FujiFilm SonoSite Edge User manual

Edge Stand
H-UniversalStand
User Guide

Edge, SonoSite, M-Turbo, MicroMaxx, SonoSite, Titan, and the SonoSite logo are trademarks or registered
trademarks of FUJIFILM SonoSite, Inc. in various jurisdictions.Value from Innovation is a trademark of FUJIFILM
Holdings America Corporation.
All other trademarks are the property of their respective owners
Patents: US 8,088,071; US 7,591,786; US 7,534,211; US 6,447,451; CA 2372158; and EP 1589878.
Part Number: P12223-04
Publication Date: November 2017
Copyright © 2017 FUJIFILM SonoSite, Inc. All Rights reserved.
Manufacturer
FUJIFILM SonoSite, Inc.
21919 30th Drive SE
Bothell, WA 98021 USA
T: 1-888-482-9449 or 1-425-951-1200
F: 1-425-951-1201
EC Authorized Representative
FUJIFILM SonoSite B.V.
Joop Geesinkweg 140
1114 AB Amsterdam,
The Netherlands
Australia Sponsor
FUJIFILM SonoSite Australasia Pty Ltd
114 Old Pittwater Road
BROOKVALE, NSW, 2100
Australia
Caution United States federal law restricts this device to sale by or on the order of a physician.

English Deutsch Español Français Italiano Português
Introduction 1
Edge Stand
H-Universal Stand
User Guide
Introduction .................................................................................................................................................... 1
Document conventions ................................................................................................................................................................. 2
Getting Help ...................................................................................................................................................................................... 2
Stand features ................................................................................................................................................. 3
Using the stand ............................................................................................................................................... 4
Inserting and removing the ultrasound system .......................................................................................... 4
Connecting the ultrasound system ............................................................................................................... 5
Using peripherals and accessories ................................................................................................................ 6
Printer or bar code scanner .......................................................................................................................................................... 6
Triple Transducer Connect ........................................................................................................................................................... 7
Troubleshooting ............................................................................................................................................. 8
Connectivity diagram ..................................................................................................................................................................... 8
Cables and connectors .................................................................................................................................................................. 9
Cleaning and disinfecting ............................................................................................................................ 11
Safety ............................................................................................................................................................. 12
Electrical safety ............................................................................................................................................................................... 12
Equipment safety ........................................................................................................................................................................... 13
Clinical safety ................................................................................................................................................................................... 14
Electromagnetic compatibility (EMC) ..................................................................................................................................... 14
Specifications ................................................................................................................................................ 14
Stand dimensions (approximate) ............................................................................................................................................ 14
Temperature, humidity, and atmospheric pressure limits ............................................................................................. 14
Introduction
This user guide explains how to use the Edge Stand (used with the SonoSite Edge II and Edge systems) or the
H-Universal Stand and peripherals or accessories. For information on a specific peripheral, see the
manufacturer’s instructions included with that product. For information on the transducers or ultrasound
systems, see the ultrasound system user guide.
If you need to disassemble the stand, refer to its assembly instructions,available with the stand at
www.sonosite.com.
WARNINGS To avoid the risk of patient injury, misdiagnosis, or user injury, read all the warnings
in the ultrasound system user guide and supplemental guides.
To avoid injury or equipment damage that could result from the stand toppling
over, do not apply undue force to the stand. Positioning the stand platform at a low
level may help to alleviate this potential hazard.

