FujiFilm ID-T741 User manual

Service Manual
FUJI COMPUTED RADIOGRAPHY
FCR ID Terminal/Console
ID-T741
Fuji Photo Film Co., Ltd.
Printed in Japan
Document No. 009-052-04
1st Edition: April
10,1998
5th Edition
: Mar.
30, 2000

009-052-04
03.30.2000 FM2762 (ITC)
2FCR ID-T741
Service Manual
All rights reserved. No part of this publication may be reproduced, stored
in a retrieval system, or transmitted in any form or by means, electronic,
mechanical, photocopying, recording or otherwise, without the prior written
permission of Fuji Photo Film Co., Ltd. Equipment Product Division,
Computed Radiography Department.
FCR®is a registered trademark of Fuji Photo Film Co., Ltd.
Copyright©1998–2000 by Fuji Photo Film Co., Ltd.
Precautions in Using the Service Manual
This manual should not be used by anyone other than the servicing
staffmembersqualified throughthe training specified byourcompany.
Since this manual contains confidential information including
engineering know-how related to product development,it is prohibited
to distribute the manual to users or to allow them free access to the
manual without obtaining prior permission from Fuji Photo Film Co.,
Ltd.
Thismanualis intended to assistthe serviceperson in thesafe andefficient
servicing of the equipment described.
Theequipment mustbe used in accordance with the procedurescontained
in this manual and must not be used for purposes other than those which
are describe herein.
The equipment should only be used by persons having recognized
qualifications and, if relevant, having adequate training on this equipment,
especiallyregarding protectivemeasure such as laser radiationprotection.
It is the responsibility of the user to ensure that existing legal regulations
regarding installation of the building with respect to the equipment are
observed. Moreover the legal regulations regarding the operation of the
equipment are to be observed.
Incorrect operation, or failure of the user to maintain the equipment in
accordance with the schedule of maintenance, relieves the manufacturer
orhis agentfrom all responsibilityforconsequentnon-compliance, damage,
injury, defects and/or other malfunction.

009-052-01
12.20.98 FM2370 (ITC) FCR ID-T741
Service Manual
1
ABOUT THIS MANUAL
AboutThis Manual
■Product denomination
In this manual, the product code name, IDT-IV is used for simplification purposes, instead of the
official product name, ID-T741.
Also, IDC-IV is used instead of ID-T742 for the same purposes.
■Menu or button selection on the display
To select a menu or a button (icon) displayed on the screen while in the service mode, lightly touch it
with the tip of a finger.You need not press hard.
Because this screen detects a selected position using infrared rays, pressing the menu or the button
with a finger or hand flat down or touching the whole palm of the hand on the screen may result in
selecting a different button or even disabling any selection.
Safety
Observe the following safety precautions to avoid possible hazards and accidents during maintenance
servicing.
■Power supply
To avoid electric shock hazards and machine malfunction, be sure to turn OFF the power before
proceeding with parts replacement. Use due care even after shutting off the power because some
parts are not fully discharged immediately after power OFF.
For voltage measurement and other jobs that cannot be performed under the unenergized condition,
follow the procedures set forth in this manual and exercise due care to avoid electric shock and other
hazards.
■Safety devices
Pay attention to safety devices, such as fuses, breakers, interlock switches, panels, and covers so that
their accident prevention functions are always operative. Never make any alteration or modification
that may impair their functionality.
■Backup battery
When installing the backup battery, ensure that it is properly oriented. If the battery is incorrectly
oriented when positioned, it may burst. When it needs replacement, use the same type as that has
been used or the battery prescribed in the parts list. When putting the battery out of use, dispose of it
at site in compliance with manufacturer instructions.

009-052-03
9.24.99 FM2561-1 (ITC)
2FCR ID-T741
Service Manual
SAFETY
■Preparations
Determine the installation site, considering the work flow and necessary additional facilities. Perform
construction, electrical facility, and air-conditioning facility work in advance.
Avoid the following installation sites:
• Places where the temperature changes drastically;
• Places near heat sources such as heaters;
• Places where water leakage or equipment submersion may occur;
• Places where the machine may be exposed to any corrosive gas;
• Dusty places;
• Places where the machine is subject to constant or excessive vibration or shock;
• Places exposed to direct sunlight;
◆◆
CAUTION
◆◆
When using the interface cables for serial connections, it is necessary to specify their connection direc-
tions because of the remote power control specification.
Therefore, if cables need to be serially connected, determine in advance the direction of each interface
cable (master/slave) depending on how the devices are laid out and which unit provides power ON/OFF
control.
■Operation Precautions
Make sure to observe the following to prevent danger to human bodies or other serious incidents from
occurring while operating the system.
◆◆
WARNING
◆◆
●
Be sure to turn the power switch OFF and disconnect the power plug holding it tight by hand prior to
maintenance servicing, unless otherwise specified in the Service Manual.
If the power cable is connected, electrical hazards or short circuits can occur even though the power
switch has been turned OFF.
●
The grounding wire must be connected.
●
To avoid damages to the parts/units due to static discharge, use a wrist strap putting one end on your
wrist and the other end on the grounded section of the equipment, as illustrated below.
Wrist strap
FT100310.EPS

