
(DEFIB).
When
the
Defib
Artifact
Function
is
selected,
defib
artifact
will
be
simulated.
If
a
post
defib
rhythm
has
been
programmed,
the
defib
artifact
will
be
automatically
followed
by
the
programmed
rhythm.
24
DS
6100
Operating
and
Service
Manual
BP
Simulation
BLOOD
PRESSURE
SIMULATION
If
you
want
to
check
the
calibration
of
your
invasive
pressure
monitor,
it
is
important
to
first
zero
each
pressure
channel
of
the
monitor.
Connect
BP
cables
from
the
simulator
to
your
monitor,
then:
1.
Zero
monitor
before
turning
the
DS
6100
on.
When
you
do
turn
the
DS
6100
on,
it
will
automatically
output
an
NSR
ECG
waveform
and
the
normal
BP
waveforms.
OR
2.
Тат
the
DS
6100
on
and
select
SPEC
FUNC
62
(0
mmHg).
Up
to
three
pressure
channels
can
be
zeroed
simultaneously
via
channels
2-4.
Use
the
pressure
monitor
manufacturer’s
suggested
procedure
to
obtain
a 0
mmHg
pressure
baseline
for
each
pressure
channel.
After
successfully
adjusting
the
monitor’s
zero
point, select
SPEC
FUNC
79
to
check
calibration
of
the
pressure monitor.
Then
select
the
desired
pressure
waveform
via
special
function
or
direct
function
keys
to
output
dynamic
BP
waveforms
on
all
three
channels.
Pressure
cables
(listed
on
page
33)
are
available
to
connect
the
DS
6100
to
most
pressure
monitors.
NOTE.
Figure
5
defines
systolic,
diastolic,
and
mean
values
of
the
dynamic
pressure
waveforms
for
normal
sinus
beats.
If
your
monitor
or
recorder
does
not
closely
agree
with
these
pressure
values,
any
of
the
following
may
be true:
1)
The
monitor
or
recorder
is
not
calibrated
or
properly
zeroed.
2)
The
interconnect
cable
is
not
supplying
an
excitation
reference
voltage
required
by
DS
6100.
3)
The
interconnect
cable
is
wired
to
the
wrong
signal
output
pin.
4)
Respiration
artifact
on
the
pressures
modulate
the
pressure
values
in
sync
with
the
respiration
cycle.
This
artifact
significantly
alters
the
systolic/diastolic
values.
To
remove
respiration
artifact,
select
SPEC
FUNC
67,
(PRES
RESP)
and
decrease
to
the
minimum
level.
DS
6100
Operating
and
Service
Manual
25

