STEP FOUR
STEP TWOSTEP ONE
What you’ll need...
VECTOR FIELD CONTROL UNIT
SENSOR MAGNETIC CALIBRATION GUIDE
1.a 1.a
TAP the Button with the
Magnetic Wand (MW) to access
the Menu.
2. Calibration Cap
1.b
TAP the Buttons to highlight
“Calibrate 1” .
2.a
Connect the zero gas cylinder to the
inlet of the calibration cap.
2.b
Apply the zero gas at NO MORE than
0.5 l/min, for at least 1 Minute; wait
for gas reading to STABILIZE.
2.c
STEP THREE
TAP the Button to Set zero
point calibration.
OVERVIEW
2.c
2.a
3.a
TAP the Buttons to adjust
number of gases for span calibration
3.b
3.c
Select 1 for mid-span gas only (1 gas cylinder)
Select 2 for full & mid span gases (2 gas cylinder)
+
+
TAP the Button to Select
“Calibrate 1” and begin calibrating Sensor 1.
1.c 1.c
2.b
+
3.a
3.c
TAP the Button to Enter selection
+
4.a
TAP the Buttons to adjust value
of each digit. Refer to the gas concentra-
tion on the gas cylinder
4.b
4.c
TAP the Button after the last digit
has been edited to proceed
+
TAP the Button to Advance
cursor to the next digit.
+
BACK / CANCEL DOWN Button
UP Button
START / ENTER
4.b
4.c
1. Zero Gas: a gas that is free of any substance that will
adversely aect the operation of or cause a response of the
instrument.
2. Mid span Gas: a reference gas used in calibration of a
concentration _20% from 50% of sensor’s full range.
3. Full span Gas: a reference gas used in calibration of a
concentration higher 75% of sensor’s full range.
Tools:
Gases:
1.b
1. Magnetic Wand 3. 1/4” Clear Tubing
TAP the Magnetic Wand against glass to activate buttons.
+