009-052-04
03.30.2000 FM2762 (ITC) FCR ID-T741
Service Manual
Maintenance-0
FCR ID Terminal/Console
ID-T741 Service Manual
MAINTENANCE

009-052-04
03.30.2000 FM2762 (ITC)
Maintenance-00 FCR ID-T741
Service Manual
04/10/98 00 New release All pages
12/20/98 01 Revision (FM2370) All pages
02/20/99 02 Revision (FM2434) Pages 1, 2, 4~6, 8~44, 46~52
03/30/99 02 Revision (FM2483) Pages 00, 7, 45
09/24/99 03 Revision (FM2561) Pages 0, 00, 0.1, 3, 4, 5, 9, 10,
11, 12, 16, 23, 24, 26, 34, 35
09/24/99 03 Changes in pagination (FM2561) Pages 13~15, 17~22, 25, 27~33,
36~54
03/30/2000 04
Revision for software version B08 (FM2762)
Pages 0, 00, 9, 15, 24
Control Sheet
Revision number Reason Pages affected
Issue date

MAINTENANCE Contents
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-0.1
MAINTENANCE
Contents
1. Machine Specifications ............................................................... 1
2. Maintenance Mode and Service Mode ....................................... 2
2.1 Maintenance Mode ............................................................. 2
2.1.1 Entering the maintenance mode ............................. 2
2.1.2 Shutting down the machine from the maintenance
mode ....................................................................... 4
2.1.3 Maintenance menu tree diagram............................. 4
2.1.4 Menu option list ....................................................... 5
2.1.5 Details of menu options ........................................... 6
2.2 Service Mode...................................................................... 12
2.2.1 Entering the service mode....................................... 12
2.2.2 Shutting down the machine from the service mode 13
2.2.3 Service menu tree diagram ..................................... 14
2.2.4 Updating parameters of all menus........................... 14
2.2.5 Updating X-ray exposure parameters of all menus . 15
3. Machine Overview ...................................................................... 17
3.1 Internal Structure ................................................................ 17
3.2 Remote Power Supply Control............................................ 19
3.3 Setting the Boards .............................................................. 20
3.4 System Connection Configuration ...................................... 21
3.4.1 Date/time management function ............................. 21
3.4.2 HOST information management.............................. 21
3.4.3 IP registration information management/sharing
function .................................................................... 22
3.4.4 Connection of multiple units of image reader .......... 23
3.4.5 Patient information on-line function ......................... 27
4. Replacement Procedures ........................................................... 30
4.1 Top Cover Assembly ........................................................... 30
4.2 Touch Panel ........................................................................ 31
4.3 Power Supply Unit .............................................................. 33
4.4 Floppy Disk Drive................................................................ 33
4.5 Hard Disk Drive................................................................... 34
4.6 Main Board ......................................................................... 35
4.7 Sub-board........................................................................... 37
4.8 LCD..................................................................................... 38
4.9 Backup Battery ................................................................... 39
5. Service Parts List........................................................................ 41
5.1 Enclosure Section............................................................... 42
5.2 Base Section - Floppy Disk Drive and Sub-board .............. 44
5.3 Base Section - Hard Disk Drive, Main Board, and Power
Supply................................................................................. 46
5.4 Accessories ........................................................................ 48
5.5 Cables................................................................................. 50