Personality
Modules
PERSONALITY
MODULES:
OPTIONAL
SPECIAL
FUNCTIONS
Optional
Personality
Modules
can
be
plugged
into
the
front
panel
connector
to
increase
the
number
of
special
functions
available.
It
is
necessary
to
turn
the
simulator
power
off
before
inserting
or
removing
a
Personality
Module.
Each
Module
has
a
function
menu
listing the
functions
and
code
numbers
6070-03
Pediatric
ECG
6070-08
Sinus
Arrhythmia
Junctional
Junctional
Escape
Wandering
Pacer
Hyperkalemia
Enlarged
Atrium
CPR
Artifact
Sinus
Tachycardia
6070-09
Advanced
Pacer
Undersensing
Oversensing
Muscle
Artifact
Oversensing
Tall
T
Waves
Fusion
Pseudo
Fusion
Runaway
Pacer
PAC
with
DVI
Pacer
Retrograde
VA
Conduction
6070-10
MCLI
Atrials
Normal
Sinus
Rhythm
Bradycardia
Sinus
Arrest
6070-13
Atrial
Tach
Atrial
Flutter
Atrial
Fib
PAC
PAT
within
the
Module's
memory.
These
functions
will
be
assigned
code
numbers
between
89
and
99.
Personality
Modules
currently
available
are
listed
below.
DS6100
00
00000000000000000000000000000:00
MCL
Blocks
Normal
Sinus
Rhythm
Ist
degree
AV
block
2nd
degree
AV
block
Type
I
2nd
degree
AV
block
Type
II
3rd
degree
AV
block
RBBB
LBBB
MCLI
Ectopy/Aberrancy
Normal
Sinus
Rhythm
Right
PVC
Left
PVC
Multifocal
PVCs
RonT
Bigeminy
Atrial
Fib
w/Aberrancy
Cardiac
Failure
Sequence
MCLI
Set
(07,
08,
09)
Includes
the
three
MCLI
modules
listed
above.
12
Lead
Set
Includes
modules
6070-14,
6070-15,
and
6070-16.
DS
6100
Operating
and
Service
Manual
등등
도도
도
도
이
여도
여여
이
어겨
Personality
Modules
6070-14
12
Lead
Normal
ECG
6070-15
12
Lead
Anterior
Infarct
6070-16
12
Lead
Infertor
Infarct
Leads
I,
II,
HE
AVR,
AVL,
AVF,
V1
6070-01
IABP
Early
Inflation
Late Inflation
Early
Deflation
Late
Deflation
Proper
Timing
6070-05
=
Intra-Cranial
Pressures
Normal
ICP
Normal
with
Resp
Artifact
Damped
with
Resp
Artifact
Cough
Artifact
Hypercapnia
B
Wave
&
A
Wave
Jugular
Comp
6070-11
Left
Heart
Pressures
Pressure
Zero
Normal
AO,
LV,
LA
AO
to
LV
Insertion
LV
to
AO
Pullback
AO,
LV,
and
LA
with
PVCs
6070-12
Valve
Disease
Mitral
Stenosis
Mitral
Regurgitation
Aortic
Stenosis
Aortic
Regurgitation
LV/AO
Pullback
with
AS
LV/AO
Pullback
with
AR
6070-17
6070-18
6070-20
6070-22
6070-23
6070-25
6070-26
6070-29
6070-30
ST
Segments
2.5mm
Horizontal
ST
Depression
2.5mm
Horizontal
ST
Elevation
ST
Downslope
Depression
ST
Upslope
Elevation
ST
Upslope
Depression
45°
ST
Upslope
Depression
30°
ST
Upslope
Depression
20°
CO
Dilution
Curve
0°C
Injectate
temp
for
use
w/6070-23
CO
Dilution
Curve
25° C
Injectate
temp
for
use
w/6070-22
CO2/Cardiac
Output
25°
C
CO2/Cardiac
Output
0°
C
CO
at
5L/4L/3L
per
minute
Left
to
right
shunt
Prolonged
Curve
Interrupted
Injectate
Normal
Respiration
Embolism
Cheyne
Stokes
-
Deep
Sigh
CO2
Zero
IABP
Interactive
Kontron*
(7,
10,
K2000,
KAAT)
IABP
Interactive
Datascope*
(System
80)
IABP
Interact
Datascope
90*
(System
90)
IABP
Interact
Aries*
Early
Inflation
Late
Inflation
Early
Deflation
Late
Deflation
Proper
Timing
*See
description
on
the
following pages
DS
6100
Operating
and
Service
Manual
27

