Sensor Start-up Time / Measurement Time!
The time from application of power to the SDI-12 power bus until the sensor is ready to
receive a command is approximately 75ms. The reported measurement time in response
to the aM! measure command (where a is sensor address) is 3 seconds, measured from
the end of aM! command response (a0022<CR><LF>) where n is the number of
segments. Actual measurement times are less, and a service request is issued as soon as
the measurement is completed. Current is at active level (15-20 mA) only during
measurement time, otherwise current is less than 0.1 mA.
Testing the GroPoint Lite (4-20mA)!
If desired, you can test the GroPoint Lite sensor before deploying it. The following test is
for functional testing only, NOT determining calibration values. It is only to confirm that
the sensor output increases with moisture increases, as it lacks the precision and
controlled environment used for calibration.
Materials Required!
To perform this test, you will need the
following:
•A 12V power supply.
•A digital multimeter set to 0-30mA.
•A 100 ohm to 500 ohm sense resistor.
•A plastic or glass container (NOT metal) at
least 15cm (6”) in diameter. It should be deep
enough to completely cover the sensor with
water when the sensor is held vertically
(about 23 cm, or 9” deep), although
acceptable results will still be obtained if
sensor is covered with water when
positioned diagonally in the container (about 17 cm, or 7” deep). The readings you
obtain may differ from stated if you use a different container.
•Water to fill the container.