2Introduction
Document conventions
A WARNING describes precautions necessary to prevent injury or loss of life.
A Caution describes precautions necessary to protect the products.
A Note provides supplemental information.
Numbered and lettered steps must be performed in a specific order.
Bulleted lists present information in list format but do not imply a sequence.
Single-step procedures begin with .
Symbols and terms used on the system are explained in the ultrasound system user guide.
Getting Help
For technical support, please contact FUJIFILM SonoSite as follows:
Printed in the U.S.
Phone
(U.S. or Canada)
877-657-8118
Phone
(outside U.S. or Canada)
425-951-1330, or call your local representative
Fax 425-951-6700
Email service@sonosite.com
Web www.sonosite.com
Europe Service Center Main: +31 20 751 2020
English support: +44 14 6234 1151
French support: +33 1 8288 0702
German support: +49 69 8088 4030
Italian support: +39 02 9475 3655
Spanish support: +34 91 123 8451
Asia Service Center +65 6380-5589

English Deutsch Español Français Italiano Português
Stand features 3
Stand features
The stand provides a mobile work platform and a storage area for transducers and other supplies. It also
provides connections for accessories and peripherals.
Figure 1 Stand with accessories. H-Universal Stand shown with M-Turbo ultrasound system, and optional
basket, black-&-white printer, and Triple Transducer Connect.
*Storage bin is positioned under the printer. Refer to the Storage Bin Installation Instructions.
1 Ultrasound system 6 Black-&-white printer
2 Hook (1 of 4) for cables and other
items
7 AC adapter 1: Connector for system
power cord
3 Triple Transducer Connect 8 AC adapter 2: Connector for AC
power cord (under stand base)
4Height-adjustmentlever 9Locking lever on wheel
5 Basket or storage bin*
2
1
5
7
3
9
6
4
8

4Using the stand
Using the stand
To raise or lower the system platform
1Make sure that the area above the platform is clear.
2Facing the back of the stand, turn the height-adjustment lever counter-clockwise.
The pole may rise automatically from the column.
3Raise or lower the pole to the desired height, and then tighten the lever.
To tighten the lever, turn it clockwise.
If lowering the pole, avoid grasping its base. Skin could get caught and pinched in the gap.
4If you want to reposition the lever (for example, parallel to the pole), pull the lever out, turn it to the desired
position, and release.
To lock or unlock a wheel
Push the locking lever on the wheel:
Push down to lock.
Push up to unlock.
Inserting and removing the ultrasound system
To insert the ultrasound system
1Rotate system latches outward.
2Place the ultrasound system on the system platform, ensuring that the transducer cable is positioned
between the handles of the stand.
3Secure by rotating the system latches inward.
To remove the ultrasound system
1Disconnect any cords, cables, or TTC attached to the system.
2Rotate latches outward.
3Lift the ultrasound system.

English Deutsch Español Français Italiano Português
Connecting the ultrasound system 5
Connecting the ultrasound system
Attach the mini-dock if you use the system with AC-powered peripherals.
To attach the mini-dock
1Turn off the ultrasound system.
The mini-dock may not function properly if you connect the system while the power is on.
2Place the mini-dock on the system platform.
3Slide the mini-dock feet under the ultrasound system and firmly press the mini-dock into the I/O connector
on the back of the system.
The mini-dock is correctly attached when no gap is between the system and mini-dock.
See Figure 2.
Always verify that the mini-dock and system are correctly attached after changing a transducer. When you
change a transducer, the mini-dock may loosen.
Figure 2 System platform with system and mini-dock attached (H-Universal stand with M-Turbo system
shown).
Note Make sure that the battery is installed before you attach the mini-dock.
Note If at any time the ultrasound system should lose connectivity, turn off the system
and reattach the mini-dock and system.
Mini-dock
System
System latch

