2.2.1 Pressure Equalisation
The pressure equalisation system, which ensures that the air-pressure within the cartridge of
the condenser microphone is kept equal to the ambient atmospheric pressure, allows sufciently
rapid pressure equalisation while keeping to a minimum the exchange of air between the interior
and exterior. This also extends the working life of the dehumidier.
Pressure equalisation takes place via the 1.5 mm diameter stainless-steel tube visible on the
side of the upper cylindrical section (see Fig. 2.4). This tube is also a conduit for the signal wire
connecting the calibration oscillator with the electrostatic actuator.
2.2.2 Electrostatic Actuator
The position of the actuator relative to the microphone is maintained by a calibration ring posi-
tioned under the Teon sleeve. The calibration ring ensures the correct spacing between the
actuator and the diaphragm of the microphone.
For inspection purposes, access to the signal wire in the pressure-equalisation tube is gained
after removing the rain-protection cap (see section 2.2.4) which is held in place by a left-hand
thread mushroom-head screw. In the case of the Type 41CN, the small black plastic cone
mounted on top of the screw has to be removed rst.
2.2.3 Microphone
If the electrostatic actuator, Teon sleeve and calibration ring are removed, the microphone
can be unscrewed (see section 2.2.4). The microphone is mounted on a galvanically isolated
assembly, together with a thick-lm, high-impedance input preamplier which is miniature,
ceramic-based, and robust.
Fig. 2.4 Showing microphone assembly - note the difference between
the two rain-protection caps
Rain-protection cap and actuator
Teon sleeve
Pressure-equalisation tube
Dehumidier inspection window
Transport tube
Type 41AM
Type 41CN
Add silicone grease to this thread