MAINTENANCE 1. Machine Specifications
009-052-02
2.20.99 FM2434 (ITC) FCR ID-T741
Service Manual
Maintenance-1
1. Machine Specifications
■Machine specifications
Item Specification Remarks
CPU Pentium
TM
(166 MHz) On-board card PC
Intel 486DX4
TM
(100 MHz)
RAM
Pentium : Basic 8MB+64MB (SO-DIMM)
RTC Monthly error: 3 min;
Backup period: 6 years *1Lithium primary battery
Display 640 x 480 pixels,
256-color display (VGA) 10.4-inch, TFT color
Display controller Trident Cyber 9320
VRAM Pentium : 1MB
Intel 486DX4 : 512KB
Floppy disk drive 3.5-inch
Hard disk drive 2.5-inch, 815MB or more IDE interface
Input device Infrared touch panel RS-232C
External
interface
CRT 15-pin mini D-sub
Keyboard 6-pin mini DIN
Mouse Not available
RS-232C (barcode reader) 8-pin DIN
AT half-size slot Machine internal card-
edge connector
LAN (10Base-T) SMC91C94
External device drive contact x 2
Remote power supply shutoff signal contact
2-pin modular jack x 3
Switches/
buttons
Main switch Rocker type
Sub-switch Lock type
Brightness control
Buzzer volume control
Magnetic card reader
Barcode reader RS-232C
Standard
compliance
(Not applicable to
Pentium version)
Safety assurance:
UL1950 and IEC950 (EN60950)
Noise immunity:VCCI Class 1
The EN6100 (high-power
high-frequency control)
standard is not applicable
because the product is of a
low power consumption type.
Employed OS Windows NT 4.0
*1 : This battery life prevails while the power is OFF.
The battery power is not consumed while the power is ON. TT100151.EPS
ISO (JIS-I) or JIS (JIS-II) RS-232C
Flash memory Intel 486DX4 : 512KB
Intel 486DX4 : Basic 4MB+32MB (SO-DIMM)
Pentium : 128KB Incorporated in the card
PC
■Variations
●Magnetic card reader
None/ISO (JIS-I)/JIS (JIS-II)
●OS
Windows NT 4.0 Japanese edition/English edition
■Options
●SIF04A board (serial interface option board)
This PC oard has three channels of serial communication ports.
It is used with the RS-422 or RS-232C interface selected. One unit
of this oard can e incorporated in the IDT-IV main unit.
The following devices can e serially connected to the IDT-IV.
• One unit of the 9000 Series image reader
• RIS or radiographic equipment
• IDT-IIIB
When an RS-422 interface connection is made to a 9000 Series
image reader and IDT-IIIB, remote power ON/OFF control can e
exercised with the serial ca le internal contact signal.
●Remote power ON/OFF cable
This ca le makes remote power control possi le to other devices.
The connector type will depend on the device.
3. Machine Overview, “3.2 Remote Power Control”
“Service Parts List, 5.5 Ca les”
●Full-sized keyboard
Either the PS/2 (DOS/V) key oard or the Windows key oard can
e used.
PS/2 (DOS/V) Windows
US: 101 104
Europe: 102 105
●Wall-mounting bracket
Bracket for mounting the IDT-IV on a wall
●Stand
With this stand, it is possi le to use the IDT-IV in its tilted position.
●Optional software
The following two software options are availa le. The optional
software is to e supplied on floppy disks.
• Patient information on-line function
• Radiographic equipment connection function

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-02
2.20.99 FM2434 (ITC) FCR ID-T741
Service Manual
Maintenance-2
2.1 Maintenance Mode
This mode is to e used y the service engineers only. It provides
means of setting up the IDT-IV operating environment.
2.1.1 Entering the maintenance mode
The maintenance mode can e called up either y canceling the auto
logon sequence or from the routine mode. These two different main-
tenance mode entry procedures are descri ed elow.
■Entering the maintenance mode by canceling the auto
logon sequence [A keyboard is needed.]
The key oard should e connected ahead of time.
(1) If the main switch is OFF, turn it ON.
(2) Turn ON the sub-switch.
FT100048.EPS
Sub-switch
Main switch
(3) When the Windows NT’s blue startup screen opens, press and
hold down the <Shift> key until the logon message appears.
➔ The logon message appears.
If the routine mode starts due to a delayed execution of step (4),
switch to the maintenance mode y performing “■ Entering the
maintenance mode from the routine mode” steps (4) and eyond.
(4) Enter “fcr” as the user name.
(5) Enter “fcr” as the password.
The entered password does not appear on the display.
2. Maintenance Mode and Service
Mode
The IDT-IV has three processing modes: service mode, routine mode,
and maintenance mode. For servicing, use the maintenance mode or
service mode.
The mode transition diagram is shown elow.
Mode changes effected
by general user or service engineer
Mode changes effected by service engineer
* A keyboard should be
connected ahead of time.
Power ON
Power OFF
[OS startup]
Routine mode
(normal operation
mode) Maintenance
mode
Service mode
Service option
User service
auto logon
Canceling the auto logon sequence
(by holding down the <Shift> key
★
)
logoff
logoff
Shutdown & restart
Logging on with
the service engineer
password
Entering the
password for
the entry into
the maintenance
mode and touching
[ENTER]
Entering the service
engineer password and
touching [ENTER]
touching [ENTER]
without doing anything else
FT100226.EPS
★ Use the full key oard when starting up y canceling auto logon.

