ERTE B2 User manual

B2
Weighing Indicator
User Manual

B2 WEIGHING INDICATOR
Table of contents
1 Introduction.......................................................................................................................................3
1.1 Attention Before Start................................................................................................................4
2 Interface.............................................................................................................................................
2.1 Display.......................................................................................................................................
2.2 Keypad.......................................................................................................................................6
3 Connection Diagram..........................................................................................................................7
4 Configuration.....................................................................................................................................8
4.1 Adjustment (Calibration) switch................................................................................................8
4.2 Configuration menu.................................................................................................................10
4.3 Weighing parameters................................................................................................................11
4.4 Adjustment ( Calibration ).......................................................................................................13
4. Miscellaneous Application Settings.........................................................................................14
4.6 Serial Output Settings..............................................................................................................1
Operation.........................................................................................................................................17
.1 Weighing with tare...................................................................................................................17
.2 High resolution display............................................................................................................17
.3 Multi range weighing...............................................................................................................17
6 Appendix..........................................................................................................................................18
6.1 Error messages.........................................................................................................................18
6.2 Specifications...........................................................................................................................19
v1.00 Page 2
USER MANUAL

B2 WEIGHING INDICATOR
1 Introduction
The B2 is a universal digital weighing indicator for analog load cells, which incorporates:
○Delta-sigma analog to digital converter for accurate weighing.
○nternal alibi memory for recording weighing results.
○RS-232 ports for connections to printers, remote displays, computers or other digital
terminal equipments.
○LED display.
This manual contains the information needed to configure, adjust and operate the device. To ensure
safe operation, the instructions in this manual shall be respected.
v1.00 Page 3
USER MANUAL

B2 WEIGHING INDICATOR
1.1 Attention Before Start
●Always use Earthed/Grounded Mains socket.
●Do not connect powerful electrical machines on the same mains plug.
●Disconnect mains connector before opening the enclosure.
●Recommended load cell cable cross section is 1 mm². f the cable length is too long, use multiple
wires for ±Excitation connections.
●f the sense inputs on the load cell connector are left empty, the device will display Err 3. Make sure
all the pins are connected.
●Shielded twisted-pair cables must be used for load cell and communication.
●For best accuracy, the device shall be powered on at least 30 minutes before adjustment.
●Device setup shall be performed by authorized people.
v1.00 Page 4
USER MANUAL

B2 WEIGHING INDICATOR
2 Interface
2.1 Display
Weight
display
6 digit 7-segment led display with decimal point
(Negative values are displayed with a minus sign in the leftmost digit)
R1 , R2 Range indicator: f configured as a multi range instrument, indicates the current
weighing range.
enter-of-zero indicator: ndicates the deviation from zero is not more than ±0.25 e.
NET Net indicator: ndicates that a tare device is in operation and the displayed weight
value is a net weight.
Stability (No-motion) indicator: ndicates that stability of equilibrium has been
reached.
kg Unit of weight: ndicates the unit of the weight, if a weight value is displayed.
BAT Low battery indicator: f flashing, indicates a low battery condition.
FFunction indicator: ndicates that a function is in operation and the displayed value is
a calculated value. A sticker identifying the displayed value (e.g. unit) shall be attached
here for the configured function.
v1.00 Page 5
USER MANUAL

B2 WEIGHING INDICATOR
2.2 Keypad
Zero key: Set the indication to zero. Zeroing can be established only if the gross weight
is within ±2% of the maximum capacity.
Tare In / Tare Out key: f the indicated value is positive and stable, it is saved as tare
and net weight is indicated. f there is already a tare device in operation, it is cancelled
and gross weight is indicated.
Print key: f the indicated value is positive and stable it is saved as a weighing result
and a print out is produced (if configured).
Star key: Call up the configuration menu. Please refer to the configuration section.
Function key: Activates and deactivates a configured function. For a complete
explanation refer to the related section of this manual.
Note: For an explanation of the function of the keys while editing numerical values please refer to the
configuration chapter.
v1.00 Page 6
USER MANUAL

B2 WEIGHING INDICATOR
3 onnection Diagram
Power 8 – 264 VAC, 0.4 A, 0/60 Hz
Note: Always use Earthed/Grounded Mains socket.
Load Cell
Pin Value
1 - Signal
2 + Signal
3 - Sense (Note: If load cell is 4-wire connect -Excitation here)
4 + Sense (Note: If load cell is 4-wire connect +Excitation here)
Ground - Shield
6 + Excitation
7 - Excitation
USB
RS-232 Pin Signal PC Connection
(Type-D 9-Pin Male)
2 Tx 2 (Rx)
3 Rx 3 (Tx)
7 Gnd (Gnd)
Remote Display Pin Signal ERTE HG 7 Remote Display Connection
(Type-D 9-Pin Female)
1, 2,
3NC No connection
4, , 6, 7, 8, 9 4, , 6, 7, 8, 9 one-to-one
v1.00 Page 7
USER MANUAL

