SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
Instruction Manual
www.hannainst.com
DearCustomer,
Thank you for choosing a Hanna Product. Please read the
instructions carefully before using the chemical test kit. It
will provide you with the necessary information for correct
use of the kit.
Remove the chemical test kit from the packing material and
examine it carefully to make sure that no damage has
occurred during shipping. If there is any noticeable dam-
age, notify your Dealer or the nearest Hanna office
immediately.
Each kit is supplied with:
• HI 3844A-0 Hydrogen Peroxide Reagent, 1 bottle
(100 mL) ;
• HI3844B-0 Hydrogen Peroxide Reagent, 1 bottle
(17g) ;
• HI 3844C-0 Hydrogen Peroxide Reagent, 1 bottle
with dropper (30 mL);
• HI 3844D-0 Hydrogen Peroxide Reagent, 1 bottle
with dropper (25 mL)
• 1 plastic test tube, graduated with cap;
• 1 calibrated plastic vessel (50 mL);
• 1 plastic pipette (3 mL);
• 1 plastic pipette (1 mL), for HI 3844A-0 Reagent;
• 1 plastic spoon.
Note: Any damaged or defective item must be returned in
its original packing materials.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
HI 3844
Hydrogen
Peroxide
Test Kit
ISTR3844 12/99 PRINTED IN ITALY
Range 0 to 2 ppm H2O2Low Range
0 to 10 ppm H2O2High Range
Smallest Increment 0.25 ppm H2O2Low Range
1.0 ppm H2O2High Range
Analysis Method Drop-Count Iodometric Titration
Sample Size 25 mL Low Range
5 mL High Range
Number of Tests 100 (average)
CaseDimensions 235x175x115 mm (9.2x6.9x4.5")
Shipping Weight 450 g (15.9 oz.)
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Remove the cap from the plastic vessel and rinse it with
sample. For H2O2Low Range fill it up to the 25 mL
mark with the sample. In case of H2O2High Range fill
the vessel with 5 mL of the sample using the graduated
plastic test tube and add D.I. water to the 15 mL
mark.
• Add 1 mL of HI 3844A-0 reagent for Low Range or 0.5
mL for High Range using the plastic pipette and swirl
gently to mix.
• Using the plastic spoon provided, add 1 spoon of HI
3844B-0 reagent and swirl gently to mix.
• Wait for 15 minutes to allow
reaction to occur, leaving the
plastic vessel closed and in a dark
place.
Note: if hydrogen peroxide is present, the solution will turn
to a dark yellow color.
• Slowly add drops of the titration reagent HI 3844C-0
while swirling the solution and counting the drops.
Continue adding the titration solution until the yellow
colorofthesolutionisalmost colorless.
Hydrogen peroxide is widely used as a disinfectant and as
a bleach for textiles, wood pulp, hair, fur etc.. It is also
used as a substitute for chlorine in water and sewage
treatment. Most common commercial forms are aqueous
solutions containing about 6, 12 and 30 per cent hydrogen
peroxide and are referred to as "20-volume", "40-volume"
and "100-volume" respectively, meaning the value of oxy-
gen liberated when the solution is boiled. The Hanna test
kit can quickly and easy determine concentration in water
up to 10 ppm of hydrogen peroxide. This is due to the fact
that it is not affected by stabilizers, which are sometimes
added to commercial hydrogen peroxide solutions. The kit is
portable and can be used in the field as well as in the
laboratory.
Note:Note:
Note:Note:
Note: mg/L is equivalent to ppm (parts per million).
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USE
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTION
Hydrogen peroxide is determined by a titrimetric method. It
reactsslowlywithiodide in acid solution (Step 1); thus a 15
minutes interval is required to allow the reaction to occur
completely. The amount of iodine generated is equivalent to
the hydrogen peroxide in the sample. The liberated iodine
is then titrated with standard sodium thiosulfate solution
that reduces the iodine back to iodide ions (Step 2).
Step 1: H2O2 + 2H++ 2I–→I2+ 2H2O
Step 2: I2+ 2S2O3
2– →2I–+ S4O6
2–
REFERENCESREFERENCES
REFERENCESREFERENCES
REFERENCES
Vogel's, Textbook of quantitative chemical analysis, 5th ed.,
Longman Scientific and Technical
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETY
The chemicals contained in this test kit may be hazardous if
improperly handled. Read Health and Safety Data Sheets
before performing the test.
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONS
• Add 2 to 3 drops of HI 3844D-0 reagent and mix by
carefully swirling the plastic vessel in tight circles. The
solution will turn a blue color.
• Continue adding slowly drops of the titration reagent HI
3844C-0, while swirling and counting the drops, until
the solution changes from blue to colorless.
• To obtain the concentration in ppm of Hydrogen Perox-
ide in your sample, multiply the total number of drops
of the titration reagent HI 3844C-0 used to turn the
solution from the dark yellow color to colorless by 0.25
or 1 for Low Range or High Range, respectively.
#of DROPS*0.25=ppm HydrogenPeroxide LR
# of DROPS *1= ppm Hydrogen Peroxide HR
NoteNote
NoteNote
Note: after a few minutes, the blue color of the starch/
iodine complex may reappear after the titration has
been completed due to air oxidation of the iodine.

