
2
STEP 10 Rinse the shaft for at least 1
minute in a large volume of sterile,
deionized water (for example, 8 L). Repeat
3 separate times. Discard the rinse water
after each rinse.
STEP 11 Dry the transducer with a
sterile, non-linting towel or medical-grade
air. Examine the transducer and cable for
any cracks. DO NOT use the transducer if
any damage is found.
Disinfecting
STEP 9
APlace up to 90 cm of the shaft in an approved
disinfectant. Make sure the controller cannot
fall into the solution. CHECK the expiration
date of the disinfectant, and follow ALL
chemical manufacturer’s instructions for soak
time and temperature.
B Wipe the controller and cable with a sterile,
non-linting cloth or single-use endoscopic
sponge moistened with disinfectant.
Remove cleaner by wiping again with a
fresh sterile cloth or sponge moistened with
sterile water.
STEP 7 Set up the
equipment.
AInsert the transducer into the
transducer adapter.
BInsert the fork adapter
securely in between the
control wheels on the
controller and do not allow it
to swivel.
CConnect the splitter to the
transducer adapter, the fork adapter, and the leakage tester.
DConnect the conductivity probe to the leakage tester and insert
into 0.9% saline solution or Cidezyme/Enzol.
A
B
C
D
Testing for electrical leakage
STEP 8
Make sure the tester displays
Ready for Testing, press FULL
TEST. Repeat the test two more
times.
If any of the results are Fail, do not
use the transducer and contact
FUJIFILM SonoSite for repair.
STEP 12 Transport and store the
transducer in accordance with your
organization’s policies.
Transporting and storing
For complete instructions on cleaning, disinfecting,
transporting, and storing the TEE transducer, refer to
the TEE Transducer User Guide.
For complete instructions on using the leakage tester,
refer to the ULT2000 Series User Manual.
SonoSite is a registered trademark of FUJIFILM Sonosite, Inc. in various jurisdictions. FUJIFILM is a registered trademark of FUJIFILM Corporation. Value From Innovation is a
trademark of FUJIFILM Holdings America Corporation. All other trademarks are the property of their respective owners.
Copyright © 2019 FUJIFILM SonoSite, Inc. All rights reserved.
P05497-09
*P05497-09*
EC Authorized
Representative
FUJIFILM SonoSite B.V.
Joop Geesinkweg 140
1114 AB Amsterdam,
The Netherlands
Manufacturer:
FUJIFILM SonoSite, Inc.
21919 30th Drive SE
Bothell, WA 9802, USA
T: 1-888-482-9449 or
1-425-951-1200
F: 1-425-951-1201
Australia Sponsor
FUJIFILM SonoSite
Australasia Pty Ltd
114 Old Pittwater Road
Brookvale, NSW, 2100
Australia