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-3
■Entering the maintenance mode from the routine mode
(1) If the main switch is OFF, turn it ON.
(2) Turn ON the sub-switch.
➔The routine mode starts in a out 5 minutes.
(3) Select the utility mode.
If the “Patient ID Information Input” screen is displayed, touch
[Cancel] and then select the utility mode.
➔The “Password Input” screen opens.
The password (000000, displayed as ******) for permitting the
user to enter the service mode is displayed in the character string
input area.
FT100011.EPS
(4) Touch [AC] to delete the default password for the user’s service
mode entry.
➔The character string in the input area is then cleared.
(5) Touch [123].
➔The upper-case alpha etical character input mode then pre-
vails.
(6) Touch [OK].
➔The maintenance mode is then invoked so that the “Mainte-
nance process” menu screen appears on the display.
FT100010.EPS

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-4
(6) Touch the [7], [8], [9], [4], [5], and [6] buttons in sequence and
then touch [ENTER].
The password “ADGJMP” is then entered. The screen reads
“******”.
➔The maintenance mode then starts with the “Maintenance
process” menu screen opened.
2.1.2 Shutting down the machine from the maintenance
mode
This section descri es the procedure for shutting down the equip-
ment from the maintenance mode.
(1) While the “Maintenance process” menu screen is displayed, touch
[Close].
➔The maintenance termination confirmation screen appears.
(2) Touch [Yes].
●If control panel setup has been performed from the “system
setup” screen
➔The shutdown confirmation screen appears.
Touch [No].
➔The routine mode starts after logoff.
◆
NOTE
◆
Do not use the routine mode at this moment. The control panel setup
does not take effect unless you shut down the machine.
●If control panel setup has not been performed from the “system
setup” screen
➔The routine mode starts after logoff.
(3) Turn OFF the sub-switch.
➔The computer shutdown screen opens. After a while, the
equipment shuts down.
◆
NOTE
◆
Do not turn OFF the main switch. If you turn OFF the main switch
during remote power supply control, the remote power ON/OFF
function is disabled.
2.1.3 Maintenance menu tree diagram
Data format setup
Maintenance
mode System setup
Network setup
Hookup device
setup
Barcode setup
Magnetic card
setup
Selector setup
Operating
environment setup
Interface data setup
Inquiry time setup
Control panel setup
Host information setup
Group information setup
Barcode reader setup
Serial hookup device setup
Image processing parameter
rewriting target image reader
setup
Barcode format definition
Magnetic card type setup
Magnetic card format definition
Menu coordination setup
Selector button definition
Overall machine operation
mode setup
IP number input setup
Group information setup
Patient information display setup
Examination menu/
additional code setup
Patient information input setup
UID setup
Data type setup
Examination number field setup
FT100201.EPS
Birth date format setup
ID number setup
Reception number setup
Examination number setup
Exposure menu
parameter
environment setup
Exposure menu parameter
setting range setup
Parameter button display setup
Frontal and lateral images on
film setup

