Factair F8104 User manual

0
F8104 SAFE-AIR MONITOR
OPERATING MANUAL
FAC QAM 221 Issue B 04/20

1

2
Contents
Page No
Introduction
3
Calibration & Warranty
3
Diagram & Display Icons
4
Equipment List
5
Technical Data
6
1
-
Pre Use
7
2
-
Start Up
7
3
-
Continuous Monitoring Display
3.1 –Completing an oil test using a Dräger Impactor or
tube
8
9
4
-
Configuring the Instrument
4.1 –Selecting a test standard
4.2 –Configuring the main alarm
4.3 –Configuring the pre alarm
4.4 –Auxiliary inputs / outputs
4.5 –Sensors
10
10
11
14
14
15
5
-
Exporting the Data Log
16
Appendix 1 –Using the test port for the Dräger oil mist tube
17
Appendix 2 - Reading the Dräger Oil Impactor and Chemical Reagent Tube
19
Appendix 3 –Volatile Organic Compounds
22
Appendix 4 - Helpful Tips
23
Air Quality Testing –Why?
Medical Air Standards
24
25

3
Introduction
The F8104 Safe Air Monitor is designed to provide continuous monitoring of breathing-air and
compressed-air systems. Once set up the instrument will continue to monitor and data log the air
quality at user defined intervals, validating the air quality against a range of alternative international
standards. For medical, surgical, dental air and medical gas the F8104 is programmed to test against
the requirements of European Pharmacopoeia and in the UK, HTM 02-01.
The F8104, is designed primarily for use on airline systems up to 10 bar but can also test higher
pressure systems up to 20 bar when used in conjunction with the F3005 High Pressure Regulator.
High pressure cylinder air can also be tested with the F3002 HP regulator.
The F8104 is supplied in a plastic protective case with storage provision for accessories.
The test is carried out using built in electrochemical cells for carbon monoxide and oxygen; infrared
LED sensor for the carbon dioxide and a Photo Ionisation Detector (PID) for Volatile Organic
Compounds. The moisture is measured with an electronic dewpoint meter and the instrument also
records pressure and ambient temperature. For measuring oil aerosols, Dräger Impactors are used in
the side mounted test port (or alternatively the Dräger oil chemical reagent tube). The Dräger
Impactors are supplied in packs of 10 and show the degree of oil contamination on the cut glass baffle
plate.
Calibration and Warranty
Safe-Air Monitors leave our factory with a 12-month warranty and calibration certificate. If the
instrument is returned to Factair for annual calibration, our standard turnaround is 10-15 working
days providing there is no major damage that requires an extensive rebuild. Note:- Please download
all stored data before returning the instrument to Factair.
Temperature Parameters
Storage: -10/+50°C Operating Range: -5/+40°C
IMPORTANT –IT IS RECOMMENDED THAT YOUR SAFE-AIR MONITOR IS RECALIBRATED AND
SERVICED WITHIN 12 MONTHS FROM THE ISSUE DATE OF ITS CALIBRATION CERTIFICATE

4
Air Inlet
230V 50Hz
Mains Input
PC - USB Connection
Dräger Impactor / Oil
Mist Tube Test Port
On / Off Button
Compressed
Air Inlet
Ethernet
Connection
2 no. 4-20mA
Input
2 no. 4-20mA
Output
2 no. Alarm Volt
Free Contacts
Right Hand Side
Bottom

5
Home Screen Display Icons
Standard Equipment List
Optional Extras
F8104 Safe-Air Tester
TUB0053 –8103560 Dräger Oil Impactor –Box of 10
USB Download Cable
TUB0003 –6728371 Dräger Oil Tubes –Box of 10
Stylus Pen
F3002 High-Pressure Regulator
Wall Mounting Plate
F1946 Connection ¼ BSP inlet adaptor
ASSY0013 supply hose with rectus 21
plug and socket connections
F2193 Safe-Air Tester to Dräger/Scott compatible plug
F1946 Connection ¼ BSP inlet adaptor
with rectus 21 socket
F2194 Connection adaptor terminating in a Rectus Plug
F2195 Connection adaptor terminating in a Instantair
plug
Selected Test Standard
Settings
Battery Charge
& Charging
Indicator
Installer
Access Status
SIM Card
Signal
Red –No
Signal
Green -
Signal
Ethernet
Connectivity
status
SD Card
Memory
Status
Unlocked
Locked
Supply
Pressure
By pressing
the icon you
can change
display units
between
barg and
psig
Supply
Temperature
By pressing
the icon you
can change
display units
between °C
and °F
Time and
Date
By pressing
the icon you
can adjust
the date and
time
Sensor
Status
= OK
----------
= Sensor
Out of
Calibration
only :
Dewpoint close
to
measurement
limit
----------
= Sensor
Fault
only :
Last test failed
or not yet
tested

