Emec ECL6 User manual

OPEN AMPEROMETRIC CELLS
OPERATING MANUAL
EN
ECL6 - ECL6E - ECL12 - ECL12E

INDEX
GENERAL SAFETY GUIDELINES ...................3
PURPOSE OF USE AND SAFETY ...................4
ENVIRONMENTAL SAFETY...........................5
SPARE PARTS...............................................5
INTRODUCTION...........................................7
Open amperometric cells ................................ 7
MODELS ........................................................ 8
Packaging ...................................................... 8
ECL6.............................................................. 9
ECL6/E........................................................... 9
ECL12............................................................ 10
ECL12/E......................................................... 10
Operating principle......................................... 11
Precautions .................................................... 11
Installation..................................................... 12
pH/ORP/temperature probes installation ......... 13
Commissioning............................................... 13
Electrode replacing......................................... 14
Electrode cleaning.......................................... 14
Calibration ..................................................... 15
Decommissioning .......................................... 15
MAINTENANCE ...........................................16
Maintenance schedule.................................... 16
Maintenance inspection ................................ 16
Calibration procedure..................................... 16
TROUBLESHOOTING....................................17
Repair service................................................. 17
ECL6 ............................................................. 18
ECL6/E........................................................... 19
ECL12............................................................ 20
ECL12/E......................................................... 21

3
⎘
This operating instructions contains safety information that if ignored can endanger life or result
in serious injury.
The original instruction is in English.
Read these instructions carefully before use and keep them for future reference.
Information and specifications on this manual could be uncorrect or could have printing errors.
Specifications are subject to change without notice.
Version: R1-02-17
D.M. 7 Febbraio 2012 n.25 – D.M.6 Aprile 2004 n.174 – Regolamento UE 10/2011
Apparecchiature finalizzate al trattamento dell’acqua destinata al consumo umano/
Equipment intended to come into contact with food/
Materiales y objetos plasticos destinados a entrar en contacto con alimentos.
Danger!
Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
Warning!
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
Important - A practice not related to personal injury or additional information.
Cross reference - An instance which refers to related information elsewhere in the same
document
GENERAL SAFETY GUIDELINES
Operating, installing, or maintaining the unit in any way that is not covered in this manual could
cause death, serious personal injury, or damage to the equipment.
This manual use the following safety message icon:
ICON

4
EQUIPMENT INTENDED FOR THE MEASUREMENT OF CHLORINE IN WATER.
Do not use in explosive area (EX).
Do not use with flammable chemicals.
Do not use with radioactive chemicals.
Use the equipment in accordance with the data and specifications printed on the label.
Do not modify or use in a manner inconsistent with the provisions of the operating manual.
When using this product with aggressive chemicals observe the regulations concerning
the transport and storage of aggressive fluids.
When installing always observe national regulations.
Manufacturer is not liable for any unauthorized use or misuse of this product that may
cause injury, damage to persons or materials.
Probes must be serviced and repaired by qualified and authorized personnel only.
Before any operation:
• always read chemical Material Safety Data Sheet (MSDS);
• always wear protective clothing;
• empty and rinse the liquid end before work on a the product which has been
used with hazardous or unknown chemicals.
Avoid grinding / shock / falls / friction.
PURPOSE OF USE
AND SAFETY

5
Work area
Always keep the area clean to avoid and/or discover emissions.
Recycling guidelines
EWC code: 16 02 16
Always recycle according to these guidelines:
1. If the unit or parts are accepted by an authorized recycling company, then follow local recycling
laws and regulations.
2. If the unit or parts are not accepted by an authorized recycling company, then return them to
the nearest representative.
Waste and emissions regulations
Observe these safety regulations regarding waste and emissions:
• Dispose appropriately of all waste.
• Handle and dispose of the dosed chemical in compliance with applicable environmental
regulations.
• Clean up all spills in accordance with safety and environmental procedures.
• Report all environmental emissions to the appropriate authorities.
Environmental
safety
For spare parts orders or any other communication, refer to the pump’s label.
Code (CODE) and serial number (P / N) uniquely identify the probe.
Spare parts

6
Transportation and
storage
Storage
An unsuitable transportation or storage can cause damages.
Use original box to pack the amperometric cell.
Observe storage conditions also for transportation.
Although packed, always protect the unit against humidity and the action of
chemicals.
Before returning the probe to the manufacturer repair service, refer to ⎘
Decommissioning.
DO NOT DISCARD PACKAGING. USE IT TO RETURN THE PROBE.
To store the probe refer to ⎘Decommissioning.
1. Clean the electrode, refer to ⎘ Cleaning procedure.
2. Store the probe in the original box, protected against sun, chemicals and humidity.
Transportation and storage temperature ..... +10°C - +50°C (+32°F - +122°F)
Humidity .................................................... 95% relative humidity (not condensed)

