Hach DR 5000 User manual

1
DOC272.53.90090
Instruction Sheet
DR 5000 Pour-Thru Cell Kit Installation
Safety Information
Please read this entire document before unpacking, setting up, or operating this
equipment. Pay attention to all danger and caution statements. Failure to do so could
result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired, do not use or
install this equipment in any manner other than that specified in this document.
Use of Hazard Information
Important Note: Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
Note: Information that supplements points in the main text.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol, if noted on the instrument, will be
included with a danger or caution statement in the manual.
General Information
The Pour-Thru Cell Kit is an optional accessory for the DR 5000 Spectrophotometer. The
Pour-Thru cell improves measurement accuracy, as zeroing, measurements and
measurement comparisons are carried out under the same optical conditions. As the
samples all flow through the same cell, errors attributable to the different optical
characteristics of individual cells are excluded. The Pour-Thru is designed to allow the
samples to be introduced into the cell without having to handle the cell in any way.
There are three types of Pour-Thru cells; a 1-inch cell, 1-cm cell and a 5-cm cell. Only the
quartz cells (1 cm and 5 cm) can be used over the entire wavelength range. The 1-inch
Pour-Thru cell can be used in the visible spectrum (340 to 900 nm).
Specifications
Specifications are subject to change without notice.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Available path lengths 1-inch or 1-cm or 5-cm
Wavelength range 190 to 1100 nm
1-inch cell can only be used in the visible spectrum (340 to 900 nm)
Rinsing volume At least 20 mL for the 1-inch path length
At least 10 mL for the 1-cm path length
At least 20 mL for the 5-cm path length

2
DR 5000 Pour-Thru Cell Kit Installation
Pour-Thru cell kit components
Refer to Figure 1 and Figure 2 for a complete description of components supplied with the
Pour-Thru cell kit. If any of these items are missing or damaged, refer to the manufacturer
or a sales representative immediately.
Note: Retain the original packaging materials. Returned instruments should be shipped in the
original packaging material to protect against damage during transportation.
Figure 1 Pour-Thru cell module with adapter
1Pour-Thru module 2Locking screws 31-inch cell adapter

3
DR 5000 Pour-Thru Cell Kit Installation
Assembly
The 1-inch Pour-Thru cell requires the use of a Pour-Thru cell adapter (Figure 1, item 3).
The 1-cm and 5-cm cells use the multi-cell holder supplied with the DR 5000.
Prepare the tubing and Pour-Thru housing
1. Cut the 182 cm (6 ft) length of clear tubing into two equal lengths of tubing,
approximately 91 cm (3 ft) in length.
2. Feed both ends of the clear tubing through the two holes on the side of the housing,
leaving approximately 20 cm (8 inch) inside the housing (Figure 3).
3. Resting the housing on its left side, remove the rubber fitting if installed. Route the
external tubing through the channels molded into the exterior right side of the housing.
Replace the rubber fitting (Figure 3).
Important Note: Do not kink or bend the tubing sharply.
Figure 2 Pour-Thru cell kit components
1Stand assembly 3Glass funnel 5Pour-Thru cell—1-inch or 1-cm
(5 cm sold separately). The 1-inch
cell is shown.
2Overflow tube assembly 4Tubing

4
DR 5000 Pour-Thru Cell Kit Installation
1-inch cell assembly
1. Remove the multi-cell holder from the DR 5000 and replace it with the Pour-Thru cell
adapter (Figure 1, Item 3). Do not screw down the Pour-Thru cell adapter.
2. Place the Pour-Thru module on top of the Pour-Thru cell adapter in the DR 5000.
Make sure the locking screw holes are aligned. Secure the assembly with the two
locking screws.
3. Clean the Pour-Thru cell with a lint-free cloth. Insert the Pour-thru cell with the barb
fittings pointing left.
4. Route the tubing as shown in Figure 4 on page 5, trimming tube length as needed to
achieve smooth curves with no kinks or sharp bends.
5. Assemble the stand and connect the tubing, noting flow direction arrows molded into
the housing where the tubes emerge. The inflow tube is connected to the glass funnel;
the outflow tube is connected to the overflow tube assembly (Figure 5 on page 5).
6. Connect the drain tube (black rubber tube) to the bottom fitting of the overflow tube
assembly. Connect the other end of the drain tube to an appropriate waste vessel.
The tubes can be shortened if necessary.
Figure 3 Tubing installed in housing and channels with rubber fitting
1Tubing (8” inside housing) 2Rubber fitting 3Tubing channels