6Using peripherals and accessories
To connect AC power
1If using a mini-dock, ensure that it is connected to the system.
2Connect the cables. See “Cables and connectors” on page 9.
DC power cord (A)
System AC power cord (K)
Using peripherals and accessories
For specific information on using a peripheral, including warnings and cautions, see the manufacturer’s
operating instructions. For the bar code scanner, see also Bar Code Scanner User Guide.
To connect a peripheral or accessory, see its assembly instructions.
Printer or bar code scanner
You configure the system for the printer or serial bar code scanner on the system’s Connectivity setup page.
For instructions, see the ultrasound system user guide.
For additional information on the bar code scanner, see the Bar Code Scanner User Guide.
To print the screen
1Turn on the printer.
2(Color Printer Only) Change the input setting from S-video to video.
3Do one of the following:
Use the controls on the printer for printing images. Follow the manufacturer’s instructions.
Caution To avoid damaging the system, use only accessories and peripherals recommended
by SonoSite with the ultrasound system.
Optional Peripherals and Accessories for Stand
AC-powered peripherals USB peripherals Accessories
Black-and-white printer
Color printer
(MicroMaxx, or TITAN) Serial bar
code scanner
(SonoSite Edge II, Edge, M-
Turbo) USB bar code scanner
(SonoSite Edge II, Edge, M-
Turbo) USB footswitch
Mini-dock
Triple Transducer Connect

English Deutsch Español Français Italiano Português
Using peripherals and accessories 7
If the large Aor Bkey (SonoSite Edge II, Edge, M-Turbo) or Delta key (MicroMaxx or TITAN) is programmed
to print, press it. For instructions to program the key, see the ultrasound system user guide.
You can also save images and print them from the patient list. For instructions, see the ultrasound system user
guide.
Triple Transducer Connect
The Triple Transducer Connect (TTC) is an optional three-transducer module that lets you simultaneously
connect up to three transducers while the ultrasound system is docked.
Figure 3 Triple Transducer Connect
To connect a transducer to the TTC
1Pull the transducer latch up and rotate 90°.
2Align and insert the transducer connector to the connector on the bottom of the TTC.
3Turn the latch and press down, securing the transducer connector to the TTC.
To select a transducer
Push a transducer select button on the TTC.
When selecting another transducer, the system restarts. The light on the button flashes momentarily and then
turns solid green.
Only one transducer is accessible at a time.
Note When you use the TTC, there may be a minor reduction of transducer penetration.
Mounting post
Handle
Transducer
select buttons
Transducer
connector

8Troubleshooting
To remove the TTC
1Remove any attached transducers.
2Disconnect the transducer connector from the system.
3Pull out the knob on each of the two mounting brackets, and slide the TTC out.
Troubleshooting
Connectivity diagram
Use the following diagram for assistance disconnecting and reconnecting peripherals. For more detailed
information, refer to the assembly instructions. Additional copies are available at www. sonosite.com.
See also “Cables and connectors” on page 9.
Symptom Solution
Printer does not print. Check the printer selection on the Connectivity setup page.
See the ultrasound system user guide.
Check the cable connections. See the printer installation instructions.
Ensure that the printer is turned on and set up properly. See the printer
manufacturer’s instructions, if necessary.
Stand tips over more easily
than expected.
Lower the platform on the stand.
Avoid placing heavy objects on the platform. Too much weight on the
platform can cause the stand to become less stable.
Caution To avoid damaging the system, use only accessories and peripherals recommended
by SonoSite with the ultrasound system.

English Deutsch Español Français Italiano Português
Troubleshooting 9
Connectivity of printer and system
Cables and connectors
Table 1: Connectivity symbols on mini-docks
Symbol Definition Symbol Definition
DC input S-video in
Print control DVI video out
USB Composite video out
Ethernet Audio out
RS-232 (DVD recorder, VCR, or bar code scanner) ECG/Footswitch
S-video out
A
H
C
J
K
System
Power
Supply

10 Troubleshooting
Table 2: Cables and connectors
Letter Name
Length Illustration
Connector 1
[Illustration]
[Connects to]
Connector 2
[Illustration]
[Connects to]
Peripheral
ADC power supply
cord
6.8 ft./2 m
See “Stand
features” on
page3.
System or mini-
dock
Pre-installed n/a
CComposite
video cable
6ft./1.8m
Mini-dock Printer
Printer
HPrint control cable
6ft./1.8m Mini-dock Printer
Printer
JAC power cord
39 in./1 m
Printer AC adapter 2 on
stand base
Printer
KSystem AC
power cord
10 ft./3 m
AC adapter 1 on
stand base
Wall outlet
n/a