B2 WEIGHING INDICATOR
4 onfiguration
4.1 Adjustment ( alibration) switch
A toggle switch located on main pcb determines the mode of configuration. f the switch is in “LOCKED”
position, it is not possible to alter weighing related settings. f the switch is in “UNLOCKED” position, it
is possible to edit weighing parameters and adjustment data.
v1.00 Page 8
USER MANUAL
Main board Display board
Keypad board
Connectors
INTERNAL VIEW

B2 WEIGHING INDICATOR
v1.00 Page 9
USER MANUAL
ON
OFF
ADJUSTMENT SWIT H
1 2
ON
OFF 1 2
n this position, adjustment
is OFF / LOCKED
n this position, adjustment
is ON / UNLOCKED
Note: Position of switch #2 has no effect

B2 WEIGHING INDICATOR
4.2 onfiguration menu
The position of the adjustment switch determines the method of entering the configuration menu and its
contents. f the adjustment switch is in “UNLOCKED” position, press key to enter configuration
menu. Contents of the menu:
pAR Weighing parameters
AyAr Adjustment ( Calibration )
tFon Miscellaneous Application Settings
Ser RS232 Serial Port Settings
Ser2 Usb Settings
Fab_aY Reset All Settings
The function of the keys inside the menu are:
Exit: Return to the weighing display.
Previous: Display the previous entry.
Nest: Display the next entry.
Enter: Enter the displayed sub-menu.
ancel: Return to a higher menu level.
When user input is needed while changing a device setting, you are prompted to edit a numerical value.
For this purpose, a numerical value is displayed with a flashing digit. During the edit procedure the
function of the keys are:
lear: The whole value is reset to zero.
Next digit: The flashing moves to the next digit at right. f the rightmost digit is flashing,
the flashing moves to the leftmost digit.
Increment digit value: ncrements the value of the flashing digit by one. f the digit has
a value of 9 the value becomes 0.
Accept: Finish edit procedure by accepting the new value. f the new value is not
acceptable (e.g. out of limits) the edit procedure re-starts with the last value.
ancel: Cancel current edit procedure, return to previous parameter editing (if any).
v1.00 Page 10
USER MANUAL

B2 WEIGHING INDICATOR
4.3 Weighing parameters
Press key while “ PAr “ message is on display. Weighing parameters are edited one at a time.
Edit each value by using and keys; press to accept and proceed with the next
parameter. You may press and go back to the previous parameter while editing. When all the
paremeters are accepted, the device saves the changes automatically.The parameters are:
1. apacity 1
n single range instruments, this is the maximum capacity.
n multi (two) range instruments, this is the maximum capacity of the first
range (smaller range).
2. apacity 2
n single range instruments, this value have to be entered as 0 (zero).
n multi (two) range instruments, this is the maximum capacity of the
second range (greater range).
3. Scale interval (e)
This value is the difference between two consecutive indicated values (e).
n multi range instruments, this is the interval of the smaller range (e1). The
device selects e2 automatically.
( Accepted values: 1,2,5,10,20,50 )
4. Decimal places
This value determines the place of the decimal dot while indicating weight
values.
( Accepted values: 0,1,2,3 )
5. Stability range
Stability of equilibrium is said to be reached, when the rate of change in displayed weight values
falls below a predetermined rate. This predetermined rate depends on 'Stability range' (SR) and
'Stability delay' (SD). f the difference of the maximum and the minimum weight values
calculated in a preiod of SD is less than or equal to SR, stability condition is satisfied.
The range for displayed weight values, for determination of the stability.
( Accepted values: 0 .. 5 interval(s) )
Note: A value of zero means 0.5 interval
v1.00 Page 11
USER MANUAL
c1
c2
t 02
. 3
ht 1

B2 WEIGHING INDICATOR
6. Stability delay
The time period for displayed weight values, for determination of the
stability.
( Accepted values: 1 .. 5 second(s) )
7. Filter
This value determines the interpretation of the changes in weight. A lower
value causes fast reactions to weight changes and is suitable in relatively
stable conditions. A higher value causes slow reactions but the vibrational
flickers are eliminated.
( Accepted values: 1 .. 5 )
8. Zero at power-on
This setting determines if the indication is set to zero automatically at
power on. (initial zero setting)
( Accepted values: 0 (off) , 1 (on) )
v1.00 Page 12
USER MANUAL
hc 1
Fb 3
A 0