5
DR 5000 Pour-Thru Cell Kit Installation
Figure 4 1-inch cell tube orientation
1Barb fittings 2Locking screws
Figure 5 Final assembly—1-inch Pour-Thru cell and stand assembly
1Pour-Thru cell module 5Glass funnel
2Cell outflow tube 6Distance between inflow tube into the cell and outflow tube
from the cell into the overflow tube assembly = 5 cm
3Cell inflow tube 7Stand assembly
4Overflow tube assembly 8Drain tube

6
DR 5000 Pour-Thru Cell Kit Installation
1-cm and 5-cm cell assembly
1. Arrange the multi-cell adapter so the appropriate cell holder for the Pour-Thru cell
points towards the front of the instrument. Make sure the imprint of the cell type is
visible. Do not screw down the multi-cell holder.
2. Place the Pour-Thru cell housing on top of the multi-cell holder in the DR 5000. Make
sure the locking screw holes are aligned. Secure the assembly with the two locking
screws.
3. Clean the Pour-Thru cell with a lint-free cloth. Insert the Pour-Thru cell in the cell
holder so that the cell windows coincide with those of the cell holder.
4. Route the tubing as shown in Figure 6. Trim the tube length as needed to achieve
smooth curves with no kinks or sharp bends.
5. Assemble the stand and connect the tubing, noting flow direction arrows molded into
the housing where the tubes emerge. The inflow tube is connected to the glass funnel;
the outflow tube is connected to the overflow tube assembly.
6. Connect the drain tube (black rubber tube) to the bottom fitting of the overflow tube
assembly. Connect the other end of the drain tube to an appropriate waste vessel.
The tubes can be shortened if necessary.

7
DR 5000 Pour-Thru Cell Kit Installation
Figure 6 Pour-Thru cell 1-cm with tubes in place
1Entry/exit channel 5Overflow tube assembly
2Cell outflow tube (plastic) 6Glass funnel
3Cell inflow tube 7Stand, including mountings for glass funnel and overflow
tube assembly
41-cm or 5-cm Pour-Thru cell (1-cm is shown) 8Drain tube (rubber tube)

8
DR 5000 Pour-Thru Cell Kit Installation
Cleaning the Pour-Thru cells
Important Note: Do not use solvents (e.g. acetone) to clean the Pour-Thru cell. A dilute
acid solution can be used for cleaning. Rinse thoroughly with deionized water.
If windows are cloudy or soiled or bubbles form, add 50 mL of a detergent solution to the
cell and wait for a few minutes to allow it to take effect. Then purge the cell thoroughly with
distilled water. If necessary, the 1-inch Pour-Thru cell can be taken apart to be cleaned.
Wipe the windows of the cell with a soft cloth only—paper towels and other paper products
may scratch the glass.
Troubleshooting
Carefully inspect clear tubing for kinks or sharp bends which may restrict the flow of liquid
through the Pour-Thru cell.
Use of the Pour-Thru cell
Since there is a difference in the light path between the 1-inch square cells and the 1-inch
Pour-Thru cell it is necessary to adjust the calibration slightly when using the 1-inch
Pour-Thru cell. This is done by applying a dilution factor of 1.1 or performing a standard
adjustment. The appropriately revised method can be stored as a user program. If the use
of a Pour-Thru cell is prescribed in a Hach method, no adjustment is needed.
Be sure to use the Pour-Thru cell that corresponds with the cell path length called for in
the method. The 1-cm and 5-cm Pour-Thru cells should only be used with Hach methods
that call for those specific cells or with user programs.
Be sure to use the appropriate sample volume and reagents required for each particular
cell. A 1-inch Pour-Thru cell can be used with a Hach method if the prescribed sample
volume is 25 mL. If the prescribed sample volume is different than 25 mL, adjust the
amount of sample and reagents as necessary to achieve the stated minimum rinsing
volume (refer to Specifications on page 1).
Although the Pour-Thru cell can be used for the Hach methods listed in Table 1, the cell
must be thoroughly purged with distilled water between individual samples.
The Pour-Thru cell can also be used to run Nitrogen, Ammonia, Nessler Method and TKN
chemistries. Clean the cell by pouring a few sodium thiosulfate pentahydrate crystals into
the cell. Rinse out the crystals with deionized water.
Important Note: The Pour-Thru cell cannot be used for the Hach methods listed in
Table 2.
Table 1 Pour-Thru cell methods that require purging
Aluminum, Aluminon Chloride Dioxide, LR Cobalt, PAN
Copper, Porphyrin Hardness, Calmagite Manganese, LR, PAN
Nickel, PAN Nitrate, MR Nitrate, HR
Table 2 Hach methods that cannot be used with the Pour-Thru cell
Aluminum ECR Arsenic Barium Boron, Carmine
Cyanuric Acid Flouride Formaldehyde Lead, LeadTrak
Mercury Nickel, Heptoxime Nitrite, HR PCB
Phenols Potassium Selenium Silver
Suspended Solids Sulfate TPH Volatile Acids
Zinc Surfactants, Anionic (Detergents)