English Deutsch Español Français Italiano Português
Cleaning and disinfecting 11
Cleaning and disinfecting
To clean and disinfect a peripheral, refer to its manufacturer’s instructions. For the bar code scanner, see also
the Bar Code Scanner User Guide.
To clean and disinfect the stand or TTC
You can clean and disinfect the exterior surface of the stand or TTC using a recommended cleaner or
disinfectant. For the most recent list of approved products, see www.sonosite.com.
1Turn off the ultrasound system and any accessory equipment. Disconnect the system and accessory
equipment from AC power.
2Clean the surfaces of the stand to remove any debris or bodily fluids. Use the following procedure:
WARNINGS To avoid electrical shock, before cleaning always disconnect the system from the
power supply and remove it from the stand.
To avoid injury, always use protective eyewear and gloves when performing
cleaning and disinfecting procedures.
To avoid infection:
Ensure that the solution expiration date has not passed.
The level of disinfection required for a product is dictated by the type of tissue it
contacts during use. Ensure that the solution strength and duration of contact
are appropriate for the equipment. For information, see the disinfectant label
instructions and the recommendations of the Association for Professionals in
Infection Control and Epidemiology (APIC) and FDA.
Caution Use only recommended cleaners or disinfectants on surfaces. Immersion-type
disinfectants are not tested for use on surfaces.

12 Safety
aUse either a pre-moistened wipe or a soft cloth dampened with cleaner or disinfectant.
bRemove all gel, debris, and bodily fluids from the stand.
cWith a new wipe, clean all parts of the stand by wiping from the clean areas to the soiled areas. This
method helps to avoid cross-contamination.
dObserve the minimum wet contact time. Monitor the stand parts for wet appearance. Re-apply with a
new wipe if no longer wet.
3Allow the stand to air dry in a clean, well-ventilated space.
For information about cleaning the peripherals, see the manufacturer’s instructions.
Safety
Observe the following warnings and cautions before using accessories and peripherals with the ultrasound
system and transducers.
Electrical safety
The stand is not an electrical device and does not require ground bond testing. The stand is for facilitating the
mobility of peripherals and their connection to the ultrasound system.
Approved cleaners and disinfectants for the stand
Cleaner/disinfectant Minimum wet contact time1
SaniCloth AF3 (gray top)23 minutes
SaniCloth Plus (red top) 3 minutes
PI-Spray II 10 minutes
1For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a
minimum period of time.
2Qualified for use as an intermediate-level disinfectant for mycobacteria. For a more complete list, refer to
the cleaners and disinfection document available at www.sonosite.com

English Deutsch Español Français Italiano Português
Safety 13
Equipment safety
WARNINGS To avoid the risk of electrical shock:
Do not plug the power cord of the Edge Stand or H-Universal Stand into a
multiple portable socket outlet (MPSO) or use an extension cord.
Use only properly grounded equipment. Shock hazards exist if the power supply
is not properly grounded. Grounding reliability can only be achieved when
equipment is connected to a receptacle marked “Hospital Only” or “Hospital
Grade”or the equivalent. The grounding wire must not be removed or defeated.
Use only accessories and peripherals recommended by SonoSite, including the
power supply. Connection of accessories and peripherals not recommended by
SonoSite could result in electrical shock. Contact SonoSite or your local
representative for a list of accessories and peripherals available from or
recommended by SonoSite.
To avoid the risk of electrical shock or injury, the equipment shall be connected
to a center-tapped single phase supply circuit when users in the United States
connect the equipment to a 240V supply system.
To avoid the risk of electrical shock and fire hazard, inspect the power supply, plugs,
cables, and power cords on a regular basis for damage.
Cautions Excessive bending or twisting of cables can cause a failure or intermittent
operation.
Improper cleaning or disinfecting of any part of the accessories and peripherals
can cause permanent damage. For cleaning and disinfecting instructions, see
“Cleaning and disinfecting” on page 11 and the manufacturer’s operating
instructions.
Accessible metal on the signal inputs and outputs on the back and side of the
ultrasound system is not protectively earthed. Do not perform a high current
ground impedance test involving this part.