Personality
Modules
INTERACTIVE
IABP
PERSONALITY
MODULES
The
Interactive
Intra-Aortic
Balloon
Pump
personality
modules
are
used
with
the
DS
6100
simulator
and
Kontron,
Datascope
System
80,
Datascope
System
90,
and
Aries
monitors.
Invasive
arterial
BP
and
sync
timing
cables
are
included
with
each
module.
To
generate
the
special
waveforms:
Step
1:
Determine
the
need
for
a
balloon
according
to
monitor
used.
Datascope
90:
It
is
essential
that
a
balloon
(or
suitable
replace-
ment)
be
connected
to
the
Balloon
Output
connector
for
proper
operation
of
the
Datascope
90
pump.
Datascope
80,
Kontron,
Aries:
It
may
be
necessary
to
defeat
the
alarm
by
connecting
a
balloon
or
suitable
replacement
to
the
Balloon
Output
Connector.
Step
2:
Connect
Datascope/Kontron/Aries
ECG
cable
to
the
DS
6100
ECG
snap
connectors.
Step
3:
Insert
the
Interactive
IABP
module
into
the
DS
6100’s
Person-
ality
Module
connector.
Turn
the
DS
6100
on.
Step
4:
Connect
the
following
cables
to
their
respective
monitors:
Datascope
90:
connect
cable
#3010-0314
(GE
3100/Datascope
6F
cable)
from
the
Datascope
Pressure
Transducer
input
connector
to
the
DS
6100’s
ART
connector
Datascope
80:
cable
#3013-0314
(GE
3100/Datascope
6F
cable)
from
the
Datascope
ART
PRES
input
connector
to
the
DS
6100’s
ART
connector.
28
DS
6100
Operating
and
Service
Manual
HEBEI
Personality
Modules
Kontron
7,
10,
2000,
KAAT:
cable
#3010-0354
(Kontron
High
Level
cable)
from
the
KONTRON
AUX
input
connector
to
the
DS
6100’s
ART
connector
(Kontron
2000
ART
input)
Aries:
cable
#3010-0315
(4100-9
SpaceLabs
6M
cable)
from
the
ARIES
ART
PRES
input
connector
to'the
DS
6100’s
ART
connector.
Step
5:
Remember
to
zero
the
arterial
pressure
channel
on
the
pump.
To
simulate
atmospheric
pressure,
select
SPEC
FUNC
62
(OmmHg).
Use
the
module’s
suggested
procedure
to
obtain
a
OmmHg
pressure
reading.
When
zeroed,
select
any
cardiac
thythm
to
bring
the
arterial
waveform
on
the
screen.
Step
6:
Connect
the
following
IABP
cable
assemblies
to
the
DS
6100’s
AUX
connector:
Datascope
90:
3010-0352
from
the
DATA
COM
output
connector
Datascope
80:
3010-0351
from
the
System
INTERFACE
output
connector
Kontron:
3010-0353
from
the
ASSIST
INTERVAL
output
connector
Aries:
3010-0353
from
the
ASSIST
INTERVAL
output
connector
Step
7
:
Tum
the
balloon
pump
on
and
follow
normal
procedures
for
initializing
the
pump.
Select
a2
to
1
assist
ratio.
Step
8:
The
DS
6100
will
respond
to
the
balloon
pump’s
inflation/
deflation
signal
with
the
appropriate
augmented
arterial
pressure
waveform.
Change
ECG
rhythms
using
the
DS
6100’s
keypad.
DS
6100
Operating
and
Service
Manual
29