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-5
XCON exposure/
implementation
condition setup
IP use statistical data
Software installation
Backup/restore
Hardware maintenance utility
Exposure condition transmission
mode setup
Implementation condition
reception mode setup
Operating mode
Operation category
Data conversion
FT100202.EPS
Patient's info.
on-line DB
functions setup
RIS connection mode setup
Inquiry key setup
Serial hookup device setup
2.1.4 Menu option list
Setup item Function overview Remarks
System setup
Network setup Can be set up
at the master IDT
only.
Hookup device
setup
Barcode setup
TT100101.EPS
Back
up
•Inquiry time setup
•Network setup
•Date/time setup
•Keyboard setup
•Regional setup
•Screen saver setup
•Host information/group information
setup
•Host information/group information
transfer
•Network connection verification
•Barcode reader setup
•Serial hookup device setup
•Image processing parameter
rewriting target image reader setup
•Barcode format definition
Magnetic card
setup •Magnetic card type definition
•Magnetic card format definition
•ISO card read character code
definition
Can be set only
when the magnetic
card reader is
connected.
Selector setup • Menu coordination setup
• Selector button definition
Setup item Function overview Remarks
Operating
environment
setup
•Exposure condition transmission
mode setup
•Implementation condition reception
mode setup
IP use statistical
data –
–
–
–
–
TT100102.EPS
•IP use statistical data list display
•IP use statistical data deletion
Back
up
•Overall machine operation mode
setup
•IP number input setup
•Group information setup
•Examination menu/additional code
setup
•Patient information display setup
•Patient information input setup
Exposure menu
parameter
environment
setup
•Exposure menu parameter setting
range setup
•Parameter button display setup
•Frontal and lateral images on
film setup
Patient's info.
on-line DB
functions setup
•RIS connection mode setup
•Inquiry key setup
•Serial hookup device setup
Can be set up when
optional software is
installed.
Can be set up when
optional software is
installed.
XCON exposure/
implementation
condition setup
Hardware
maintenance
utility
Backup/restore •Backing up/restoring the mainte-
nance mode setup data
•Checking whether magnetic card
reader, touch panel and HDD are
normal or not.
Data conversion •Converting IDT-III and CSL-III
menu data for IDT-IV
Software
installation •Optional software/software upgrade
installation from floppy disks
Notes on "Backup" column
Indicates whether the setup data can be backed up and the backed-up data can be
restored for use with other units of the IDT-IV.
: The backed-up data can be used as is for the other units of the IDT-IV.
: The backed-up data needs to be restored and edited for use with the other units
of the IDT-IV.
: The setup data cannot be backed up.
•Birth date format setup
•ID number setup
•Reception number setup
•Test number setup
•UID setup
•Data type setup
•Examination number field setup
Data format
setup
Interface data
setup

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-02
2.20.99 FM2434 (ITC) FCR ID-T741
Service Manual
Maintenance-6
2.1.5 Details of menu options
This section descri es the menu options provided for the mainte-
nance mode and their usage precautions.
■System setup
To ena le the setup, it is necessary to re oot the equipment after
completion of setup.
Installation, “3.3.6 Setting the system”
●Inquiry time setup
Sets the time at which date/time setup is to e performed to ensure
that all the networked units of the IDT-IV agree in the date and time.
This inquiry time must e set up for each slave IDT unit.
Master IDT: IDT-IV or IDC-IV for date/time and host information
management. One master IDT unit must e set up for
the network.
Of all the IDT-IV and IDC-IV units, the master should
e the device that is operated the longest.
Slave IDT: All IDT-IV and IDC-IV units other than the master IDT.
●Control panel setup
❍Network
Installation, “3.3.6 Setting the system, ■ Setting the network"
• Computer name
For identification purposes, a computer name is given to each
device installed in the network.
Each device should e given a different computer name. In addi-
tion, names should agree with the host names set in the section
entitled "■ Network setup."
Ex.: IDT4-1
• Work group
The group of computers displayed, for convenience, under the same
work group name when a network configuration is referred to.
Leave the default as it is.
Ex.: FCRIDT4
• IP address
Num er assigned to network devices for identification purposes.
Set in the format "xxx.xxx.xxx.xxx," using a num er from 0 to 255
for "xxx" portions.
Ex.: 172.16.0.3
• Su net mask
In IP addresses, assigns "1" to its handled as network nos. and "0"
to portions handled as host num ers and is set as a decimal.
Format is "xxx.xxx.xxx.xxx," using a num er from 0 to 255 for "xxx"
portions.
Ex.: 255.255.255.0 (class C)
• Default gateway
Address of router and other devices. Set as necessary.
❍Date and Time
Set the date, time and time zone.
◆
NOTE
◆
Make sure to make the time zone for all IDT-IV and IDC-IV units in the
network the same.
The IDT-IV and IDC-IV date/time synchronization function uses Green-
wich Standard Time in the form of numerical data. For this reason,
discrepancies in the time will result if the time zones for IDT-IV and IDC-
IV units are different.
❍Key oard
Set the type of key oard. In Japan, this does not need to e set.
❍Region
In Japan, confirm that "Japanese" has een selected.
❍Screen Saver
Confirm that "Screen Saver" has een selected for the screen saver
and "12 min." selected for the time it will take to appear.