14 Specifications
Clinical safety
Electromagnetic compatibility (EMC)
See the safety section of the ultrasound system user guide and supplemental guides for information on
electromagnetic compatibility with IEC 60601-1-2, including the Manufacturer’s Declaration.
Specifications
For information on the peripherals, see the manufacturer’s specifications.
Stand dimensions (approximate)
Width: 20.25 in. (51.4 cm)
Depth: 22 in. (55.9 cm)
Height: 45 in. (114.3 cm) maximum, 33 in. (83.8 cm) minimum
Height range: 12 in. (30.5 cm)
Weight: 40.0 lbs. (18.1 kg)
Safe working load per storage bin: 5 kg
Total stand weight with system and peripherals: 50 kg maximum
Temperature, humidity, and atmospheric pressure limits
Operating: 10–40°C; 15–95% R.H., 700 to 1060 hPa
Shipping/Storage: -35–65°C; 15–95% R.H., 500 hPa to 1060 hPa
WARNING:
To avoid misdiagnosis, do not use non-medical (commercial) grade peripheral
monitors. These monitors have not been verified or validated by SonoSite as
being suitable for diagnosis.

English Deutsch Español Français Italiano Português
Einführung 15
Norsk
Edge-Stativ
H-Universal-Stativ
Benutzerhandbuch
Einführung .................................................................................................................................................... 15
Dokumentkonventionen ............................................................................................................................................................ 16
Weiterführende Informationen ................................................................................................................................................ 17
Optionen des Stativs .................................................................................................................................... 18
Verwenden des Stativs ................................................................................................................................. 19
Einsetzen und Entfernen des Ultraschallsystems ...................................................................................... 19
Anschließen des Ultraschallsystems ........................................................................................................... 20
Verwenden von Peripheriegeräten und Zubehör ..................................................................................... 21
Drucker oder Barcode-Leser ...................................................................................................................................................... 21
Triple Transducer Connect ......................................................................................................................................................... 22
Fehlersuche ................................................................................................................................................... 23
Anschlussübersicht ....................................................................................................................................................................... 23
Kabel und Anschlüsse .................................................................................................................................................................. 24
Reinigung und Desinfektion ........................................................................................................................ 26
Sicherheit ...................................................................................................................................................... 27
Elektrische Sicherheit ................................................................................................................................................................... 27
Gerätesicherheit ............................................................................................................................................................................. 28
Klinische Sicherheit ....................................................................................................................................................................... 29
Elektromagnetische Verträglichkeit (EMV) ........................................................................................................................... 29
Technische Daten ......................................................................................................................................... 29
Stativabmessungen (ungefähr) ................................................................................................................................................ 29
Temperatur, Feuchtigkeit und Luftdruckgrenzwerte ....................................................................................................... 29
Einführung
Dieses Benutzerhandbuch beschreibt die Verwendung des Edge-Stativs (mit den Systemen SonoSite Edge II
und Edge verwendet) oder des H-Universal-Stativs und der Peripheriegeräte und Zubehörteile. Weitere
Informationen zu einem bestimmten Peripheriegerät entnehmen Sie der mit dem jeweiligen Produkt
gelieferten Bedienungsanleitung des Herstellers. Informationen zu den Schallköpfen und Ultraschallsystemen
finden Sie im Benutzerhandbuch für das Ultraschallsystem.