B2 WEIGHING INDICATOR
4.4 Adjustment ( alibration )
Press key while “ AyAr “ message is on display. There are 3 options for proceeding.
Dead load adjustment:
This option may be used when only dead load (zero point) adjustment is
needed.
While this message is displayed, unload the platform and press key
when there is no motion. The display will become blank until the device
calculates and saves the calibration data. After that, the device exits the
menu and returns to the weight display.
Span adjustment:
This option may be used when a dead load adjustment is not needed, but a
span adjustment is needed. (The device assumes the last zero point for a
reference for the adjustment weight.)
While this message is displayed, press key. The dead load
adjustment (zero calibration ) is skipped.
Enter (edit) the value of the weight which will be used for adjustment.
Accept with key.
The value you entered will be displayed constantly. Load the platform with
the adjustment weight you entered and press key when there is no
motion. The display will become blank until the device calculates and
saves the calibration data. After that, the device exits the menu and returns
to the weight display.
Full calibration: This option is performing the dead load adjustment and the span
calibration consecutively.
While this message is displayed, unload the platform and press key
when there is no motion. The display will become blank until the device
calculates and saves the calibration data.
Then Enter (edit) the value of the weight which will be used for calibration.
Accept with key.
The value you entered will be displayed constantly. Load the platform with
the calibration weight you entered and press key when there is no
motion. The display will become blank until the device calculates and
saves the adjustment data. After that, the device exits the menu and
returns to the weight display.
v1.00 Page 13
USER MANUAL
SıFır
SıFır
015000
15000
SıFır
015000
15000

B2 WEIGHING INDICATOR
4.5 Miscellaneous Application Settings
This setting is used to configure miscellaneous application settings like automatic tare, automatic
printing etc.
Press key while “ tFon “ message is on display.
Automatic Tare
0 – Off
1 – On
Automatic Tare Cancel
0 – Off
1 – On
Automatic Printing
( This option is only for indicators with integrated mini printer )
0 – Off
1 – On
Automatic Tare Limit (e)
0 – Off or
# of e divisions
Automatic Printing Limit (e)
( This option is only for indicators with integrated mini printer )
0 – Off or
# of e divisions
v1.00 Page 14
USER MANUAL
od 0
di 0
oy 0
dL 0
yL 0

B2 WEIGHING INDICATOR
4.6 Serial Output Settings
Press key while “ ser “ or “ ser2 “ message is on display. Device parameters are edited one
at a time. Edit each value by using and keys; press to accept and proceed with the
next parameter. You may press and go back to the previous parameter while editing. When all the
paremeters are accepted, the device saves the changes automatically.
RS232 Output Mode
0 – Continuous Output Mode
1 – Terminal Mode
* See below for details on Continuous and Terminal Mode protocols.
Baud Rate
0 – 9600
1 – 4800
2 – 2400
3 – 1200
4 – 19200
Output Mode Type
( f you don't need different serial output format, don't change the
defaults )
0 – ERTE B2 series output format
v1.00 Page 15
USER MANUAL
f 0
baud 0
tip 0

B2 WEIGHING INDICATOR
* ontinuous Mode Output Format
Protocol: 9600 baud, 8 data bits, 1 stop bit, no parity (9600-8-N-1) ( Default )
Data: (8 characters)
Character Açıklama
1. character
Astable, no tare
Bstable, have tare
Cno stability, no tare
Dno stability, have tare
Eunderload / overload
2-7. characters 1 2 3 4 6 display value with 6 digits ASC , no decimal point, no
unit (e.g. ' 1500')
8. character ↵Carriage Return (ASCII 13)
* Terminal Mode Output Format
9600 baud, 8 data bit, 1 stop bit, no parity (9600-8-N-1)
n this mode, the indicator responds to specific commands via the serial interface. The commands and
the responses are as below :
Command Response
d↵Sends the indication value as seen in the display in the
form of continuous mode protocol
z↵Zero Key No response
t↵Tare Key No response
p↵Print Key Outputs weighing result in COM1 alibi record output format
if saving is successful.
v1.00 Page 16
USER MANUAL