6
Technical Data
Model
F8104
Minimum Working Pressure
2 barg
Maximum Working Pressure
10 barg
Air Inlet Connection
Rectus Series 21 Plug
Internal Rechargeable Batteries
Lithium Ion 3.7V Capacity : 15,600 mAh
Supply Voltage
230V/1ph/50Hz 13A supply
Internal Memory
8 GB
Sensors
Range
Sensor Type
Oxygen (O2)
0 –25%
Electrochemical
Carbon Monoxide (CO)
0 - 20 PPM
Electrochemical
Carbon Dioxide (CO2)
0 –2000 PPM
Non-Dispersive Infrared
Volatile Organic Compounds (VOC)*
Nitrogen Oxide (NOₓ)
Sulphur Dioxide (SO₂)
0 –40 PPM
0 –10 PPM
0 –10 PPM
10.6 eV Photoionisation Detector
Electrochemical (NO and NO2sensors)
Electrochemcial
Moisture
-100°C to
+20°C
PDP
Ceramic Moisture Sensor Dewpoint
Meter
Pressure
0 –10.0 barg
Transducer
Test Port –Compatible with the
following Dräger Impactors / Tubes
8103530
TUB0053
Oil Impactor
6728371
TUB0003
Oil Chemical Reagent Tube
Dimensions
315 mm (h) x 330 mm (w) x 105 mm (d)
Weight
5 kg
* Optional 0-3 PPM VOC sensor also available, with this sensor fitted the unit can be pre-
set at Factair to display in mg/m3referenced against isobutylene calibration gas

7
1. Pre-Use
The F8104 is supplied with a mounting plate which enables the instrument to be wall mounted at the
most suitable location to test your breathing-air / compressed-air supply. The test location needs to
be as close to the breathing-air supply point as possible, in a well ventilated area free from dust and
with a 230V mains power supply available. To install the back plate mark the hole locations on the
wall and check with a spirit level. Ensure you use the correct fixings to suit the wall.
Once you have installed the back plate mount the F8104 in place and connect the power supply. It is
important the unit is free from vibration during use. The F8104 Safe-Air Monitor requires a
230V/1ph/50Hz mains power supply.
The maximum compressed air supply pressure for the F8104 Safe-Air Monitor is 10 barg. If the
instrument is intended for use on a high pressure system contact Factair for advice on suitable
regulators and connection arrangements.
2. Start-Up
To turn the instrument on press and hold the “on” button, located on the right-hand side of the unit,
for approximately 3 seconds.
A stylus pen is provided for use with the touch
screen. The display will show the calibration
date, serial number and check the status of
each sensor. Once each sensor status has been
checked follow the touch screen prompt on
screen to proceed.
The compressed-air supply can then be
connected to the instrument
(minimum 2 barg, maximum 10 barg).

8
3. Continuous Monitoring Display
By tapping the calibration date screen, the
instrument will then display the home screen.
In the top right hand corner a countdown will
commence for a commissioning delay during
which time the sensor readings are stabilising
and the alarms will be disabled.
The duration of the commissioning delay can
be later adjusted in the installer menu.
If the sensor readings have already stabilised
you can commence immediate sensor
monitoring by pressing the “Commissioning
Delay” countdown icon.
Once commissioning is complete the home
screen will display all live sensor readings.
Individual sensor readings can be further
interrogated by pressing on their readings bar.
The number and order of the displayed
sensors can be changed in the installer menu.