16 Einführung
Falls Sie das Stativ demontieren müssen, entnehmen Sie entsprechende Informationen den
Montageanweisungen,, die dem Stativ beigefügt sind oder unter www.sonosite.com zur Verfügung stehen.
Dokumentkonventionen
Ein WARNHINWEIS beschreibt die notwendigen Vorsichtsmaßnahmen zur Vermeidung einer Verletzung
oder eines tödlichen Unfalls.
Ein Vorsichtshinweis beschreibt die notwendigen Vorsichtsmaßnahmen zum Schutz der Produkte.
Ein Hinweis enthält ergänzende Informationen.
Nummerierte und mit Buchstaben versehene Schritte müssen in einer bestimmten Reihenfolge ausgeführt
werden.
Listen mit Gliederungspunkten stellen Informationen in einem Listenformat dar, schreiben jedoch keine
Reihenfolge vor.
Verfahren mit nur einem Schritt beginnen mit .
Die im Zusammenhang mit dem System verwendeten Symbole und Begriffe werden im Benutzerhandbuch
des Ultraschallsystems erklärt.
WARNHINWEIS Um eine Verletzung des Patienten, eine Fehldiagnose oder eine Verletzung
des Bedieners zu vermeiden, müssen die Warnhinweise im
Benutzerhandbuch des Ultraschallsystems und in den ergänzenden
Handbüchern sorgfältig gelesen werden.
Um Verletzungen oder Geräteschäden zu vermeiden, die durch ein Umkippen
des Stativs entstehen könnten, darf keine übermäßige Kraft auf das Stativ
angewandt werden. Das Einstellen der Stativplattform auf eine tiefe Position
kann dazu beitragen, diese potenzielle Gefahr zu verringern.

English Deutsch Español Français Italiano Português
Einführung 17
Norsk
Weiterführende Informationen
Der technische Kundendienst von FUJIFILM SonoSite ist wie folgt erreichbar:
Gedruckt in den USA.
Telefon
(USA oder Kanada)
+1-877-657-8118
Telefon
(außerhalb USA oder
Kanadas)
+1-425-951-1330, oder wenden Sie sich an Ihren
zuständigen Kundendienstvertreter
Fax +1-425-951-6700
E-Mail ser[email protected]
Web www.sonosite.com
Europäisches
Servicezentrum
Hauptstelle: +31 20 751 2020
Kundendienst auf Englisch: +44 14 6234 1151
Kundendienst auf Französisch: +33 1 8288 0702
Kundendienst auf Deutsch: +49 69 8088 4030
Kundendienst auf Italienisch: +39 02 9475 3655
Kundendienst auf Spanisch: +34 91 123 8451
Asiatisches
Servicezentrum
+65 6380-5589

18 Optionen des Stativs
Optionen des Stativs
Das Stativ bietet eine mobile Arbeitsplattform und ermöglicht die Unterbringung der Schallköpfe und anderer
Zubehörkomponenten. Außerdem bietet es Anschlussmöglichkeiten für Zubehör und Peripheriegeräte.
Abbildung 1 Stativ mit Zubehör. H-Universal-Stativ abgebildet mit M-Turbo-Ultraschallsystem, optionalem
Korb, Schwarzweißdrucker und Triple Transducer Connect.
*Der Aufbewahrungsbehälter befindet sich unter dem Drucker. Beachten Sie die Montageanleitung
für den Aufbewahrungsbehälter.
1 Ultraschallsystem 6 Schwarzweißdrucker
2 Haken (1 von 4) für Kabel und
Sonstiges
7 Netzadapter 1: Anschluss für das
Stromversorgungskabel des
Systems
3 Triple Transducer Connect 8 Netzadapter 2: Anschluss für das
Netzstromkabel (unterhalb der
Stativbasis)
4 Hebel zur Höhenverstellung 9 Sperrhebel für Rad
5 Korb oder Aufbewahrungsbehälter*
2
1
5
7
3
9
6
4
8
Other manuals for SonoSite Edge
2
This manual suits for next models
1
Table of contents
Languages:
Other FujiFilm Rack & Stand manuals