7
ECL Series is designed for measuring free chlorine (both organic and inorganic).
Open amperometric cells comprise an Off-line probe holders, a sensing electrode and a flow electrode.
Probe holders can contain up to three probes (temperature, pH and ORP).
The flow of water within this cell must remain constant and within 40 l/h. A pressure stabilizer is
available for areas subject to sudden pressure changes.
It is recommended to instal a filter before the probe holder.
Suitable for non-clean water thanks to the presence of cleaning balls.
Fig. 1. Description
SEPR
Water inlet valve
Water sampling
valve
Water sampling point
WATER INLET
WATER OUTLET
ELECTRODE
Temperature
pH or ORP probe
brown (+ supply)
blue (- supply)
black (PNP transistor
collector output)
RED (+)
BLACK (-)
Open
amperometric
cells
INTRODUCTION

8
MODELS • ECL6 for free chlorine (organic and inorganic) - pH and ORP probe-holder
• ECL6/E for free chlorine (organic and inorganic)
• ECL12 for free chlorine (organic and inorganic) for salt water - pH and ORP probe-holder
• ECL12/E for free chlorine (organic and inorganic) for salt water
Packaging Packaging includes:
• AMPEROMETRIC CELL
• COD. 11070081: KIT SCREW + WASHER +SCREW COVERS
• COD. 11067291: STANDARD KIT (1/2” PP+FP NIPPLE; 1/2” PVC GREY BALL VALVE M/M; FLAT
GASKET 1/2” F 1/4” M EPDM)
• COD. 02502121: 1/2” 6X8 HOSE HOLDING KIT FOR BLACK PP VALVE
• 4 MT 6X8 PE HOSE

9
ECL6
ECL6/E ECL6 ECL6/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) / BROMINE
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/copper; on request gold/copper)
Ph working range 6-8 pH
Run-in-time First polarization: 2 h about
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
adjustment See Operating manual: “Calibration”
Alcalinity 100 ppm
Working
temperature 5-40° C (41-104°F)
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
holder for pH, ORP and temperature temperature
Fittings for
connection to the
sample pipeline
6x8
Material Electrode: platinum/copper
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Maintenance
Regular control of the signal
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

10
ECL12
ECL12/E ECL12 ECL12/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/silver)
Ph working range 6-8 pH
Run-in-time First polarization: 2 h about
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
adjustment See Operating manual: “Calibration”
Slope calibration See Operating manual: “Calibration”- DPD1 method
Alcalinity 100 ppm
Working
temperature 5-40° C (41-104°F)
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
holder for pH, ORP and temperature temperature
Fittings for
connection to the
sample pipeline
6x8
Material Electrode: platinum/silver
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Maintenance
Regular control of the signal
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

11
ECL amperometric cells have an amperometric measuring system.
Measurement method is an electrochemical technique: the two electrodes (platinum and copper, or
platinum and silver) produces a current directly proportional to the concentration of chlorine.
Before any operation (preparation, cleaning and replacements) and before handly the
amperometric cell you MUST FOLLOW these PRECAUTIONS.
Operating
principle
Precautions
WEAR UNPOWDERED NITRILE GLOVES.
Avoid contact of the electrolyte with the skin.
In case of contact with skin, rinse immediately with plenty of water.
WEAR EYE PROTECTION
Avoid contact of the electrolyte with the eyes.
In case of contact with eyes, rinse immediately with plenty of water
and seek medical advice.
DANGER
DANGER
DANGER
AUTHORIZED AND QUALIFIED PERSONNEL
Installation and maintenance tasks should be carried out by
AUTHORIZED AND QUALIFIED PERSONNEL only in accordance with
local regulations.