9
DR 5000 Pour-Thru Cell Kit Installation
Replacement parts
Description QTY Item no.
Pout-Thru cell kit for the DR 5000; 1-inch cell
includes: 1-inch Pour-Thru cell (all other cells should be purchased separately) 1LZV479
Pour-Thru cell kit for the DR 5000; 1-cm cell
includes: 1-cm Pour-Thru cell (all other cells should be purchased separately) 1 LZV789
Pour Thru Cell, 1x1 cm (Quartz/160 μL center height 10 mm) 1A24209
Pour Thru Cell, 1x1 cm (Quartz/450 μL center height 10 mm) 1 LZV510
Pour Thru Cell, 5x1 cm (Quartz/370 μL center height 10 mm) 1LZV649
Pour Thru Cell, 1" Square (1" path length) 1 5913700
Tubing set Pour-Thru 1LZV569

Contact Information
HACH Company
World Headquarters
P.O. Box 389
Loveland, Colorado
80539-0389 U.S.A.
Tel (800) 227-HACH
(800) -227-4224
(U.S.A. only)
Fax (970) 669-2932
www.hach.com
Repair Service in the
United States:
HACH Company
Ames Service
100 Dayton Avenue
Ames, Iowa 50010
Tel (800) 227-4224
(U.S.A. only)
Fax (515) 232-3835
Repair Service in Canada:
Hach Sales & Service
Canada Ltd.
1313 Border Street, Unit 34
Winnipeg, Manitoba
R3H 0X4
Tel (800) 665-7635
(Canada only)
Tel (204) 632-5598
Fax (204) 694-5134
Repair Service in
Latin America, the
Caribbean, the Far East,
Indian Subcontinent, Africa,
Europe, or the Middle East:
Hach Company World
Headquarters,
P.O. Box 389
Loveland, Colorado,
80539-0389 U.S.A.
Tel +001 (970) 669-3050
Fax +001 (970) 669-2932
HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf
Tel. +49 (0)2 11 52 88-320
Fax +49 (0)2 11 52 88-210
www.hach-lange.de
HACH LANGE LTD
Pacific Way
Salford
GB-Manchester, M50 1DL
Tel. +44 (0)161 872 14 87
Fax +44 (0)161 848 73 24
www.hach-lange.co.uk
HACH LANGE LTD
Unit 1, Chestnut Road
Western Industrial Estate
IRL-Dublin 12
Tel. +353(0)1 46 02 5 22
Fax +353(0)1 4 50 93 37
www.hach-lange.ie
HACH LANGE GMBH
Hütteldorferstr. 299/Top 6
A-1140 Wien
Tel. +43 (0)1 9 12 16 92
Fax +43 (0)1 9 12 16 92-99
www.hach-lange.at
DR. BRUNO LANGE AG
Juchstrasse 1
CH-8604 Hegnau
Tel. +41(0)44 9 45 66 10
Fax +41(0)44 9 45 66 76
www.hach-lange.ch
HACH LANGE FRANCE
S.A.S.
33, Rue du Ballon
F-93165 Noisy Le Grand
Tél. +33 (0)1 48 15 68 70
Fax +33 (0)1 48 15 80 00
www.hach-lange.fr
HACH LANGE SA
Motstraat 54
B-2800 Mechelen
Tél. +32 (0)15 42 35 00
Fax +32 (0)15 41 61 20
www.hach-lange.be
DR. LANGE NEDERLAND
B.V.
Laan van Westroijen 2a
NL-4003 AZ Tiel
Tel. +31(0)344 63 11 30
Fax +31(0)344 63 11 50
www.hach-lange.nl
HACH LANGE APS
Åkandevej 21
DK-2700 Brønshøj
Tel. +45 36 77 29 11
Fax +45 36 77 49 11
www.hach-lange.dk
HACH LANGE AB
Vinthundsvägen 159A
SE-128 62 Sköndal
Tel. +46 (0)8 7 98 05 00