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-02
3.30.99 FM2483 (ITC) FCR ID-T741
Service Manual
Maintenance-7
■Network setup
The network information can e set up at the master IDT only.
Installation, “3.3.5 Setting the network information”
●Group information setup
Maintenance, “3.4.3 IP registration information management/
sharing function”
• Group name
• Gateway host name
To e set up after host information setup.
• Comment
●Host information setup
• Host name
Set the same computer name as was set for "● Control panel
settings" under "■ System settings."
• Device ID
• IP address
• Group
• Comment
When you touch the [Ping] utton in the “Host Information Edit”
screen, you can verify the network connection.
■Hookup device setup
Installation, “3.7.1 Setting the hookup devices”
●Barcode reader setup
Specifies whether or not to use the arcode reader.
●Serial hookup device setup
Specifies the channels and devices to e connected.
◆
NOTE
◆
When connecting to RIS, make settings via the "Setting the Patient
Information Online Function " screen.
RIS cannot be selected from this screen.
●Image processing parameter rewriting target image reader setup
When making settings for 9000 series, 3000 series or 9501/2 series
image readers, first make the settings specified under “■ Setting the
device to e connected serially.”
When making the settings for 5000/R and 5501 image readers, first
register the image readers in the host information under "■ Network
setup."
❍Connection mode
Specifies whether or not to make a connection.
❍Button caption
Sets the name of the utton used for the "Image processing param-
eter rewriting" function availa le in service mode. Failure to do so
will generate an error.
❍Hookup model
Sets the connection format, "Network" or "Serial," and the model of
image reader used.
❍Parameters
• GR/RP
These image processing parameters pertain to gradation/frequency
processing.
•DRC/TAS
These image processing parameters pertain to DR compression
filter / DR compression processing.
Sets to [ON] for the 5000/R and 5501.
• MFP
These image processing parameters are to ena le enhancement
processing with artifact suppression or dynamic range compres-
sion.
Sets to [OFF] for 9000 series image readers.
• PEM
These image processing parameters are for granularity suppression
and sharpness enhancement processing exclusively for mammogra-
phy.
Sets to [OFF] for 9000 series image readers.

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-02
2.20.99 FM2434 (ITC) FCR ID-T741
Service Manual
Maintenance-8
■Barcode setup
Sets the arcode type (NW7/Code39) and format.
■Magnetic card setup
Installation, “3.7.3 Setting the magnetic card”
●Magnetic card type setup
Sets the magnetic card type (Fuji card type).
●Magnetic card format definition
When a hospital’s card is to e used, its format must e defined.
• Data type
• Format
When "Sex," "Date of Birth" or "Year or Month of Date of Birth" is
selected, select the format from the ones availa le.
• Card offset
Sets the data start position on the card. Set the position after
"STX" (with a FUJI card, it will e the letter "F") to 0.
Do not set "SI/SO" or "Kanji IN/OUT" code portions as starting
points.
• Data length
Sets the length of data to e read in terms of the num er of letters
from the card offset position.
Do not include "SI/SO" or "Kanji IN/OUT" code portions in data
length.
For kanji data, set an even num er for data length.
Data lengths for patient ID nos., reception nos. and examination
nos. should not exceed the valid num er of digits set in “Setting the
data formats." Excesses will result in errors.
If the valid num er of digits is not satisfied, the padding process
set in "■ Setting the data formats" will e conducted.
• Storage offset
Sets the storage position for read data.
For example, if "2" is set, 2 spaces will e added to the front of the
data when it is stored.
◆
NOTE
◆
The format, card offset, and data length cannot always be set up de-
pending on the data type.
■Selector setup
This item should e set when using multiple processing modes (IP
registration processing mode and console processing mode).
Installation, “3.7.4, Setting the selector”
●Menu coordination setup
Specifies whether or not to change the processing mode in coordina-
tion with the exposure menu selection.
●Selector button definition
• Button caption
Sets the name of the selector utton. Failure to do so will generate
an error.
• X-ray tu e and technique codes
• Operating mode
• Exposure mode transition request
Sets when using the IDT-IV as the console. For cassette exposure,
set whether to use the IP withdrawal utton.
• Selector at time of startup
Sets whether to display the selector utton on the HOME screen
after starting up.
●Length of time-out for exposure mode transition request
When "Exposure mode transition request" has een set to [ON], sets
a num er for the time-out pertaining to requests for IP withdrawal
(transition to cassette exposure mode) and IP setting (transition to
uilt-in exposure mode). Leave the default as it is.