Personality
Modules
CARDIAC
OUTPUT
PERSONALITY
MODULES
The
DS
6100
can
be
used
with
either
of
two
Cardiac
Output
Personal-
ity
Modules,
0
Degrees
C
injectate
temperature
(model
6070-18)
and
25
Degrees
C
injectate
temperature
(model
6070-20).
Both
simulate
the
Edwards
style
blood
temperature
catheter
in
generating
several
dilution
curves.
The
dilution
curve
at
0
degrees
C
injectate
temperature
assumes
the
following
conditions:
10cc
Injectate
volume
Edwards
93-131-7F
Catheter
Injectate
Bath
Probe,
90.22
k
ohms
(compatible
with
Switchcraft
connector
SL-40-47F)
Computation
Coefficient
.542*
The
dilution
curve
at
25
degrees
C
injectate
temperature
assumes
the
following
conditions:
10cc
Injectate
volume
Edwards
93-131-7F
Catheter
Injectate
Bath
Probe,
62.34
k
ohms
(compatible
with
Switchcraft
connector
SL40-47F)
Computation
Coefficient
.595*
*The
use
of
other
computation
LM10
coefficients
could
result
in
CO
values
other
than
specified.
30
DS
6100
Operating
and
Service
Manual
Personality
Modules
Follow
these
steps
when
using
the
CO
module:
Step
1:
Insert
the
Cardiac
Ouput
Module
into
the
DS
6100’s
Personal-
ity
Module
connector.
Turn
on
the
DS
6100.
Step
2:
Connect
the
CO
computer’s
BT
(blood
temperature)
cable
to
the
CO
Module’s
3
pin
connector.
(The
BT
cable
in
the
clinical
setting
is
attached
to
the
thermistor
on
the
Swan-Ganz
catheter.)
Step
3:
Connect
the
CO
computer’s
injectate
temperature
cable
to
the
black
injectate
temperature
box
which
simulates
0
or
25
degrees
C
temperature.*
Step
4:
Set
up
CO
computer
per
the
manufacturer’s
instructions.
When
ready
to
inject
select
SPEC
FUNC
and
number
for
CO
curve
desired.
(Curves
listed
on
module)
*
The
internal
resistor
value
is
selected
to
meet
the
input
requirements
of
the
American
Edwards
or
SpaceLabs
bath
probe
injectate
temperature.
We
suggest
using
the
actual
bath
probe
and
bath
to
meet
the
input
requirements
for
other
desired
cardiac
output
computer
systems
or
modules.
Some
patient
monitoring
systems
will
default
to
an
iced
injectate
(0°
C)
ifno
bath
probe
is
attached
to
the
system.
DS
6100
Operating
and
Service
Manual
31

Personality
Modules
CAPNOGRAPHY
PERSONALITY
MODULE
The
DS
6100’s
Capnography
Personality
Module
is
available
in
combination
with
Cardiac
Output
Models
6070-22
and
6070-23.
The
capnography
function
simulates
the
graphic
portrayal
of
the
varying
CO?
levels
during
the
respiratory
cycle.
Capnography
simulation
is
possible
only with
selected
patient
monitors.
Compatible
monitors
(see
list
below)
require
the
following
interconnect
cables:
MONITOR
CO2
Cable
to
order
Spacelabs
Alpha
9,
14,
PC
#6085-03,
P/N
3010-0333
Spacelabs
500,
600
#6085-04,
P/N
3010-0334
Mennen
#6085-05,
P/N
3010-0335
NOTE:
The
Capnography
option
is
not
compatible
with
other
manufacturers’
CO2
monitors
because
they
lack
analog
CO2
input.
Step
1:
Insert
the
CO/CO2
personality
module
into
the
DS
6100
personality
module
connector.
Turn
the
DS
6100
on.
Step
2:
Connect
the
CO2
cable
from
CO2
monitor
Auxiliary
Input
to
the
DS
6100
AUX
channel.
Step
3:
Remember
to
calibrate
the
CO2
line.
Select
SPEC
FUNC
99
(OmmHg).
Use
the
CO2
monitor
manufacturer’s
suggested
procedure
for
calibration.
Step
4:
Select
SPEC
FUNC
95
through
98
to
obtain
the
specific
capnography
waveform
desired.
32
DS
6100
Operating
and
Service
Manual
Нанна
BP
Cables
BLOOD
PRESSURE
CABLES
To
ensure
proper
operation,
be
sure
the
interface
cables
you
are
using
between
simulator
pressure
outputs
and
your
equipment
include
the
proper
connectors
and
wiring.
If
you
wish
to
assemble
cables,
please
refer
to
the
list
below.
Figures
9
through
13
on
the
following
page
show
the
proper
wiring
for
cables
connecting
channels
2-4
to
most
pressure
monitors.
For
assistance
in
wiring
pressure
cables,
please
contact
DNI.
NOTE:
A
special
interface
cable
is
not
necessary
for
either
ECG
or
respiration
since
the
simulator’s
ECG
snap
outputs
are
directly
compatible
with
most
standard
patient
cable
lead
wires.
MODEL
ITEM
MODEL
ITEM
4100-01
Unterminated
4100-20
Kontron/Rouche
(6M)
4100-02
HP
SuV
(SF)
4100-23
Marquette
(8M)
4100-03
HP
40uV
(SF)
4100-24
Burdick
DataSim
4100-04
HP
5uV
(12M)
4100-25
Nihon/Kohden
(5M)
4100-05
HP
40uV
(12M)
4100-37
Honeywell/Meddars(9M)
4100-06
SpaceLabs
(5M)
-
4100-40
Fukuda
Denshi
(8M)
(700,
Alpha
9,14)
4100-42
Siemens
Mingo
7
(15M)
4100-08
GE
3100/Datascope
(6F)
4100-50
Armstrong
High
Level
4100-09
SpaceLabs
Output
(400,500,600,PC)
4100-60
Marquette
Eagle
Series
4100-09
Physiocontrol
(6M)
4100-61
Fukuda
Denshi
(12M)
4100-09
MDE
Escort
(6M)
4100-09
Mennen
(6M)
4100-09
Ivy
(6M)
4100-11
EforM/Honeywell
(6F)
4100-12
BD/Datamedix/Ohio
(9M)
4100-14
Abbott
(7M)
4100-15
Gould/Care
(SM)
4100-18
Litton
(4M)
4100-19
Siemens
(10M)
DS
6100
Operating
and
Service
Manual
33