B2 WEIGHING INDICATOR
5 Operation
After applying power to a properly installed and configured device, the power-on tests are performed.
During these tests all the segments (indicating devices) on the display are turned on. After completition
of these tests the device performs initial zero setting, if configured. At the end, the weight display
appears.
The weight display is the primary display of the device. The measured weight values are displayed with
the unit of weight sign turned on (e.g. kg indicator). Other indicating devices which accompany the
weight value are explained in the introduction chapter of this manual.
5.1 Weighing with tare
n order to tare the weight on the platform, press the key. The indicated weight value is saved as
tare value and the indication switches to net mode. The “NET” symbol indicates this situation. To save
the weighing result and get a printout press the key. To cancel the tare and return to gross
weighing mode press the key again or unload the platform and press the key. Unstable and
negative weight values can not be tared.
5.2 High resolution display
This function needs the function key configured for high resolution display.
t is possible to temporarily display measured weight values in 10-fold increased resolution. To perform
this function press the key. The indicated weight value is displayed in high resolution for 5
seconds. At the end of this period, the indication returns to normal resolution mode automatically. t is
also possible to turn off high resolution mode by pressing the key without waiting for 5 seconds.
5.3 Multi range weighing
f the device is configured as a multi-range instrument, the range which is currently active is indicated
on the display by “R1” or “R2” indicators. Range selection from the smaller range (R1) to the greater
range (R2) is performed automatically when the indicated load exceeds the maximum capacity of the
smaller range. Range selection from the greater range (R2) to the smallest range (R1) is performed
automatically when indication is zero or at a negative net value equal to gross zero, or manually when
the key is pressed when there is no load. Tare is cancelled automatically.
v1.00 Page 17
USER MANUAL

B2 WEIGHING INDICATOR
6 Appendix
6.1 Error messages
Message Meaning Solution
****$ Overload. Platform is loaded beyond
maximum capacity + 9e.
Unload platform.
%____& Underload. Weight is less than zero setting
limit. Deadload may be changed.
Check the platform.
hata 1 Error in non-volatile parameter memory.
Weighing parameters/adjustment data are
missing or corrupted.
Configure the indicator. f the error
repeats call service.
hata 2 Possible hardware failure in analog circuit. Call service.
hata 3 Signal error. Possible disconnection or
erroneous connection of load cell cable.
Check load cell connection.
hata 4 nitial zero setting error. Weight value is
above or below zero setting range.
Check the platform.
hata 92 (when performing an adjustment)
Calculations show that load cell signal may
be out of range for weighings near capacity
(max).
The adjustment will be accepted, this
is just a warning.
hata 99 (when performing an adjustment) Calculated
load cell signal per scale interval is too low.
Check weighing parameters and/or
adjustment weight.
Press any key to close error message displays.
Note: Some errors can not be cancelled until the cause of error disappears. (e.g. hata 3)
v1.00 Page 18
USER MANUAL

B2 WEIGHING INDICATOR
6.2 Specifications
Power Supply 85 – 264 VAC, 50/60 Hz, maximum 15 Watt
Enclosure Stainless Steel
Display 6 digit, 25 mm bright red display
nterval, zero, stability, net, unit and function indications
Keypad 5 mechanical keys
(Zero, tare / tare cancel, print and function keys)
Loadcell 4 or 6 wires
Excitation Voltage: 5V DC
mpedance : 43 ... 1100 Ω (Ex: 8 x 350 Ω)
Cable Length : Maximum 350m (≥0.75mm², 6 wire cable)
Analog
Specifications
O ML Class 3000e, Class 1000e, single range / multi range
ndustrial applications 40000e
Minimum input-voltage per verification scale interval: ≥0.8μV/e (approved)
≥0.2μV/e (industrial)
nput Signal Range: 0 ... 20 mV
Measurement Method: Delta-sigma A/D, 20ms ... 1s update rate
nternal Resolution: 6,700,000 adım
Linearity: <0.007%
nterfaces RS-232 serial port
USB connection (PC)
ERTE HG External Display
Environmental Operating Temperature: -10°C ... +40°C
Storage Temperature: -30°C ... +80°C
Electromagnetic Compatibility: EC 61000-4-2 (Electrostatic discharge)
EC 61000-4-3 (Radiated radio frequency field)
EC 61000-4-4 (Electrical fast transient burst)
EN 55011 (Emmision)
Dimensions 210mm (width) x 185mm (length) x 110mm (height)
Approvals
EU Directive EN45501 , EMC - EN 61326-1:2006 and EN 55011:2007
v1.00 Page 19
USER MANUAL
Table of contents
Popular Accessories manuals by other brands

Chamberlain
Chamberlain 915GA install guide

PCB Piezotronics
PCB Piezotronics IMI Sensors CS640B92 Installation and operating manual

Honeywell
Honeywell PPMINI user manual

logika
logika WL-ZSSMBPW-FD-01 product manual

DuroZone
DuroZone MB Series Multi-Blade Damper installation instructions

Grunbeck
Grunbeck GENO-UV-75 Operation manual

Joolz
Joolz Comfort cover instruction manual

turck
turck BI10-G30-Y2X 7M manual

NetPeppers
NetPeppers VLP50 Quick reference guide

Moog
Moog Tritech Super SeaKing 700 V7 Series product manual

Stearns
Stearns 57,500 Series Installation and service instructions

Salus
Salus SS912ZB Installation & quick start guide