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-04
03.30.2000 FM2762 (ITC) FCR ID-T741
Service Manual
Maintenance-9
■Operating environment setup
Installation, “3.7.5 Setting the operating environment”
●Overall machine operation mode setup
Because this setup is effective only under Japanese environments,
leave this column OFF.
(DBCS: Dou le Byte Character System dedicated for Japanese kanji
characters)
●IP number input setup
Because this setup is effective only for the magazine type image
reader, leave this column OFF.
Specifies whether or not to use the [NO ID] utton (which can e
used when the IP arcode cannot e read y a magazine type reader).
●Group information setup
Defines the range of group num ers that are to e used for the group
process and multi-frame output process.
Using an IDT-IV in the same network, make the setting so that the
range of group nos. is not duplicated. Leave the default as it is.
●Examination menu/extension code settings
Set "Examination menu" to [ON] if using an examination menu.
Setting "Examination menu" to [ON] also makes it possi le to use the
extension menu (to register multiple menus under one MPM code).
If "Examination menu" has een set to [ON], set whether to auto-
matically have menu codes generated and the color of the menu's
letters.
◆
NOTE
◆
Normally, “Code automatic generate” is set to [OFF].
When it is set to [ON], if one IDT-IV is operating alone, it will automati-
cally generate extension menu codes when menus are created.
However, if an ID is shared by multiple IDT-IVs, it will be necessary to
register the same examination menu code at each IDT. In keeping,
inputting in a mode that does not automatically generate these codes is
safer.
If printing frontal and lateral images on film when using an IDT-IV
as the console for the 9501/2, set "Examination Menu" to [ON], as
this is the only place where frontal and lateral dou le-image output
can e specified.
●Patient information display setup
• Patient name display mode
Select [Main]. The [Main/Su ] setup is effective only for Japanese.
If [OFF] is selected in step I. a ove, [Main] will e selected auto-
matically.
Select [Expansion] when expanding the num er of characters to e
displayed.
• Name of the department which requests for the service
Formulates the display/input definition.
• Name of the department displayed in Japanese kanji characters.
Formulates the display/input definition.
• Examination num er
Formulates the display/input definition.
●Patient information input setup
• Automatic window opening
Specifies whether or not to automatically open the patient informa-
tion input window on the routine mode HOME screen.
• Japanese kanji name input
Specifies whether or not to use Japanese kanji characters for name
input purposes. If the use of kanji is deselected in the “● Overall
system operation mode” setup sequence, the system automatically
inhi its the input of kanji names.
■Data format setup
Sets the irth date and other data formats.
Installation, “3.7.6 Setting the data formats”
●Birth date format setup
Specifies the irth date display/input format.
●ID number setup
• Character string length
• Padding method

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-10
●Reception number setup
• Character string length
• Padding method
●Examination number setup
• Character string length
• Padding method
■Interface data setup
Installation, “3.7.7 Setting the Interface Data”
◆
NOTE
◆
Make sure to set the same contents for all IDT-IV and IDC-IV units in the
network.
●UID setup
• RIS UID Used
When connected to the RIS, sets whether or not to use the study
information issued y the RIS.
• UID Issue
Sets whether or not the IDT-IV issue the study information or
series information.
After setting the UID Issue to ON, set the UID issue system.
●Data type setup
• Network IR I/F
Defines interface etween the IDT-IV and 5000/R or 5501/2.
“FINP110” is availa le from software version A07. For this inter-
face, the restriction fixing the patient ID num ers and examination
num ers at 10 figures has een released and the data length
extended.
• Serial ID online
Sets whether to use previous interface or extended interfaces for
each patient information online interface.
Extended interface is availa le from software version A07. When
using extended interface, select “LONG” ecause the data field
length for the patient ID num ers or examination num ers has
een extended.
●Examination number field setup
Sets which of the examination num er, reception num er or order
num er that was input in the IDT-IV is to e handled as the study
num er.
The num er specified as the study num er will e handled as the
“Accession Num er” when an image is output via DICOM from the
Image Reader or QA-WS.
■Exposure menu parameter environment setup
Installation, “3.7.8 Setting the exposure menu parameters”
●Exposure menu parameter setting range setup
• IP logical size
Sets the size for logical reading.
• Image reversal mode
Sets modes for moving images 90° upward/downward/clockwise/
counterclockwise.
• Filing mode
Sets image transmission mode to the ODF.
• Image format
Sets the format in which images are output on film.
• EDR mode
Sets the method for setting image reading sensitivity.
• Film output count upper limit
Sets a num er etween 1 and 9 for the upper limit of film output.
• Function processing
Sets which of the 8 function codes, other than RT, to display.
RT : Routine processing (routine exposure)
TS : Temporal su traction processing
ES : Energy su traction processing
WS : Whole-spine processing
PS : IVP processing
TM : Tomographic processing
TX : Collective temporal su traction processing
EX : Collective energy su traction processing
WX : Collective whole-spine processing