12
Installation 1. Secure amperometric cell on a stable wall by screwing the two fixing screws on the body of
the amperometric cell (⎘ Dimensional).
2. Connect water inlet and outlet to the pipe.
3. We recommend the installation of a filter before the amperometric cell.
4. Install pH, ORP and temperature probes (if any) in their seats.
5. Connect SEPR and electrode to the instrument terminal board (refer to the manual of the
instrument).
Fig. 2. Amperometric cell installation
SEPR
Connect to the
instrument
WATER INLET
ELECTRODE
Connect to the
instrument
Temperature
pH or ORP probe
Water inlet valve
WATER OUTLET

13
Referring to the drawing below, follow the assembly sequence.
Fig. 3. pH, ORP, temperaure porbes installation.
pH/ORP/
temperature
probes
installation
1. Open water flow into the cell turning water inlet valve.
2. Turn until the float reaches the black mark on the cell.
3. SEPR LED lights to indicate the proper flow into the cell.
Fig. 4. Correct float position
Commissioning
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Progettato da: Controllato da: Data:
Edizione Foglio
1 / 1
ECL6 CELLA AMPEROMETRICA
Esploso_ECL6
Massimo_F Massimo_F 07/02/2017
00
Toll. Gen.
±0.05
Peso Lordo:
Peso Netto:
Materiale:
pH / ORP probe temperature probe

14
DANGER
AUTHORIZED AND QUALIFIED PERSONNEL
Installation and maintenance tasks should be carried out by AUTHORIZED AND QUALIFIED
PERSONNEL only in accordance with local regulations.
DANGER
POWER SUPPLY DISCONNECTION
Always disconnect power to the equipment before you perform any installation or maintenance
tasks. Failure to disconnect power will result in serious physical injury.
1. Follow precautions described in paragraph ⎘ “Precautions”.
2. Close water inlet valve.
3. Disconnect electrode’s wirings on the instruments terminal (refer to the manual of the
instrument).
4. Unscrew the screw at the top right that secures the amperometric cell to the wall / panel.
5. Pivoting on bottom left screw, gently rotate the cell to the left approximately 30 degrees so
as not to fall outside the cleaning balls.
6. Unscrew the electrode and replace it with a new one or clean as specified in paragraph
⎘ “Electrode cleaning”.
7. Screw electrode in its seat.
8. Tighten right screw onto the amperometric cell in order to secure it on the wall / panel.
9. Reconnect wires on the instrument terminal (refer to the manual of the instrument).
10. Proceed to the calibration of amperometric cell.
Electrode cleaning To clean electrode, follow from step 1 to 6 of paragraph ⎘ “Electrode replacing”, then
1. Unscrew electrode.
2. Clean the copper with a cotton ball soaked in acid solution, gently rub with a fine-grain
abrasive paper and rinse with water.
3. Screw electrode in its seat.
4. Reconnect wires on the instrument terminal (refer to the manual of the instrument).
5. Proceed to the calibration of amperometric cell.
Electrode replacing
DANGER
Do not USE alcohol to clean the amperometric cell.

15
Calibration Perform calibration of amperometric cell on a monthly basis or more, if necessary.
Amperometric cell must be calibrated with the instrument to which it is connected.
Calibration is carried out on two points: 0 and a second point near the work value.
For calibration procedure refer to the manual of the instrument.
Decommissioning Decommission the cell before any maintenance operation, shipping and long downtimes.
DANGER
AUTHORIZED AND QUALIFIED PERSONNEL
Installation and maintenance tasks should be carried out by AUTHORIZED AND
QUALIFIED PERSONNEL only in accordance with local regulations.
DANGER
POWER SUPPLY DISCONNECTION
Always disconnect power to the equipment before you perform any installation or
maintenance tasks. Failure to disconnect power will result in serious physical injury.
1. Disconnect power supply.
2. Stop water feeding by turning water inlet valve.
3. Keep a basin under sampling point and open water sampling valve to drain the liquid inside.
4. Disconnect electrode’s wirings on the instruments terminal (refer to the manual of the
instrument) and uscrew electrode (keep cleaning balls in a basin).
5. Disconnect hydraulic connections.
6. Unscrew amperometric cell from panel / wall.
7. Unscrew SEPR:
8. Wash the PMMA body from chlorine residues.
9. Air dry.