Other Hanna Instruments Chemical Test manuals

Hanna Instruments HI 3880 User manual

Hanna Instruments

Hanna Instruments HI 3880 User manual

Hanna Instruments HI 38015 User manual

Hanna Instruments

Hanna Instruments HI 38015 User manual

Hanna Instruments HI 38067 User manual

Hanna Instruments

Hanna Instruments HI 38067 User manual

Hanna Instruments HI 3859 User manual

Hanna Instruments

Hanna Instruments HI 3859 User manual

Hanna Instruments HI 3820 User manual

Hanna Instruments

Hanna Instruments HI 3820 User manual

Hanna Instruments HI 3826 User manual

Hanna Instruments

Hanna Instruments HI 3826 User manual

Hanna Instruments HI 3881/0 User manual

Hanna Instruments

Hanna Instruments HI 3881/0 User manual

Hanna Instruments HI 3898 User manual

Hanna Instruments

Hanna Instruments HI 3898 User manual

Hanna Instruments HI764 User manual

Hanna Instruments

Hanna Instruments HI764 User manual

Hanna Instruments HI 3851 User manual

Hanna Instruments

Hanna Instruments HI 3851 User manual

Hanna Instruments HI 3839 User manual

Hanna Instruments

Hanna Instruments HI 3839 User manual

Hanna Instruments HI 3811 User manual

Hanna Instruments

Hanna Instruments HI 3811 User manual

Hanna Instruments HI 3824 User manual

Hanna Instruments

Hanna Instruments HI 3824 User manual

Hanna Instruments HI 3810 User manual

Hanna Instruments

Hanna Instruments HI 3810 User manual

Hanna Instruments HI 3815 User manual

Hanna Instruments

Hanna Instruments HI 3815 User manual

Hanna Instruments HI 38051 User manual

Hanna Instruments

Hanna Instruments HI 38051 User manual

Hanna Instruments HI 3880/0 User manual

Hanna Instruments

Hanna Instruments HI 3880/0 User manual

Hanna Instruments HI 3857 User manual

Hanna Instruments

Hanna Instruments HI 3857 User manual

Hanna Instruments HI 713 User manual

Hanna Instruments

Hanna Instruments HI 713 User manual

Hanna Instruments HI 38014 User manual

Hanna Instruments

Hanna Instruments HI 38014 User manual

Hanna Instruments HI 38017 User manual

Hanna Instruments

Hanna Instruments HI 38017 User manual

Hanna Instruments HI 38049 User manual

Hanna Instruments

Hanna Instruments HI 38049 User manual

Hanna Instruments HI 3849 User manual

Hanna Instruments

Hanna Instruments HI 3849 User manual

Hanna Instruments HI 3822 User manual

Hanna Instruments

Hanna Instruments HI 3822 User manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.