9
3.1 Completing an oil test using a Dräger Impactor or tube
The instrument displays the date and result from
the last oil Impactor or tube test.
To undertake a new test tap the result button.
Choose between using a Dräger oil Impactor or
Dräger chemical reagent tube in the test port.
For breathing-air tests the most commonly used
option is to select a Dräger oil Impactor.
Instructions for using the Dräger tube can be
found in appendix 1.
The instrument will now commence a 5-minute
purge sequence, during which time an airflow
will be passed through the test port. This clears
contaminants from previous tests and ensures
that a representative sample is being tested.
Once the purge has finished, the Dräger
Impactor/oil tube should be inserted into the
test port and tightened firmly; ensure the
Impactor has an intact protective seal. This seal
must remain in place for the duration of the
test. Press the arrow on the right hand side to
start the test.
At the end of the test the Impactor result screen
will be shown. The Impactor should now be
taken out of the test port, sticker removed and
evaluated.
To enter the results press the corresponding icon
on the screen.
Note:- Instructions on reading the Impactor can
be found in Appendix 2 at the back of the
manual.
The test result and date of test will then be
displayed on the home screen.

10
4 Configuring the instrument
4.1 Selecting a test standard
To enter the installer mode press this icon,
which is located in the top right hand corner of
the screen near the battery icon.
To change the test standard, from the home
screen, press on the current standard.
Enter user log in password “installer”.
The standards menu will then display the
available test standards.
As well as medical standards the F8104 can also
test breathing-air to EN12021:2014. This
standard has 4 different requirements
depending if you are testing an airline system
<40 bar; Cylinder <200 bar; Cylinder >200bar or
direct from a high pressure charging
compressor. By toggling the icon at the top of
the screen the default limits will be displayed.

11
If you also want to test for Volatile Organic
Compounds then by clicking on the adjacent box
you can set a maximum limit. For further
guidance on VOCs see appendix 3.
Click on the tick to confirm the VOC value you
have entered as a test parameter.
Once you have selected the appropriate
standard press the save button to update the
change or alternatively to discard press the
waste basket icon.
4.2 Configuring the main alarm
To enter the set up mode press this icon,
which is located in the top right hand corner
of the screen near the battery icon.
Enter user log in password “installer”.
The F8104 has 2 configurable alarms, a main
and pre alarm. Both of the configurations
pages are located in the settings pages.
The alarm values are based on the selected
alarm standard. The values can be changed
from this screen.
If a maximum limit for VOC has been inputted
then the alarm will also activate if this is
exceeded.

12
Relay 1 and 2 are used to set each alarm.
These correspond with the alarm output
connections located at the bottom of the
instrument.
If Relay 1 is selected for Main Alarm only,
Relay 2 can be selected for Pre Alarm.
The ratings for volt free contact is 1A at 24V
AC/DC.
2 matching plugs are provided with each
F8104.
The action of the contacts can be adjusted in
the software from active to passive by
selecting the Inverse tick box.
This is an optional, factory fitted feature
which allows a sounder to be fitted inside or
on the unit itself. By selecting the Sounder, an
audible tone will be emitted in the event of
any of the readings exceeding their minimum
and maximum value.
This is an optional, factory fitted feature
which allows a beacon to be fitted on the unit
itself.
If Power Fail is selected, then in the event of a
failure of the mains power supply, it will
activate the alarm.
To function correctly the F8104 requires a
minimum 2 barg supply pressure (10 barg
maximum). In the event the supply pressure
falls below 2 barg, the unit will activate its
alarm. For situations where the drop in
pressure may only be temporary a delay can
be added (in seconds) before the alarm is
activated.
1
Normally closed
2
Common
3
Not Used
4
Normally open
1
Sensor Out
1
-
2
Sensor V-
2
+
3
Sensor V+
3
Not Used
4
Not used
4
Not Used

13
An Alarm delay (in seconds) may be set,
between the sensor values exceeding their
minimum/maximum values and when the
alarms will be triggered.
A commissioning delay may be set to prevent
the unit activating the alarms when the
instrument is first turned on, allowing time
for the sensors to acclimatise.
If a SIM card is fitted in the instrument the
SMS Message feature can be enabled by
selecting this tick box. When activated a
message will be sent when any of the
readings reach their alarm values.
The SMS alarm message can be altered by
pressing on the text.
The SMS contact number can be set by
clicking on the existing number.
Once you have configured the “Main Alarm”
press the arrow button in the top right
section of the screen to access the Pre Alarm
menu. Alternatively you can save or discard
the changes and return to the home screen.