Wiring
Charts
150
ohm
1%
red
EXC
REF
PIN
4
V
GND
REF
150
ohm
1%
white
+SIG
PIN
1
|
+
SIG
black
pair
SIG
GND
PIN
6
|
-
SIG
white
+EXC
PIN
5
+
EXC
FIGURE
9:
Pressure
Cable
Wiring
for
Monitor
with
SuV/VmmHg
sensitivity
150
ohm
1%
red
EXC
REF
PIN
4
AMW-
GND
REF
$
150
ohm
1%
|
white
+SIG
PIN
2
|
+SIG
black
pair
SIG
GND
PIN
6
-
SIG
white
+EXC
PINS
+EXC
FIGURE
10:
Pressure
Cable
Wiring
for
Monitor
with
40uV/VmmHg
sensitivity
NOTE:
To
minimize
noise,
the +/-
SIG
wires
should
be
a
twisted
pair,
as
should
the
+/-
EXC
wires.
34
DS
6100
Operating
and
Service
Manual
Connector
Interface
CONNECTOR
INTERFACE
PIN
DEFINITIONS
PIN#*
Function
N/C
N/C
HIGH
LEVEL
OUT
+9V
DEFIB
IN
GND
PACER
FLAG
(5V,
10msec)
JO
Aa
YN
r
Figure
11:
ECG
Connector
PIN
#*
Function
+
Signal
Out
(SuV/VmmHg)
+
Signal
Out
(40uV/VmmHg)
+
High
Level
Out
_
EXC
REF
+
Excitation
In
-
Signal
+7,15
V.
REF
SI
OO
D
BUN
=
Figure
12:
Pressure
Connectors
*Cable
connector
is
#2710-0278
or
Switchcraft
#15GM7M
1
4
2
5/69
90
Figure
13:
Connector
0
5
Pin
Assignments
oY
DS
6100
Operating
and
Service
Manual
35