16
DANGER
MAINTENANCE SCHEDULE
In order to ensure the requirements of potable drinking water treated and the maintenance of
the improvements as declared by the manufacturer, this equipment must be checked at least
once a week.
DANGER
OPERATOR PROTECTION
Use safety equipment according to the company regulations.
Use this safety equipment within the work area during installation, service and when handling
chemicals:
• protective mask
• unpowdered nitrile gloves
• safety goggles
• further security device, if necessary.
DANGER
POWER SUPPLY DISCONNECTION
Always disconnect power to the equipment before you perform any installation or maintenance
tasks. Failure to disconnect power will result in serious physical injury.
DANGER
AUTHORIZED AND QUALIFIED PERSONNEL
Installation and maintenance tasks should be carried out by AUTHORIZED AND QUALIFIED
PERSONNEL only in accordance with local regulations.
IMPORTANT
Use original spare parts.
INTERVAL MAINTENANCE INSPECTIONS REFERENCE
Weekly Check probe reading (DPD1 or colorimetric
method)
Calibration procedure
Monthly Check electrical wiring and cable -
Periodically Check electrode Cleaning procedure
Shorten the inspection intervals appropriately if necessary.
Maintenance
schedule
Maintenance
inspection
MAINTENANCE

17
TROUBLESHOOTING
PROBLEM CAUSES SOLUTIONS
Sensor cannot be calibrated
and measured value
greater than DPD measurement
Run-in period too short Observe the run-in period
Troublesome substances in
the water
Check water for troublesome
substances and contact supplier
Short circuit in the measuring
line
Identify short circuit and
eliminate the cause.
Contact supplier
Electrode is dirty Clean or replace electrode
Sensor measured value is
unstable
Dirty / damaged electrode Clean or replace electrode
Unstable pressure Check installation
Nessun segnale di lettura Insufficient flow into the cell Check correct position of the float.
If necessary, open water flow into
the cell
Desoldering wires in the electrode cable Contact supplier
Black and red wires reversed Check for proper connection of the
wires on the instrument terminal
Before return the amperometric cell to the manufacturer Repair service, follow
⎘ Decommisioning procedure.
If there is the possibility that residual corrosive liquid into pump head could cause
damages, declare it on REPAIR FORM.
Complete the PRODUCT SERVICE REPAIR FORM and send it with the amperometric cell.
Repair service is not accepted if PRODUCT SERVICE REPAIR FORM is missing.
Repair service

18
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Progettato da: Controllato da: Data:
Edizione Foglio
1 / 1
ECL6 CELLA AMPEROMETRICA
ECL6
Massimo_F Massimo_F 21/07/2015
00
Toll. Gen.
±0.05
Peso Lordo:° Peso Netto:
Materiale:
60.00
201.70
223.20
9.00
106.00
9.00
100.00
ECL6
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Progettato da: Controllato da: Data:
Edizione Foglio
1 / 1
ECL6 CELLA AMPEROMETRICA
ECL6
Massimo_F Massimo_F 21/07/2015
00
Toll. Gen.
±0.05
Peso Lordo:° Peso Netto:
Materiale:
60.00
201.70
223.20
9.00
106.00
9.00
100.00

19
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Progettato da: Controllato da: Data:
Edizione Foglio
1 / 1
ECL6E CELLA AMPEROMETRICA
ECL6E
Massimo_F Massimo_F 21/07/2015
00
Toll. Gen.
±0.05
Peso Lordo: Peso Netto:
Materiale:
9.00
81.00
177.30
60.00
171.70
9.00
100.00
ECL6/E
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Progettato da: Controllato da: Data:
Edizione Foglio
1 / 1
ECL6E CELLA AMPEROMETRICA
ECL6E
Massimo_F Massimo_F 21/07/2015
00
Toll. Gen.
±0.05
Peso Lordo: Peso Netto:
Materiale:
9.00
81.00
177.30
60.00
171.70
9.00
100.00

20
DIMENSIONS
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Progettato da: Controllato da: Data:
Edizione Foglio
1 / 1
ECL12 CELLA AMPEROMETRICA
ECL12
Massimo_F Massimo_F 22/07/2015
00
Toll. Gen.
±0.05
Peso Lordo:° Peso Netto:
Materiale:
60.00
201.70
223.20
9.00
106.00
9.00
100.00
ECL12
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Progettato da: Controllato da: Data:
Edizione Foglio
1 / 1
ECL12 CELLA AMPEROMETRICA
ECL12
Massimo_F Massimo_F 22/07/2015
00
Toll. Gen.
±0.05
Peso Lordo:° Peso Netto:
Materiale:
60.00
201.70
223.20
9.00
106.00
9.00
100.00
This manual suits for next models
3
Table of contents
Other Emec Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Racal-Dana
Racal-Dana 488 Operator's manual

Keysight Technologies
Keysight Technologies N432A user guide

Agilent Technologies
Agilent Technologies U1731B quick start guide

Sony
Sony DVCAM DSR-80 Service manual

LI-COR
LI-COR LI-200R installation guide

Endress+Hauser
Endress+Hauser Cerabar S Series operating instructions