14
4.3 Configuring the pre alarm
The Pre Alarm screen provides the option to
set a lower alarm value which can be used to
notify the user before the main alarm values
are exceeded.
As with the main alarm the pre-alarm has a
range of configurable options including an
alarm output, sounder, pre alarm delay and if
a SIM card is fitted, an SMS message.
4.4 Auxiliary Inputs and Outputs
The F8104 can display readings from external
sensors which have a 2 wire, 15V, 4 –20 mA
output.
The description, units of measure and the 4
and 20mA parameters, for the sensors need
to be set to match the external device(s).
It is also possible to program 4-20mA outputs
for 2 of the F8104 sensors. These can be used
if you need remote indication for these
sensor values.
By selecting the Auxiliary Output 1 and
Auxiliary Output 2 fields, you can choose
which sensors you wish to output.

15
4.5 Sensors
The order in which sensors are displayed on
the home screen can be set, you can also
choose which sensors you want to be
displayed.
Any changes need to be saved to take effect.

16
5. Exporting Test Results and PC Software.
To transfer the datalog from the F8104 Safe-Air Monitor to the PC software you must first install the
software on your PC. This is available in the download section of Factair’s website:
http://www.factair.co.uk/downloads
To transfer the datalog from the Safe-Air Tester to the PC software, turn on the F8104, connect the
USB cable between the F8104 and your PC and open the Safe-Air Tester Results F8000 series
software.
With the F8104 successfully connected the
status light will turn green.
The unit will then automatically read the
instruments firmware version, saved test
results and the size of the data monitoring file.
The data log file can be downloaded by
clicking on this button. Results are stored in a
CSV format.
The monitoring log file on the F8104 SD card
can also be deleted using this software.
When complete you can save the test result file.

17
Appendix 1 - Using the test port for the Dräger oil tube
As well as the Dräger oil Impactor the F8104 test port can also be used with the Dräger oil tube.
Test
Dräger ref.
Factair Part
No:
Sensitivity
Oil (chemical reagent tube)
6728371
TUB0003
0.1 mg/m3
To prepare the tube Factair recommends the F2187 Dräger tube tip cutter.
If you have another Dräger tube, which you wish to use with the instrument, contact Factair for
technical advice on its suitability.
To complete a tube test, select a tube test
option as shown, press the button to proceed.
If you can determine with a high degree of
certainty the specific compressor lubricant
being used, then refer to the “test times” chart
provided on Factair’s website marked for the
F8104. If you have no knowledge of the type of
compressor lubricant being used in the
breathing-air system then Factair recommends
15 minutes as this covers the vast majority of
lubricants.
Once the instrument has completed its purge
sequence you will be prompted to insert the
tube into the test port and tighten firmly.
Note: Both ends of the tube should only be cut
immediately prior to the test. Ensure the arrow
printed on the tube is pointing away from the
instrument.
Press the right hand arrow to proceed.

18
The test will proceed as described in section 4
and at the end of it you will be prompted to
enter the result from the tube.
The result will then be displayed on the final
test result screen and stored in the instrument’s
datalog file.

19
Appendix 2 - Reading the Dräger Oil Impactor and Oil Chemical Reagent Tube
Oil Impactor
The Oil Impactor’s protective seal must be kept in place for the duration of the test and only
removed after the test has been completed.
With a standard measurement range of 0.1 to 1.0 mg/m3the Impactor has a limit of detection of
0.05 mg/m3. The Impactor can detect all mineral and synthetic oil mists, it features a series of 3
horizontal lines, each formed by a series of precision manufactured nozzles. These horizontal lines of
nozzles are calibrated to a different concentration of oil. When air is passed through these nozzles
any oil present in the air is deposited on the glass plate prior to the air being exhausted by vents
around its circumference, this allows the user to easily and quickly identify minute quantities of oil.
For tests that pass with an oil concentration of less than 0.05mg/m3, the screen will remain blank. For
tests above that read as follows:
The bottom line represents an oil concentration of 0.1 mg/m3. When the oil deposited forms a
continuous line then the concentration is in excess of 0.1 mg/m3. The middle line represents a
concentration of 0.5 mg/m3, again when the oil deposited forms a continuous line the concentration
is in excess of 0.5 mg/m3. The top line represents a concentration of 1.0 mg/m3, again when this
forms a continuous line the concentration is in excess of 1.0 mg/m3.
* –The above pass and fail criteria is based on breathing-air standard EN12021
Pass* -
greater
than
0.05mg/m3
but less
than
0.1mg/m3
Pass* -
greater
than
0.1mg/m3
but less
than
0.5mg/m3
Fail* -
greater than
0.5mg/m3
but less than
1.0mg/m3
Fail* -
greater
than
1.0mg/m3
Table of contents
Other Factair Measuring Instrument manuals