Battery
Care
CHARGING
THE
BATTERY
The
rechargeable
battery
used
by
the
DS
6100
is
a
1.9
AH
sealed
lead
acid
battery
which
powers
the
unit
for
up
to
20
hours.
The
integrated
charger
requires
up
to
10
hours
to
charge
the
battery
if
its
charge
is
significantly
depleted.
A
Low
Bat
message
is
displayed
when
the
battery
needs
to
be
recharged.
To
charge
the
battery,
plug
the
DS
6100’s
battery
charger
into
the
unit
and
an
AC
line
power
source.
CAUTION:
The
battery
should
never
be
discharged
completely.
Repeated
complete
discharging
can
result
in
damage
to
the
battery.
To
avoid
this
situation:
*Recharge
the
battery
every
6
months
if
the
unit
has
not
been
used
*Immediately
begin
charging
when
the
battery
level
is
LOW.
A
fully
charged
battery
will
retain
at
least
95%
of
its
charge
for
7
days
with
the
unit
turned
OFF.
The
maximum
charge
time
to
reach
95%
battery
capacity
is
approximately
10
hours.
REPLACING
THE
BATTERY
Only
qualified
service
personnel
should
be
permitted
to
replace
the
battery.
When
replacing
the
battery
follow
these
instructions:
1.
Turn
power
off
and
disconnect
unit
from
AC
line
power.
2.
Remove
the
6
front
panel
screws.
3.
Remove
the
front
panel
assembly
exposing
the
battery.
4.
Detach
the
battery
wires
and
remove
the
battery
bracket.
5.
Replace
with
1.9
AH
12
Volt
sealed
lead
acid
battery,
part
#1001-
4004
or
equivalent.
36
DS
6100
Operating
and
Service
Manual
USERRA
oo
Parts
List
Ref.
Des.
Part
No.
Description
Qty
DS
6100
Shipping
System
(#203301,
Rev
A)
ZZ01
9519-0313
DS
6100
FINAL
ASSY
1
2202
9508-0213
DS
6100
OPERATOR
MANUAL
1
ZZ03
1201-0088
BATTERY
CHARGER
1
DS
6100
Final
Assembly
(#203305,
Rev
A)
2201
1001-4004
|
12V
BATTERY
DATASIM
ZZ02 3010-0347
CABLE
SPIRAL
25FT
DS2000
ZZ03
4704-2435
SCRW
PPH
6-32
X
3/8
2204
4709-0017
SCRW
PBH
46-1/4
SELF
TAP
2205
4709-0026
SCRW
SLF
TP
(PENDANT
RL.)
ZZ06
4709-0027
SCRW
SLF
TP
(PANEL
MT.)
ZZ07 4711-9720
NUT,
KEP
6-32
X
1/4
ZZ08
4720-0013
RIVET
BLACK
ANODIZED.
ZZ09
5001-0426
DS
6100
TOP
DATA
PANEL
ZZ10
5001-0434
DS
6100
SERIAL
NO.
LABEL
2211
5001-0435
DS
6100
SERIAL
NO.
TAG
2712
5001-0437
DS
6100
CASE
LABEL
2213
5008-0418
DS
6100
PCB
SHIELD
BASE
ZZ14
5023-0138
DATASIM
PENDANT
RAIL
ZZ15 5027-0205
DS
6100
CASE
MODIF
ZZ16
5205-0294
DS
6100
MAIN
PCB
ASSY
ZZ17
5210-0240
PARTIAL
PENDANT
ASSY
ZZ18
5008-0400
DATASIM
BATTERY
BRACKET
ZZ19
5215-0298
DS6100
TOP
PANEL
MECH
ASSY
2220
5215-0299
DS6100
BATTERY
PLATE
ASSY
2221
5215-0300
DS6100
PCB
SHLD
TOP
ASSY
ZZ22
9530-0037
DATASIM
ACCESS
POUCH
Z223
2719-0213
PWRJACK
PLASTIC
PANEL
MT.
ZZ24
4910-0150
SPACER
#6
CLR
X
.485
LONG
DS
6100
Operating
and
Service
Manual
TR
He
HH
E
RH
ee
EAU
DO
A
Qu
mm
いっ
>
NI
Table of contents
Popular Medical Equipment manuals by other brands

DeVilbiss Healthcare
DeVilbiss Healthcare drive Novaltis user guide

Konica Minolta
Konica Minolta FLEXDR C30 brochure

Atos Medical
Atos Medical Provox Vega XtraSeal Clinician manual

Invacare
Invacare Medley Ergo user manual

Wellspect
Wellspect Navina Classic quick guide

Rotaid
Rotaid SOLID PLUS HEAT installation manual