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-11
●Parameter button display setup
Defines the parameter utton display in the routine mode and
service mode.
●Frontal and lateral double-image output setting
This is set when printing frontal and lateral images on film with the
9501/2.
■Patient’s info. on-line DB functions setup
This setup can e performed when the optional software is installed.
●RIS connection mode setup
“3.4.5 Patient information on-line function”
●Inquiry key setup
Defines the key for patient information inquiry.
●Serial hookup device setup
Sets the channels and devices to e connected.
●Inquiry time-out setting
Valid when the RIS connection mode is either [Type B] connection or
[IDC4].
Leave the default (30 sec.) as it is.
■XCON exposure/implementation condition setup
This setup can e performed when the optional software is installed.
●Exposure condition transmission mode setup
Specifies whether or not to transmit exposure condition data to the
X-ray controller.
●Implementation condition reception mode setup
Specifies whether or not to receive implementation condition data
from the X-ray controller.
■IP use statistical data
Used to confirm the IP use status.
●IP use statistical data list display
●IP use statistical data deletion
■Software installation
Used to install the optional software or software update from floppy
disks.
■Backup/restore
Backs up the maintenance mode setup data onto or restores it from
floppy disks.
●Operating mode setup
Chooses etween the ackup and restore functions.
●Operation category setup
Chooses etween machine-specific data and common system data.

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-12
■Hardware maintenance utility
Checks the following for precision or whether normal or not.
Installation, “Appendix 2 Hardware Maintenance Utility”
• MCR (magnetic card reader)
• BCR ( arcode reader)
•LCD
• Buzzer
• Touch (touch panel)
• SIF04A
• Network
• HDD
• FDD
• KB (key oard)
• RTC (date and time)
■Data conversion
Converts the IDT-III and CSL-III menu data for the IDT-IV.
Installation, “Appendix 3 Data Conversion Procedures”
2.2 Service Mode
The service mode is to e used y the service engineers and general
users. However, the “Updating Parameters of All Menus” and
“Updating X-ray Exposure Parameters of All Menus” functions
offered y the service mode are for the service engineers only.
For the functions to e used y oth the service engineers and
general users, refer to the operation manual.
2.2.1 Entering the service mode
To enter the service mode, proceed as directed elow.
(1) If the main switch is OFF, turn it ON.
(2) Turn ON the sub-switch.
➔The routine mode starts in a out 5 minutes.
(3) Select the utility mode.
If the “Patient ID Information Input” screen is displayed, touch
[Cancel] and then select the utility mode.
➔The “Password Input” screen opens.
The password (000000, displayed as ******) for permitting the
user to enter the service mode is displayed in the character string
input area.
FT100011.EPS

MAINTENANCE 2. Maintenance Mode and Service Mode
009-052-03
9.24.99 FM2561 (ITC) FCR ID-T741
Service Manual
Maintenance-13
(4) Touch [AC] to delete the default password for user’s service
mode entry.
➔ The character string in the input area is then cleared.
(5) Type in “111111” as the password and then touch [ENTER]. The
screen then reads “******”.
➔ The service mode is then invoked so that the “Service Process-
ing” screen opens.
FT100204.EPS
2.2.2 Shutting down the machine from the service mode
This section descri es the procedure for shutting down the equip-
ment from the service mode.
(1) While the service main screen is displayed, touch [Close].
➔ The servicing confirmation screen opens.
●To shut down the equipment after saving your service mode setup
data editing
Touch [Yes].
➔ The routine mode starts.
●To shut down the machine without saving your service mode
setup data editing
Touch [No].
➔ The routine mode starts.
◆
IMPORTANT
◆
To shut down the equipment without saving your data editing, be sure
to perform step (2). Your service mode setup data editing is retained
until the machine shuts down.
(2) Turn OFF the sub-switch.
➔ The computer shutdown screen opens. The machine shuts
down after a while.
◆
NOTE
◆
Do not turn OFF the main switch. If you turn OFF the main switch
while remote power supply control is exercised, the remote power
ON/OFF function is disabled.