Fax +46 (0)8 7 98 05 30
www.hach-lange.se
HACH LANGE S.R.L.
Via Riccione, 14
I-20156 Milano
Tel. +39 02 39 23 14-1
Fax +39 02 39 23 14-39
www.hach-lange.it
HACH LANGE S.L.U.
Edif. Arteaga Centrum
C/Larrauri, 1C- 2ª Pl.
E-48160 Derio/Vizcaya
Tel. +34 94 657 33 88
Fax +34 94 657 33 97
www.hach-lange.es
HACH LANGE LDA
Av. do Forte nº8
Fracção M
P-2790-072 Carnaxide
Tel. +351 214 253 420
Fax +351 214 253 429
www.hach-lange.pt
HACH LANGE SP.ZO.O.
ul. Opolska 143 a
PL-52-013 Wrocław
Tel. +48 (0)71 342 10-83
Fax +48 (0)71 342 10-79
www.hach-lange.pl
HACH LANGE S.R.O.
Lešanská 2a/1176
CZ-141 00 Praha 4
Tel. +420 272 12 45 45
Fax +420 272 12 45 46
www.hach-lange.cz
HACH LANGE S.R.O.
Roľnícka 21
SK-831 07 Bratislava –
Vajnory
Tel. +421 (0)2 4820 9091
Fax +421 (0)2 4820 9093
www.hach-lange.sk
HACH LANGE KFT.
Hegyalja út 7-13.
H-1016 Budapest
Tel. +36 (06)1 225 7783
Fax +36 (06)1 225 7784
www.hach-lange.hu
HACH LANGE S.R.L.
Str. Leonida, nr. 13
Sector 2
RO-020555 Bucuresti
Tel. +40 (0) 21 201 92 43
Fax +40 (0) 21 201 92 43
www.hach-lange.ro
HACH LANGE
8, Kr. Sarafov str.
BG-1164 Sofia
Tel. +359 (0)2 963 44 54
Fax +359 (0)2 866 04 47
www.hach-lange.bg
HACH LANGE SU
ANALİZ SİSTEMLERİ
LTD.ŞTİ.
Hilal Mah. 75. Sokak
Arman Plaza No: 9/A
TR-06550 Çankaya/ANKARA
Tel. +90 (0)312 440 98 98
Fax +90 (0)312 442 11 01
www.hach-lange.com.tr
HACH LANGE D.O.O.
Fajfarjeva 15
SI-1230 Domžale
Tel. +386 (0)59 051 000
Fax +386 (0)59 051 010
www.hach-lange.si
ΗΑCH LANGE E.Π.Ε.
Αυλίδος 27
GR-115 27 Αθήνα
Τηλ. +30 210 7777038
Fax +30 210 7777976
www.hach-lange.gr
HACH LANGE E.P.E.
27, Avlidos str
GR-115 27 Athens
Tel. +30 210 7777038
Fax +30 210 7777976
www.hach-lange.gr
© Hach Company, 2008. Hach Lange GmbH, 2008. All rights reserved. Printed in the U.S.A and Germany Edition 2Sept 2008
Other manuals for DR 5000
2
Table of contents
Other Hach Measuring Instrument manuals

Hach
Hach Polymetron 9523sc pH Quick manual

Hach
Hach LOC 100 User manual

Hach
Hach DR300 User manual

Hach
Hach 2100N User manual

Hach
Hach pHD Sensor User manual

Hach
Hach ANATEL PAT700 User manual

Hach
Hach 9525sc User manual

Hach
Hach Micro Dist User guide

Hach
Hach FT660 sc Quick manual

Hach
Hach 2100Q User manual

Hach
Hach QuikChem 8500 2 Series Instruction Manual

Hach
Hach DR2400 User manual

Hach
Hach ISENA381 Series User manual

Hach
Hach HQ1110 Quick manual

Hach
Hach TSS Portable User manual

Hach
Hach 2100P Reference guide

Hach
Hach DR/700 Reference guide

Hach
Hach orbisphere C1100 User manual

Hach
Hach POCKET COLORIMETER II LCK 360 Zinc User guide

Hach
Hach POLYMETRON 8810 ORP Quick manual